K131866 · Hollender Sustainable Brands, LLC · HIS · May 1, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K131866
Device Name
SUSTAIN
Applicant
Hollender Sustainable Brands, LLC
Product Code
HIS · Obstetrics/Gynecology
Decision Date
May 1, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
The Sustain Tailored Fit, Ultra Thin and Comfort Fit condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
Device Story
SUSTAIN Natural Rubber Latex Condoms are male condoms designed for contraception and prophylaxis against sexually transmitted infections. The device consists of a natural rubber latex sheath, available in three variants: Tailored Fit (straight-walled), Ultra Thin (straight-walled), and Comfort Fit (bulbous-shaped). All variants are transparent, teat-ended, and lubricated with silicone oil. The device is intended for over-the-counter use by individuals. It functions as a physical barrier to prevent the exchange of bodily fluids during intercourse. Safety and effectiveness are established through adherence to established material and performance standards, ensuring the device provides a reliable barrier for the user.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including compliance with ASTM D3492-08 (Standard Specification for Rubber Contraceptives) and ISO 4074 (Natural Rubber Latex Condoms Requirements and Test Methods), as well as biocompatibility testing per ISO 10993 (cytotoxicity, irritation, sensitization, and systemic toxicity).
Technological Characteristics
Materials: Natural rubber latex. Lubricant: Silicone oil. Additives: Magnesium carbonate and calcium carbonate dusting powder. Design: Cylindrical, straight-walled or bulbous-shaped, teat-ended. Dimensions: Length 186±4mm; Width 49-52±2mm; Thickness 0.055-0.065±0.005mm. Standards: ASTM D3492-08, ISO 4074, ISO 10993. Connectivity: None. Energy source: None. Sterilization: Not specified.
Indications for Use
Indicated for use by individuals for contraception and the prevention of pregnancy and sexually transmitted infections.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Durex Colors and Scents, Latex Condoms with Lubricant and Fragrance Additives (K980174)
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(k) SUMMARY
| Submitted by: | Hollender Sustainable Brands LLC<br>212 Battery Street<br>Burlington, VT 05401 | | 807.92(a)(1) |
|----------------------|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------|--------------|
| Contact Person: | Jeffrey Hollender, CEO<br>Telephone:<br>Fax:<br>Email: | 802-846-3443<br>802-863-4724<br>jeffrey@jeffreyhollender.com | |
| Date Prepared: | April 21, 2014 | | |
| Proprietary Name: | SUSTAIN Natural Rubber Latex Condoms:Tailored Fit,<br>Ultra Thin, and Comfort Fit<br>807.92(a)(2) | | |
| Common Name: | Male Latex Condoms | | |
| Classification Name: | Condom (21 CFR §884.5300) Code HIS | | |
| Predicate Devices: | | 807.92(a)(3) |
|-------------------------------------|----------------------------------------------------------------------------------------|---------------|
| Manufacturer | Brand Name | 510(k) Number |
| London International LLC | Ultra Comfort<br>Latex,Rubber Condoms | K980319 |
| London International US<br>Holdings | Durex Colors and Scents,<br>Latex Condoms with<br>Lubricant and Fragrance<br>Additives | K980174 |
#### Description of the Device:
807.92(a)(4)
The SUSTAIN Natural Rubber Latex Condoms are made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. These condoms conform to ISO 4074 Natural Rubber Latex Condoms - Requirements and Test Methods, ASTM D3492-08 Standard Specification for Rubber Contraceptives (Male Condom), and ISO 10993 Biological Evaluation of Medical Devices.
#### Product Smooth Surface - Sustain Tailored Fit
The Sustain Tailored Fit Natural Rubber Latex Condoms are parallel straight walled teat ended with smooth surface, transparent, having a nominal length of 186 ±4 mm, a nominal width of 49±2mm and a nominal thickness of 0.065 ± 0.005mm.
#### Product Smooth Surface - Sustain Ultra Thin
The Sustain Ultra Thin Natural Rubber Latex Condoms are parallel straight walled teat ended with smooth surface, transparent, having a nominal length of 186 ±4 mm, a nominal width of 52±2 mm and a nominal thickness of 0.055 ± 0.005mm.
#### Product Smooth Surface - Bulbous - Sustain Comfort Fit
The condoms are bulbous shaped teat ended with smooth surface, transparent, having a nominal length of 186 ±4mm, a nominal width of 52±2mm and a nominal thickness of 0.065 ± 0.005mm.
{1}------------------------------------------------
#### Intended Use of the Device:
## 807.92(a)(5)
The Sustain Tailored Fit, Ultra Thin and Comfort Fit condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
#### Predicate Product Comparison
## 807.92(a)(6)
| Table I | Predicate Product Comparisons - Sustain Tailored Fit Natural Rubber Latex Condom | | |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Parameters | Hollender Sustainable | London International | Similarities and |
| | Brands LLC | LLC | Differences |
| 510(k) Number | K 131866 | K980174 | |
| Intended Use | The Sustain Tailored Fit<br>Natural Rubber Latex<br>Condomsare used for<br>contraception and for<br>prophylactic purposes (to<br>help prevent pregnancy and<br>the transmission of sexually<br>transmitted infections). | A Natural Rubber Latex<br>Condom is intended to be<br>worn over the male penis<br>during acts of vaginal<br>intercourse to aid in the<br>prevention of pregnancy<br>and spread of sexually<br>transmitted infections. | Equivalent |
| Straight walled teat<br>ended | Yes | Yes | Identical |
| Nominal length | 186 ±4mm | 180±10mm | Both meet<br>ASTM D3492-<br>08 length<br>specifications |
| Nominal width | 49±2mm | 52±2mm | Both meet<br>ASTM D3492-<br>08 width<br>specifications |
| Nominal thickness | 0.065 ± 0.005mm | 0.060 - 0.068mm | Both meet<br>ASTM D3492-<br>08 thickness<br>specifications |
| Smooth surface | Yes | Yes | Identical |
| Magnesium carbonate<br>and Calcium carbonate<br>dusting powder | Yes | Yes | Identical |
| Lubricated with Silicone<br>base | Lubricated with Silicone Oil | Lubricated with Silicone<br>Base Zeus Odor Masker<br>and flavors | Predicate uses<br>the identical<br>silicone oil<br>base and<br>incorporates an<br>odor masker<br>and flavors into<br>the lubricant |
| Transparent | Yes | Colored | Predicate offers<br>different<br>flavors and |
{2}------------------------------------------------
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AND CHANAL CHARACT LAND | A LED OR AND LEASE I FAR IN AND LINER FRANCE CHANAL CONSULTER OF LEASE | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
Technological Characteristics Sustain Tailored Fit Natural Rubber Latex Condom 807.92(a)(6)
The Sustain Tailored Fit Natural Rubber Latex Condom and the predicate device (K980174) are made from natural rubber latex, with biocompatible materials and designs in conformance with the requirements of ASTM D 3492-08 Standard Specification for Rubber Contraceptives (Male Condom). Both are cylindrical, straight-walled, teat ended, and use silicone oil as a lubricant. The Sustain Tailored Fit Natural Rubber Latex Condom is transparent and has no flavors or colors, whereas the predicate device (K980174) has flavors and colors. The absence of flavors and colors has no material impact on safety and effectiveness.
| Parameters | Hollender Sustainable<br>Brands LLC | London International LLC | Similarities<br>and<br>Differences |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K131866 | K980319 | |
| Intended Use | The Sustain Ultra Thin<br>Natural Rubber Latex<br>Condomsare used for<br>contraception and for<br>prophylactic purposes (to<br>help prevent pregnancy and<br>the transmission of sexually<br>transmitted infections). | A Natural Rubber Latex<br>Condom is intended to be<br>worn over the male penis<br>during acts of vaginal<br>intercourse to aid in the<br>prevention of pregnancy<br>and spread of sexually<br>transmitted infections. | Similar |
| Parallel teat ended | Yes | Bulbous teat ended | Both meet<br>ASTM D3492-<br>08<br>specifications |
| Nominal length | 186±4mm | 200±10mm | Both meet<br>ASTM D3492-<br>08 length<br>specifications |
| Nominal width | 52±2mm | 56±2mm | Both meet<br>ASTM D3492-<br>08 width<br>specifications |
| Nominal thickness | 0.055±0.005 mm | 0.055±0.005 mm | Identical |
| Smooth surface | Yes | Yes | Identical |
| Magnesium carbonate<br>and Calcium carbonate<br>dusting powder | Yes | Yes | Identical |
| Lubricated with Silicone<br>base | Lubricated with Silicone Oil | Lubricated with Silicone<br>Base Zeus Odor Masker | Predicate uses<br>the identical<br>silicone oil<br>base and<br>incorporates an<br>odor masker<br>into the<br>lubricant |
Predicate Product Comparison Sustain Ultra Thin Natural Rubber Latex Condom Table 2
{3}------------------------------------------------
| | | . | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| POMONO POM<br>. | <br>State for Jan 2017<br>1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1<br>1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1 |
| | | . | |
Technological Characteristics - Sustain Ultra Thin Natural Rubber Latex Condom 807.92(a)(6)
The Sustain Ultra Thin Natural Rubber Latex Condom and the predicate device (K980319) are made from natural rubber latex, with biocompatible materials and designs in conformance with the requirements of ASTM D 3492-08 Standard Specification for Rubber Contraceptives (Male Condom). Both are smooth surface, use identical dusting powder, use silicone oil as a lubricant, and are transparent. The predicate offers a bulbous teat ended feature and the Sustain Ultra Thin Natural Rubber Latex Condom offers a teat ended feature. The difference between the two features has no material impact on the safety and effectiveness as both condoms meet ASTM D 3492-08 Standard Specification for Rubber Contraceptives (Male Condom) recommendations. The differences in the nominal length and nominal width has no material impact on safety and effectiveness as both condoms fall within ASTM D 3492-08 Standard Specification for Rubber Contraceptives (Male Condom) length and width specifications.
| Sustain Comfort Fit<br>Natural Rubber latex<br>Condom Parameters | Hollender Sustainable<br>Brands LLC | London International<br>LLC | Similarities and<br>Differences |
|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K131866 | K980319 | |
| Intended Use | The Sustain Comfort Fit<br>Natural Rubber Latex<br>Condomsare used for<br>contraception and for<br>prophylactic purposes (to<br>help prevent pregnancy and<br>the transmission of sexually<br>transmitted infections). | A Natural Rubber Latex<br>Condom is intended to be<br>worn over the male penis<br>during acts of vaginal<br>intercourse to aid in the<br>prevention of pregnancy<br>and spread of sexually<br>transmitted infections. | Identical |
| Bulbous shaped teat<br>ended | Yes | Yes | Identical |
| Nominal length | $186\pm4mm$ | Length $200\pm10mm$ | Both meet<br>ASTM D3492-<br>08 length<br>specifications |
| Width | The closed end of the<br>condom has a Bulbous with<br>slightly increased width of<br>$57mm - 59mm$<br>Nominal width of $52\pm2mm$ | The closed end of the<br>condom has a Bulbous<br>with slightly increased<br>width of $59mm-61mm$<br>Nominal width of<br>$56\pm2mm$ | Nominal width<br>meets ASTM<br>D3492-08<br>specifications.<br>Bulbous width<br>is not a safety<br>concern. |
| Average thickness | $0.065 \pm 0.005mm$ | $0.055 \pm 0.005mm$ | Both meet<br>ASTM D3492-<br>08 thickness<br>specifications |
| Smooth surface | SmoothSurface | Smooth Surface | Identical |
| Magnesium carbonate<br>and Calcium carbonate<br>dusting powder | Yes | Yes | Identical |
Prodicate Product Comparison Sustain Comfort Fit Noturel Rubber Lotox Condom Tahle 3
{4}------------------------------------------------
| Lubricated with Silicone<br>base | Lubricated with Silicone Oil | Lubricated with Silicone<br>Base Zeus Odor Masker | Predicate uses<br>the identical<br>silicone oil but<br>incorporates an<br>odor masker<br>into the<br>lubricant |
|----------------------------------|------------------------------|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Transparent | Yes | Yes | Identical |
#### Technological Characteristics Sustain Comfort Fit Natural Rubber latex Condom
The Sustain Comfort Fit Natural Rubber Latex Condom and the predicate device (K980319) are made from natural rubber latex, with biocompatible materials and designs in conformance with the requirements of ASTM D 3492-08 Standard Specification for Rubber Contraceptives (Male Condom). Both are cylindrical, bulbous shaped teat ended, smooth surface; use identical dusting powder, use silicone oil as a lubricant, and are transparent. The differences in the nominal length, nominal width and nominal thickness has no material impact on safety and effectiveness as both condoms fall within ASTM D 3492-08 Standard Specification for Rubber Contraceptives (Male Condom) length, width and thickness specifications.
### Nonclinical Testing
#### 807.92(b)
The following tests were performed on the Natural Rubber Latex Condoms:
- ASTM D3492-08 Standard Specifications for Rubber Contraceptives (Male Condoms) ●
- ISO 4074:2002 Natural Rubber Latex Condoms Requirements and Test Methods ●
- ISO 10993-5 [2009] Biological Evaluation of Medical Devices Part 5: Tests for In Vitro ● Cytotoxicity
- ISO 10993-10 [2010] Biological Evaluation of Medical Devices Part 10: Tests for . Irritation and Skin Sensitization (Penile Irritation)
- . ISO 10993-10 [2010] Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization (Primary Irritation)
- . ISO 10993-10 [2010] Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization (Vaginal Irritation)
- . ISO 10993-10 [2010] Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization (Skin Sensitization Study (GPMT) of Refined Cotton Seed Oil Extract in Guinea Pigs)
- . ISO 10993-10 [2010] Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization(Skin Sensitization Study (GPMT) of Normal Saline Extract in Guinea Pigs)
- ISO 10993-11 [2009] Biological Evaluation of Medical Devices - Part 11: Tests for ● Systemic Toxicity
### Clinical Testing
No clinical testing was performed.
{5}------------------------------------------------
#### Conclusion
## 807.92(b)(3)
The Sustain Tailored Fit, Ultra Thin and Comfort Fit Natural Rubber Latex Condoms are substantially equivalent to the predicate device in indications for use, materials and design. Safety and performance testing to ASTM D 3492-08 Standard Specifications for Rubber Contraceptives (Male Condoms)and ISO 10993 Biological Evaluation of Medical Devices has concluded that the Sustain Tailored Fit, Ultra Thin and Comfort Fit Natural Rubber Latex Condoms are substantially equivalent to the predicate devices (K980174 and K980319) in terms of safety and effectiveness.
{6}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 15, 2014
Hollender Sustainable Brands, LLC % E.J. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, MD 21114
Re: K131866
> Trade/Device Name: SUSTAIN Tailored Fit, Ultra Thin and Comfort Fit Natural Rubber Latex Condoms Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated (Date on orig SE Itr): April 4, 2014 Received (Date on orig SE Itr): April 4, 2014
Dear E.J. Smith.
This letter corrects our substantially equivalent letter of May 1, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - E.J. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K131866
#### Device Name
SUSTAIN Tailored Fit, Ultra Thin and Comfort Fit Natural Rubber Latex Condoms
Indications for Use (Describe)
The Sustain Tailored Fit, Ultra Thin and Comfort Fit condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.