The Vigileo™ APCO/Oximetry Monitor is indicated for continuously measuring hemodynamic parameters such as cardiac output and oximetry to assess oxygen delivery and consumption. When connected to an Edwards oximetry catheter, the monitor measures oximetry in adults and pediatrics. The monitor also displays parameters, such as stroke volume and stroke volume variation, used to assess fluid status and vascular resistance. The Vigileo™ APCO/Oximetry Monitor may be used in all setting in which critical care is provided.
Device Story
Vigileo™ Monitor is a microprocessor-based instrument for continuous hemodynamic monitoring. It processes input signals from connected Edwards FloTrac sensors (arterial pressure) or Edwards oximetry catheters (venous oxygen saturation). The device calculates parameters including Cardiac Output (CO), Cardiac Index (CI), Stroke Volume (SV), Stroke Volume Variation (SVV), Systemic Vascular Resistance (SVR/SVRI), and Oxygen Delivery (DO2/DO2I). Used in critical care environments (OR, ICU) by clinicians to assess patient fluid status and vascular resistance. Output is displayed on the monitor screen, assisting healthcare providers in clinical decision-making regarding patient hemodynamic management and oxygenation status.
Clinical Evidence
Evidence includes side-by-side bench testing, pre-clinical animal studies, and comparative analysis of clinical cases. No new clinical trials were required; the device demonstrated performance and functionality equivalent to the predicate device in critical care environments.
Technological Characteristics
Microprocessor-based diagnostic computer. Interfaces with external sensors (FloTrac) and oximetry catheters. Connectivity for hemodynamic parameter monitoring. Software-based processing of arterial pressure and oximetry signals. Class II device.
Indications for Use
Indicated for continuous hemodynamic monitoring (cardiac output, oximetry, stroke volume, stroke volume variation, systemic vascular resistance) to assess oxygen delivery, consumption, and fluid status in adult and pediatric patients in critical care settings.
Regulatory Classification
Identification
A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.
{0}------------------------------------------------
Traditional 510(k) - Vigileo™ Monitor, ver 3.08
# KAY 2 2 2014
# SECTION 5 - 510(K) SUMMARY
.
| VIGILEO™ MONITOR 510(k) SUMMARY | |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter | Edwards Lifesciences |
| Contact Person | Sally L. Maher<br>One Edwards Way<br>Irvine, CA 92614-5686 |
| Date Prepared | May 6, 2014 |
| Trade Name | Vigileo™ Arterial Pressure Cardiac Output (APCO)/Oximetry Monitor |
| Common Name | Cardiac Output/Oximetry Computer |
| Classification<br>Name | Single-Function, Preprogrammed Diagnostic Computer<br>(21 CFR 870.1435, product codes DXG, DQE ) |
| Regulation Class/<br>Product Code | Class II/<br>DXG, DQE |
| Predicate Device | K103094 (cleared 17 March 2011), Edwards Lifesciences, Vigileo™<br>Arterial Pressure Cardiac Output/Oximetry Monitor |
| Device<br>Description | The Vigileo™ Arterial Pressure Cardiac Output (APCO)/Oximetry<br>Monitor (Vigileo™ Monitor) is a microprocessor-based instrument.<br>When used with an Edwards FloTrac sensor, the Vigileo™ Monitor is<br>capable of continuously monitoring the following parameters:<br>Cardiac Output (CO);<br>●<br>Cardiac Index (CI):<br>●<br>Stroke Volume (SV):<br>Stroke Volume Variation (SVV);<br>●<br>Systemic Vascular Resistance (SVR);<br>●<br>Systemic Vascular Resistance Index (SVRI);<br>Oxygen Delivery (DO2); and<br>Oxygen Delivery Index (DO2I).<br>When used with Edwards Oximetry catheters, the Vigileo™ Monitor is<br>capable of continuously monitoring the following parameters:<br>Central venous oxygen saturation (ScvO2); and,<br>●<br>Mixed venous oxygen saturation (SvO2). |
| VIGILEO™ MONITOR 510(k) SUMMARY, con't. | |
| Indications for<br>Use/Intended Use | The Vigileo™ APCO/Oximetry Monitor is indicated for continuously<br>measuring hemodynamic parameters such as cardiac output and oximetry<br>to assess oxygen delivery and consumption. When connected to an<br>Edwards oximetry catheter, the monitor measures oximetry in adults and<br>pediatrics. The monitor also displays parameters, such as stroke volume<br>and stroke volume variation, used to assess fluid status and vascular<br>resistance. The Vigileo™ APCO/Oximetry Monitor may be used in all<br>setting in which critical care is provided. |
| Comparative<br>Analysis | Verification and validation testing was conducted to compare the<br>performance and functionality of the subject and predicate<br>devices. This testing regimen included side-by-side bench, pre-<br>clinical studies, and comparative analysis of clinical data. The<br>Vigileo™ Monitor has been shown to be safe, effective, and<br>substantially equivalent to the cited predicate device for its<br>intended use critical care environments. |
| Functional/ Safety<br>Testing | The Vigileo™ Monitor has successfully passed functional and<br>performance testing, including software verification and validation,<br>mechanical and electrical testing, bench studies, pre-clinical animal<br>studies, comparison testing of clinical cases, and clinical utility. |
| Conclusion | The Vigileo™ Monitor has been shown to be safe, effective, and is<br>substantially equivalent to the cited predicate devices for their intended<br>use in the OR and ICU environments. |
{1}------------------------------------------------
## K131588
.
.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white seal for the Department of Health & Human Services - USA. The seal is circular with text around the perimeter. The right side of the seal contains a symbol of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-00002
May 22, 2014
Edwards Lifesciences LLC Mr. Stephen M. Enos Director, Regulatory Affairs One Edwards Way Irvine, CA 92614
Re: K131588
> Edwards Lifesciences™ Vigileo™ Arterial Pressure Cardiac Output/Oximetry Monitor Regulation Number: 21 CFR 870.1435 Regulation Name: Single-function, Preprogrammed Diagnostic Computer Regulatory Class: Class II (two) Product Code: DXG, DQE Dated: May 6, 2014 Received: May 7, 2014
Dear Mr. Enos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
#### Page 2 - Mr. Stephen M. Enos
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
Image /page/3/Picture/7 description: The image shows a logo with the letters FDA. The letters are in a bold, outlined font. The logo appears to be crossed out with a thick, dark line. The image is in black and white.
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K131588
Device Name
Edwards Lifesciences™ Vigileo™ Arterial Pressure Cardiac Output/Oximetry Monitor
Indications for Use (Describe)
The Vigileo™ APCO/Oximerry Monitor is indicated for continuously measuring hemodynamic parameters such as cardiac output and oximetry to assess oxygen delivery and connected to an Edwards oximetry catheter, the monitor measures oximetry in adults and pediatrics.
The monitor also displays parameters, such as stroke volume variation, used to assess fluid status and vascular resistance. The Vigileo™ APCO/Oximetry Monitor may be used in all settings in which critical care is provided.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
2014.05.22 11:18:12
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY -- -- -- T Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
J. Russell Blein
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of Information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.