K131433 · Isys Medizintechnik GmbH · JAK · Mar 11, 2014 · Radiology
Device Facts
Record ID
K131433
Device Name
ISYS 1 (U.S),ACCESSORY, NON-STERILE
Applicant
Isys Medizintechnik GmbH
Product Code
JAK · Radiology
Decision Date
Mar 11, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The iSYS 1 device is a user-controlled electromechanical arm with a needle guide. It is intended to assist the surgeon in the positioning of a needle or electrode where both computed tomography (CT) and fluoroscopic imaging can be used for target trajectory planning and intraoperative tracking. The needle or electrode is then manually advanced by the surgeon. Trajectory planning is made with software that is not part of the iSYS device.
Device Story
iSYS 1 is a modular, 4-DOF electromechanical robotic arm for interventional radiology. It functions as a remote-operated needle/electrode guidance platform. Input consists of trajectory planning data generated by external third-party navigation or imaging software (CT, cone-beam CT, fluoroscopy). The device uses fiducial markers for registration and localization within the imaging field. The system is mounted to the CT table or side rail via adapters. A physician operates the device via a cable-connected handheld control unit to position the needle guide. The device provides mechanical guidance for the needle, which the surgeon then manually advances. It is used in clinical settings to assist in procedures such as biopsies or fluid drainage. Benefits include precise, remote-controlled needle positioning and orientation, reducing manual effort and improving localization safety compared to traditional methods.
Clinical Evidence
Bench testing only. Testing included electrical and mechanical safety (EN 60601-1, EN 60601-1-8, ISO 60601-1-2), electromagnetic compatibility (FCC Part 15), and performance validation. Performance tests verified needle placement accuracy under clinical conditions, imaging modality compatibility, artifact production, mechanical stability under payload, and alarm system effectiveness. Accuracy was validated using Philips Allura Xper Guide FD 20 and Siemens Axiom Artis Zeego systems.
Technological Characteristics
4-DOF electromechanical arm; materials include metal, electronics, Polyamide, Polyethylene, and Prolen. Powered (115VAC/230VAC). Connectivity via cable-connected control unit. Mechanical accuracy < 1 mm. Fiducial marker-based registration. Sterile covers used for patient contact. Complies with IEC/EN 60601-1 and ISO 60601-1-2.
Indications for Use
Indicated for use by trained physicians to assist in the positioning of a needle or electrode during interventional procedures, utilizing CT or fluoroscopic imaging for trajectory planning and intraoperative tracking.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of several curved lines.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Decument Control Center - WO66-6608 Silver Spring, MD 20993-0002
March 11, 2014
iSYS Medizintechnik GmbH c/o Mrs. Cornelia Damsky 56 Wesicon Road Stamford. CT 06902
Re: K131433
Trade/Device Name: iSYS I · Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Date: February 11, 2014
Dear Mrs. Damsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordunce with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore. market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 331-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Page 2 - Mrs. Cornelia Damsky
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
### Joyce M. Whang -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
d
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K131433
Device Name isys 1
Indications for Use (Describe)
The ISYS I device is a user-controlled electromechanical arm with a needle guide. It is intended to assist the surgeon in the positioning of a needle or electrode where both computed tomography (CT) and fluoroscopic imaging cony planning and intraoperative tracking. The needle or electrode is the surgeon. Trajectory planning is made with software that is not part of the iSYS device.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
[] Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Joyce M. Whang -S
FORM FDA 3881 (1/14)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden lime for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gether and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Section |
|---------|
| 005 |
Image /page/4/Picture/2 description: The image shows a logo for iSYS Medizintechnik GmbH. The logo features the letters "SYS" in a stylized font, with the word "Medizintechnik" written below it in a smaller font. The letters are underlined with a curved line. The letters "GmbH" are written below the word "Medizintechnik".
### 510(k) SUMMARY
| 1. | Applicant: | iSYS Medizintechnik GmbH |
|-----|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Address: | Bergwerksweg 21<br>6370 Kitzbühel / Austria |
| 3. | Contact Person: | Dr. Michael Vogele<br>Tel. +43 (0) 664 2411140 |
| 4. | Preparation Date: | January 20, 2014 |
| 5. | Device Submitted: | iSYS 1 (U.S.) |
| 6. | Proprietary Name: | iSYS 1 (U.S.) |
| 7. | Common Name: | Robotic Positioning Unit |
| 8. | Classification Name: | System, X-ray, Tomography, computed<br>Product Code JAK, MAXX, Reg.No. 892.1750 |
| 9. | Substantial Equivalence: | The iSYS 1 is substantially equivalent to the following<br>legally marketed devices:<br>PinPoint manufactured by Philips Medical.<br><br>The characteristics of this device are similar to those of<br>the predicate devices identified on the comparison<br>chart, which follows. It is our opinion that the iSYS 1<br>does not have technological characteristics that raise<br>additional types of questions related to terms of safety<br>and effectiveness. |
| 10. | Device Description: | The iSYS 1 is a modular needle guidance platform for<br>interventional radiology and related fields. Core<br>components are a 4 DOF micro positioning unit which<br>allows the submillimetric needle positioning from<br>simple needle angulations up to positioning and<br>angulations with adjustable pivot point and a control<br>unit which is directed by a cable connected control<br>panel. The passive macro positioning unit and different<br>table adapters allow different setups of the system<br>around the patient in the region of interest. The needle-<br>guide-kit (manufactured by ECOLAB) provides<br>disposable components that ensure precise and sterile<br>needle guidance.<br><br>Planning of the tool position/orientation as well as<br>validation of the correctness of the tool position must<br>be performed with an external planning and<br>measurement system which is not part of the iSYS-1<br>Interventional Platform. The position of iSYS 1 is<br>visible for most imaging systems due to the used<br>markers. During treatment the tool is controlled by the |
| | | ------------------<br>-------------<br>---------------------------------- | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|---------------------------------------|
| | ----------------------------------------------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1 | | |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | with county county and parts and<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | <br>------------------- | | ------------------------------------- |
{5}------------------------------------------------
| Section<br>005 | | 510k Summary<br>510(k) Number: K131433<br>Image: [logo] |
|----------------|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 11. | Intended Use: | user.<br>The intended use of the iSYS1 device is to function as a remote-operated positioning and guidance system during interventional procedures. Positioning is done in remote control manner; planning of the position/angulation is done based on 2D/3D patient data (CT, cone-beam CT, fluoroscopy) by external planning software - for example using an external navigation system, or planning software coming with the used imaging device. Also verification of the correct position and orientation of the tool prior to/during/after the intervention is done by means of these external devices. The iSYS1-System is then acting as a guideway during the manual insertion of the interventional tool - usually a needle type device, and the like. |
| 12. | Indications for Use | The iSYS 1 device is a user-controlled electromechanical arm with a needle guide. It is intended to assist the surgeon in the positioning of a needle or electrode where both computed tomography (CT) and fluoroscopic imaging can be used for target trajectory planning and intraoperative tracking. The needle or electrode is then manually advanced by the surgeon. Trajectory planning is made with software that is not part of the iSYS device |
| 13. | Clinical Use | The Side Rail Adapter is attached to the side rail for the setup; the Table Top Adapter is attached directly onto the CT table. Both Adapters are equipped with a "starburst" connector, to which the Multifunctional Arm (MFA) can be attached. The MFA is equipped with one "starburst" connector at the bottom and one "spoon" connector at the top. The "spoon" connector has its counterpart on the Robotic Positioning Unit for the setup. The Control Unit is attached directly to the side rail. The Robotic Positioning Unit is connected by a cable to the Control Unit, which has cable connections to both the electric power supply and to a Handheld Control Unit. The Needle Guide Extensions are screw-fixed onto the Robotic Positioning Modules. The Sterile Cover is directly fixed onto the Needle Guide Joints and can be drawn over the robotic parts. The Needle Insert is placed into the appropriate connector of the Needle Guide, to prepare needle usage. The Handheld Control Unit can be clamped onto the side rail and covered with a drape. Planning of the tool position/orientation as well as validation of the correctness of the tool position must be performed with an external planning and measurement system |
| Doc No: PD_75_11535 | Revision 06 from: 10.03.2014 | 2/9 | | |
|---------------------|------------------------------|----------------|------------|----------------|
| Created by: | Approved by: | Invalid since: | Source: | Approval Date: |
| Oliver Spitz | Dr. Michael Vogele | | FO_42_1996 | 06.06.2011 |
{6}------------------------------------------------
| Section<br>005 | 510k Summary<br>510(k) Number: K131433 | Image: SYS Medizintechnik GmbH logo |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| 14. Biocompatibility: | (imaging or navigation software).<br>The iSYS 1 is not in contact with patient. At any time when in use a sterile cover sheet is to be placed between the patient and the system. Additionally there are no new materials introduced in the manufacture of the iSYS 1. Therefore, no biocompatibility studies were performed for this device. | |
| 15. Performance Data: | Operating temperature/humidity range: 15 to 30°C; 30 to 70% relative humidity with no condensation<br><br>Storage temperature/humidity range: 10 to 50°C; 30 to 70% relative humidity with no condensation<br><br>Power Supply: 50W; 115VAC/230VAC; 50-60Hz; cable: C14 according to IEC/EN 60320-1, US498, CSA C22.2 no. 42;<br><br>Sterilization of the sterile accessories has been validated by Bioseal and Preferred Medical, the companies responsible for packaging and sterilizing the iSYS accessories.<br><br>The accuracy of the intervention depends on the resolution and capabilities of the imaging device or software. The mechanical accuracy of iSYS1 is below 1 mm, but that gives no accuracy value for treatments | |
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
| Doc No: PD_75_11535 | Revision 06 from: 10.03.2014 | | | 3/9 |
|---------------------|------------------------------|----------------|------------|----------------|
| Created by: | Approved by: | Invalid since: | Source: | Approval Date: |
| Oliver Spitz | Dr. Michael Vogele | | FO_42_1996 | 06.06.2011 |
{7}------------------------------------------------
### Device Comparison Summary:
| Reference | PinPoint | iSYS 1 (for new submission) | Equivalence ? |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|---------------|
| 510k-Number | K974513 | K131433 | |
| Manufacturer | PHILIPS MEDICAL SYSTEMS<br>(CLEVELAND), INC.<br>595 Miner rd.<br>Cleveland, OH 44143<br>USA | iSYS Medizintechnik<br>GmbH<br>Bergwerksweg 21<br>A-6370 Kitzbuehel<br>Austria | |
| Design | | | |
| General device description | Five jointed, position-sensing stereotactic arm, mounted on the CT gantry, and indicated for invasive procedures | Computer controlled electromechanical multi-joined arm indicated for invasive procedures | Yes |
| Localization means | Robot arm absolute encoders | Fiducial markers on tool holder. | *3) |
| Image-guided | Yes | Yes | Yes |
| Planning software | No, not cleared under this 510k (K955268) | No (third party) | Yes |
| Registration method | During installation | Fiducial markers | *4) |
| Instrumentation | Cannula<br>Laser needle guide | Marker<br>Tool Holder<br>Sterile Covers<br>(third party) | Yes |
| Instrument fixation | Cannula attached to the arm (optional: laser needle guidance) | Special tool holders for several applications mounted to the Robot | Yes |
| Instrument calibration | Factory Calibration<br>Optional: intra operative | Intra operative | *1) |
| System immobilization between patient and device | Yes | Yes | Yes |
| Planning and Navigation Software | | | |
| N/A. | | | |
| System Operation | | | |
| Fiducial markers registration with | No | No | Yes |
| Doc No: | PD 75 11535 | Revision 06 from: | 10.03.2014 | | 4/9 |
|-------------|--------------|-------------------|--------------------|----------------|------------|
| Created by: | Oliver Spitz | Approved by: | Dr. Michael Vogele | Invalid since: | |
| | | | | Source: | FO 42 1996 |
| | | | | Approval Date: | 06.06.2011 |
-
{8}------------------------------------------------
| Section |
|---------|
| 005 |
#
Image /page/8/Picture/2 description: The image shows a logo for iSYS Medizintechnik GmbH. The logo features the text "iSYS" in a stylized font, with two diagonal lines above the "i". Below "iSYS" is the word "Medizintechnik" in a smaller, sans-serif font. Underneath that is "GmbH", also in a smaller font.
A
| Reference | PinPoint | iSYS 1 (for new<br>submission) | Equivalence<br>? |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| pointer probe | | | |
| Optical<br>registration | No | No | Yes |
| Ultrasound<br>Registration | No | No | Yes |
| Accuracy<br>verification | Yes, external anatomical<br>verification | Yes, performed by user | Yes |
| Provide<br>mechanical<br>guidance for tools | Yes | Yes | Yes |
| Instrument guide<br>position<br>adjustment | Manual | Manual | Yes |
| Physician carries<br>out final gesture<br>through tool guide | Yes | Yes | Yes |
| Indication for use | The PinPoint is an<br>accessory to a CT<br>system intended to<br>provide the radiologist<br>with a means of<br>simulating and initiating<br>interventional<br>procedures by<br>interactively relating the<br>patient's CT image<br>volume to the actual<br>target field.<br>This accessory includes<br>mounting the<br>stereotactic arm on a<br>CT-Gantry, a flat panel<br>TV monitor, cabling,<br>biopsy phantom and<br>software.<br>In addition the CT is<br>intended to produce<br>cross-sectional images<br>of the body by computer<br>reconstruction of X-ray<br>transmission data from | The iSYS 1 device is a<br>user-controlled<br>electromechanical arm<br>with a needle guide. It is<br>intended to assist the<br>surgeon in the<br>positioning of a needle or<br>electrode where both<br>computed tomography<br>(CT) and fluoroscopic<br>imaging can be used for<br>target trajectory planning<br>and intraoperative<br>tracking. The needle or<br>electrode is then<br>manually advanced by<br>the surgeon. Trajectory<br>planning is made with<br>software that is not part<br>of the iSYS device. | Yes |
| Doc No: | PD_75_11535 | Revision 06 from: 10.03.2014 | | 5/9 |
|--------------|--------------------|------------------------------|------------|----------------|
| Created by: | Approved by: | Invalid since: | Source: | Approval Date: |
| Oliver Spitz | Dr. Michael Vogele | | FO_42_1996 | 06.06.2011 |
{9}------------------------------------------------
## Section
#
Image /page/9/Picture/2 description: The image shows a logo for iSYS Medizintechnik GmbH. The logo features the letters "SYS" in a stylized font, with the word "Medizintechnik" written in a smaller font below it. The letters are accompanied by a small circle above the "I" and two diagonal lines to the left of the letters. The letters "GmbH" are written in a smaller font below the word "Medizintechnik".
| Reference | PinPoint | iSYS 1 (for new<br>submission) | Equivalence<br>? |
|--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| | the same axial plane<br>taken at different angles.<br>This device may include<br>signal analysis and<br>display equipment,<br>patient and equipment<br>supports, components<br>and accessories.<br>PinPoint is not limited to<br>any particular region of<br>the body. It is equally<br>viable for planning head<br>and spine interventions<br>as it is in the abdomen.<br>It is expected that its<br>major use will be in the<br>planning of biopsies of<br>abdominal organs and<br>drainage of fluid<br>collections in the<br>abdomen. However,<br>more complicated<br>procedures such as<br>Brachy therapy and<br>bone pinnings will be<br>planned using the<br>PinPoint. | | |
| Intended Use | The PinPoint is an<br>accessory to a CT<br>system intended to<br>provide the radiologist<br>with a means of<br>simulating and initiating<br>interventional<br>procedures by<br>interactively relating the<br>patient's CT image<br>volume to the actual<br>target field.<br>The system produces<br>cross sectional images<br>by computer<br>reconstruction of x-ray<br>transmission data from<br>the same axial plane<br>taken at different angles.<br>A stereotactic arm on a | The intended use of the<br>iSYS1 device is to<br>function as a remote-<br>operated positioning and<br>guidance system during<br>interventional<br>procedures. Positioning<br>is done in remote control<br>manner; planning of the<br>position/angulation is<br>done based on 2D/3D<br>patient data (CT, cone-<br>beam CT, fluoroscopy)<br>by external planning<br>software - for example<br>using an external<br>navigation system, or<br>planning software<br>coming with the used<br>imaging device. Also | Yes |
| Doc No: PD_75_11535 | Revision 06 from: 10.03.2014 | | | 6/9 |
|---------------------|------------------------------|----------------|------------|----------------|
| Created by: | Approved by: | Invalid since: | Source: | Approval Date: |
| Oliver Spitz | Dr. Michael Vogele | | FO_42_1996 | 06.06.2011 |
{10}------------------------------------------------
| Section<br>005 | 510k Summary<br>510(k) Number: K131433 | | |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Reference | PinPoint | iSYS 1 (for new<br>submission) | Equivalence<br>? |
| | CT Gantry, a flat panel<br>TV monitor, and biopsy<br>phantom are used as<br>accessories to produce<br>these images. The<br>primary intend of the<br>system is to provide<br>guidance for biopsies in<br>the abdomen, spine and<br>head. | verification of the correct<br>position and orientation<br>of the tool prior<br>to/during/after the<br>intervention is done by<br>means of these external<br>devices. The iSYS1-<br>System is then acting as<br>a guideway during the<br>manual insertion of the<br>interventional tool –<br>usually a needle type<br>device, and the like. | |
| Anatomical site | Total Body | Total body | Yes |
| User | Radiologist | Physician | *2) |
| Accessory | V-Channel Cannula<br>Cylindrical Cannula | Sterile Covers<br>Table Adapters<br>Cable Sets | Yes |
| Real-time<br>instrument<br>position | Yes | Yes | Yes |
| Mechanical<br>Guidance of<br>instruments | Yes | Yes | Yes |
| Technology | | | |
| Powered | Yes | Yes | Yes |
| CE-Conformity | No | Yes | Yes |
| Computer-<br>controlled | Yes | Yes | Yes |
| Materials | Metal, electronics and<br>plastics | Metal, electronics<br>Polyamide<br>Polyethylene<br>Prolen | Yes |
.
:
| Doc No: PD 75_11535 | Revision 06 from: 10.03.2014 | | | 7/9 |
|---------------------|------------------------------|----------------|------------|----------------|
| Created by: | Approved by: | Invalid since: | Source: | Approval Date: |
| Oliver Spitz | Dr. Michael Vogele | | FO_42_1996 | 06.06.2011 |
{11}------------------------------------------------
Image /page/11/Picture/2 description: The image shows a logo for a company called "SYS Medizintechnik GmbH". The logo features the letters "SYS" in a stylized font, with the word "Medizintechnik" written below it in a smaller font. The letters are partially encircled by a curved line. The letters are also preceded by two diagonal lines.
### Similarities and Differences:
All three devices are intraoperative instruments used by physician for assisting the spatial positioning and orientation of a surgical tool.
iSYS 1 is equivalent to:
- . PinPoint for assisting the spatial positioning and orientation of a surgical tool.
- PinPoint for being localized by the navigation or imaging software,
- PinPoint for being guided by an imaging device .
- PinPoint for providing a registration method for the imaging device. .
- PinPoint for instrumentation and instrument fixation. ●
- . PinPoint for being immobilized in relation to the patient.
- PinPoint for providing accuracy verification, tool guidance and position ● adiustment.
- . PinPoint for the physician making the final gesture.
- . PinPoint for having the same indications for use and region of interest.
### Differences:
*1) PinPoint provides a factory instrument calibration of the encoders. During intervention the physician using iSYS1 and PinPoint verifies the position of the needle or needle-type instrument (tool). Verification of the tool position requires greater physical effort but provides for exact (depending on the imaging system) tool localization. The localization of the tool is not required for the iSYS 1 and, thus, this has no effect on safety or effectiveness.
*2) While the PinPoint is indicated to be used by radiologists only, the iSYS 1 is indicated to be used by any trained physician. This has no effects on safety and effectiveness.
*3) PinPoint provides encoders for localization, whereas iSYS uses fudicial markers on the tool. Using fiducial markers provides better localization safety.
*4) While the Pinpoint is registered during installation, the iSYS 1 uses fiducial markers for registration. Again, using fiducial markers provides better localization safety.
#### Non-Clinical Performance Data:
Non-clinical testing has mainly been performed to prove electrical and mechanical safety of the devices. Similar to the predicate devices, iSYS 1 was tested according to EN ISO 60601-1 2nd edition:
- EN 60601-1:1990 + A1:1993 + A2:1995 .
- EN 60601-1-8:2004 + A1:2006 .
The Electromagnetic Compatibility of the iSYS1 System has been tested according to FCC Part15 (Edition 1st October 2010) and ISO 60601-1-2:2007 and meets the acceptance criteria of both.
Performance tests have been performed to prove:
- the accuracy of the needle placement under clinical conditions, .
- the compatibility with the imaging modality, .
- the tendency to produce artifacts, .
| Doc No: PD_75_11535 | Revision 06 from: 10.03.2014 | | 8/9 | |
|---------------------|------------------------------|----------------|------------|----------------|
| Created by: | Approved by: | Invalid since: | Source: | Approval Date: |
| Oliver Spitz | Dr. Michael Vogele | | FO_42_1996 | 06.06.2011 |
{12}------------------------------------------------
Image /page/12/Picture/2 description: The image shows the logo for iSYS Medizintechnik GmbH. The logo features the letters "iSYS" in a stylized font, with two diagonal lines above the letters. Below the letters, the word "Medizintechnik" is written in a smaller font, and below that, the letters "GmbH" are written in an even smaller font. A curved line is drawn below the letters, giving the impression of a base or foundation.
- the mechanical stability of the system despite external payload under worst ● conditions
- the mechanical stability of the table adapters despite external payload, .
- . the accuracy in remote-control mode
- the accuracy of needle guidance ●
- functionality of hardware modules
- . effectiveness of hardware safety measures for the software,
- and the effectiveness of the alarm systems .
The results of these performance tests support that the predicate devices and iSYS 1 are substantially equivalent.
### Conclusion:
It is our opinion that the iSYS 1 System (U.S.) does not have technological characteristics that raise additional types of safety or effectiveness questions, and we consider them to be an enhancement to the existing devices.
To validate the accuracy of the iSYS 1 an external reference system has to be used. We have used and validated the accuracy of the iSYS 1 System (U.S.) with the Philips Allura Xper Guide FD 20 X-ray system and the Axiom Artis Zeego from Siemens Healthcare, and passed all tests.
| Doc No: PD 75 11535 | Revision 06 from: 10.03.2014 | | | 9/9 |
|---------------------|------------------------------|----------------|------------|----------------|
| Created by: | Approved by: | Invalid since: | Source: | Approval Date: |
| Oliver Spitz | Dr. Michael Vogele | | FO 42 1996 | 06.06.2011 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.