# PREBENT PLATING WITH VIRTUAL SURGICAL PLANNING (K131674)

_Biomet Microfixation · JEY · Aug 8, 2013 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K131674

## Device Facts

- **Applicant:** Biomet Microfixation
- **Product Code:** [JEY](/productcode/JEY.md)
- **Decision Date:** Aug 8, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

For use in the stabilization and fixation of mandibular fractures and mandibular reconstructive surgical procedures.

## Device Story

Device consists of titanium bone plates for internal fixation in mandibular reconstruction. Plates are available in various geometries (straight, angled, curved, Y-shape, matrix) and thicknesses. Plates are pre-bent by manufacturer based on surgeon specifications or patient-specific anatomical models. System integrates with Medical Modeling VSP System (K120956), which utilizes patient CT scans to generate anatomical models, patient-specific surgical guides, and instrumentation. These guides and models assist in the precise contouring and bending of plates to match patient anatomy. Used by surgeons in clinical settings for fracture stabilization and reconstruction. Output is a patient-matched, pre-contoured implant that facilitates surgical alignment and reduces intraoperative bending time.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on identical design, materials, and indications to predicate devices.

## Technological Characteristics

Titanium bone plates; various geometries (straight, angled, curved, Y-shape, matrix). Non-sterile, single-use. Manual bending process based on patient-specific anatomical models and VSP System guides. No electronic or software components integral to the plate itself.

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Predicate Devices

- Lorenz Titanium Fracture/ Reconstruction Devices and Pre-Bent Plates ([K063052](/device/K063052.md))
- VSP System ([K120956](/device/K120956.md))
- Biomet Microfixation Mandibular Fracture/ Reconstruction Devices and Pre-Bent Plates System ([K113512](/device/K113512.md))
- Fracture/Reconstruction System (TraumaOne) ([K081067](/device/K081067.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) Summary

| Date Prepared:                       | August 1, 2013                                                                                                                                                                                                                                                                               |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:                        | Biomet Microfixation<br>1520 Tradeport Drive<br>Jacksonville, FL 32218-2480                                                                                                                                                                                                                  |
| Contact:                             | Elizabeth Wheeler<br>Regulatory Affairs Specialist<br>904-741-4400 x 9558                                                                                                                                                                                                                    |
| Proprietary Name:                    | Biomet Microfixation PreBent Plates with Virtual Surgical<br>Planning                                                                                                                                                                                                                        |
| Common or Usual Name:                | Bone Plate                                                                                                                                                                                                                                                                                   |
| Device Classification:               | Class II                                                                                                                                                                                                                                                                                     |
| Device Product Code:                 | 76 JEY (21 CFR 872.4760)                                                                                                                                                                                                                                                                     |
| Secondary Device<br>Product Code(s): | 76 DZJ<br>90 LLZ                                                                                                                                                                                                                                                                             |
| Predicate Device:                    | K063052- Lorenz Titanium Fracture/ Reconstruction Devices and<br>Pre-Bent Plates<br>K120956- VSP System (Medical Modeling)<br>K113512- Biomet Microfixation Mandibular Fracture/<br>Reconstruction Devices and Pre-Bent Plates System<br>K081067- Fracture/Reconstruction System (TraumaOne) |
| AUG 0 8 2013                         |                                                                                                                                                                                                                                                                                              |

#### Device Description:

The Biomet Microfixation PreBent Plates with Virtual Surgical Planning is comprised of a variety of titanium plates with shapes and sizes designed for internal fixation for mandible reconstruction procedures. The plates include variations of straight, angle, curved, Y-shape, double Y-shape, and matrix, options with various lengths and thickness. Plates are offered flat or pre-bent. Pre-bent plates are contoured by Biomet Microfixation per surgeon specifications or patient specific anatomical model and include patient specific guides and instruments, of the VSP System (K120956, cleared December 12, 2012). The VSP System is used by Medical Modeling to create the guides, instrumentation and a patient specific anatomical model from CT scans provided by a surgeon. Medical Modeling uses the surgeon information to create the anatomical patient specific model. The anatomical patient specific model is then used to aide in proper alignment and bending of the plates previously cleared under K063052 (cleared 01/12/07), K113152 (cleared 06/29/12), and K081067 (cleared 05/09/08). This bending process

{1}------------------------------------------------

is conducted at Biomet Microfixation based on our process specification and inspection criteria previously cleared with K063052.

#### Indications for Use:

For use in the stabilization and fixation of mandibular fractures and mandibular reconstructive surgical procedures.

#### Possible Risks:

- 1. Poor bone formation, Osteoporosis, Osteolysis, Osteomyelitis, inhibited revascularization, or infection can cause loosening, bending, cracking or fracture of the device.
- 2. Nonunion or delayed union, which may lead to breakage of the implant.
- 3. Migration, bending, fracture or loosening of the implant.
- 4. Metal sensitivity or allergic reaction to a foreign body.
- 5. Decrease in bone density due to stress shielding.
- 6. Pain, discomfort, abnormal sensation, or palpability due to the presence of the device.
- 7. Increased fibrous tissue response around the fracture site and/or the implant.
- 8. Necrosis of bone.
- 9. Inadequate healing.

### Technological Characteristics:

The subject PreBent Plates with Virtual Surgical Planning devices are identical to the predicate devices in terms of indications, use, material, and design. The only difference between the subject device and the predicate device is that the subject device, is not only shaped based on an anatomical specific model, but also the provided patient specific guides and instruments of the VSP System (K 120956, cleared December 12, 2012), where the predicate pre-bent plates are shaped based on an anatomical specific model only, Biomet Microfixation has a process specification for bending plates and an inspection criteria to verify the plate matches the anatomical model. Because the plates are verified to a patient specific model they are 100% verified to the Medical Modeling output, anatomical model, for every case. Therefore, no additional testing was conducted on the subject plates.

#### Sterility Information:

The plates will be marketed as non-sterile, single-use devices.

#### Clinical Testing:

Clinical testing was not performed to support this submission.

#### Non-Clinical Testing:

Non-clinical testing was not performed, as the predicate plates have been previously cleared and the subject devices are identical. We have provided the process specification for bending the plates and inspection criteria for inspecting the output to ensure the proper alignment to the anatomical model.

{2}------------------------------------------------

#### Substantial Equivalence Conclusion:

When compared to the predicate devices, Lorenz Titanium Fracture/ Reconstruction Devices and Pre-Bent Plates (K063052, cleared 01/12/07), Fracture/Reconstruction System (TraumaOne) (K081067, cleared 05/09/08) and Biomet Microfixation Mandibular Fracture Reconstruction Devices and Pre-Bent Plates System (K113152, cleared 06/29/12), substantial equivalence is based on the same indications for use and technological characteristics. The subject devices are identical to the predicate devices, but an update to the labeling will be made to give the option of the prebent plates with Medical Modeling VSP System to include patient specific guides, instruments and anatomical model.

Based on the determination that no clinical or non-clinical testing was deemed necessary, it can be concluded that the subject devices, PreBent Plates with Virtual Surgical Planning, are safe and effective and perform as well as the predicate devices. Therefore, we conclude that the subject devices are substantially equivalent.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Gibth Silver Spring, MD 20993-0002

#### August 8, 2013

Biomet Microfixation Ms. Elizabeth Wheeler Regulatory Affairs Specialist 1520 Tradeport Drive JACKSONVILLE FL 32218-2480

Re: K131674

Trade/Device Name: Biomet Microfixation PreBent Plates with Virtual Surgical Planning Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY, DZJ, LLZ Dated: July 8, 2013 Received: July 11, 2013

Dear Ms. Wheeler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ase saled in the May 28, 1976, the enactment date of the Medical Device Amendments, or 10 conninered processified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may merelor symbols of the Act include requirements for annual registration. Ilisting of general controls provide in of active, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warrantics. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Wheeler

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# Kwame O. Ulmer -S

Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Biomet Microfixation Special 510(k)-PreBent Plates with Virtual Surgical Planning

:

## Indications for Use

510(k) Number (if known): |૮٦ | ٢٠٦ | ١٦ |

Device Name: Biomet Microfixation PreBent Plates with Virtual Surgical Planning

Indications For Use: For use in the stabilization and fixation of mandibular fractures and mandibular reconstructive surgical procedures.

Prescription Use xx AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Andrew I. Steen -S 2013.08.06 09:33:55 -04'00'

Division Sign-Off) rivision of Anesthestology, General Hospital ifection Control, Dental Devices

Page 1 of 1

510(k) Number: الح الك اك اك

---

**Source:** [https://fda-staging.innolitics.com/device/K131674](https://fda-staging.innolitics.com/device/K131674)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com