K131822 · Toshiba America Medical Systems, In.C · LLZ · Jul 23, 2013 · Radiology
Device Facts
Record ID
K131822
Device Name
ULTRAEXTEND USWS-900A V2.1 AND V3.1
Applicant
Toshiba America Medical Systems, In.C
Product Code
LLZ · Radiology
Decision Date
Jul 23, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
This software is intended for displaying and analyzing ultrasound images for medical diagnosis in cardiac and general examinations.
Device Story
UltraExtend USWS-900A v2.1/v3.1 is post-processing software installed on general-purpose PCs. It enables loading and analysis of ultrasound data acquired from Toshiba Aplio diagnostic systems (Aplio XG, MX, Artida, 300, 400, 500). The software provides a seamless image reading environment, replicating the functionality and operability of the source ultrasound system for diagnosis on a PC. Key features include CHI-Q, TDI-Q, and 2D wall motion tracking. Clinicians use the software to display and analyze images to support medical decision-making. The device benefits patients by allowing remote or offline review of diagnostic ultrasound examinations.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, software verification, and validation documentation per FDA guidance and IEC 62304 processes.
Technological Characteristics
Software-based post-processing system for PCs. Operates on Windows XP (v2.1) or Windows 7 (v3.1). Implements image processing, CHI-Q, TDI-Q, and 2D wall motion tracking. Developed under ISO 13485 and IEC 62304 standards.
Indications for Use
Indicated for patients undergoing cardiac and general ultrasound examinations for medical diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Aplio 300, Aplio 400 and Aplio 500 Diagnostic Ultrasound Systems (K121422)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a handwritten string of characters. The string appears to be a combination of letters and numbers. The string starts with the letter 'K', followed by the numbers '131822'. The characters are written in a cursive style, with some of the numbers connected to each other.
ТОЯНІВА АМЕRICA MEDICAL БУБТЕМБ, INC. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-5000
### 510(k)-SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with 21 CFR 807.92, the following summary of the information is provided.
- 1. Submitter's Name: Toshiba America Medical Systems, Inc.
- 2. Submitter's Address: 2441 Michelle Drive Tustin, CA 92781-2068
- 3. Establishment Registration Number: 2020563
- 4. Contact Person: Charlemagne Chua Manager Regulatory Affairs (714) 730-5000
- 5. Date Prepared: February 25, 2013
- 6. Device Proprietary Name: UltraExtend USWS-900A, v2.1 UltraExtend USWS-900A, v3.1
#### 7. Common Name: System, Picture Archiving and Communications
- 8. Classification: 21 CFR §892.2050 Class II
Product Code: LLZ
- 9. Predicate Devices:
| Product | Marketed by | 510(k) Number | Clearance Date |
|-----------------|-----------------|---------------|--------------------|
| UltraExtend | Toshiba America | K082596 | September 23, 2008 |
| USWS-900A | Medical Systems | | |
| UltraExtend FX | Toshiba America | K121076 | October 9. 2012 |
| TUW-U001S v2.02 | Medical Systems | | |
JUL 2 3 2013
{1}------------------------------------------------
#### 10. Reason for Submission:
Modification of the cleared device UltraExtend USWS-900A, K082596
### 11. Device Description:
UltraExtend USWS-900A v2.1 and v3.1 is a software package that can be installed in a general-purpose personal computer (PC) to enable data acquired from Aplio diagnostic ultrasound systems (Aplio XG, Aplio MX, Aplio Artida, Aplio 300, Aplio 400 and Aplio 500), to be loaded onto a PC for image processing with other application software product. UltraExtend USWS-900A v2.1 and v3.1 is a postprocessing software that implements functionality and operability equivalent to that of the diagnostic ultrasound system the data was acquired from, providing a seamless image reading environment from examination using the diagnostic ultrasound system to diagnosis using the PC.
# 12. Indications for Use:
This software is intended for displaying and analyzing ultrasound images for medical diagnosis in cardiac and general examinations.
#### 13. Determination of Substantial Equivalence:
UltraExtend USWS-900A v2.1 and v3.1 functions in a manner similar to and is intended for the same use as the predicate devices UltraExtend, USWS-900A (K082596) and UltraExtend FX (K121076), marketed by Toshiba America Medical Systems.
UltraExtend USWS-900A v2.1 and v3.1 include modifications to the previously cleared device, UltraExtend USWS-900A, which allows data acquired by Aplio 300, Aplio 400 and Aplio 500 Diagnostic Ultrasound Systems (K121422) to be accessible on a PC running UltraExtend USWS-900A v2.1 and v3.1. Additionally, two applications, CHI-Q and TDI-Q, as well as the new feature, 2D wall motion tracking for the images acquired by the Aplio 300, Aplio 400 and Aplio 500 were added. UltraExtend USWS-900A v2.1 runs under Windows XP and v3.1 runs under Windows 7.
UltraExtend USWS-900A v2.1 and v3.1 employs the same fundamental scientific technology as the predicate devices.
A complete comparison of the subject device versus the predicate devices is available in TAB 3 of this submission.
# 14. Safety:
This device is designed and manufactured under the Ouality System Regulations as outlined under 21 CFR§820 and ISO 13485 Standards.
{2}------------------------------------------------
# 15. Summary of Testing:
#### Software Documentation a.
Risk Analysis, Verification/Validation testing conducted through bench testing, as well as software validation documentation as required by the FDA Guidance Document titled, "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices," are presented in TABS 5 and 7. This documentation demonstrates that the device meets established performance and safety requirements and is therefore deemed safe and effective. Additionally, IEC 62304 processes were implemented in the development of the subject device.
#### b. Clinical Test
UltraExtend USWS-900A v2.1 and v3.1 did not require clinical studies to support substantial equivalence.
#### 16. Conclusion:
The modifications implemented in the subject device do not change the the intended use of the device. Based upon the safety and effectiveness data/information provided, the clinical performance of UltraExtend USWS-900A v2.1 and v3.1 are deemed to be substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 23, 2013
Toshiba America Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street N W BUFFALO MN 55313
Re: K131822
Trade/Device Name: UltraExtend USWS-900A v2.1 and v3.1 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ and IYN Dated: June 19, 2013 Received: June 20, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Job
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.ltm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131822
Device Name: UltraExtend, USWS-900A, v2.1 and v3.1
Indications for Use:
This software is intended for displaying and analyzing ultrasound images for medical diagnosis in cardiac and general examinations.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Sm(i,j)
(Division Sign-Off) Division of Radiological Health Office of In Viro Diagnostics and Radiological Health
510(k) K131822 ___________
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.