U2 TIBIAL BASEPLATE- CMA TYPE, TIBIAL INSERT AND AUGMENT
K131864 · United Orthopedic Corporation · JWH · Jan 15, 2014 · Orthopedic
Device Facts
Record ID
K131864
Device Name
U2 TIBIAL BASEPLATE- CMA TYPE, TIBIAL INSERT AND AUGMENT
Applicant
United Orthopedic Corporation
Product Code
JWH · Orthopedic
Decision Date
Jan 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion contraction. This device may also be indicated in the salvage or previously failed surgical attempts or for knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. This device system is designed for cemented use only.
Device Story
U2 Tibial Baseplate-CMA, Tibial Insert, and Augment are components for total knee arthroplasty; designed for primary or revision surgery requiring augmentation or stem extensions due to inadequate bone stock. Baseplate manufactured from titanium alloy; inserts from UHMWPE; augments from Ti-6Al-4V alloy. System provides articulating bearing surface fixed in metal tibial baseplate. Used in clinical/surgical settings by orthopedic surgeons. Device replaces bearing/articulating surfaces of proximal tibia to restore joint function and stability. Output is physical structural support for knee joint; aids in pain reduction and improved mobility. System is for cemented use only.
Clinical Evidence
Bench testing only. Performance evaluated via mechanical testing: locking strength, contact area/pressure, wear simulation, range of motion, and fatigue testing of tibial baseplate and tibial insert spine. No clinical data presented.
Indicated for skeletally mature patients requiring knee arthroplasty for pain relief/function improvement due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, pseudogout, or posttraumatic joint loss. Applicable for cases with patellofemoral erosion, dysfunction, prior patellectomy, or moderate valgus/varus/flexion contraction. Indicated for salvage of failed surgeries or knees lacking flexion stability. Cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
### (9) U2 Tibial Baseplate- CMA type, Tibial Insert and Augment 510(k) Summary
# 510(k) Summary of Safety and Effectiveness
JAN 1 5 2014
| Submitted by: | United Orthopedic Corporation |
|-----------------------|-----------------------------------------------------------------------------------------------------|
| Address: | No 57, Park Ave 2, Science Park, Hsinchu 300, Taiwan |
| Phone Number: | +886-3-5773351 ext. 2212 |
| Fax Number: | +886-3-577156 |
| Date of Summary: | June 21, 2013 |
| Contact Person | Fang-Yuan Ho<br>Regulation and Document Management |
| Proprietary Name: | U2 Tibial Baseplate- CMA type, Tibial Insert and Augment |
| Common Name: | Semi-constrained total knee prostheses |
| Device Classification | Knee joint patellofemorotibial polymer/metal/polymer |
| Name and Reference: | semi-constrained cemented prosthesis per 21CFR 888.3560.<br>This falls under the Orthopedics panel. |
| Device Class | Class II |
| Panel Code | Orthopaedics Device |
| Device Product Code: | JWH |
| Predicate Device: | 1. "UNITED" U2 Total Knee System (K051640) |
| | 2. "UNITED" U2 XPE Total Knee System (K103733) |
| | 3. "UNITED" U2 Total Knee System-PSA Type (K082424) |
| | 4. "UNITED" Augment and Screw, PSA Type (K122183) |
| | 5. "ZIMMER" NexGen Complete Knee Solution L.P.S. Flex |
## Device Description:
The subjected device is designed for the replacement of the bearing and/or articulating surfaces of the proximal tibia composed of an articulating bearing surface fixed in a metal tibial baseplate, and is used for patients in primary or revision surgery who require
Fixed Bearing Knee (K991581)
.UOC-FDA-027
Page: 7-1/3
{1}------------------------------------------------
### (9) U2 Tibial Baseplate- CMA type, Tibial Insert and Augment 510(k) Summary
augmentation and/or stem extensions due to inadequate bone stock. This premarket notification includes the following components: U2 Tibial Baseplate-CMA type, #7 Tibial Insert (CR/PS), #7 XPE Tibial Insert (CR/PS) and #7 Tibial Augment.
U2 Tibial Baseplate-CMA (Cemented Modular Augmentation) is manufactured from titanium alloy forging (ASTM F620) which is forged by titanium alloy bars (ASTM F136) with seven proportional sizes (#1 to #7). Tibial Insert and XPE insert, including PS type and CR type, is an additional size extension to the previously cleared "UNITED" U2 Total Knee System (K051640, K103733). It is manufactured from UHMWPE (ASTM F648/ISO 5834) with five available thicknesses (9, 11, 13, 15, and 18 mm). Tibial Augment is an additional size extension to the previously cleared "UNITED" U2 Total Knee System-PSA Type (K082424, K122183). It is manufactured from Ti-6Al-4V alloy (ASTM F136/ISO 5832-3) with three available thicknesses (5, 10, 15mm).
### Indications:
This device is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion contraction. This device may also be indicated in the salvage or previously failed surgical attempts or for knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. This device system is designed for cemented use only.
### Basis for Substantial Equivalence:
#### Tibial Baseplate- CMA a.
The Tibial Baseplate- CMA has the same basic design, intended use, materials and the same manufacturing method as the tibial baseplate of the "United" U2 Total Knee System
{2}------------------------------------------------
### (9) U2 Tibial Baseplate- CMA type, Tibial Insert and Augment 510(k) Summary
and "United" U2 Total Knee System-PSA. Like the "United" U2 Total Knee System-PSA, the Tibial Baseplate- CMA can be used with extension stem and augments. Besides, the Tibial Baseplate- CMA has the same basic design, intended use, material as the tibial plate of "Zimmer" NexGen Complete Knee Solution L.P.S. Flex Fixed Knee.
#### Tibial Insert b.
Tibial Insert, including PS type and CR type, is an additional size extension of tibial insert of the previously cleared "UNITED" U2 Total Knee System (K051640), while the subjected XPE insert is an additional size extension of the previously cleared "UNITED" U2 XPE Total Knee System (K103733). The materials, design, manufacturing process of the tibial insert and XPE insert are identical to the cleared tibial inserts of "UNITED" U2 Total Knee System and "UNITED" U2 XPE Tibial Inset, respectively.
#### Tibial Augment c.
Tibial augment—screw locking is an additional size extension to the previously cleared "UNITED" U2 Total Knee System - PSA Type (K082424, K122183). The materials, design, manufacturing process of this device are identical to the cleared tibial inserts of "UNITED" U2 Total Knee System-PAS type.
### Performance Data:
Tests as follows were conducted to evaluate the safety and effectiveness of the subjected device, and the test results demonstrated that this device is safe and effective.
- Locking Strength of Tibial Insert a.
- Contact Area and Contact Pressure b.
- Wear Simulation C.
- d. Range of Motion
- e. Fatigue Test of Tibial Baseplate
- f. Fatigue Test of Tibial Insert Spine
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2014
United Orthopedic Corporation Fang-Yuan Ho Regulatory Affairs Manager No. 57, Park Avenue 2, Science Park Hsinchu, 300 TAIWAN
Re: K131864
Trade/Device Name: U2 Tibial Baseplate-CMA type, Tibial Insert and Augment Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: December 5, 2013 Received: December 11, 2013
Dear Fang-Yuan Ho:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
### Page 2 -- Fang-Yuan Ho
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also.plcase note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincercly yours.
# Ronald和剑ean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
K131864 510 (k) Number (if known):
### Device Name: U2 Tibial Baseplate- CMA type, Tibial Insert and Augment
### Indications for Use:
This device is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion contraction. This device may also be indicated in the salvage or previously failed surgical attempts or for knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. This device system is designed for cemented use only.
Prescription Use ___x Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabeth | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
Page 1 of 1 Page: 6-1/1
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.