LOTUS

K131427 · Spinal Elements, Inc. · KWP · Oct 24, 2013 · Orthopedic

Device Facts

Record IDK131427
Device NameLOTUS
ApplicantSpinal Elements, Inc.
Product CodeKWP · Orthopedic
Decision DateOct 24, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Spinal Elements Lotus Posterior Cervical/Thoracic Spinal System is intended for posterior fixation of the cervical and thoracic spine (C1-T3) for the following conditions: degenerative disc disease (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies): spondylolisthesis; spinal stenosis; trauma (fracture/dislocation); failed previous fusion; tumors. The Spinal Elements Lotus Posterior Cervical/Thoracic Spinal System may be used in conjunction with the Spinal Elements Mercury system for transition at T1 - T3. The use of polyaxial pedicle screws is limited to placement in TI-T3 for treating thoracic conditions only. The screws are not intended to be placed in the cervical spine.

Device Story

Lotus Posterior Cervical/Thoracic Spinal System consists of fixation hardware including screws, hooks, and rods; used to stabilize spine segments. Surgeons assemble constructs to match patient anatomy and pathology. System components mechanically join to provide fixation. Polyaxial pedicle screws are limited to T1-T3 thoracic use. Device may be used with Spinal Elements Mercury system for T1-T3 transitions. Intended for use in clinical settings by surgeons. Benefits include spinal stabilization for degenerative disease, trauma, or tumors.

Clinical Evidence

Bench testing only. Testing included static compression bending, dynamic compression bending, and static torsion per ASTM F1717, as well as torsion and axial grip testing per ASTM F1798.

Technological Characteristics

Constructs consist of titanium alloy (Ti-6Al-4V) components conforming to ASTM F1472 and F136. System utilizes mechanical fixation via screws, hooks, and rods. Dimensions vary by configuration. No software or electronic components.

Indications for Use

Indicated for patients requiring posterior fixation of the cervical and thoracic spine (C1-T3) for degenerative disc disease, spondylolisthesis, spinal stenosis, trauma (fracture/dislocation), failed previous fusion, or tumors. Polyaxial pedicle screws are restricted to T1-T3 for thoracic conditions only and are contraindicated for cervical spine placement.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ OCT 2 4 2013 Spinal Elements, Inc. Premarket Notification - Spinal Elements Lotus® Posterior Cervical/Thoracic Spinal System ### 510(k) Summary Lotus" #### 510(k) Number K131427 ### Manufacturer Identification Submitted by: Spinal Elements, Inc. 3115 Melrose Drive, Suite 200 Carlsbad, CA 92010 760-607-0121 **Contact Information:** Benjamin A. Kimball Regulatory Affairs Manager Spinal Elements, Inc. 3115 Melrose Drive, Suite 200 Carlsbad. CA 92010 760-607-1816 Bkimball@spinalelements.com Date Prepared: Proprietary Name Device Classification Device Product Code Common Name May 15, 2013 # Lotus® Spinal Interlaminal Fixation and Spinal Intervetebral Fixation Orthosis and/or Pedicle Screw System 21 CFR 888.3050, and / or 21 CFR 888.3070 Proposed Regulatory Class Class II KWP, MNI. MNH #### Purpose of this Traditional 510(k) This 510(k) seeks clearance for a line addition to the previously cleared (K120467) Lotus system and a minor modification to the indication for use by adding language relative to the device's ability to be used in conjunction with Spinal Elements' Mercury® system. #### Device Description The Spinal Elements Lotus Posterior Cervical/Thoracic System consists of a variety of fixation devices that are attached to the spine by means of screws and hooks placed in/or on the pedicles or posterior elements of the various vertebrae, rods that span the distance between the screws/hooks. Screws are intended for attachment to the thoracic (T1-T3) spine only. The system achieves fixation by the mechanical joining of the rods, screws, and hooks. A variety of constructs may be assembled to suit the individual pathology and anatomy of the patient. {1}------------------------------------------------ #### Spinal Elements, Inc. ## Premarket Notification - Spinal Elements Lotus® Posterior Cervical/Thoracic Spinal System #### Connectors Connectors are offered in a variety of sizes and configurations and are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F 1472 and / or F 136. The following connectors are being added to the system: - Offset connector . - Round offset connector . - Axial rod to rod connector . - 03.5mm Axial rod to rod connector o - 0 5.5mm Axial rod to rod connector . - Parallel rod to rod connector . - . 0 3.5mm Parallel rod to rod connector - 0 5.5mm Parallel rod to rod connector ● - Transition rods . #### Intended Use of the Device The Spinal Elements Lotus Posterior Cervical/Thoracic Spinal System is intended for posterior fixation of the cervical and thoracic spine (C1-T3) for the following conditions: degenerative disc disease (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies): spondylolisthesis; spinal stenosis; trauma (fracture/dislocation); failed previous fusion; tumors. The Spinal Elements Lotus Posterior Cervical/Thoracic Spinal System may be used in conjunction with the Spinal Elements Mercury system for transition at T1 - T3. The use of polyaxial pedicle screws is limited to placement in TI-T3 for treating thoracic conditions only. The screws are not intended to be placed in the cervical spine. #### Performance Data The following testing was performed: - Static Compression Bending per ASTM F1717 . - Dynamic Compression Bending per ASTM F1717 . - Static Torsion testing per ASTM F1717 . - Torsion testing per ASTM F1798 o - . Axial grip testing per ASTM F1798 #### Substantial Equivalence Spinal Elements Posterior Cervical/Thoracic Spinal System is substantially equivalent to the previously cleared Spinal Elements Posterior Cervical/Thoracic Spinal System (Lotus) cleared in 510(k) K 120467. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 24, 2013 Spinal Elements, Incorporated Mr. Benjamin A. Kimball Regulatory Affairs Manager 3115 Melrose Drive, Suite 200 Carlsbad, California 92010 Re: K131427 Trade/Device Name: Lotus® Posterior Cervical/Thoracic Spinal System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP, MNI, MNI, MNI-I Dated: September 25. 2013 Received: September 26. 2013 Dear Mr. Kimball: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Benjamin A. Kimball forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Sincercly yours, # Ronald P. Jean -S for - Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use #### K131427 510(k) K: Spinal Elements Lotus® Posterior Cervical / Thoracic Spinal System Device Name: ## Indications for Use: The Spinal Elements Lotus Posterior Cervical/Thoracic Spinal System is intended for posterior fixation of the cervical and thoracic spine (CI-T3) for the following conditions: degenerative disease (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies); spondylolisthesis; spinal stenosis; trauma (fracture/dislocation); failed previous fusion; tumors. The Spinal Elements Lotus Posterior Cervical/Thoracic Spinal System may be used in conjunction with the Spinal Elements Mercury system for transition at TI - T3. The use of polyaxial pedicle screws is limited to placement in TI-T3 for treating thoracic conditions only. The screws are not intended to be placed in the cervical spine. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Ronald P. Jean -S Page 1 of 1 (Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K131427
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...