K131838 · Beijing Honkon Technologies Co., Ltd. · GEX · Apr 24, 2014 · General, Plastic Surgery
Device Facts
Record ID
K131838
Device Name
DIODE LASER FOR HIAR REMOVAL
Applicant
Beijing Honkon Technologies Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The diode laser for hair removal is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
Diode laser system for hair removal; utilizes selective photothermolysis to heat melanin and hair follicles while sparing surrounding tissue. Device components include semiconductor solid-state lasers, switching power supply, treatment handle, and contact cooling system. Operated by clinicians in professional settings. Input parameters include wavelength (808nm), pulse duration (5-300ms), frequency (1-10Hz), and fluence (<120 J/cm²). Output is laser energy delivered via sapphire-tipped handle to axilla, facial, and neck regions. Clinical benefit is long-term, stable reduction in hair regrowth.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with electrical and radiation safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, 21 CFR 1040) was used to establish substantial equivalence.
Technological Characteristics
Diode laser system; 808nm wavelength; sapphire contact cooling; output power 30W; fluence <120 J/cm²; pulse duration 5-300ms; frequency 1-10Hz. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), and IEC 60825-1. Standalone device.
Indications for Use
Indicated for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# K131838
| Image: HONKON logo and Beijing company name | | |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| | 510(k) Premarket Notification Submission | |
| | 510(k) Summary | APR 2 4 2014 |
| Date:<br>Submitter: | In accordance with 21 CFR 807.92 the following summary of information is provided:<br>18 June, 2013<br>Beijing Honkon Technologies Co., Ltd.<br>Address: No.3 Building, No.11 Yard, Kangding Street, BDA,<br>100176, Beijing, P.R.China | |
| Primary Contact Person: | Mike Gu<br>Regulatory Affairs Manager<br>OSMUNDA Medical Device Consulting Co., Ltd.<br>Tel: (+86) 20 6232 1333<br>Fax: (+86) 208633 0253 | |
| Secondary Contact Person: | Li Zhao<br>Management Representative<br>Beijing Honkon Technologies Co., Ltd.<br>Tel: (+86) 10-56370050<br>Fax: (+86) 10-56370076-610 | |
| Trade Name:<br>Common/Usual Name:<br>Classification Names: | Diode Laser for Hair Removal<br>Diode Laser<br>Powered Laser Surgical Instrument | |
| Regulation number | 878.4810 | |
| Product Code: | GEX | |
| Predicate Device(s):<br>Device Description: | K112031<br>The diode laser for hair removal is generally included components such as semiconductor solid-state lasers, switching power supply, laser power, treatment handle, key switch, cooling system and accessories. Its models are 808CL, 808CM, 808CH, 808EH, 808BL, 808BM, 808BH, 808FH, 808HGG and 808DH. Theirs differences are between the size of device which is | |
described in the table below:
| Model | Dimension (width* length* high) mm | |
|-------|------------------------------------|---------------|
| | Before package | After package |
| 808CL | 815*530*1420 | 895*590*1510 |
| 808CM | 790*540*1300 | 870*600*2200 |
| 808CH | 775*510*1395- | 855*570*1485 |
| 808EH | 810*530*1060 | 890*570*1960 |
| 808BL | 810*550*1300 | 890*610*1390 |
| 808BM | 820*590*1210 | 900*650*1300 |
| 808BH | 820*590*1280 | 900*650*1370 |
| 808DH | 780*620*1110 | 860*680*1200 |
Confidential and Privileged. This document confidential and privileged trade secrets and other information of Beijing Honkon Technologies Co., Ltd. and as such may not be disclosed to others not employed by Beijing Honkon Technologies Co., Ltd. All rights reserved
{1}------------------------------------------------
**HONKON**
气安星激光技术有限公司
510(k) Premarket Notification Submission
Proposed device
Technologies Co., Ltd
808CH, 808EH, 808BL,
808FH, 808HGG
Honkon
808CM,
808BH,
-
Modified
Treatment
Module
Beijing
808CL,
808BM,
| 808FH | 785*620*1180 | arrange a datt - a sua man a manufacturer and any final possible of the province<br>865*680*1270 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|--------------------------------------------------------------------------------------------------|
| 808HGG<br>The first from the may in the county of the county of the county of the county of the county of the county of the county of the county of the county of the county of the coun | 780*630*1290 | 860*690*1380 |
| Intended Use: | The Honkon's diode laser for hair removal is intended for hair<br>removal, permanent hair reduction on all skin types (Fitzpatrick<br>skin type I-VI).<br><br>Permanent hair reduction is defined as the long-term, stable<br>reduction in the number of hairs regrowing when measured at 6,<br>9 and 12 months after the completion of a treatment regime |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology: | The principle of laser hair removal is selective photothermolysis, the<br>matching of a specific wavelength of light and pulse duration to obtain<br>optimal effect on a targeted tissue with minimal effect on surrounding<br>tissue. Laser can cause localized damage by selectively heating melanin<br>and follicle while not heating the rest of the skin. |
| Determination of | Specification | Predicate Device |
|--------------------------|---------------|-------------------|
| Substantial Equivalence: | | |
| | K number | K112031 |
| | Manufacturer | Alma Lasers, Inc. |
Model
.
Confidential and Privileged. This document confidential and privileged trade secrets and other information of Beijing Honkon Technologies Co., Ltd. and as such may not be disclosed to others not employed by Beijing Honkon Technologics Co., Ltd. All rights reserved
Alma
Diode
with
Mode for
Lasers
HR
Laser
{2}------------------------------------------------
510(k) Premarket Notification Submission
| Intended Use | | |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The HR Mode is<br>intended for hair<br>removal, permanent hair<br>reduction. (Fitzpatrick<br>skin types I-VI), including<br>tanned skin. | The diode laser for<br>hair removal is<br>intended for hair<br>removal, permanent<br>hair reduction on all<br>skin types (Fitzpatrick<br>skin type I-VI).<br>Permanent hair<br>reduction is defined<br>as the long-term,<br>stable reduction in<br>the number of hairs<br>regrowing when<br>measured at 6, 9 and<br>12 months after the<br>completion of a<br>treatment regime |
| Wavelength<br>(nm) | 810 | 808 |
| Output power<br>(W) | 30 | 30 |
| Fluence (Energy<br>Density) | <120 J/cm² | <120 J/cm² |
| Pulse Duration | 5-200ms | 5-300 ms |
| Frequency | Up to 10Hz | 1-10Hz |
| Spot Size | 12cm² | 12mm² |
| Material | Sapphire | Sapphire |
| Cooling | Contact cooling | Contact cooling |
| Anatomical Sites | Axilla, Facial, Neck | Axilla, Facial, Neck |
| Electrical Safety | Comply with IEC 60601-<br>1 and IEC 60601-1-2 | Comply with IEC<br>60601-1 and IEC<br>60601-1-2 |
| Radiation Safety | Comply with IEC 60601-<br>2-22, IEC 60825-1 and 21<br>CFR 1040<br>PERFORMANCE<br>STANDARDS FOR LIGHT-<br>EMITTING PRODUCTS | Comply with IEC<br>60601-2-22, IEC<br>60825-1 and 21 CFR<br>1040 PERFORMANCE<br>STANDARDS FOR<br>LIGHT-EMITTING<br>PRODUCTS |
Confidential and Privileged. This document confidential and priviteged trade secrets and other information of Beijing Connection and Printegor. The Locallion Called on thers not printers and soles and Ministman of Colling Honkon i'cchnologies Co., Ltd. All
rights reserved
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "HONKON" in bold, black letters. Below the word "HONKON" is a line of Chinese characters. The characters appear to be a company name.
#### 510(k) Premarket Notification Submission
Conclusion:
The Diode Laser for Hair Removaland its application comply with standards as detailed in section 9, 11 and 17 of this premarket notification. Therefore, Beijing Honkonstatesthat the non-clinical tests determined that the Diode Laser for Hair Removal to be as safe, as effective and performance is substantially equivalent to the predicate device(s).
Confidential and Privileged. This document contains confidential and privileged trade secrets and other information of Beijing Honkon Technologies Co., Ltd. and as such may not employed by Beijing Honkon Technologies Co., Ltd. All rights reserved
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or a caduceus-like design.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2014
Osmunda Medical Device Consulting Company, Ltd. Mr. Mike Gu Regulatory Affairs Manager 7th floor, Jingui Business Building, 982 Congyun Road Baiyun District, 510420, Guangzhou, China
Re: K131838
Trade/Device Name: Diode Laser for Hair Removal Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: March 25, 2014 Received: March 27, 2015
Dear Mr. Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{5}------------------------------------------------
Page 2 - Mr. Mike Gu
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S.
- Acting Director
- Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K131838
Device Name Diode Laser for Hair Removal
#### Indications for Use (Describe)
The diode laser for hair removal, permanent hair reduction on all skin types (Fitzpatick skin type I-V).
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Neil R Ogden 2014.04.23 15:34:14
FORM FDA 3881 (1/14)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
.
ﺍﻟﻤﺮﺍﺟﻊ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.