K131661 · Innovative Medical Manufacturing Co. · FMF · Nov 5, 2013 · General Hospital
Device Facts
Record ID
K131661
Device Name
INFUSION SYRINGE SET
Applicant
Innovative Medical Manufacturing Co.
Product Code
FMF · General Hospital
Decision Date
Nov 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
IMM™ Infusion Syringe Set is intended for delivering controlled volume of fluids from reservoirs consistent with the labeled route of administration of fluids by manual operation.
Device Story
IMM Infusion Syringe Set is a single-use, sterile, manual fluid delivery system. Components include a spike inlet for fluid reservoir connection, PVC tubing, a dual check valve, and a 12ml control syringe. The device is operated manually by clinicians to deliver controlled volumes of fluids from a solution bag to an access device (e.g., needle). The distal end features a 6% taper luer fitting for secure connection. The device facilitates fluid administration in clinical settings. Benefits include controlled, manual fluid delivery and ease of use/replacement. The device is disposed of after single patient use.
Clinical Evidence
Bench testing only. No clinical data provided. Functional performance testing demonstrated compliance with ISO8536-4 (infusion sets) and ISO7886-1 (sterile hypodermic syringes). Biocompatibility testing conducted in accordance with ISO10993.
Technological Characteristics
Materials: PP, PVC, PC, ABS. Sensing/Actuation: Manual piston syringe with dual check valve. Energy: Manual. Connectivity: None. Sterilization: Ethylene Oxide (EtO). Standards: ISO8536-4, ISO7886-1, ISO10993. Form factor: Single-use infusion set with 12ml syringe and 6% taper luer fitting.
Indications for Use
Indicated for patients requiring controlled volume fluid delivery from reservoirs via manual operation. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
# Innovative Medical Manufacturing Company
107, Lane 181, Sec.1, Yong Jane Rd., Chunan, Miaoli 350, Taiwan, ROC TEL:037-620236 FAX:037-620239 E-mail:service@immc.com.tw Unified Code:16980249
#### 510(k) Summary of Safety and Effectiveness
#### 1. General Information
| Submitted by: | Innovative Medical Manufacturing Company<br>107, Lane 181, Sec. 1., Yong Jane Road<br>Chunan, Miaoli 350, Taiwan (ROC) |
|----------------------|------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Lucy Huang, Regulatory Specialist<br>Tel: +886-37-620236<br>Fax: +886-37-620239 |
| Date of preparation: | 10/14/2013 |
| | NOV 05 2013 |
## 2. Device Information
| Common Name: | Syringe Kit |
|-------------------|--------------------------------------------------|
| Proprietary Name: | IMMTM Infusion Syringe Set |
| Classification | Piston Syringe; Intravascular Administration Set |
| Name: | |
| Product code: | FMF(880.5860)<br>FPA (880.5440) |
#### 3. Predicate Device
| Trade Name | 510(k) Number | Decision Date |
|--------------------------------|---------------|---------------|
| Vascular Auto-Fill Syringe Kit | K063851 | 2007/1/26 |
| Multi-Ad Dispensing system | K792227 | 1979/12/20 |
## 4. Device Description
IMM 110 Infusion Syringe Set is single use product which contains disposable parts of one spike infusion set for connecting to fluid reservoir and fluid delivery, and a control syringe for delivery fluid supply from a solution bag to access device for purpose of delivering controlled volume of fluids from reservoirs consistent with the labeled route of administration of fluids by manual operation. The distal end of the set is a dual valve with 6% taper design for connecting to other access device such as needles before use. The IMM Infusion Syringe Set is provided sterile in fluid path, and to be disposed after single patient use.
{1}------------------------------------------------
# Innovative Medical Manufacturing Company
107, Lane 181, Sec.1, Yong Jane Rd., Chunan, Miaoli 350, Taiwan, ROC TEL:037-620236 FAX:037-620239 E-mail:service@immc.com.tw Unified Code:16980249
# 5. Indication for Use
IMM™ Infusion Syringe Set is intended for delivering controlled volume of fluids from reservoirs consistent with the labeled route of administration of fluids by manual operation.
## 6. Technological Characteristics
The IMM™ Infusion Syringe Set is single use device, and can be delivered in unit sterile package or in bulk non-sterile shipment for further packaging. It allows easy operation and replacement before use.
## 7. Substantial equivalence
The IMM™ Infusion Syringe Set has the same indication for use and similar characters with the predicated device. The technological characteristics of the IMM device are as same as the predicate device.
| Element of<br>Comparison | Subject Device<br>K131661 | Vascular<br>K063851 | B. Braun<br>K792227 |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|----------------------------------------------------------|
| Intended Use | For delivering<br>controlled volume<br>of fluids from<br>reservoirs of<br>labeled route of<br>administration of<br>fluids by manual<br>operation | Same | Similar |
| Construction | Spike inlet, PVC<br>tubing, dual check<br>valve, syringe | Similar, except<br>additional<br>extension | Spike inlet, PVC<br>tubing, dual check<br>valve, syringe |
| Transparency | Tubing of fluid line<br>is transparency | Same | Same |
| Tubing/fitting<br>bonding<br>strength | Tensile strength<br>>15 N | Similar | Similar |
| Flow regulator<br>Clamping<br>efficacy | From zero to<br>maximum in<br>gravity use | Same | Same |
| Syringe | 12ml | 12ml | 10ml |
| Interface | Luer fitting | Luer fitting | Luer fitting |
| Material | PP, PVC, PC, ABS | PP, PVC, PC, ABS | Similar |
| Sterility | Single use | Same | Same |
| Sterilization<br>method | EtO | EtO | EtO |
{2}------------------------------------------------
# Innovative Medical Manufacturing Company
107, Lane 181, Sec.1, Yong Jane Rd., Chunan, Miaoli 350, Taiwan, ROC TEL: 037-620236 FAX:037-620239 E-mail:service@immc.com.tw Unified Code:16980249
## 8. Performance Summary
The non-clinical functional and performance tests demonstrated that IMM™ Infusion Syringe Kit meets the specific requirements established in voluntary standards: ISO8536-4 and ISO7886-1.
#### 9. Material
The Infusion syringe set is composed of materials that have been tested in accordance with ISO10993 and have been determined to be suitable for the intended use of this product.
#### 10. Safety and Effectiveness
All verification and validation test data indicated that the IMM Infusion syringe set can perform as intended use and is substantially equivalent to the predicate device.
5-3
{3}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 5,2013
Innovative Medical Manufacturing Company Ms. Lucy Huang Regulation Specialist No .. 107, Lane 181, Sect 1., Yong Jane Road Chunan, MIAOLI CHINA 350
Re: K131661
Trade/Device Name: IMM™ Infusion Syringe Set Regulation Number: 21 CFR 880.5860, 21 CFR 880.5840 Regulation Name: Piston Syringe, Intravascular Administration Set Regulatory Class: II Product Code: FMF, FPA Dated: September 13, 2013 Received: September 25, 2013
Dear Ms.Huang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Huang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/7 description: The image shows the text "Mary Surfammer -S". The text is in a bold, sans-serif font. The word "Mary" is in a larger font size than the rest of the text. The word "Surfammer" is stylized with a wave-like design. The "-S" is in a smaller font size than the rest of the text.
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
K131661 510(k) Number (if known):
Device Name: IMM™ Infusion Syringe Set
Indications for Use:
IMM™ Infusion Syringe Set is intended for delivering controlled volume of fluids from reservoirs consistent with the labeled route of administration of fluids by manual operation.
Prescription Use Over-The-Counter Use (Part 21 CFR 801 Subpart AND/OR (21 CFR 801 Subpart C) D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _1_ of __ 1_
Image /page/5/Picture/9 description: The image shows the logo of the Food and Drug Administration (FDA). The logo is in black and white and features the letters "FDA" in a stylized, geometric font. The letters are outlined with multiple lines, giving them a three-dimensional appearance. The "A" in FDA is shaped like a triangle.
Richard C. Chapman 2013.11.04 16:10:11 -05'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.