Durex® RealFeel™ Pleasure Gel Personal Lubricant is intended for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease of comfort of intimate sexual activity and supplement the body's natural lubrication. This lubricant is compatible with polyurethane condoms. It is not compatible with natural rubber latex and polyisoprene condoms.
Device Story
Durex® RealFeel™ Pleasure Gel is a non-sterile, silicone-based personal lubricant composed of cyclopentasiloxane and dimethiconol. Intended for over-the-counter use, it is applied topically to the penis or vagina to supplement natural lubrication and enhance comfort during sexual activity. The device is used by consumers in a home or private setting. It functions as a physical lubricant; it is not compatible with natural rubber latex or polyisoprene condoms, but is compatible with polyurethane condoms. The healthcare provider or user identifies the product's compatibility profile via labeling to ensure safe use with specific condom types, thereby preventing potential condom failure or breakage during use.
Clinical Evidence
Bench testing only. Biocompatibility studies (cytotoxicity, systemic toxicity, maximization, vaginal irritation) conducted per ISO 10993-1 and GLP requirements confirmed biocompatibility. Condom compatibility testing performed per ASTM D7661-2010. Stability testing (accelerated and real-time) confirmed shelf-life specifications.
Technological Characteristics
Silicone-based lubricant containing cyclopentasiloxane and dimethiconol. Non-sterile. Tested for compatibility with polyurethane condoms per ASTM D7661-2010. Specifications include viscosity, specific gravity, and microbial limits (total aerobic, yeast, mold, and absence of pathogens).
Indications for Use
Indicated for over-the-counter use for penile and/or vaginal application to moisturize and lubricate during intimate sexual activity. Compatible with polyurethane condoms; contraindicated for use with natural rubber latex and polyisoprene condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features the department's name encircling a stylized emblem. The emblem consists of three abstract human profiles facing right, stacked on top of each other, with flowing lines extending from the bottom profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2014
Reckitt Benckiser, LLC Shahper Rahman Regulatory Strategic Lead Morris Corporate Center IV 399 Interpace Parkway Parsippany, NJ 07054
Re: K131643
Trade/Device Name: Durex® RealFeel™ Pleasure Gel Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: November 18, 2014 Received: November 19, 2014
Dear Shahper Rahman,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K131643
Device Name: Durex ® RealFeel™ Pleasure Gel Personal Lubricant
Indications For Use:
Durex® RealFeel™ Pleasure Gel Personal Lubricant is intended for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease of comfort of intimate sexual activity and supplement the body's natural lubrication. This lubricant is compatible with polyurethane condoms. It is not compatible with natural rubber latex and polyisoprene condoms.
AND/OR Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use X (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Reckitt Benckiser. The logo consists of a pink triangle with the letters "rb" inside, followed by the words "Reckitt Benckiser" in a dark gray sans-serif font. The words are stacked on top of each other, with "Reckitt" on top and "Benckiser" on the bottom.
### 510(k) Summary K131643
#### Submitted by:
| Submitted by: | Reckitt Benckiser, LLC<br>Morris Corporate Center IV<br>399 Interpace Parkway<br>Parsippany, NJ 07054<br>973-404-2781 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Shahper Rahman, Regulatory Strategic Lead, Reckitt Benckiser, LLC |
| Date Prepared: | November 18, 2014 |
| Proprietary Name: | Durex® Personal Lubricant-Silicone |
| Trade Name: | Durex® RealFeel™ Pleasure Gel Personal Lubricant |
| Common Name: | Personal Lubricant |
| Classification Name: | Condom (21CFR 884.5300) |
| Product Code: | NUC (lubricant, personal) |
| Predicate Devices: | ONE® Silicone Lubricant<br>One<br>510(k) Control Number: K110690 |
### Description of Device:
Durex® RealFeel™ Pleasure Gel is a non-sterile, personal silicone base lubricant. The lubricant is comprised of cyclopentasiloxane and dimethiconol. Durex® RealFeel™ Pleasure Gel contains a blend of silicone fluid ingredients similar to ingredients found in the predicate device The specifications for Durex® RealFeel™ Gel include appearance, odor, viscosity, specific gravity, total aerobic microbial count, total yeast and mold count, and absence of pathogenic organisms (Pseudomonas aeroginosa, Staphylococcus aureus, and Candida albicans).
### Intended Use of the Device:
Durex® RealFeel™ Pleasure Gel Personal Lubricant is intended for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease of comfort of intimate sexual activity and supplement the body's natural lubrication. This lubricant is compatible with polyurethane condoms. It is not compatible with natural rubber latex and polyisoprene condoms.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Reckitt Benckiser. The logo consists of a pink triangle with the letters "rb" in white inside of it. To the right of the triangle are the words "Reckitt Benckiser" in a dark gray sans-serif font, with "Reckitt" on the top line and "Benckiser" on the bottom line.
# Comparison of Technological Characteristics with the Predicate Device
### Intended Use Comparison
The table below shows the indication statements for the Durex® RealFeel™ Pleasure Gel and the predicate devices:
| | Durex® RealFeel™ Pleasure Gel | ONE Silicone Lubricant |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K131643 | K110690 |
| Indication for Use | Durex® RealFeel™ Pleasure Gel<br>Personal Lubricant is intended for penile<br>and/or vaginal application and to<br>moisturize and lubricate, to enhance the<br>ease and comfort of intimate sexual<br>activity and supplement the body's<br>natural lubrication. This product is<br>compatible with polyurethane condoms.<br>It is not compatible with natural<br>rubber latex and polyisoprene<br>condoms. | ONE Silicone Lubricant is an over the<br>counter personal lubricant for penile and/or<br>vaginal application, intended to moisturize<br>and lubricate, to enhance the ease of comfort<br>of intimate sexual activity and supplement the<br>body's natural lubrication. This product is<br>compatible with natural rubber latex,<br>polyisoprene, and polyurethane condoms |
The only difference in the intended use Durex® RealFeel™ Pleasure Gel in comparison to predicate device is condom compatibility. This difference does not represent a new intended use. Therefore, to the Durex® RealFeel™ Pleasure Gel and the predicate device have the same intended use.
### Technological Characteristics:
The Durex® RealFeel™ Pleasure Gel contains a blend of silicone fluid ingredients similar to ingredients found in the predicate device.
The Durex® Embrace Pleasure Gels and the predicate device have technological characteristics.
### Performance Data:
The performance data provided in the 510(k) submission are summarized below.
### Biocompatibility:
Four biocompatibility studies were conducted on the subject device. The endpoints evaluated included cvtotoxicity, systemic toxicity in mice, maximization in guinea pigs, and vaginal irritation in rabbits. Each study was conducted in accordance with GLP requirements and the applicable ISO 10993-1 standard. The results demonstrate that the subject device is biocompatible.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Reckitt Benckiser. The logo consists of a pink triangle with the letters "rb" in white inside of it. To the right of the triangle is the company name, "Reckitt Benckiser", in a dark gray sans-serif font.
# Condom Compatibility:
The condom compatibility testing was conducted in accordance with ASTM D7661-2010. The results demonstrate that the Durex® RealFeel™ Pleasure Gel is compatible with polyurethane condoms but in not compatible with natural rubber latex and polyisoprene condoms.
# Stability:
The results of accelerated and real time studies demonstrate that the subject lubricant maintains its specifications over the duration of its shelf life.
# Conclusion:
Based on the results of biocompatibility testing and nonclinical performance testing, Durex® RealFeel™ Pleasure Gel is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.