SEMIAUTOMATIC EXTERNAL DEFIBRILLATOR AND ACCESSORIES
Applicant
Defibtech, LLC
Product Code
MKJ · Cardiovascular
Decision Date
Apr 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The DDU-1000 Semiautomatic External Defibrillator (AED) is indicated for use on victims of sudden cardiac arrest (SCA) who are: - Unconscious and unresponsive - Not breathing For patients under 8 years old, or less than 55 pounds (25kg), use child/infant electrode pads. Do not delay therapy to determine exact age or weight. The DDU-1000 AED must be used by or on the order of a physician.
Device Story
Portable semiautomatic external defibrillator (AED) for sudden cardiac arrest (SCA) victims. Device inputs: ECG signals via patient-applied pads. Patient Analysis System ensures pad connection; analyzes ECG rhythm to determine shock necessity. If indicated, delivers 150J (50J pediatric) impedance-compensated, biphasic truncated exponential shock without user intervention. Used in clinical or public settings; operated by trained personnel or on physician order. Provides voice prompts and LED-illuminated pictogram guidance. Records event data including ECG, audio, and shock recommendations. Benefits patient by providing automated, timely defibrillation therapy.
Clinical Evidence
Bench testing only. Performance verified via hardware verification, software validation, and defibrillation waveform comparison. ECG analysis and shock advisory system validated against American Heart Association recommendations for arrhythmia analysis algorithm performance. Compliance with IEC 60601-2-4 and IEC 60601-1-2 standards confirmed.
Technological Characteristics
Portable AED; impedance-compensated, biphasic truncated exponential waveform. User-replaceable battery and single-use pads. LED-illuminated pictogram user interface. Complies with IEC 60601-2-4 (cardiac defibrillator safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for victims of sudden cardiac arrest (SCA) who are unconscious, unresponsive, and not breathing. Pediatric use (under 8 years or <55 lbs/25kg) requires child/infant electrode pads.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Predicate Devices
DDU-2300 Semiautomatic External Defibrillator and Accessories (K081259)
Submission Summary (Full Text)
{0}------------------------------------------------
## Attachment 3
# 510(k) Summary
# # K131525
Page 1 of 3
# 510(k) Summary
APR 2 4 2014
Date Summary Prepared: April 22, 2014
510(k) Owner Information: Defibtech, LLC 741 Boston Post Road Guilford, CT 06437
| Contact Information: | Mr. Ed Horton |
|----------------------|-------------------------------------|
| | 203-453-6651, extension 138 (phone) |
| | 203-453-6657 (fax) |
| | ehorton@defibtech.com (e-mail) |
Trade (Proprietary) Name: DDU-1000 Semiautomatic External Defibrillator and Accessories
Common Name: Semiautomatic External Defibrillator
Classification Name: Automated External Defibrillator (21 CFR 870.5310, Product Code MKJ)
# Predicate Device
The DDU-1000 is a modification to the DDU-2300 predicate device. The design and intended use of the DDU-1000 AED is substantially equivalent in performance and safety to the DDU-2300 AED and accessories cleared under the following Proprietary Name:
| Proprietary Name | Manufacturer | 510(k) Number |
|---------------------------------------------------------------------|----------------|---------------|
| DDU-2300 Semiautomatic<br>External Defibrillator and<br>Accessories | Defibtech, LLC | K081259 |
{1}------------------------------------------------
# # K131525
Page 2 of 3
# Device Description
The DDU-1000 is a portable, Semiautomatic External Defibrillator (AED) intended for use on victims of sudden cardiac arrest (SCA). It is powered by a user-replaceable nonrechargeable battery pack and supports both adult and pediatric user-replaceable singleuse defibrillation/monitoring pads.
The DDU-1000 employs a Patient Analysis System that ensures proper pad/patient connection and analyzes the patient's ECG rhythm to determine whether a shock is required. If needed, the DDU-1000 provides a 150 J (50J pediatric) impedance compensated, biphasic truncated exponential defibrillation shock to the patient without user intervention. The cardiac rhythm analysis algorithm and defibrillation energy and waveform utilized are the same as the predicate device.
Voice prompts provide simple instructions for the operator. The DDU-1000 AED is capable of recording event information including electrocardiogram (ECG), audio data and SHOCK/NO SHOCK recommendations.
### Intended Use
The DDU-1000 Semiautomatic External Defibrillator (AED) is indicated for use on victims of sudden cardiac arrest (SCA) who are:
- · Unconscious and unresponsive
- · Not breathing
For patients under 8 years old, or less than 55 pounds (25kg), use child/infant electrode pads. Do not delay therapy to determine exact age or weight.
The DDU-1000 AED must be used by or on the order of a physician.
### Comparison of Technology Characteristics
The DDU-1000 AED design characteristics are the same as those of the predicate device including the same battery and pads. Both devices employ the same underlying scientific technology for patient analysis and defibrillation therapy. Both user interface designs guide the user with voice prompts and visual guidance. The DDU-1000 AED user interface includes a simple stationary pictogram, which is illuminated with light emitting diodes (LEDs) to guide the user. In comparison, the user interface of the DDU-2300
{2}------------------------------------------------
#### Attachment 3
## 510(k) Summary
## # K131525
Page 3 of 3
AED (predicate device), which consists of a color liquid crystal display (LCD) screen that displays dynamic imagery and text prompts to guide the user through the event. In addition, the intended users are the same for both AEDs.
## Performance testing
The DDU-1000 AED uses the same underlying technologies to provide functionally equivalent performance characteristics as the predicate device. Testing, including hardware verification, software validation, design validation and defibrillation waveform comparison, demonstrates that the DDU-1000 meets functional and/or performance specifications. Safety testing, including IEC 60601-2-4 for the "Particular Requirements for the Safety of Cardiac Defibrillators," IEC 60601-1-2 for "General Requirements for Safety - Collateral Standard: Electromagnetic compatibility - Requirements and Tests" and the ECG analysis and shock advisory system to the "American Heart Association's Automatic External Defibrillators for Public Access Defibrillation: Recommendation for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms and Enhancing Safety," assures compliance with applicable industry safety standards.
## Conclusion Summary of Safety and Effectiveness
Testing and performance evaluations demonstrate that the Defibtech DDU-1000 AED is substantially equivalent to the predicate device. The introduction of the DDU-1000 AED does not present new issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health and Human Services (HHS) in the United States. The logo features the HHS symbol, which is a stylized representation of a person embracing another person. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2014
Defibtech, LLC c/o Mr. Ed Horton Vice President, Quality and Regulatory 741 Boston Post Road Guilford, CT 06437
Re: K131525
Trade/Device Name: DDU-1000 Semiautomatic External Defibrillator and Accessories , Regulatory Number: 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: II (two) Product Code: MKJ Dated: March 11, 2014 Received: March 12, 2014
Dear Mr. Horton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
. . .:
{4}------------------------------------------------
Page 2 - Mr. Ed Horton
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.html
. .
. .
Image /page/4/Picture/7 description: The image shows the phrase "Sincerely yours," followed by a logo. The logo appears to be for the FDA, but it is crossed out with black lines. The logo is a stylized version of the letters FDA.
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# K131525
# Indications for Use
510(k) Number (if known):
Device Name: DDU-1000 Semiautomatic External Defibrillator and Accessories
Indications For Use:
The DDU-1000 Semiautomatic External Defibrillator (AED) is indicated for use on victims of sudden cardiac arrest (SCA) who are:
- · Unconscious and unresponsive
- · Not breathing
For patients under 8 years old, or less than 55 pounds (25kg), use child/infant electrode pads. Do not delay therapy to determine exact age or weight.
The DDU-1000 AED must be used by or on the order of a physician.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Date
2014.04.24
07:38:46 -04'00'
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.