K131414 · GE Healthcare Finland Oy · MHX · Aug 28, 2013 · Cardiovascular
Device Facts
Record ID
K131414
Device Name
CARESCAPE MONITOR B850
Applicant
GE Healthcare Finland Oy
Product Code
MHX · Cardiovascular
Decision Date
Aug 28, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 is indicated for monitoring of: hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation). respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange) and neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission. The CARESCAPE Monitor B850 also provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. The CARESCAPE Monitor B850 is not intended for use during MRI.
Device Story
High-acuity multi-parameter patient monitor; processes physiological signals including ECG, invasive/non-invasive pressure, pulse oximetry, cardiac output, temperature, gas exchange, EEG, and neuromuscular transmission. System comprises host processing unit (ESP V2 software), display, keyboard, and parameter measurement modules. Used in professional healthcare facilities by trained clinicians; provides real-time monitoring, alarms, trends, and diagnostic analysis. Facilitates clinical decision-making through continuous patient status visualization and data integration. Supports standalone operation or network connectivity for remote viewing and data management. Benefits patients through continuous, high-acuity physiological surveillance.
Clinical Evidence
No clinical data required; substantial equivalence supported by bench testing, including module verification, system integration, safety, and performance testing against recognized standards (IEC 60601 series, AAMI EC11/13/57/SP10, ISO 9919/21647).
Technological Characteristics
Multi-parameter monitor; host processing unit with ESP V2 software. Connectivity via network for remote viewing/data management. Standards compliance: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62304 (software lifecycle), IEC 62366 (usability). Features EK-Pro v13 arrhythmia detection algorithm. Modular design allows various parameter measurement inputs.
Indications for Use
Indicated for monitoring hemodynamic, respiratory, and neurophysiological status in adult, pediatric, and neonatal patients in professional healthcare facilities. Contraindicated for use during MRI.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
K131414 Page 1 of 5
GE Healthcare Finland Oy
Kuortaneenkatu 2, P.O. Box 900 FI-00031 GE
Finland T: +358 10 39411
F: +358 9 1463310
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
## Owner/Contact/Date (807.92(a)(1):
| Date:<br>Owner/Submitter: | August 20, 2013<br>GE Healthcare Finland Oy.<br>Kuortaneenkatu 2<br>00510 Helsinki<br>FINLAND<br>Phone: +358 10 39411<br>Fax: +358 9 1463310 | AUG 2 8 2013 |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Primary Contact Person: | Joel Kent<br>Manager, Quality and Regulatory Affairs<br>GE Healthcare<br>Phone: 781-449-8685<br>Fax: 781-433-1344<br>E-mail: joel.kent@ge.com | |
| Secondary Contact Person: | Rauno Ruoho<br>Regulatory Affairs Manager<br>GE Healthcare Finland Oy<br>Kuortaneenkatu 2<br>00510 Helsinki<br>Finland<br>Phone: + 358 10 394 3624<br>Fax: +358-92726532<br>E-mail: rauno.ruoho@ge.com | |
| Device names (807.92(a)(2)): | | |
| Trade Name:<br>Common/Usual Name:<br>Classification Names: | CARESCAPE Monitor B850<br>multi-parameter patient monitor<br>21 CFR 870.1025 Arrhythmia detector and alarm (including ST-segment<br>measurement and alarm) | |
| Primary Product Code:<br>Secondary Product Codes: | MHX<br>BZK, BZL, BZQ, CAP, CBQ, CBR, CBS, CCK, CCL, DPS, DPZ, DQA,<br>DQK, DRT, DSI, DSJ, DSK, DXG, DXN, FLL, GWJ, GWQ, KOI, KRB,<br>MLD, NHO, NHP, NHQ, OLT, OLW, OMC, ORT | |
Image /page/0/Picture/7 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular frame. The letters and the frame are in black, contrasting with the white background. The logo is a well-known symbol associated with the General Electric company.
{1}------------------------------------------------
Predicate Device(s) (807.92(a)(3)::
Device Description (807.92(a)(4))::
#### K092027 CARESCAPE Monitor B850
The CARESCAPE Monitor B850 includes both new and existing subsystems interconnected to form a high acuity, multi-parameter patient monitoring system. A typical configuration would be a CARESCAPE Monitor B850 host processing unit running the CARESCAPE Monitoring platform software (in this submission the ESP V2 software) , a display with integrated keyboard and a frame for the insertion of parameter measurement modules. Many times a secondary display with keyboard is also used. The CARESCAPE Monitor B850 is designed to give the user maximum flexibility to choose the particular options for the particular care environment at the healthcare facility. A variety of options are available to the customer including additional displays, various input devices (keyboard, mouse, bar code reader), additional modules and frames and software options. Some of these major subsystems include non-patient contact accessory items (e.g. cables and mounting hardware). The CARESCAPE Monitor B850 interfaces to a variety of existing physiological parameter measurement modules. In addition, the CARESCAPE Monitor B850 interfaces to a variety of existing OEM medical devices via the existing network infrastructure. There are also various care area specific main software options and software licenses for specific features the customer can choose from for the CARESCAPE Monitor B850. There are also various upgrade programs offered to existing customers via the CARESCAPE Life Upgrade Programs to provide new CARESCAPE Monitor B850 hardware and/or software to the customers, but offer the benefit that the customers can continue to use the majority of their parameter measurement modules.
{2}------------------------------------------------
K131414 Page 3 of 5
#### Intended Use: (807.92(a)(5):
The indications for use are:
The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility.
The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time
The CARESCAPE Monitor B850 is indicated for monitoring of:
· hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation).
· respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange) and
· neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission.
The CARESCAPE Monitor B850 also provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices.
The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.
The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility,
The CARESCAPE Monitor B850 is not intended for use during MRI.
The CARESCAPE Monitor B850 with ESP V2 software is a new version of an existing monitor, introducing new features and improvements to existing features. Hardware used is essentially the same hardware as in the predicate. The predicate device used in this comparison is the CARESCAPE Monitor B850 with ESP VI software (K092027). The CARESCAPE Monitor B850 has the EK-Pro Arrhythmia Detection Algorithm (EK-Pro v13) that employs the same functional technology as the predicate device in the monitoring of ECG parameter data. The predicate arrhythmia algorithms are EK-Pro arrhythmia detection algorithm (EK-Pro v11, K031320) and Datex-Ohmeda bedside arrhythmia detection algorithm, both used with the predicate CARESCAPE Monitor B850 (K092027).
#### Summary of Non-Clinical Tests:
The CARESCAPE Monitor B850 and its applications comply with voluntary standards as detailed below. The following quality assurance measures were applied to the development of the system:
- � Risk Analysis
- . Requirements Reviews
- . Design Reviews
- . Testing on unit level (Module verification)
Technology (807.92(a)(6))::
Determination of Substantial Equivalence (807.92(b)(1))::
{3}------------------------------------------------
- . Integration testing (System verification)
- . Final acceptance testing (Validation)
- . Performance testing (Verification)
- . Safety testing (Verification)
The CARESCAPE Monitor B850 was designed and tested for compliance to the following standards:
IEC 60601-1:1988, A1:1991, A2:1995, Corr1:1995, Medical 1. Electrical Equipment Part 1: General Requirements for Safety - Second Edition
2. IEC 60601-1-1:2000, Medical Electrical Equipment - Part 1-1: General Requirements for Safety - Collateral Standard: Safety Requirements for Medical Electrical Systems - Edition 2.0
IEC 60601-1-2:2001 + A1:2004, Medical electrical equipment - Part 3. 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests - Edition 2.1
4. IEC 60601-1-4:1996 + A1:1999 (AKA ed 1.1:2000), Medical electrical equipment - Part 1: General requirements for safety - 4 - Collateral standard: Programmable electrical medical systems, Edition 1.1
IEC60601-1-6:2006, Medical electrical equipment - Part 1-6: న్. General requirements for basic safety and essential performance - collateral Standard: Usability - Edition 2
IEC 60601-1-8:2006, Medical electrical equipment - Part 1-8: ર્ણ રાજ્યના વિદ્યાર તાલુકામાં આવેલું એક ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests, and guidance for alarm systems in medical electrical equipment and medical electrical systems, Second Edition
7. IEC 60601-2-10:1987 + A1:2001, Medical Electrical Equipment Part 2: Particular Requirements for the Safety of Nerve and Muscle Stimulators - First Edition
8. IEC 60601-2-25:1993 + A1:1999, Medical Electrical Equipment Part 2: Particular requirements for the safety of electrocardiographs - First edition
ல். IEC 60601-2-26:2002, Medical electrical equipment - Particular requirements for the safety of electroencephalographs
10. IEC 60601-2-27:2005, Medical electrical equipment Part 2-27: Particular requirements for the safety, including essential performance, of electrocardiographic monitoring equipment
IEC 60601-2-30:1999, Medical electrical equipment - Part 2-30: 11. Particular requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressure monitoring equipment-Second Edition
IEC 60601-2-34:2000, Medical electrical equipment - Part 2-34: l 2. Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment-Edition 2
IEC 60601-2-40:1998, Medical electrical equipment - Particular 13. requirements for the safety of electromyographs and evoked response equipment
14. IEC 60601-2-49:2001, Medical electrical equipment - Part 2-49: Particular requirements for the safety of multifunction patient monitoring equipment-Edition 1
lEC 60601-2-51:2003, Medical electrical equipment Part 2-51: 15. Particular requirements for safety, including essential performance, of
{4}------------------------------------------------
# K131414
## Page 5 of 5
recording and analyzing single channel and multichannel electrocardiographs-Edition 1
AAMI EC11:1991/(R)2001/(R)2007, Diagnostic 16. Electrocardiographic Devices
17. AAMI EC13: 2002/(R)2007, Cardiac monitors, heart rate meters, and alarms,
AAMI EC-57:1998, A1:2003, Testing and reporting performance 18. results of cardiac rhythm and ST segment measurement algorithms
AAMI SP10:2002 + A1:2003 + A2:2006, Manual, electronic, or 19. automated sphygmomanometers
20. EN1041:2008, Information supplied by the manufacturer with medical devices
21. EN1060-1:1995 +A1:2002, Non-invasive sphygmomanometers- Part 1: General requirements
EN1060-3:1997 +A1:2005, Non-invasive sphygmomanometers- Part 22. 3: Supplementary requirements for electro-mechanical blood pressure measuring systems
Except for the following clause:
o 7.9 for PDM module: Testing performed in accordance with EN 1060-4
EN 12470-4:2000, A1:2009, Clinical Thermometers - Part 4: 23. Performance of Electrical Thermometers for Continuous Measurement Except for the following clauses:
6.3 b) Temperature measurement error with single use probes 0 exceeded maximum permissible error.
6.4: The response time of the Esophageal stethoscope with O temperature probe exceeds 150s for the probe sizes 18F and 24F.
24. ISO 21647:2004 + C1:2005, Medical electrical equipment -Particular requirements for the basic safety and essential performance of respiratory gas monitors
ISO9919:2005, Medical electrical equipment Particular requirements 25. for the safety and essential performance of pulse oximeter equipment for medical use - Second Edition
26. IEC62304:2006, Medical device software - Software life cycle processes
IEC62366:2007, Medical Devices - Application of usability 27. engineering to medical devices (General)
Clinical (807.92(b)(2)):
Summary of Clinical Tests: The subject of this premarket submission, CARESCAPE Monitor B850 did not require clinical studies to support substantial equivalence.
Conclusion (807.92(b)(3))::
GE Healthcare considers the CARESCAPE Monitor B850 to be as safe, as effective, and performance is substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, a symbol often associated with healthcare, with a staff entwined by a serpent.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
August 28, 2013
GE Healthcare Finland Oy Mr. Joel Kent Manager, Quality and Regulatory Affairs Kuortaneenkatu 2 Helsinki, FIN-00510 FI
Re: K131414 Trade/Device Name: Carescape Monitor B850 Regulation Number: 21 CFR 870.1025 Regulation Name: Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms) Regulatory Class: Class II Product Code: MHX, BZK, BZL, BZQ, CAP, CBQ, CBR, CBS, CCK, CCL, DPS, DPZ, DQA, DQK, DRT, DSI, DSJ, DSK, DXG, DXN, FLL, GWJ, GWQ, KOI, KRB, MLD, NHO, NHP, NHQ, OLT, OLW, OMC, ORT Dated: August 8, 2013 Received: August 9, 2013
Dear Mr. Joel Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{6}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Owen P. Faris -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: CARESCAPE Monitor B850
Indications for use:
The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility.
The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time
The CARESCAPE Monitor B850 is indicated for monitoring of:
· hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation).
· respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange) and
· neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission.
The CARESCAPE Monitor B850 also provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices.
The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.
The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The CARESCAPE Monitor B850 is not intended for use during MRI.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEFDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page *_ of _*
Digitally signed by
Digitally signed by
Owen P. Faris -S
Date: 2013.08.28
15:18:36-04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.