K131737 · Liaoning Kampo Medical System Co., Ltd. · LNH · Oct 23, 2014 · Radiology
Device Facts
Record ID
K131737
Device Name
SUPERNOVA C5 MRI SYSTEM
Applicant
Liaoning Kampo Medical System Co., Ltd.
Product Code
LNH · Radiology
Decision Date
Oct 23, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Supernova C5 is intended for use as a diagnostic imaging device transyerse, sagittal, coronal and oblique cross-sectional images of the internal structure of the human head, body or extremities. The images produced by MRI system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination
Device Story
Supernova C5 is a 0.35T permanent magnet MRI system; inputs include magnetic resonance signals from human anatomy; system transforms signals via gradient coils, RF coils, and spectrometer into cross-sectional images; output is diagnostic images for physician interpretation; used in clinical settings; system includes magnet, patient couch, RF coils, and computer workstation; software manages scan control, image reconstruction, and DICOM archiving; physician uses images for diagnostic determination; benefits include non-invasive visualization of internal structures.
Clinical Evidence
No formal clinical trial data provided. Evidence consists of sample clinical images obtained for all receive coils and pulse sequences to demonstrate diagnostic capability. Bench testing included SNR, geometric distortion, image uniformity, slice thickness, and acoustic noise measurements per NEMA MS standards.
Technological Characteristics
0.35T permanent magnet MRI system. Components: Magnet, gradient coils, RF coils, patient couch, system cabinet. Connectivity: DICOM compliant. Software: Windows XP/7 based. Standards: IEC 60601-1, IEC 60601-2-33, NEMA MS 1-2008, MS 2-2008, MS 3-2008, MS 4-2010, MS 5-2010, MS 6-2008, MS 8-2008, MS 9-2008, NEMA PS3.1-3.20.
Indications for Use
Indicated for diagnostic imaging of human head, body, or extremities to produce transverse, sagittal, coronal, and oblique cross-sectional images based on proton density, T1, T2, and flow properties. For use by trained physicians.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, resembling a bird-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 23, 2014
Liaoning Kampo Medical System Co., Ltd. % Mr. Shao Kai Quality Engineer No. 9. Yaodu Street. Benxi Economic Development Area Benxi, Liaoning 117004 CHINA
Re: K131737
Trade/Device Name: Supernova C5 MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: July 19, 2013 Received: September 25, 2013
Dear Mr. Kai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K131737
Device Name Supernova C5 MRI System
The Supernova C5 is intended for use as a diagnostic imaging device transyerse, sagittal, coronal and oblique cross-sectional images of the internal structure of the human head, body or extremities.
The images produced by MRI system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the text "KMPO" in a stylized font. The letters are primarily gray, except for the "M", which is orange. The logo has a modern and geometric design.
# 510 (k) Summary
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92
# 1.0 Contact Information
Title : Quality Engineer Tel : 86-414-5555322 Fax : 86-24-45689287 E-Mail : shaok@kampo.cn
### 2.0 Establishment Name and Address
Liaoning Kampo Medical System Co.,Ltd. No.9, Yaodu Street, Benxi Economic Development Area,Liaoning,P.R.China Post Code : 117004
## 3.0 Submission Date
May 31, 2013,revised September 19,2014
### 4.0 Device Information
| Trade Name: | Supernova C5 MRI System |
|----------------------|--------------------------------------------------------------|
| Product Model: | Supernova C5 |
| Classification name: | 21 CFR Part 892.1000<br>Magnetic resonance diagnostic device |
| Product Code: | LNH |
| Device Class: | Class II |
## 5.0 Predicate Device(s)
| 510(k)# | Device | Original Applicant | Decision Date |
|---------|-----------------|-------------------------------------|---------------|
| K071154 | Superstar 0.35T | NEUSOFT MEDICAL SYSTEMS<br>CO., LTD | 05/10/2007 |
Page 1 of 5
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters "KMPO" in a stylized font. The letters are a combination of orange and gray colors. The "K" and "P" are gray, while the "M" is orange. The letters are bold and have rounded edges, giving the logo a modern and professional look.
# 6.0 Device Description
The Supernova C5 is a 0.35T permanent magnet MRI system. It is composed of Magnet and Magnet Enclosure, RF Coils , Patient Couch, System Cabinet, computer system. The system software based on Windows XP or Windows 7 is an interactive program with user-friendly interface. Its functions cover scanning control, image reconstruction and image/archive management and maintenance.
Supernova C5 has been designed to comply with the FDA Recognized Consensus Standards listed in the table below, as applicable to device features and components:
| Reference # | Title |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC60601-1 | Medical electrical equipment - Part 1: General<br>Requirements for Safety |
| IEC60601-2-33 | Medical electrical equipment--Part 1:General<br>requirements for safety-2. Collateral Standard:<br>Electromagnetic compatibility—Requirements and<br>tests |
| MS 1-2008 | Determination of Signal-to-Noise Ratio (SNR) in<br>Diagnostic Magnet Resonance Imaging |
| MS 2-2008 | Determination of Two-Dimensional Geometric<br>Distortion in Diagnostic Magnetic Resonance Images. |
| MS3-2008 | Determination of Image Uniformity in Diagnostic<br>Magnetic Resonance Images |
| MS4-2010 | Acoustic Noise Measurement Procedure for<br>Diagnostic Magnetic Resonance Imaging Devices |
| MS5-2010 | Determination of Slice Thickness in Diagnostic<br>Magnetic Resonance Imaging. |
| MS6-2008 | Determination of Signal-to-Noise Ratio and Image<br>Uniformity for Single-Channel, Non-Volume Coils in<br>Diagnostic Magnetic Resonance Imaging (MRI) |
| MS8-2008 | Characterization of the Specific Absorption Rate<br>(SAR) for Magnetic Resonance Imaging Systems |
| MS9-2008 | Characterization of Phased Array Coils for Diagnostic<br>Magnetic Resonance Images (MRI) |
| NEMA PS3.1 -<br>3.20 (2011) | Digital Imaging And Communications In Medicine<br>(DICOM) Set |
7.0 Indications for Use
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for KMPRO. The letters are in a sans-serif font, with the letters K, M, P, R, and O. The letter M is orange, while the other letters are gray. The letters are bold and have rounded corners.
The Supernova C5 is intended for use as a diagnostic imaging device which can produce transverse, sagittal, coronal and oblique cross-sectional images of the internal structure of the human head, body or extremities. The images produced by MRI system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination
# 8.0 Technological Characteristics Comparison to Predicate Device
The Supernova C5 is comprised of the following functional modules, and the predicate device contains the same functional modules:
- Magnet - responsible for providing a homogeneity magnetic field in specified space.
- Gradient Coil generates a linear gradient field in specified space when excited by electric current.
- Gradient Amplifier provides high electric current for gradient coil ロ
- Transmit Coil generate a homogeneity RF field in specified space when driven by RF power
- RF Amplifier provides RF power for transmit coil ロ
- RX Coils receives the FID signal which has excited by the transmit coil and encoded by the gradient coil.
- Spectrometer receives parameters transferred from scan control to generate time sequence to control the operation of gradient amplifier, RF amplifier and RX coil. The signal from RX coil will be digitized and transferred to SW for further processing
- User Interface (SW) user can get desired images by proper operation in UI.
- � Acquisition - Receive MR raw data from spectrometer
- Reconstruction responsible for the efficient processing of raw data to � generate MR images.
- � Analysis - contains the image post-processing tools
- � Protocols - Supernova C5's each protocol is comprised at a pulse sequence, user parameters.
- � Scan Control - responsible for the real-time transfer of controlling orders for protocols, protocols parameters modifications, and dynamic information from the MR host in response to user or program requests
Page 3 of 5
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo with the letters "KMPO" in a stylized font. The "M" is orange, while the other letters are gray. The letters are bold and have rounded edges. The logo appears to be for an organization or company with the initials KMPO.
- Information System the central repository of all relevant MRI system � configuration, patient, study, scan, etc., parameters associated with the current patient study
- � Storage - obtains current patient and scan information, performs non-volatile local storage, exports images and data in DICOM format.
- Visualization implements all aspects of the user interface, including � scan protocols selection, controls to modify protocol parameters, image display, graphical slice prescription, and image review, save, and export.
- Patient couch provides a support for patient during scanning.
- Cover provides a shell for magnet
There is no differences between the Supernova C5 and the predicate device's intended use.
Both Supernova C5 and the predicate device can offer pulse sequences. Supernova C5 are optimized for whole body applications, and offer the imaging capabilities same as predicate device.
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
## 9.0 Performance Data - Discussion of Non-Clinical Tests
Design controls quality assurance measures during the development of Supernova C5 include:
- Customer requirement analyze
- System requirement analyze
- Detail design
- Design reviews
- Unit and integration level testing
- Verification testing, including System and Manual testing
- Safety testing, including SAR, dB/dt, and acoustic noise
- Performance testing, including SNR and uniformity
- Validation testing
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the text '510(k)'. This likely refers to a type of premarket submission to the FDA (Food and Drug Administration) to demonstrate that a medical device is safe and effective. The 510(k) pathway is a common route for medical device approval in the United States.
Image /page/7/Picture/1 description: The image shows the logo for KMPO. The letters are in a sans-serif font, with the 'A' in orange and the rest of the letters in gray. The 'A' has a small orange drip coming from the bottom left corner. The logo is simple and modern.
Risk management, compliant with ISO 14971:2007, identified hazards, sequences of events, and resultant harms: developed, implemented, and tested risk-controlling mitigations: and evaluated residual risks.
# 10.0 Safety Parameters
| | Supernova C5 | Predicate Device |
|--------------------------|--------------------------------|-----------------------------|
| Magnet field strength | 0.35T | 0.35T |
| Operating Modes | Normal Level | 1st Level Operating |
| IEC 60601-2-33 (2010-03) | Operating Mode | Mode |
| Safety Parameter Display | SAR, dB/dt | SAR, dB/dt |
| Max SAR | <2W/kg whole-body | <2W/kg whole-body |
| Max dB/dt | Normal Level<br>Operating Mode | 1st Level Operating<br>Mode |
# 11.0 Performance Data - Discussion of Clinical Tests
Sample clinical images were obtained for all receive coils and pulse sequences included with the system.
# 12.0 Conclusions
The conclusions drawn from the nonclinical and clinical tests that demonstrate that Supernova C5 is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.