← Product Code [LKF](/productcode/LKF) · K131781

# CLEARVIEW TOTAL (K131781)

_Clinical Innovations, LLC · LKF · May 28, 2014 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K131781

## Device Facts

- **Applicant:** Clinical Innovations, LLC
- **Product Code:** [LKF](/productcode/LKF.md)
- **Decision Date:** May 28, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4530
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The ClearView Total is intended for use in laparoscopic procedures where it is desirable to delineate the vaginal fornices and the surgeon intends to remove or access intraperitonial tissue through the vagina by use of a colpotomy or culdotomy incision; such as laparoscopically assisted vaginal hysterectomies, total laparoscopic hysterectomies, while maintaining pneumoperitoneum by sealing the vagina while a colpotomy is performed.

## Device Story

ClearView Total is a single-use, sterile uterine manipulator used during laparoscopic pelvic surgeries. Device features a mechanical handle for uterine manipulation and a plastic ColpoCup accessory, available in three sizes, which screws into the manipulator to delineate vaginal fornices. A pre-attached, inflatable saline-filled balloon occluder at the base of the cup seals the vagina to maintain pneumoperitoneum during colpotomy or culdotomy. Operated by surgeons in clinical/OR settings. Device provides visual guidance via high-contrast cup colors and maintains surgical field integrity. Benefits include improved access to intraperitoneal tissue and maintenance of abdominal pressure during minimally invasive hysterectomies.

## Clinical Evidence

Bench testing only. No clinical data. Integrity testing included accelerated aging, balloon leak/burst testing, and cup security/break testing. Biocompatibility testing included cytotoxicity, intracutaneous reactivity irritation, and sensitization. All tests met acceptance criteria.

## Technological Characteristics

Single-use, sterile, plastic uterine manipulator. Features include mechanical handle, screw-in ColpoCup (3.0cm, 3.5cm, 4.0cm), and inflatable saline-filled balloon occluder. Compatible with standard electrosurgical and harmonic tools. No electronic components or software.

## Regulatory Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

## Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

## Predicate Devices

- Clinical Innovations ClearView ([K940681](/device/K940681.md))
- ConMed VCARE ([K071907](/device/K071907.md))
- The Koh Colpotomizer System ([K954311](/device/K954311.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Summary (As Required by 21 CFR 807.92)

22 May 2014

This 510(k) Summary of safety and effectiveness for the Clinical Innovations ClearView Total is submitted with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(k) summary.

# Submitter Information

Company Name: Company Address:

Company Phone: Company Fax:

#### Device Identification

Device trade name: Device Common Name: Classification: Product Code: Device Class: Establishment Registration Number: Clinical Innovations 747 West 4170 South Murray, UT 84123 801-268-8200 801-266-7373

ClearView Total Cannula, Manipulator/Injector, Uterine Unclassified per Pre-Amendment LKF Unclassified 1722684

### Contact Person

Tom Haueter Director, Quality Assurance and Regulatory Affairs 801-260-6028 T.Haueter@clinicalinnovations.com

### Predicate Devices:

Clinical Innovations ClearView® (K940681)

ConMed VCARE (K071907)

The Koh Colpotomizer™ System (K954311)

### Device Description:

Clinical Innovations' ClearView Total is a single-use sterile device used for uterine manipulation. Uterine manipulation is essential for laparoscopies involving the female pelvic organs (uterus, tubes, ovaries) when a uterus is present. Uterine manipulators may be helpful when clinicians perform tubal ligations, diagnostic laparoscopies for evaluating pelvic pain and infertility, treatment of endometriosis, removal of pelvic scars (adhesions) involving the uterus, fallopian tubes and ovaries, treatment of ectopic pregnancy, removal of uterine fibroids, removal of ovarian cysts, removal of ovaries, tubal repair, laparoscopic hysterectomy, laparoscopic repair of pelvic bowel or bladder, sampling of pelvic lymph nodes, laparoscopic bladder suspension procedures for treatment of incontinence, and biopsy of pelvic masses.

{1}------------------------------------------------

The ColpoCup accessory is a plastic cup which is mechanically screwed into the uterine manipulator. The ColpoCup is compatible with typical surgical devices, including harmonics and electrosurgical tools. Three different sizes of ColpoCups will be included with the device; 3.0cm, 3.5cm. and 4.0cm. Each ColpoCup will be a high contrast color in order to provide the surgeon with clear visibility during laparoscopic dissection.

At the base of the ColpoCup, past the tip pivot point, is a pre-attached Occluder constructed of an inflatable balloon and will be included to seal off the vagina and prevent pnuemoperitoneum loss. The Occluder Balloon is connected to a separate inflation valve which is located proximally from the balloon and allows for inflation after placement.

### Indications for Use:

The ClearView Uterine Manipulator Device with ColpoCup and Occluder is intended for use in laparoscopic procedures where it is desirable to delineate the vaginal fornices and the surgeon intends to remove or access intraperitoneal tissue through the vagina by use of a colpotomy or culdotomy incision; such as laparoscopically assisted vaginal hysterectomies, total laparoscopic hysterectomies, while maintaining pneumoperitoneum by sealing the vagina while a colpotomy is performed.

### Indications for Use Comparison:

| Property                                                                                                                                                                                                                         | Result                                                                                                                    |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Laparoscopic procedures where it is<br>desirable to delineate the vaginal<br>fornices and the surgeon intends to<br>remove or access intraperitoneal tissue<br>through the vagina by use of a<br>colpotomy or culdotomy incision | Similar to predicate device, The Koh Colpotomizer™ System<br>Similar to predicate device, Clinical Innovations® ClearView |
| Laparoscopically assisted vaginal<br>hysterectomies.                                                                                                                                                                             | Similar to predicate device, the ConMed VCARE<br>Similar to predicate device, Clinical Innovations® ClearView             |
| Total laparoscopic hysterectomies                                                                                                                                                                                                | Similar to predicate device, the ConMed VCARE                                                                             |

# Technical Characteristics Comparison:

| Property                                   | Result                                                        |
|--------------------------------------------|---------------------------------------------------------------|
| Sterile, Single-use                        | Similar to predicate device, ConMed VCARE                     |
|                                            | Similar to predicate device, Clinical Innovations * ClearView |
| Distal holes that allow suture fixation if | Similar to predicate device, ConMed VCARE                     |
| desired                                    |                                                               |
| Three different cup sizes; 3.5cm, 4.0cm,   | Similar to predicate device, The Koh Colpotomizer™ System     |
| 4.5cm                                      |                                                               |
| Distal handle for uterus manipulation      | Similar to predicate device, The Koh Colpotomizer™ System     |
|                                            | Similar to predicate device, Clinical Innovations ClearView   |
| Occluding balloon to maintain proper       | Similar to predicate device, The Koh Colpotomizer™ System     |
| peumoperitoneum pressure during            |                                                               |
| surgery.                                   |                                                               |
| Inflatable (saline-filled) balloon for     | Similar to predicate device, ConMed VCARE                     |
| insertion into uterus                      | Similar to predicate device. Clinical Innovations "ClearView  |

{2}------------------------------------------------

### Summary of Studies

Clinical Innovations performed device integrity testing to support the claim that the device is substantially equivalent to the predicate devices. All device integrity tests for the ClearView Total met the specified requirements, which consisted of:

- Accelerated Age Testing ●
- Balloon Leak/Burst Testing .
- . Cup Security and Cup Break

Clinical Innovations conducted the following biocompatibility testing in accordance to relevant industry standards:

- Cytotoxicity .
- Intracutaneous Reactivity Irritation .
- . Sensitization

All testing met the appropriate acceptance criteria.

### Basis for Substantial Equivalence:

Clinical Innovations believes that the ClearView Total is substantially equivalent based on the similarities to the predicate devices in indications and technical properties. Further, the ClearView Total introduces no new intentions, indications, or technical properties than exist currently in the previously cleared predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized emblem of an abstract human figure with three flowing lines, representing life, knowledge, and service. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 28, 2014

Clinical Innovations, LLC Tom Haueter Director, Quality and Regulatory Affairs 747 West 4170 South Murray, UT 84123

Re: K131781

> Trade/Device Name: ClearView Total Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: LKF Dated: May 1, 2014 Received: May 2, 2014

Dear Tom Haueter,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Tom Haueter

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the . Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/4/Picture/7 description: The image shows the name "Herbert P. Lerner - S" in a bold, sans-serif font. The letters are large and clearly legible. There is a circular emblem between the "P." and "Lerner" which appears to be a seal or logo, adding a formal touch to the presentation of the name. The overall impression is clean and professional.

for Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use 510(k) Number (if known): K131781 Device Name: ClearView Total

Indications for Use:

The ClearView Total is intended for use in laparoscopic procedures where it is desirable to delineate the vaginal fornices and the surgeon intends to remove or access intraperitonial tissue through the vagina by use of a colpotomy or culdotomy incision; such as laparoscopically assisted vaginal hysterectomies, total laparoscopic hysterectomies, while maintaining pneumoperitoneum by sealing the vagina while a colpotomy is performed.

> Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D)(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1

> > Herbert P. Lerner -S 2014.05.28 13:13:53 -04'00'

---

**Source:** [https://fda-staging.innolitics.com/device/K131781](https://fda-staging.innolitics.com/device/K131781)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
