K131916 · Healthstats International Pte. , Ltd. · DXN · Mar 11, 2014 · Cardiovascular
Device Facts
Record ID
K131916
Device Name
BPROSOFT CASP PC SOFTWARE
Applicant
Healthstats International Pte. , Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Mar 11, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The BPro® system is a noninvasive ambulatory central blood pressure monitoring system based on tonometry at the radial artery of the wrist. The system consists of a BPro® wrist-mounted tonometric monitor, and a BPCalibrator MC3000™ oscillometric blood pressure monitor for calibration, and BProSOft CASP® Software, a PC-based computer program. Before each measurement session, the wrist-mounted tonometric device is calibrated using the oscillometric monitor. Once calibrated, the BPro wrist device is capable of recording and displaying up to 96 measurements of systolic and diastolic blood pressure, pulse rate, and radial pulse waveform over a 24-hour period. The central aortic systolic pressure (CASP) is calculated utilizing the recorded radial pulse waveform and BProSOft CASP® PC software. BProSOft CASP® PC software is used to provide data to qualified medical personnel for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to 24-hour period. The BPro system is intended for use on patients who are eighteen (18) years and older and who have a palpable radial pulse. The BPro system is intended only for measurement, recording, and display. It makes no diagnosis. Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner.
Device Story
System uses wrist-mounted arterial tonometry to capture radial artery pulse waveforms over 24 hours. Requires pre-session calibration via BPCalibrator MC3000 oscillometric monitor. Data transferred to BProSOft CASP PC software; software performs arterial pulse waveform analysis to calculate central aortic systolic pressure (CASP). Used by clinicians to assess cardiac health via blood pressure profiles. Provides measurements, recordings, and displays; makes no automated diagnosis. Benefits include noninvasive 24-hour ambulatory monitoring of central blood pressure to assist in cardiovascular condition evaluation.
Clinical Evidence
Clinical study compared BProSOft CASP PC software with a listed predicate device for derivation of central aortic systolic pressure (CASP). Results confirmed accuracy of CASP measurements consistent with predicate device claims. Previous clinical validation for systolic/diastolic pressure and pulse rate measurements was leveraged from prior clearance (K060315).
Technological Characteristics
Wrist-mounted tonometric monitor with pressure transducer; embedded microcontroller for measurement scheduling and serial communication. PC-based software for CASP calculation via arterial pulse waveform analysis. Biocompatible skin-contact materials. Electrical safety, EMC, shock, and vibration testing performed per applicable standards.
Indications for Use
Indicated for patients 18 years and older with a palpable radial pulse for noninvasive ambulatory monitoring of systolic/diastolic blood pressure, pulse rate, and central aortic systolic pressure (CASP).
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
BPro Ambulatory Blood Pressure Monitoring with BProSOft application software (K060315)
A-PULSE CASP PC application software -- Central Aortic Systolic System (K072593)
Mobil-O-Graph 24h PWA with Hypertension Management Software Client Server (HMS-CS 4.3) (K110603)
{0}------------------------------------------------
MAR 1 1 2014
Page 1 of 4
510(k) Summary of Safety and Effectiveness in accordance with 21 CFR 807.92
- (a) Submitted by: HealthSTATS International Pte. Ltd. 6 New Industrial Road #04-01/02 Singapore 536199 Phone: +65-6858 3248 Fax: +65-6858 0148 E-mail: cmting@healthstats.com.sg Contact Person: Dr. Choon Meng TING, M.D. Position/Title: President/CEO Date of Preparation: June 20, 2013 HealthSTATS BPro® Ambulatory Blood Trade Name: Pressure Monitoring System with BProSOft CASP® PC software Common/Classification Name: System, Measurement, Blood-pressure, Central, Non-invasive Product Code: 74 DXN, 21 CFR & 870.1130 Class: Class II
- (3) Predicate Device(s):
| K060315 | BPro Ambulatory Blood Pressure Monitoring with BProSOft<br>application software |
|---------|-------------------------------------------------------------------------------------------|
| K072593 | A-PULSE CASP PC application software -- Central Aortic<br>Systolic System |
| K110603 | Mobil-O-Graph 24h PWA with Hypertension Management<br>Software Client Server (HMS-CS 4.3) |
Reason for Submission: New Device
#### (4) Description of Device:
HealthSTATS BPro® Ambulatory Blood Pressure Monitoring System with BProSOft CASP® PC software is a noninvasive ambulatory blood pressure monitoring system based on arterial tonometry at the radial artery of the wrist. The system consists of three main elements:
{1}------------------------------------------------
Page 2 of 4
- . BPro wrist-mounted tonometric monitor (listed device K060315),
- . BProSoft CASP report management software, a PC-based computer program (new/added measurement function).
- . BPCalibrator MC3000 oscillometric blood pressure monitor [listed device, K051546] for calibration of BPro monitors.
Once the BPro is calibrated using the BPcalibrator oscillometric device, it is capable of measuring and recording systolic pressure, diastolic pressure and pulse rate over a period of 24 hours. Arterial pulse waveforms obtained over the 24 hour period are used by the BProSoft CASP PC software to calculate central aortic systolic pressure (CASP).
#### (5) Intended use:
Blood pressure is measured in millimeters of mercury (mmHg) and is typically represented by two values: systolic pressure and diastolic pressure. The systolic pressure represents the pressure in the blood vessels when the heart contracts (pumps), while the diastolic pressure is the pressure when the heart relaxes and blood fills the heart. The calculation of central aortic systolic pressure (CASP) which is determined from arterial pulse waveform analysis can provide additional useful information on the condition of the heart.
Blood pressure fluctuates throughout the day. Specific ranges are associated with normal, hyper- and hypotension. Ambulatory blood pressure measurement devices provide useful 24 hour blood pressure profiles This submission includes the addition of ambulatory central aortic systolic pressure measurement. Evaluation of blood pressure profiles can assist with diagnosis of specific cardiovascular conditions.
## Indications for Use:
The BPro® system is a noninvasive ambulatory central blood pressure monitoring system based on tonometry at the radial artery of the wrist. The system consists of a BPro® wrist-mounted tonometric monitor, and a BPCalibrator MC3000™ oscillometric blood pressure monitor for calibration, and BProSOft CASP® Software, a PC-based computer program.
Before each measurement session, the wrist-mounted tonometric device is calibrated using the oscillometric monitor. Once calibrated, the BPro wrist device is capable of recording and displaying up to 96 measurements of systolic and diastolic blood pressure, pulse rate, and radial pulse waveform over a 24-hour period. The central aortic systolic pressure (CASP) is calculated utilizing the recorded radial pulse waveform and BProSOft CASP® PC software.
{2}------------------------------------------------
## Page 3 of 4
BProSOft CASP® PC software is used to provide data to qualified medical personnel for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to 24-hour period.
The BPro system is intended for use on patients who are eighteen (18) vears and older and who have a palpable radial pulse.
The BPro system is intended only for measurement, recording, and display. It makes no diagnosis.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner.
#### (6) Technological Characteristics:
The BPro Ambulatory Blood Pressure Monitoring System combines two established measurement methods used by the referenced predicate devices:
- . BPro ambulatory blood pressure monitoring at the wrist using arterial tonometry
- . Arterial pulse waveform analysis to determine the CASP.
The BPro monitor is applied to the wrist with a pressure transducer placed over the radial artery. The monitor has a size and weight similar to a sports type wrist watch.
An embedded microcontroller supervises the actions of the monitor including scheduled measurements, status reporting, and serial data communications with the host PC.
The BProSOft CASP® PC software utilizes arterial pulse waveform analysis to calculate ambulatory central aortic systolic pressure (CASP), which is in addition to the cleared blood pressure parameters of the BPro Ambulatory Blood Pressure Monitoring System.
#### (b) (1) Non-Clinical Tests Submitted:
The BPro monitor device meets applicable standards for medical device electrical safety, electromagnetic compatibility, shock and vibration, and environment (temperature and humidity).
Materials utilized in skin contact surfaces were reviewed for conformance with biocompatibility requirements. The materials met the requirements.
The BProSOft CASP PC software was verified to requirements and validated to meet intended use. System level risk, hazard, and failure
{3}------------------------------------------------
## Page 4 of 4
mode analysis was performed and residual risks were determined to be acceptable.
#### (2) Clinical Tests Submitted:
The BPro Ambulatory Blood Pressure Monitoring System has been previously clinically validated and cleared for the measurement of ambulatory systolic and diastolic pressure and pulse rate measurements.
The BProSOft CASP PC Software is an upgraded PC Software application that adds the CASP measurement function. BProSOft CASP PC Software has been clinically studied with the BPro monitor in comparison to a listed predicate device for the derivation of central aortic systolic pressure (CASP). Test results confirm the accuracy of the BProSOft CASP measurement consistent with the predicate device measurements within the stated accuracy claims.
#### (3) Conclusions from Tests:
As described in (b)(1) and (b)(2) above, the testing demonstrates that the HealthSTATS BPro® Ambulatory Blood Pressure Monitoring System with BProSOft CASP® PC software is as safe and effective as, and functions in a manner equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure embracing a globe, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
March 11, 2014
HealthSTATS International Pte. Ltd. c/o Mr. Stephen Gorski Imagenix. Inc. S65 W35739 Piper Road, Eagle, WI 53119
Re: K131916
> Trade/Device Names: HealthSTATS BPro® Ambulatory Blood Pressure Monitoring System with BProSOft CASP® PC Software Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: January 29, 2014 Received: January 31, 2014
Dear Mr. Gorski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Stephen Gorski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K131916
### HealthSTATS BPro® Ambulatory Blood Pressure Monitoring System Device Name: with BProSOft CASP® PC software
## Indications for use:
The BPro® system is a noninvasive ambulatory central blood pressure monitoring system based on tonometry at the radial artery of the wrist. The system consists of a BPro® wrist-mounted tonometric monitor, and a BPCalibrator MC3000™ oscillometric blood pressure monitor for calibration, and BProSOft CASP® Software, a PC-based computer program.
Before each measurement session, the wrist-mounted tonometric device is calibrated using the oscillometric monitor. Once calibrated, the BPro wrist device is capable of recording and displaying up to 96 measurements of systolic and diastolic blood pressure, pulse rate, and radial pulse waveform over a 24-hour period. The central aortic systolic pressure (CASP) is calculated utilizing the recorded radial pulse waveform and BProSOft CASP® PC software.
BProSOft CASP® PC software is used to provide data to qualified medical personnel for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to 24-hour period.
The BPro system is intended for use on patients who are eighteen (18) years and older and who have a palpable radial pulse.
The BPro system is intended only for measurement, recording, and display. It makes no diagnosis.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner
× Prescription Use (Part 21 CFR 801 Subpart D) AND / OR
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Conci 2014.03.11 00 -04'00' for Bram Zuckerman
Page 1 of _1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.