Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Cardiac (adult and pediatric); Musculoskeletal Conventional and Superficial; Peripheral Vascular; Transvaginal; Transrectal
Device Story
Voluson i/e is a portable, notebook-style diagnostic ultrasound system with integrated keyboard and color LCD. It supports linear, curved linear, and mechanical scanning transducers for B, M, PWD, Color, Color M, Amplitude Doppler, Harmonic Imaging, and Coded Pulse modes. Specialized motor-driven transducers and software enable 3D/4D volume imaging. Used in clinical settings by physicians or trained sonographers for general-purpose ultrasound imaging. The system includes comprehensive networking and DICOM capabilities. Output is displayed on the integrated screen for real-time visualization, measurement, and analysis, aiding clinicians in diagnosis and clinical decision-making across various anatomical regions. Benefits include high-performance imaging and volume analysis for improved diagnostic assessment.
Clinical Evidence
Bench testing only. No clinical studies were required to support substantial equivalence. The device was evaluated for acoustic output, biocompatibility, and electrical/mechanical safety, conforming to relevant standards (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2/3, ISO 10993-1).
Technological Characteristics
Portable notebook-style console with integrated keyboard and color LCD. Supports linear, curved linear, and mechanical scanning transducers. Modes: B, M, PWD, Color, Color M, Amplitude Doppler, Harmonic Imaging, Coded Pulse, 3D/4D. Connectivity: Networked, DICOM. Biocompatible patient-contact materials. Software includes SRI, XTD View, VCI, 4D-biopsy, STIC, HD-Flow, HD-Zoom, Volume Cine, SonoVCAD, SonoAVC, SonoNT, and SonoRenderStart.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in fetal/OB, abdominal (including GYN, pelvic, infertility monitoring), pediatric, small organ (breast, testes, thyroid, etc.), adult/pediatric cardiac, musculoskeletal (conventional/superficial), peripheral vascular, transvaginal, and transrectal applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Voluson E6/E8 E8 Expert/E10 Diagnostic Ultrasound System (K122327)
Voluson S6/S8 Diagnostic Ultrasound System (K120741)
Voluson P6/P8 Diagnostic Ultrasound System (K122387)
Submission Summary (Full Text)
{0}------------------------------------------------
K131937
page 1 of 3
GE Healthcare
510(k) Premarket Notification Submission
### 510(k) Summary
| In accordance with 21 CFR 807.92 the following summary of information is provided: | |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | June 26. 2013 |
| Submitter: | GE Healthcare [GE Healthcare Austria GmbH & Co OG]<br>Tiefenbach 15<br>Zipf, Austria 4871 |
| Primary Contact Person: | Bryan Behn<br>Regulatory Affairs Manager<br>GE Healthcare<br>T:(414)721-4214<br>F:(414)918-8275 |
| Secondary Contact Person: | Thomas Reisenberger<br>Regulatory Affairs Specialist<br>GE Healthcare Austria GmbH & Co OG<br>T:(++43)7682-3800-332<br>F:(++43)7682 3800-47<br>SEP 2 4 2013 |
| Device: Trade Name: | Voluson i/e Diagnostic Ultrasound System |
| Common/Usual Name: | Voluson i/e |
| Classification Names: | Class II |
| Product Code: | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-1YN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Predicate Device(s): | K053435 Voluson i Diagnostic Ultrasound System<br>K122327 Voluson E6/E8 E8 Expert/E10 Diagnostic Ultrasound System<br>K120741 Voluson S6/S8 Diagnostic Ultrasound System<br>K122387 Voluson P6/P8 Diagnostic Ultrasound System |
| Device Description: | The Voluson i/e consists of a portable, notebook type console with integrated keyboard controls and color LCD display. It supports a variety of. It utilizes a variety of linear, curved linear transducers including mechanical scanning transducers supporting all standard acquisition modes (B, M. PWD, Color Color M and Amplitude Doppler modes, Harmonic Imaging and Coded Pulse). The Voluson i/e also utilizes motor driven transducers and image processing software that are specialized for 3D/4D volume imaging. It provides high performance ultrasound imaging and analysis and has comprehensive networking and DICOM capability. |
Image /page/0/Picture/4 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized in a cursive font, and the entire logo is presented in black and white.
:
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. The letters are stylized and connected, with the "G" looping around the "E". The logo is in black and white.
Additional probes and software features cleared on the predicate Device Modification: devices have been added.
> Addition of transducers - IC5-9W-RS (similar to IC5-9-D cleared on Voluson E Scries K122327), RAB2-6-RS (Cleared on Voluson P6/P8 K122387), 8C-RS (Cleared on Voluson S6/S8 K120741). Transducers added via Appendix E of "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers", issued September 9. 9L-RS, RAB4-8-RS, RNA5-9-RS and RIC5-9W-RS, 2008" AB2-7-RS and SP10-16-RS.
Software improvements migrated from predicate K122327:
SRI (Speckle Reduction Imaging), XTD View (Extended View) and VCI (Volume Contrast Imaging), 4D-biopsy, STIC (Spatial Temporal Image Correlation) HD-Flow (High Density Flow), HD- Zoom (High Density Zoom), Volume Cine, Sono VCAD, Sono AVC, SonoNT and SonoRenderStart
- The device is a general purpose ultrasound system. Specific Intended Use: clinical applications remain the same as previously cleared: Fetal/OB: Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Cardiac (adult and pediatric); Musculoskeletal Conventional and Superficial; Peripheral Vascular; Transvaginal: Transrectal
- The Voluson i/e employs the same fundamental scientific Technology: technology as its predicate devices.
Determination of Summary of Non-Clinical Tests:
Substantial Equivalence: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform with applicable medical device safety standards. The Voluson i/e and its applications comply with voluntary standards:
- 1. AAMI/ANSI ES60601-1. Medical Electrical Equipment - Part 1: General Requirements for Safety
- 2. IEC60601-1-2, Medical Electrical Equipment -Part 1-2:General Requirements for Safety - Collateral
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are in black, contrasting with the white background.
K131937
Page 3 of 3
Standard: Electromagnetic Compatibility Requirements and Tests
- 3. IEC60601-2-37, Medical Electrical Equipment -Part 2-37:Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
- 4. NEMA UD 3. Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- 5. ISO10993-1, Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing- Third Edition
- 6. NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- 7. ISO14971. Application of risk management to medical devices
- 8. NEMA. Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology)
The following quality assurance measures were applied to the development of the system: ·
- . Risk Analysis
- . Requirements Reviews
- . Design Reviews
- Testing on unit level (Module verification)
- . Integration testing (System verification)
- . Final Acceptance Testing (Validation)
- . Performance testing (Verification)
- Safety testing (Verification)
Transducer materials and other patient contact materials are biocompatible.
#### Summary of Clinical Tests:
The subject of this premarket submission. Voluson i/e, did not require clinical studies to support substantial equivalence.
GE Healthcare considers the Voluson i/e to be as safe, as Conclusion: effective, and performance is substantially equivalent to the predicate device(s).
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) of the United States. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is the HHS symbol, which features a stylized caduceus-like design with three parallel lines curving upwards.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 24, 2013
GE Healthcare % Mr. Bryan Behn Regulatory Affairs Manager 9900 W. Innovation Drive WAUWATOSA WI 53226
Re: K131937
Trade/Device Name: Voluson i/e Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Il Product Code: IYN, IYO, ITX Dated: June 26, 2013 Received: June 27, 2013
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
This determination of substantial equivalence applies to the following transducers intended for use with the Voluson i/e, as described in your premarket notification:
#### Transducer Model Number
| RAB2-5-RS | RAB4-8-RS |
|------------|------------|
| RIC5-9-RS | RNA5-9-RS |
| RSP6-16-RS | AB2-7-RS |
| E8C-RS | 9L-RS |
| 4C-RS | SP10-16-RS |
| 12L-RS | RAB2-6-RS |
| RIC5-9W-RS | 8C-RS |
| IC5-9W-RS | |
{4}------------------------------------------------
Page 2-Mr. Behn
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.html Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Pari 803), please go to
http://www.fda.gov/MedicalDevices/SafetyReportalProblem/default.htm for the CDRF's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
lor
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a black and white logo of General Electric (GE). The logo consists of the letters 'G' and 'E' intertwined in a stylized, circular design. The letters are surrounded by a decorative border, giving the logo a classic and recognizable appearance.
510(k) Number (if known): K131937
Voluson i/e Diagnostic Ultrasound System Device Name:
Indications for Use:
The device is a general purpose ultrasound system. Specific clinical applications include: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal; Transrectal
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_NA (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Page I of 17
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a black and white logo of General Electric (GE). The logo consists of the letters 'G' and 'E' intertwined in a stylized, circular design. The letters are surrounded by a decorative border, giving the logo a classic and recognizable appearance.
### Indications for Use Forms
The following forms represent indications with clinical applications and exam types along with the modes of operation for the Voluson i/e system and for all of its probe/mode combinations. Combinations identified by "N" are new while "P" represents those previously cleared with the unmodified Voluson i/e, P* indicates those added in this submission, but have clearance for the same modes and applications on another GE Ultrasound System. In a similar manner, "E" represents combinations added to the unmodified Voluson i/e via Appendix E of the FDA Ultrasound Guidance. This modification did not alter the previously cleared system level indications or clinical applications.
{7}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form GE Voluson i/e Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [5,6] |
| Pediatric | P | P | P | | P | P | P | P | P | P | [5,6] |
| Small Organ[2] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | [5] |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal<br>Conventional | P | P | P | | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal<br>Superficial | P | P | P | | P | P | P | P | P | P | [5,6] |
| Other | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [5,6] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication: P = previously deared by FDA; E = added under Appendix E
Notes: [1] Abdominal Includes renal, GYN/Pelvic. Urology
[2] Small organ includes breast, thyroid, salivary gland, lymph nodes, podiatric and neonalal patients [3] Cardiac is Adull and Pediatric.
[5] 3D/4D Imaging Mode
[6] Includes Imaging of guidance of biopsy (2D/3D/4D)
[7] Includes infertility monitoring of follicle development
[8] Includes urology/prostate
["] Combined modes are BM, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 2 of 17
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a black and white logo. The logo is circular and contains the letters 'GE' intertwined in the center. The letters are stylized and appear to be hand-drawn. The logo is surrounded by a thin border.
#### Diagnostic Ultrasound Indications for Use Form
### GE Voluson i/e with RAB2-5-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | P | P | P | | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[4] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication: P = previously cleared by FDA; E = added under Appendix E
Notes: (1) Abdominal includes renal, GYN/Pelvic, Urology
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatic and neonatal patients
[3] Cardiac is Adult and Pediatric.
(5) 3D/4D Imaging Mode
(6) Includes imaging of guidance of biopsy (2D/3D/4D)
[7] Includes infertility monitoring of follicle development
[8] Includes urology/prostate
['] Combined modes are BM, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 3 of 17
਼ੁ
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo with the letters 'GE' intertwined within a circular border. The letters are stylized and appear to be hand-drawn or calligraphic. The border is also decorative, with a swirling pattern that surrounds the letters. The logo is black and white.
### Diagnostic Ultrasound Indications for Use Form GE Voluson i/e with RIC5-9-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[1] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[6] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [5,6] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology
[2] Small organ includes breast, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adull and Pediatric.
[5] 3D/4D Imaging Mode
[6] Includes Imaging of guidance of biopsy (2D/3D/4D)
171 Includes infertill;y monitoring of follicle development
{8} Includes urology/prostate
[1] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 4 of 17
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a black and white logo of General Electric (GE). The logo consists of the letters "G" and "E" intertwined within a circular border. The design is intricate and stylized, with the letters having a flowing, cursive appearance. The logo is a recognizable symbol of the General Electric company.
# Diagnostic Ultrasound Indications for Use Form
### GE Voluson ile with RSP6-16-RS Transducer
Intended Use: Diagnostic ultrasound Imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|----------------|------------------|-------------|---------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes | Harmonic Imaging | Coded Pulse | Other [Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[1] | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | [5,6] |
| Small Organ[2] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal<br>Conventional | P | P | P | | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | [5,6] |
| Other | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[6] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication: P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes ranal, GYN/Pelvic, Urology
[2] Small organ includes breast, thyroid, salvary gland, lymph nodes, pediative and neonatal palients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode
(6) Includes imaging of guldance of biopsy (20/3D/4D)
[7] Includes Infertility monitoring of follicle development
[8] Includes urology/prosiale
[*] Combined modes are B/M. B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OJR)
Prescription User (Per 21 CFR 801.109)
Page 5 of 17
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, circular design. The logo is black and white and has a vintage look.
#### Diagnostic Ultrasound Indications for Use Form
#### GE Voluson i/e with E8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[1] | P | P | P | | P | P | P | P | P | P | [6] | |
| Abdominal[1] | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [6] | |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [6] | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: . [1] Abdominal Includes renal, GYN/Pelvic, Urology
12) Small organ includes breast, testes. Ihyrold, salivary gland, lymph nodes, pediatic and naonatal patients
[3] Cardiac is Adult and Pediatric.
{5} 3D/4D Imaging Mode
(6) Includes imaging of guidance of biopsy (2D/3D/4D)
[7] Includes infertility monitoring of follicle development
(8) Includes urology/prostate
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE · CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of in Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 6 of 17
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows a black and white logo. The logo is circular and contains the letters 'GE' intertwined in a stylized, cursive font. The letters are surrounded by a decorative border, also in black and white, which gives the logo a vintage or classic appearance.
#### Diagnostic Ultrasound Indications for Use Form
#### GE Voluson i/e with 4C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of Interest | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] |
| Pediatric | P | P | P | | P | P | P | P | P | P | [6] |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [8] |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[4] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
N = new indication: P = previously deared by FDA; E = added under Appendlx E
Notes: [1] Abdominal Includes renal, GYN/Pelvic, Urology
(2) Small organ includes breast, lestes, thyrold, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac Is Adult and Pediatric.
[5] 3D/4D Imaging Mode
161 Includes imaging of guidance of biopsy (20/3D/4D)
[7] Includes infertility monitoring of follicle development
[8] Includes urology/prostate
['] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OJR)
Prescription User (Per 21 CFR 801.109)
Page 7 of 17
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows a logo with the letters 'GE' intertwined within a circular border. The letters are stylized and ornate, with flowing curves and decorative elements. The border surrounding the letters is also decorative, featuring a pattern of small, repeating shapes. The logo is black and white.
#### Diagnostic Ultrasound Indications for Use Form
### GE Voluson i/e with 12L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | | |
|----------------------------------|-------------------|---|---------|---------|---------|---------|---------|----------|----------|-------|---------|--|--|
| Clinical Application | ਉ | ખ | PW | CW | Color | Color M | Power | Combined | Harmonic | Coded | Other | | |
| Anatomy/Region of Interest | | | Doppler | Doppler | Doppler | Doppler | Ooppler | Modes | Imaging | Pulse | (Notes) | | |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal / Obsteirics 17 | | | | | | | | | | | | | |
| Abdominal 11 | | | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | b | (1 | | |
| Small Organ 21 | P | P | P | | P | P | P | P | P | P | ( €) | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiada | | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | ા | | |
| Musculo-skeletal<br>Conventional | p | P | P | | P | P | P | p | P | P | ા છે | | |
| Musculo-skelatal<br>Superficial | P | P | P | | P | P | P | P | D | P | ાં છો | | |
| Other | | | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectare) | | | | | | | | | | | | | |
| Transvagina! | | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
N = new Indication: P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology
(2) Small organ includes breast, thyroid, salivary gland, lymph nodes, pediatric and neonalal pattents [3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode
[6] Includes (maging of guidance of biopsy (2D/3D/4D)
[7] Includes infertility monitoring of follicle development
[8] Includes urology/prostate
["] Combined modes are BM, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 8 of 17
K131937
ટરૂ
{14}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
### GE Voluson i/e with RIC5-9W-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[1] | E | E | E | | E | E | E | E | E | E | [5,6] |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[4] | E | E | E | | E | E | E | E | E | E | [5,6] |
| Transvaginal | E | E | E | | E | E | E | E | E | E | [5,6] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N a new indication; P = previously cleared by FDA; E = added under Appendix E (previously cleared Voluson S6/SB K120741)
Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients |3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode
[6] Includes imaging of guidance of biopsy (2D/3D/4D)
[7] Includes infertifity monitoring of follicle development
[8] Includes urology/prostate
["] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 9 of 17
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows a black and white logo of General Electric (GE). The logo consists of the letters 'G' and 'E' intertwined in a stylized, circular design. The letters are surrounded by a decorative border, giving the logo a vintage or classic appearance.
#### Diagnostic Ultrasound Indications for Use Form
#### GE Voluson ife with IC5-9W-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics(7) | N | N | N | | N | N | N | N | N | N | [6] |
| Abdominal(1) | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ(2) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac(3) | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type. Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal(4) | N | N | N | | N | N | N | N | N | N | [6] |
| Transvaginal | N | N | N | | N | N | N | N | N | N | [6] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
N = new Indicallon; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology
[2] Small organ includes breast, testes, thyroid, salivery gland, lymph nodes, pediatic and neonatal patients
[…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.