The EUM100Pro (Electro Uterine Monitor) is a transabdominal electromyography (EMG) monitor intended to non-invasively measure intrapartum uterine activity. The EUM100Pro acquires the signal from surface EMG electrodes placed on the patient abdomen. The EUM100Pro is intended for use on women in term (>36 completed weeks of gestation) labor, with singleton pregnancies. The EUM100Pro is intended for use by healthcare professionals in a clinical setting.
Device Story
EUM100Pro measures intrapartum uterine activity via transabdominal electromyography (EMG). System inputs: electrical signals from surface EMG electrodes placed on maternal abdomen. Components: multi-channel surface EMG sensor, 3D position sensor, and PC for data analysis/recording/GUI. Operation: system senses EMG activity, processes signals, and displays uterine contraction graphs similar to tocodynamometers. Used in clinical settings by trained medical personnel. Output: visual tracing of uterine activity on PC screen. Clinical utility: assists healthcare providers in monitoring labor progress; non-invasive alternative to tocodynamometry or intrauterine pressure catheters (IUPC).
Clinical Evidence
Clinical study enrolled 43 women at term in active labor. Simultaneous monitoring performed with EUM100Pro, tocodynamometry (TOCO), and intrauterine pressure catheter (IUPC). Three blinded physicians independently evaluated two sets of tracings per subject (first and second stage of labor). Physicians annotated contractions and assessed interpretability. Results demonstrated EUM100Pro performed substantially equivalently to predicate devices for detecting and displaying uterine activity.
Technological Characteristics
Transabdominal EMG sensing via surface electrodes. System includes multi-channel EMG sensor, 3D position sensor, and PC. Electrical safety per IEC 60601-1 and IEC 60601-2-40. Biocompatibility: electrodes are off-the-shelf (K990356). Power: mains power with laptop charger. Connectivity: PC-based system.
Indications for Use
Indicated for women at term (>36 weeks gestation) in labor with singleton pregnancies. Contraindicated for patients with implanted electronic devices, open wounds, or irritated skin.
Regulatory Classification
Identification
An external uterine contraction monitor (i.e., the tokodynamometer) is a device used to monitor the progress of labor. It measures the duration, frequency, and relative pressure of uterine contractions with a transducer strapped to the maternal abdomen. This generic type of device may include an external pressure transducer, support straps, and other patient and equipment supports.
{0}------------------------------------------------
| Date Prepared: | |
|----------------|--|
|----------------|--|
#### 21-Apr-14
#### APR 2 3 2014
OB-Tools Ltd. Mifaley HaEmek Industrial Zone Yozmot Building P.O Box 667 2310502 Migdal HaEmek, Israel Tel - +972-4-6040162 Fax - +972-4-6040114
| Official Contact: | Dr. Gal Ben-David<br>CEO | 510(k) number | Predicate<br>RRT SureCALL® | Predicate<br>Monica AN24 | Proposed device<br>EUM 100Pro |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|--------------------------------------|
| Proprietary or Trade Name: | EUM 100Pro (Electro Uterine Monitor) | K090145 | OSP | K112390 | OSP |
| Common/Usual Name: | External uterine contraction monitor | Procode | OSP | OSP | External uterine contraction monitor |
| Classification Name: | External uterine contraction monitor<br>OSP - CFR 884.2720<br>Class II | Name | External uterine contraction monitor | External uterine contraction monitor<br>External Fetal Heart Rate monitor | |
| Predicate Devices: | K090145 – RRT – SureCALL®<br>K112390 – Monica Healthcare – AN24 | CFR | 884.2740 | 884.2720 | 884.2720 |
| Indications for<br>Use | SureCALL® EMG labor Monitor is<br>a transabdominal electromyography<br>(EMG) monitor intended to measure<br>intrapartum uterine activity. | The Monica AN24 is an intrapartum<br>maternal-fetal monitor that non-invasively<br>measures and displays fetal heart rate<br>(FHR), uterine activity (UA), and maternal<br>heart rate (MHR). The AN24 acquires and<br>displays the FHR tracing from abdominal<br>surface electrodes that pick up the fetal<br>ECG (fECG) signal. Using the same<br>surface electrodes, the AN 24 also acquires<br>and displays the UA tracing from the<br>uterine electromyography (EMG) signal and<br>the MHR tracing from the maternal ECG<br>signal (mECG). | The EUM100Pro (Electro Uterine<br>Monitor) is a transabdominal<br>electromyography (EMG) monitor<br>intended to non-invasively measure<br>intrapartum uterine activity. The<br>EUM100Pro acquires the signal from<br>surface EMG electrodes placed on the<br>patient abdomen. | | |
| Patient population | It is intended for use on women who<br>are at term (>36 completed weeks),<br>in labor, with singleton pregnancies,<br>using surface electrodes on the<br>maternal abdomen | It is intended for use on women who are at<br>term (>36 completed weeks), in labor, with<br>singleton pregnancies, using surface<br>electrodes on the maternal abdomen | It is intended for use on women who are<br>at term (>36 completed weeks), in labor,<br>with singleton pregnancies, using surface<br>electrodes on the maternal abdomen | | |
| Prescriptive<br>Environments of<br>use | Trained medical personnel<br>Clinical settings | Trained medical personnel<br>Clinical settings | Trained medical personnel<br>Clinical settings | | |
| Power source | Mains power with laptop computer<br>charger | Battery | Mains power with isolation transformer | | |
#### Device Description:
The EUM100Pro System is designed to present and record the electrical activity of the uterus. The activity is shown as graphs similar to the commonly use toco-dynamometer.
The EUM100Pro is built around an EN- 60950 certified computer. Analog signals are obtained from CE certified amplifier box from Delsys Inc., Boston, MA. Delsys also supply the input modules, power supply, electrodes and disposable stickers.
The system is comprised of a multi-channel surface electromyogram operative to sense electromyographic (EMG) activity, a three-dimensional position sensor and a personal computer providing data analysis, recording media and a graphical user interface.
#### Indications for Use:
The EUM100Pro (Electro Uterine Monitor) is a transabdominal electromyography (EMG) monitor intended to non-invasively measure intrapartum uterine activity. The EUM100Pro acquires the signal from surface EMG electrodes placed on the patient abdomen.
The EUM100Pro is intended for use on women in term (>36 completed weeks of gestation) labor, with singleton pregnancies.
The EUM100Pro is intended for use by healthcare professionals in a clinical setting.
{1}------------------------------------------------
K131889
Page 2 of 5
.
510(k) Summary
.
omparison to Predicate
{2}------------------------------------------------
く131889
Page 3 of 5
## 510(k) Summary
| | | 1-Apr-14 | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|----------------------------------------|--------------------------------------------------------------------------------------|
| | | | roposed devic EUM 100Pro |
| | Predicate RRT SureCALL® | Predicate Monica AN24 | |
| | 6090145 | K112390 | |
| الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموق | EMG electrode<br>xternal surface l | xternal surface EMG electrode | xternal surface EMG electrod |
| | | | |
| | | | |
| | Graphica | Graphica | Graphica |
| | | | |
| | કર<br>urface electrod | urface electrode | urface electrod |
| iingle patient use Lisposable Contraindication Contraindication Ind Warnings | Yes | Yes | Yes |
| | | | |
| | None | None | atient with implanted electronic device pen wounds or irritated skin |
| | | | |
| | | | |
| afety Testing | ectrical safety | | |
| | | | |
| | | EC 60601-1 EC 60601-1-2 EC 60601-1-2-4 | Nergies to silve! EC 60601-1 EC 60601-11-2 EC 60601-11-2 EC 60601-11-2 EC 60601-2-40 |
| linical Testin | Comparison to Tocodynamometer Tocodynamometer UPC | omparison to | omparison to |
| | | | |
| | | Tocodynamomen IUPC_ | Tocodynamome IUPC |
| | | | |
# Substantial Equivalence Discussi
he EUM 100Pro is viewed as substantially equivalent to the predicate devices becaus
- Indications =
{3}------------------------------------------------
#### 510(k) Summary Page 4 of 5 21-Apr-14
Discussion - This is identical to the predicates, except that the EUM 100Pro does not monitor additional features, i.e., FHR, MHR as does K112390 -- Monica AN24, nor does it connect to other traditional sensors which can be displayed on its screen as does K090145 -- RRT SureCALL® for HR, FHR, IUPC, or TOCO sensors.
#### Patient Population -
- It is intended for use on women who are at term (>36 completed weeks), in labor, with . singleton pregnancies, using surface electrodes on the maternal abdomen Discussion - The patient population is identical to the predicates - K 1 12390 - Monica AN24 and K090145 - RRT SureCALL®
#### Environment of Use -
- For use in clinical settings by trained medical personnel . Discussion -- The environments of use and personal are identical to the predicates -K 112390 - Monica AN24 and K090145 - RRT SureCALL®
#### Technology -
- The use of transabdominal electromyography (EMG) signals to sense uterine activity via . an array of surface electrodes placed on the maternal abdomen.
- Discussion This technology is identical to the predicates K112390 Monica AN24 and K090145 - RRT SureCALL®
#### Non-clinical Testing Summary -
We have performed a number of tests appropriate for the proposed device. These tests include:
#### Biocompatibility of Materials -
- The only materials in contact with the patient are the EMG electrodes which are off-the-. shelf (K990356).
Discussion - The EMG electrodes have been cleared for the intended use under K990356.
#### Electrical, EMC. EMI testing -
- We have evaluated the proposed device per IEC 60601-1 and IEC 60601-2-40 and the . device passed the requirements.
Discussion - The proposed device met the requirements of the standards.
#### Clinical Testing Summary -
The sponsor conducted a clinical study enrolling 43 women at term gestation in active labor who agreed to simultaneous uterine activity monitoring with the EUM 100Pro, tocodynamometry (TOCO), and intrauterine pressure catheter (IUPC). For each study subject, two sets of tracings (one during the first stage and the other during the second stage of labor) consisting of EUM, TOCO, and IUPC were evaluated independently by three physicians who were masked to the technology used to acquire the tracings. Physicians were asked to annotate the tracings above
{4}------------------------------------------------
#### 510(k) Summary Page 5 of 5 21-Apr-14
each deflection above baseline they considered to be a contraction. They were also asked to indicate the total number of uterine contractions on the tracing as well as to comment on the interpretability of the tracing. The study provided evidence that the EUM 100Pro performed substantially equivalently to the legally marketed predicate devices for detecting and displaying uterine activity in women at term gestation in active labor.
#### Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing a globe or sphere, representing the department's mission to protect and promote the health of all Americans.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 23, 2014
OB Tools Ltd. % Paul E. Dryden President ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, FL 34134
Re: K131889
> Trade/Device Name: EUM100Pro (Electro Uterine Monitor) Regulation Number: 21 CFR§ 884.2720 Regulation Name: External uterine contraction monitor and accessories Regulatory Class: II Product Code: OSP Dated: March 20, 2014 Received: March 27, 2014
Dear Paul E. Dryden,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be may or subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{6}------------------------------------------------
Page 2 - Paul E. Dryden
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/6/Picture/6 description: The image shows the name "Herbert P. Lerner -S" in a bold, sans-serif font. The name is written in black ink on a white background. The letters are evenly spaced and the name is easy to read. There is some noise around the "P. Lerner" portion of the name.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K131889
Device Name EUM100Pro (Electro Uterine Monitor)
Indications for Use (Describe)
The EUM100Pro (Electro Uterine Monitor) is a transabdominal electromyography (EMG) monitor intended to non-invasively measure intrapartum uterine activity. The EUM100Pro acquires the signal from surface EMG electrodes placed on the patient abdomen.
The EUM100Pro is intended for use on women in term (>36 completed weeks of gestation) labor, with singleton pregnancies.
The EUM100Pro is intended for use by healthcare professionals in a clinical setting.
Type of Use (Select one or both, as applicable)
ﺗ
> Prescription Use (Part 21 CFR 801 Subpart D)
[] Over-The-Counter Use (21 CFR 801 Subpart C)
.
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
... Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
### Herbert P. Lerner 2014.04.23 10:
FORM FDA 3881 (9/13)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.