ALDOLASE CONTROLS LEVEL 2, AND ALDOLASE CONTROL LEVEL 3

K131776 · Randox Laboratories, Ltd. · JJX · Jul 18, 2013 · Clinical Chemistry

Device Facts

Record IDK131776
Device NameALDOLASE CONTROLS LEVEL 2, AND ALDOLASE CONTROL LEVEL 3
ApplicantRandox Laboratories, Ltd.
Product CodeJJX · Clinical Chemistry
Decision DateJul 18, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

The Randox Aldolase Control Level 2 and Aldolase Control Level 3 are intended for in vitro diagnostic use as assayed quality control materials for the quantitative determination of aldolase on the clinical chemistry analyzers indicated in the package insert. This device is for prescription use only.

Device Story

Randox Aldolase Control Level 2 and Level 3 are lyophilized quality control materials derived from human serum with added human-origin constituents, chemicals, stabilizers, and preservatives. Used in clinical laboratories to monitor the precision of aldolase testing procedures on clinical chemistry analyzers. Requires reconstitution with distilled water before use. Stability is 5 days refrigerated after reconstitution. Healthcare providers use the control results to verify the accuracy and precision of patient sample testing. Benefits include ensuring reliable diagnostic performance of aldolase assays.

Clinical Evidence

Bench testing only. Stability studies confirmed 24-month shelf life for unopened vials and 5-day stability for reconstituted vials stored at 2-8°C. Value assignment performed via nest testing across multiple systems (RX Daytona, Beckman Coulter AU640) with precision (CV%) meeting acceptance criteria of ≤3% (or ≤6% for concentrations <10 IU/L).

Technological Characteristics

Lyophilized human serum matrix; requires reconstitution with distilled water. Analyte: Aldolase (rabbit muscle origin). Storage: 2-8°C. Shelf life: 24 months. Open-vial stability: 5 days at 2-8°C. Used as a quality control material for clinical chemistry analyzers.

Indications for Use

Indicated for use as assayed quality control materials for the quantitative determination of aldolase on clinical chemistry analyzers. For prescription use only.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION ## TRIAGE-QUICK REVIEW DECISION SUMMARY 510(k) #: k131776 This 510(k) was reviewed under OIR’s Triage-Quick Review Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA. The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant’s 510(k) summary for a summary of the information that supports this SE determination.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...