The VS3 Stereoscopic High Definition Vision System, Model VS3-NE is intended for viewing internal surgical sites during general surgical procedures and for use in visualization of ventricles and structures within the brain during neurological surgical procedures.
Device Story
VS3-NE is a stereoscopic high-definition vision system for surgical visualization. System components include a rigid stainless steel endoscope, light source, camera control unit (CCU), camera, and display monitors. The device utilizes a proximal HD camera concept with a stereoscopic camera block at the endoscope handle to capture 3D video. The optical array captures right and left images from different angles; the CCU processes these to generate a stereoscopic signal for display. The system also supports a 2D coupler for use with third-party 2D endoscopes. Used in OR settings by surgeons to view internal surgical sites, the system provides enhanced depth perception to assist in surgical navigation and decision-making. The device is intended for general and neurological surgical procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, including adherence to recognized consensus standards for electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and endoscopic equipment safety (IEC 60601-2-18).
Technological Characteristics
Rigid stainless steel endoscope (4-5.5 mm diameter, 175-300 mm length); proximal stereoscopic HD camera block; CCU for signal processing; 2D coupler for third-party scope compatibility. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-18 (endoscopic equipment), ISO 14971 (risk management), ISO 8600 (optics).
Indications for Use
Indicated for visualization of internal surgical sites during general surgical procedures and visualization of ventricles and brain structures during neurological surgical procedures.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
{0}------------------------------------------------
## K 131434
### 510(k) Summary
### Date Prepared: June 17, 2013
### Submitter
Visionsense Ltd 20 Hamaqshimim Street Petach Tikva 49348 lsrael
### Contact Person(s)
Sponsor:
Azi Ben-Yishai 20 Hamagshimim Street Petach, Tikva 49348 lsrael +972 (3) 9244339 Phone: Facsimile: +972 (3) 9244335
### OR
Regulatory Counsel: Gerard J. Prud'homme Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington, DC 20004 Tel: 202-637-5735 Fax: 202-637-5910
### Device Information
Trade name: VS3 Stereoscopic High Definition Vision System, Model VS3-NE Common name: Endoscope Classification Name: Endoscope, Neurological Review Panel: Neurology Product Code: GWG Device Class: Class II Regulation: 21 C.F.R. §882.1480
### Predicate Device Information
Visionsense Ltd VS』(K081102) Visionsense Ltd VS3 (K123467)
#### Intended Use/Indications for Use
The VS3 Stereoscopic High Definition Vision System, Model VS3-NE is intended for viewing internal surgical sites during general surgical procedures and for use in visualization of ventricles and structures within the brain during neurological surgical procedures.
### JUL 1 6 2013
{1}------------------------------------------------
### Technological Characteristics
The VS3 Stereoscopic High Definition Vision System, Model VS3-NE consists of the following components:
- . Endoscope
- Light source .
- Camera Control Unit (CCU) .
- Camera .
- Display monitors
- 2D Endoscope coupler ●
The VS3 Stereoscopic High Definition ("3DHD") Vision System, Model VS3-NE is based on the proximal HD camera concept with a stereoscopic camera block located on the proximal side of the endoscope (the handle). This allows high resolution capture of the 3D video stream. The stereoscopic images are transmitted from the visual field at the distal tip of the endoscope to the proximal camera block containing the HD sensor module. The VS3 Stereoscopic High Definition Vision System, Model VS3-NE allows for separation of the camera module with image sensor module and electronics from the endoscope shaft housing optical relay components and light fibers. The VS3 Stereoscopic High Definition Vision System, Model VS3-NE also includes 2D coupler capability that allows the VS3 Stereoscopic High Definition Vision System, Model VS3-NE to be used with FDA-cleared, third party 2D scopes at user sites to display monocular video.
### Principles of Operation
During surgical procedures, the surgeon inserts the endoscope into the surgical site, which is illuminated using the internal or external illumination source. The optical array then functions by capturing both right and left images of the surgical site from different angles. Both images are detected by the camera and transmitted to the CCU. Once the images are received by the CCU. the VS3 Stereoscopic High Definition Vision System. Model VS3-NE generates a stereoscopic signal of both the right and left images that can be sent to the display monitor.
### Performance
No performance standards or special controls have been developed under Section 514 of the Federal Food, Drug, and Cosmetic Act ("FD&C Act") for endoscopes. However, the VS3 Stereoscopic High Definition Vision System, Model VS3-NE and its components follow FDA recognized consensus standards for electrical safety, electromagnetic compatibility, and biocompatibility:
- 1. IEC 60601-1, Medical electrical equipment Part 1: General requirements for basic safety and essential performance (1998); Amendment 1, 1991-11, Amendment 2, 1995.
- IEC 60601-1-1, Medical electrical equipment Part 1-1: General requirements for 2. safety. Collateral standard: Safety requirements for medical electrical systems (2000).
- 3. IEC 60601-1-2, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance. Collateral standard: Electromagnetic compatibility requirements and tests. (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004).
- 4. IEC 60601-2-18, Medical electrical equipment Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment (1996). Amendment 1, 2000.
- 5. ISO 14971. Medical devices Application of risk management to medical devices (2007).
- 6. ISO 8600, Optics and optical instruments -- Medical endoscopes and
{2}------------------------------------------------
endoscopic accessories. Part 1:2005; Part 3:1997; Part 4:1997; Part 5:2005; Part 6:2005.
### Substantial Equivalence
The Visionsense VS3 Stereoscopic High Definition Vision System, Model VS3-NE has the same intended use and similar indications, principles of operation, and technological characteristics as the cleared VS, (K081102) and VS3 (K123467). Thus, the VS3 Stereoscopic High Definition Vision System, Model VS3-NE is substantially equivalent to its predicate devices.
The following table shows the similarities between the VS3-NE and predicate devices.
| | VS3-NE<br>(subject device) | VSII System (K081102)<br>(primary predicate) | VS3 for General Surgery<br>(K123467) (secondary<br>predicate) |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Visionsense Ltd. | Visionsense Ltd. | Visionsense Ltd. |
| Classification | Endoscope, Neurological<br>21 C.F.R. §882.1480<br>Product code GWG | Endoscope, Neurological<br>21 C.F.R. §882.1480<br>Product code GWG | Laparoscope, General &<br>Plastic Surgery<br>21 C.F.R. §876.1500<br>Product code GCJ |
| Indications for Use | Intended for viewing<br>internal surgical sites<br>during general surgical<br>procedures and for use in<br>visualization of ventricles<br>and structures within the<br>brain during neurological<br>surgical procedures | Intended for viewing<br>internal surgical sites<br>during general surgical<br>procedures and for use in<br>visualization of ventricles<br>and structures within the<br>brain during neurological<br>surgical procedures | Intended for viewing internal<br>surgical sites during general<br>endoscopic and laparoscopic<br>surgical procedures |
| Endoscope type | Rigid Stainless Steel | Rigid Stainless Steel | Rigid Stainless Steel |
| Endoscope<br>diameter | 4 - 5.5 mm | 4 - 5 mm | 4 - 5.5 mm |
| Endoscope length | 175 - 300 mm (±5 mm) | 175 - 300 mm (±5 mm) | 175 - 300 mm (±5 mm) |
| Field of view | 70°-95° | 70° | 70°-95° |
| Direction of view | 0° - 70° | 0° - 70° | 0° - 70° |
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized abstract symbol to the right and text arranged in a circular pattern to the left. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA".
#### July 16, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Visionsense Ltd c/o Mr. Gerard J. Prud'homme Hogan Lovells US LLP 555 Thirteenth Street. NW Washington, DC 20004
Re: K131434
Trade/Device Name: VS3 Stereoscopic High Definition Vision System, Model VS3-NE
Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological endoscope Regulatory Class: Class II Product Code: GWG Dated: June 17, 2013 Received: June 17, 2013
Dear Mr. Prud'homme
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Gerard J. Prud'homme
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Victor Krauthamer -S
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): _ K131434
Device Name: Visionsense VS3 Stereoscopic High Definition Vision System, Model VS3-NE
Indications For Use:
Intended for viewing internal surgical sites during general surgical procedures and for use in visualization of ventricles and structures within the brain during neurological surgical procedures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Victor Krauthamer -S 2013.07.15 18:35:33-04'00'
(Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
K131434 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.