INTEGRA JARIT HEPARIN CANNULAS, INJECTORS, AND NEEDLES
Applicant
Integra LifeSciences Corporation
Product Code
DQR · Cardiovascular
Decision Date
Oct 28, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1300
Device Class
Class 2
Indications for Use
Integra® Jarit® Heparin Cannulas, Injectors, and Needles are used in conjunction with various syringe sizes (not supplied by Integra®) for flushing, irrigation and solution injection into a vessel or cavity during vascular, bypass or other cardiovascular surgical procedures.
Device Story
Manual stainless steel cannulas, injectors, and needles; used during cardiovascular surgeries (e.g., coronary artery bypass). Device connects to user-supplied syringe containing heparinized saline; used to confirm lack of leakage and ensure patency of harvested or grafted veins. Operated by surgeons in clinical/OR settings. Reusable; steam sterilizable. Provides physical conduit for fluid delivery to surgical site.
Clinical Evidence
No clinical data. Bench testing only: manual/mechanical cleaning validation (protein analysis), pre-vacuum steam sterilization validation, repeated autoclave/boiling water/corrosion/thermal testing (ISO 13402, ASTM F1089), and mechanical strength testing of laser-welded and brazed joints.
Technological Characteristics
Manual surgical instruments constructed of 303 and 304 stainless steel (ASTM F899-11, ISO 7153-1). Reusable; steam sterilizable per ANSI/AAMI ST79. Mechanical design; no energy source or software.
Indications for Use
Indicated for flushing, irrigation, and solution injection into vessels or cavities during vascular, bypass, or other cardiovascular surgical procedures. Used in conjunction with user-supplied syringes.
Regulatory Classification
Identification
A catheter cannula is a hollow tube which is inserted into a vessel or cavity; this device provides a rigid or semirigid structure which can be connected to a tube or connector.
{0}------------------------------------------------
K131446
# 510(k) Summary
| Submitted by: | Integra York PA, Inc.<br>589 Davies Drive<br>York, PA 17402 USA |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | OCT 2 8 2013 |
| Contact Person: | Stephanie Sheesley<br>Sr. Regulatory Affairs Manager<br>Integra York PA, Inc.<br>589 Davies Drive, York, PA 17402 USA<br>Phone: (717) 717-840-3522<br>Facsimile: (717) 840-9347 |
| Date Prepared: | August 15, 2013 |
| Device Trade Name: | Integra® Jarit® DeBakey Heparin Cannula, Integra® Jarit® Stoney<br>Heparin Injector, and Integra® Jarit® Heparin Flushing Needle |
| Device Common Name: | Cardiovascular cannulas, injectors, and needles |
| Part Numbers: | 310-379, 310-382, 310-383, 310-384 |
| Classification Name: | Cannula, Catheter |
| Device Class: | Class II |
| Product Code: | DQR |
| CFR Classification: | 21 CFR 870.1300 |
#### Device Description:
Integra® Jarit® Heparin Cannulas, Injectors, and Needles are manual, stainless steel catheter cannulas provided in various sizes, shapes, and lengths for use during cardiovascular surgical procedures such as coronary artery bypass surgery. The cannulas are intended to be connected to a syringe filled with heparinized saline to confirm there is no leakage as well as ensure patency of harvested and/or grafted veins. The syringes are not being supplied with the cannulas nor are the syringes offered by Integra; therefore they are not part of this submission. These reusable devices are packaged non-sterile and are steam sterilizable.
#### Indications For Use:
Integra® Jarit® Heparin Cannulas, Injectors, and Needles are used in conjunction with various syringe sizes (not supplied by Integra®) for flushing, irrigation and solution injection into a vessel or cavity during vascular, bypass or other cardiovascular surgical procedures.
{1}------------------------------------------------
### Predicate Devices:
| 510(k) # | Device | Manufacturer |
|--------------|-------------------------|-----------------------------|
| Preamendment | DeBakey Heparin Cannula | Pilling Co. |
| Preamendment | Stoney Heparin Injector | Pilling Co. |
| Preamendment | Heparin Flushing Needle | V. Mueller |
| K030788 | DeBakey Vessel Dilator | Geister Medizintechnik GmbH |
There are no differences in technology, materials, intended use, or design between the subject devices and the above predicates.
#### Performance Standards:
No performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, the Integrato Cannulas, Injectors, and Needles conform to the following standards:
- ANSI/AAMI ST79:2010 & A1:2010 Comprehensive guide to steam sterilization . and sterility assurance in health care facilities
- · ANSI/AAMI/ISO 14937:2009 Sterilization of health care products-General requirements for characterization of a sterilizing agent and the development, validation, and routine control of a sterilization process
- ASTM TIR 30:2003 A Compendium of processes, materials, test methods, and . acceptance criteria for cleaning reusable medical devices
- ASTM F 1089-02 Standard test method for corrosion of surgical instruments
- 0 ISO 13402:1995 Surgical and dental hand instruments - Determination of resistance against autoclaving, corrosion and thermal exposure
- . ISO 7153-1:2001 Surgical instruments- Metallic Materials- Part 1: Stainless Steel
- . ASTM F899-11 Standard Specification for Wrought Stainless Steels for Surgical Instruments
#### Non-Clinical Tests Submitted:
| Testing Performed | Results |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Manual Cleaning Validation (Protein Analyses) per AAMI TIR30:2003. | Pass |
| Mechanical Cleaning Validation (Protein Analyses) per AAMI TIR30:2003. | Pass |
| Pre-Vacuum (wrapped) Steam Sterilization Validation per ANSI/AAMI ST79:2010<br>& A1:2010 and ANSI/AAMI/ISO 14937:2009 at 270°F (132°C) with an Exposure<br>Time of 4 minutes and Drying Time of 30 minutes | Pass |
| Repeated Autoclave Testing, Boiling Water Testing, Copper Sulfate Corrosion<br>Testing and Thermal Testing per ISO 13402:1995 and ASTM F 1089-02. | Pass |
| Mechanical Strength Testing - Laser Weld Joint | Pass |
| Mechanical Strength Testing - Brazed Joint | Pass |
{2}------------------------------------------------
Integra® Jarit® Heparin Cannulas, Injectors, and Needles
No biocompatibility testing was performed on the proposed devices as the 304 and 303 Stainless Steel Materials are recognized biocompatible materials per ASTM F899-11 and ISO 7153-1 and have a long history of safe and effective use as shown in the predicate devices.
## Conclusions drawn from Non-Clinical Data:
All necessary testing has been performed on the Integra® Jarit® Heparin Cannulas, Injectors, and Needles and the results support the conclusion that the subject devices are substantially equivalent to the legally marketed predicate devices based on intended use, materials, technology, and design and as such, do not raise any concerns of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized emblem with three curved lines, resembling a human figure, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the emblem. The emblem is black, and the text is also black.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 28, 2013
Integra Lifesciences Corporation c/o Stephanie Sheesley Sr. Regulatory Affairs Manager Integra York PA, Inc. 589 Davies Drive York, PA 17402
Re: K131446
> Trade/Device Name: Integra® Jarit® DeBakey Heparin Cannula, Integra® Jarit® Stoney Heparin Injector, and Integra® Jarit® Heparin Flushing Needle
Regulation Number: 21 CFR 870.1300 Regulation Name: Cannula Catheter Regulatory Class: Class II Product Code: DQR Dated: August 28, 2013 Received: August 29, 2013
Dear Ms. Sheesley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Stephanie Sheesley
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M. A. Ellison
for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications For Use
510(k) Number (if known):
K131466
Device Name: Integra® Jarit® Heparin Cannulas, Injectors, and Needles
Indications for Use:
Integra® Jarit® Heparin Cannulas, Injectors, and Needles are used in conjunction with various syringe sizes (not supplied by Integra®) for flushing, irrigation and solution injection into a vessel or cavity during vascular, bypass or other cardiovascular surgical procedures.
Prescription Use > (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
M.A. Hillerman
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.