K131468 · Green Dentech Co, Ltd. · NHA · May 6, 2014 · Dental
Device Facts
Record ID
K131468
Device Name
DENRACLE DENTAL ABUTMENT
Applicant
Green Dentech Co, Ltd.
Product Code
NHA · Dental
Decision Date
May 6, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Denracle Dental Abutment for Osstem GS Implant is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. This device is compatible with the following implant system: OSSTEM GS Fixture System Ø 4.0 mm, Ø 4.5 mm and Ø 5.0 mm.
Device Story
Denracle Dental Abutment is a titanium component used to connect dental implants to prosthetic restorations. The device consists of the abutment and a Ti-6A1-4V ELI titanium screw. It is used in a clinical dental setting by a dentist to support cement-retained single or multi-unit restorations. The abutment is secured to the OSSTEM GS Fixture System via the screw. By providing a stable interface between the endosseous implant and the prosthetic crown or bridge, the device restores masticatory function and aesthetics for edentulous patients.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including mechanical fatigue testing per ISO 14801 to ensure strength, and compatibility testing (dimensions, tolerances, rotation parameters) against the OSSTEM GS Fixture System. Biocompatibility and sterilization validation were also performed with acceptable results.
Technological Characteristics
Material: Ti-6A1-4V ELI titanium. Principle: Mechanical abutment-to-implant interface secured by a screw. Form factor: Dental abutment for OSSTEM GS Fixture System (Ø 4.0, 4.5, 5.0 mm). Connectivity: None. Sterilization: Provided sterile.
Indications for Use
Indicated for use with dental implants to support single or multiple tooth prostheses in partially or fully edentulous patients (maxilla or mandible). Compatible with OSSTEM GS Fixture System (Ø 4.0, 4.5, 5.0 mm).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
# reen DenTech Co., Ltd
2F. No. 17, Deyuanpi Blvd., Da Nong Li, Liouying District, Tainan City, 73659, Taiwan.
MAY 0 6 2014
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is prepared in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 on May 06, 2014.
The assigned 510(k) number is: K131468
## 1. APPLICANT's NAME AND ADDRESS
| Applicant' Name: | Green DenTech Co., Ltd |
|-------------------|-----------------------------------------|
| Address: | 2F, No. 17, Deyuanpi Blvd., Da Nong Li, |
| | Liouying District, Tainan City, 73659, |
| | Taiwan. |
| Telephone: | 886-6-6230999 |
| Fax: | 886-6-6230789 |
| Official contact: | Henry H.J. Shih |
| Date Prepared: | May 06, 2014 |
### 2. DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Denracle Dental Abutment |
|-----------------------------|-----------------------------------------------------------------|
| Common Name: | Dental implant abutment |
| Classification Regulations: | Endosseous dental implant abutment<br>Class II, 21 CFR 872.3630 |
| Product Code: | NHA |
| Classification Panel: | Dental Products Panel |
| Reviewing Branch: | Dental Devices |
## 3. LEGALLY MARKETED DEVICES TDS Abutment for Friadent Xive - K103339.
### 4. DEVICE DESCRIPTION
Denracle Dental Abutment for Osstem GS Implant are titanium abutments designed to be used in conjunction with specific dental implants utilizing the Dental Abutment screw, which is made of Ti-6A1-4V ELI titanium and is used to secure the abutment to the implant. In combination with the implant, the abutments support single or multi-unit cement-retained restorations in the maxillary and/or mandibular arch. Denracle Dental Abutment for Osstem GS Implant is compatible with the following implant systems: OSSTEM GS Fixture System 04.0 mm, Ø4,5 and Ø5,0mm.
#### INTENDED USE OF THE DEVICE 5.
Denracle Dental Abutment for Osstem GS Implant is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
This device is compatible with the following implant system: OSSTEM GS Fixture System 0 4.0 mm, Ø 4.5 mm and Ø 5.0 mm.
{1}------------------------------------------------
# T Green DenTech Co., Ltd
2F, No. 17, Deyuanpi Blvd., Da Nong Li, Liouying District, Tainan City, 73659, Taiwan.
# 6. TECHNOLOGICAL CHARACTERISTICS
Denracle Dental Abutment for Osstem GS Implant has the following similarities to the predicate devices, TDS Abutment for Friadent Xive, which have been determined by FDA:
- · Has the same intended use,
- . Use the same operating principle,
- . Incorporates the same basic design,
- . Incorporates the same materials, and
- Is produced using the same processes.
The basis for Green DenTech Co., Ltd. Belief that TDS Abutment for Friadent Xive is substantially equivalent to the predicate devices is summarized in the following table.
| | Predicate Devices | Subject Device |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Denracle Dental Abutment for | Pou Yu Biotechnology Co. |
| | Osstem GS implant | TDS Abutment for Friadent Xive<br>K103339 |
| Intended use | Denracle Dental Abutment for<br>Osstem GS Implant is intended for<br>use with dental implants as a<br>support for single or multiple tooth<br>prostheses in the maxilla or<br>mandible of a partially or fully<br>edentulous patient.<br><br>Denracle Dental Abutment for<br>Osstem GS Implant is compatible<br>with the following implant system:<br>OSSTEM GS Fixture System Ø 4.0<br>mm, Ø 4.5 mm and Ø 5.0 mm. | TDS Abutment for Friadent Xive is<br>intended for use with dental<br>implants as a support for single or<br>multiple tooth prostheses in the<br>maxilla or mandible of a partially or<br>fully edentulous patient.<br>TDS Abutment for Friadent Xive is<br>compatible with the following<br>implant systems which have an<br>internal hex 1.78mm or greater:<br>Firadent: FRIALIT Implant, Xive<br>Implant; 3i: Internal connect Type;<br>Astra: Osseospeed Implant,<br>Osseospeed TX Implant;<br>BioHorizons: Internal Implant<br>system, Tapered Internal Implant<br>System, Single-Stange Implant<br>System, Laser-lok 3.0 implant<br>RENOVA ™ Internal Hex Implant<br>System; Zimmer: Tapered<br>Screw-Vent Implant System,<br>Screw-Vent Implant System,<br>AdVent Implant System; Osstem:<br>GS system; Noble Biocare: Active<br>Impalnt. |
| Custom Design | | |
| Attachment | Implant level | Implant level |
| Restoration | Cement-retained | Cement-retained |
| CAD/CAM<br>processing | Not CAD/CAM millable | Yes |
| Material | | |
| Abutment | Ti-6A1-4V ELI | Ti-6A1-4V ELI<br>Y-TZP Zinconia |
| Screw | Provided by Green Dentech Co.,<br>Ltd. | Provided by Pou Yu Biotechnology<br>Co. Ltd |
{2}------------------------------------------------
reen DenTech Co., Ltd
2F, No. 17, Deyuanpi Blvd., Da Nong Li, Liouying District, Tainan City, 73659, Taiwan.
## 7. NON-CLINICAL TESTING DATA
Mechanical testing, according to ISO 14801 Dentistry - Fatigue test for endosseous dental implants, was conducted on a worst-case scenario to ensure that the strength Denral Abutment for Osstem GS Implant is appropriate for its intended use.
Compatibility testing was conducted on the abutments and corresponding dental implants (which have had previous 510(K) clearance in K072896) with designated screws, the dimensions, tolerances and rotation parameters were evaluated in determining appropriate fit.
Biocompatibility testing and sterilization was performed on the device with acceptable results.
These testing results show that Denracle Dental Abutment for Osstem GS Implant made of titanium, for their respective dental implant system have sufficient mechanical strength for their intended clinical application and are compatible with the implant system for which they are indicated for use.
## 8. CONCLUSION
Green DenTech Co., Ltd demonstrated that, for the purposes of FDA's regulation of medical devices, GO series Dental Abutment for Osstem GS Implant is substantially equivalent in indications and design principles to predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA".
Public Health Service
Food and Drug Administration 10903 New Harnoshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 6, 2014
Green DenTech Company Limited Ms. Sue Ru Official Correspondent 2F, No. 17. Deyuanpi Boulevard Danung Li, Liouying District Tainan 73659 TAIWAN
Re: K131468
Trade/Device Name: Denracle Dental Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: February 13, 2014 Received: March 24, 2014
Dear Ms. Ru:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Ru
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
MarySBunner-S
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
# Indications for Use
510(k) Number (if known) K131468
Device Name Denracle Dental abutment
Indications for Use (Describe)
manage Dental Abument for OSSTEM GS Implant is intended for use with dental implant as a support for single or multiple tooth prosthoses in maxilla or mandible of a partially or fully edentulous patient. This device is compatible with the following implant system:OSSTEM GS Fixture System 04.0mm > 04.5mm > 05.0mm •
Type of Use (Select one or both, as epplicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995. •DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of Information Is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chlef Information Officer Paparwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
PJC Tablinkang Sermans (J0 | ) 443-4740
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{6}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K131468
.
Device Name Denracle Dental abutment
Indications for Use (Describe)
morale Dental Abutunent for OSSTEM GS implant is intended for use with destablimplant as a support for single or multiple tooth prostheses in maxilla or mandible of a partially or fully edentulous patient. This device is compatible with the following implant system:OSSTEM GS Fixture System 64.0mm > ø4.5mm > ø5.0mm •
| Type of Use (Select one or both, as applicable) | <span> <b>Prescription Use (Part 21 CFR 801 Subpart D)</b> </span> <span> <b>Over-The-Counter Use (21 CFR 801 Subpart C)</b> </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | Sheena A. Green -S |
|------------------------------------------------------------------------------|---------------------|
| | 2014.05.06 12:54:45 |
| | -04'00' |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 78 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (1/14)
Page 1 of 1
PIC Publica: Bertuu (201) 417-674)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.