K131415 · Neurotronics, Inc. · OLV · Aug 9, 2013 · Neurology
Device Facts
Record ID
K131415
Device Name
LIVE VIEW PANEL (LVP)
Applicant
Neurotronics, Inc.
Product Code
OLV · Neurology
Decision Date
Aug 9, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The Live View Panel is an accessory to EEG and PSG systems, intended to transmit EEG or PSG data to a remote location where the data can be viewed by a clinician to aid in diagnosis of epilepsies, sleep disorders, and other related disorders. The Live View Panel is intended for use by medical personnel within a medical facility, clinic or nursing home, or outside a medical facility under the supervision of a medical professional. The Live View Panel is available for use on all patient populations, including adults and children as determined by the medical professional. This device is not to be used to monitor life threatening situations. This device is not to be used to control life supporting or life sustaining devices.
Device Story
Live View Panel (LVP) is a software-only accessory for EEG/PSG systems; enables remote, real-time viewing of digitized physiological waveforms and patient data. System architecture includes: Acquisition Interface Server (AIS) on the recording machine; client program for grid-based display; Monitor Controller for managing multi-monitor banks. Operates on off-the-shelf computers; transmits data over networks to remote locations. Used by clinicians in medical facilities, clinics, or nursing homes to track multiple patients simultaneously. Facilitates remote review of EEG/PSG data; aids clinical diagnosis of epilepsy and sleep disorders. Does not connect directly to patients; relies on cleared EEG/PSG hardware for data acquisition. Not for life-critical monitoring.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Software-only device; runs on off-the-shelf 32-bit Windows computers. Functions as a networked interface for remote display of EEG/PSG waveforms and vital signs (ECG, EMG, Respiration, SpO2, CO2). Supports up to 64 EEG or 32 PSG waveforms simultaneously. Connectivity via standard network infrastructure. No patient-contacting components.
Indications for Use
Indicated for use by medical personnel to remotely view EEG or PSG data to aid in the diagnosis of epilepsy, sleep disorders, and related conditions. Suitable for all patient populations, including adults and children. Contraindicated for monitoring life-threatening situations or controlling life-supporting/sustaining devices.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Nihon Kohden EEG-1200A with JE-120A Multi Channel Electrode Junction Box (K113117)
Submission Summary (Full Text)
{0}------------------------------------------------
# પાડ
# 510(k) Summary
| SUMMARY DATE: | | AUG 09 2013 |
|-------------------------------------------------------------------------------|-------------------------------------------------------|-------------------------------------------|
| Monday May 13, 2013 | | |
| 510K SUBMITTER: | | |
| Neurotronics, Inc.<br>3600 NW 43rd Street, Suite F1<br>Gainesville, FL, 32606 | | Phone: 352.372.9955<br>Fax : 815.550.2871 |
| Primary Contact: | | |
| Name: David Pezet<br>Email: david.pezet@neurotronics.com | Title: Quality Manager<br>Phone: 352.372.9955 Ext 355 | |
| Establishment Registration Number:<br>1063925 | | |
DEVICE TYPE (COMMON NAME):
Non-Normalizing Quantitative Electroencephalograph Software
# PROPRIETARY NAME OF THE DEVICE:
Live View Panel (LVP)
{1}------------------------------------------------
# CLASSIFICATION:
:
Product Code:
| Product Code | Device | Regulation Description | Regulation Number |
|--------------|------------------------------------------------|------------------------|-------------------|
| GWQ | Full-Montage Standard<br>Electroencephalograph | Electroencephalograph | 882.1400 |
:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# Subsequent Product Code:
| Product<br>Code | Device | Regulation<br>Description | Regulation<br>Number |
|-----------------|----------------------------------------------------------------------------------------|---------------------------|----------------------|
| OLZ | Automatic Event Detection Software For<br>Polysomnograph With<br>Electroencephalograph | Electroencephalograph | 882.1400 |
| OLV | Standard Polysomnograph With<br>Electroencephalograph | Electroencephalograph | 882.1400 |
| OLT | Non-Normalizing Quantitative<br>Electroencephalograph Software | Electroencephalograph | 882.1400 |
## PREDICATE DEVICES
| Submitter/Holder | Device Name | Model | 510(K) |
|----------------------------|---------------------------------------------------------------------------------|-----------|---------|
| Nihon Kohden America, Inc. | Nihon Kohden EEG-1200A Series<br>Neurofax | EEG-1200A | K080546 |
| Neurotronics, Inc. | Polysmith Sleep System | NTI5498 | K062943 |
| Nihon Kohden Corp. | PSG-1100 Sleep Diagnostic System | PSG-1100 | K120888 |
| Nihon Kohden America, Inc. | Nihon Kohden EEG-1200A with JE-<br>120A Multi Channel Electrode<br>Junction Box | JE-120A | K113117 |
{2}------------------------------------------------
#### DESCRIPTION
The Live View Panel (LVP) consists of several software components that allow remote viewing of digitized physiological waveforms and general patient information for the purpose of viewing EEG (electroencephalography) or PSG (polysomnography) patients in real time or in review. The software uses off-the-shelf and custom software components to manage multiple computer displays to facilitate the viewing of several patients simultaneously.
The Live View Panel consists of three software components. The Live View Panel Acquisition Interface Server (AIS) located on the acquisition system collects patient data from a recording machine and transmits this information over a network to a client machine. The Live View Panel client program receives this information and displays basic information in grid pattern along with other patients. The Live View Panel Monitor Controller program manages a bank of displays, each displaying a single patient. The Live View Panel grid mimics the physical locations of the monitors to allow the clinician manage and keep track of multiple patients displayed on the bank of displays.
The Live View Panel is software only and does not connect to the patient except through a cleared and currently marketed EEG or PSG device. A software module resides on the EEG or PSG device and sends patient data to a remote computer where another software module displays the data on an array of monitors. The software runs on off-the-shelf computers.
Image /page/2/Figure/4 description: The image shows a diagram of a system with multiple components. On the left side, there are four acquisition stations connected to a network switch. The data from the network switch is then transmitted to two live view panel (LVP) stations, each with its own network switch. Additionally, the data is sent via the internet to branch locations.
{3}------------------------------------------------
#### INTENDED USE
The Live View Panel is an accessory to EEG and PSG systems, intended to transmit EEG or PSG data to a remote location where the data can be viewed by a clinician to aid in diagnosis of epilepsies, sleep disorders, and other related disorders.
The Live View Panel is intended for use by medical personnel within a medical facility, clinic or nursing home, or outside a medical facility under the supervision of a medical professional. The Live View Panel is available for use on all patient populations, including adults and children as determined by the medical professional.
This device is not to be used to monitor life threatening situations.
This device is not to be used to control life supporting or life sustaining devices.
{4}------------------------------------------------
#### PREDICATE COMPARISON
### Intended Use Comparison
The Live View Panel accessory improves upon the predicates' current capabilities for remote access and review of on-line or previously recorded acquisitions. This accessory does not change the indications for use for the devices which it is an accessory. This accessory is intended to aid in diagnosis of epilepsies, sleep disorders, and other related disorders.
#### Technical Comparison
The Live View Panel contains a subset of features found in the predicate. The device's main function is to facilitate the viewing of multiple patients' data from a location physically removed from the patient area. The Live View Panel extends the predicates' ability to view recordings remotely by providing an intuitive interface to manage multiple recordings. Therefore Neurotronics believes the Live View Panel is substantially equivalent to the Nihon Kohden EEG-1200A with JE-120A Multi Channel Electrode Junction Box (K113117), Nihon Kohden EEG-1200A Series Neurofax (K080546), PSG-1100 Sleep Diagnostic System (K120888), and Polysmith Sleep System, Model NTI5498 (K062943).
| Live View Panel | EEG Predicates<br>(K113117, K080546) | PSG Predicates<br>(K062943, K120888) |
|-------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--------------------------------------|
| Display Electrical activity of the brain and other<br>physiological signals on a monitor | Same | Same |
| Device uses Windows™ on a 32-bit computer | Same | Same |
| Display waveforms in acquisition and review | Same | Same |
| Maximum number of waveforms on screen is 64<br>(EEG) and 32 (PSG) | Same | Same |
| Simultaneously view waveforms from multiple<br>EEG/PSG instruments connected to a network<br>and display values from external instruments | Same | Same |
| Variable waveform display duration | Same | Same |
| Change waveform parameters (pattern, montage,<br>amplifier conditions, AV induction) | Same | Same |
| Add, edit, delete on screen events | Same | Same |
| The device can measure vital signs (including<br>ECG, EMG, Respiration, ocular motility, SpO2,<br>and CO2) | Same | Same |
| Display patient video along with waveforms | Same | Same |
| For use by medical personnel in a medical<br>facility, physician's office, laboratory, clinic, or<br>nursing home | Same | Same |
| Any patient population including adults and<br>children as determined by a trained professional | Same | Same |
{5}------------------------------------------------
#### PERFORMANCE DATA
### . Testing Data Testing included: Software testing
#### CONCLUSION
Based on the results of the Intended Use Comparison, the Technical Comparison, and Testing Data, it is believed that the Live View Panel (LVP) accessory presents no new questions of safety and effectiveness and, is substantially equivalent to the features provided by the Nihon Kohden EEG-1200A with JE-120A Multi Channel Electrode Junction Box (K113117), Nihon Kohden EEG-1200A Series Neurofax (K080546), Polysmith Sleep System, Model NTI5498 (K062943), and PSG-1100 Sleep Diagnostic System (K120888).
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure embracing a bird, often interpreted as representing the department's mission of protecting and promoting the health of Americans. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
August 9, 2013
Neurotronics, Inc. c/o Mr.David Pezet Quality Manager 3600 NW 43td Street, Suite Fl Gainesville, FL 32606
Re: K131415
> Trade/Device Name: Live View Panel Regulation Number: 21 CFR 882.1400 Regulation Name: Standard polysomnograph with electroencephalograph Product Code: OLV, GWQ, OLZ, and OLT Dated: June 16, 2013 Received: June 17, 2013
Dear Mr. Pezet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing
{7}------------------------------------------------
Page 2 - Mr. David Pezet
practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131415
Device Name:
Indications For Use:
The Live View Panel is an accessory to EEG and PSG systems, intended to transmit EEG or PSG data to a remote location where the data can be viewed by a clinician to aid in diagnosis of epilepsies, sleep disorders, and other related disorders
The Live View Panel is intended for use by medical personnel within a medical facility, clinic or nursing home, or outside a medical facility under the supervision of a medical professional. The Live View Panel is available for use on all patient populations, including adults and children as determined by the medical professional.
This device is not to be used to monitor life threatening situations.
This device is not to be used to control life supporting or life sustaining devices.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sian Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number ________________________________________________________________________________________________________________________________________________________________ K131415
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.