CO2/O2 NASAL CANNULA (WITH 4 (10CM) O2 AND CO2 TUBE), CO2/O2 NASAL CANNULA (WITH 8 (240CM) O2 AND 4 (10CM) CO2 TUBE),
K131410 · Southmedic, Inc. · CCK · Jan 13, 2014 · Anesthesiology
Device Facts
Record ID
K131410
Device Name
CO2/O2 NASAL CANNULA (WITH 4 (10CM) O2 AND CO2 TUBE), CO2/O2 NASAL CANNULA (WITH 8 (240CM) O2 AND 4 (10CM) CO2 TUBE),
Applicant
Southmedic, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Jan 13, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
This device is intended to provide a means for sampling end tidal carbon dioxide with the option to deliver supplemental O2 therapy to patients for up to 24 hours.
Device Story
Nasal cannula for simultaneous end-tidal CO2 (ETCO2) sampling and supplemental oxygen delivery; features dual-lumen tubing and nasal prongs. Oxygen delivered via central port between prongs to nose and mouth; ETCO2 sampled through nasal prongs to external capnograph. Used in clinical settings; operated by healthcare professionals. Provides real-time respiratory gas monitoring and oxygenation support; aids clinical decision-making regarding patient ventilation and oxygen status. Benefits include non-invasive respiratory monitoring and therapy delivery.
Clinical Evidence
Bench testing only. Comparative performance study conducted using a Harvard respiratory pump and mannequin head across three respiratory settings (Adult/Pediatric) and four oxygen flow rates (0-5 LPM). 144 total tests performed. Primary endpoint: ETCO2 measurement accuracy compared to predicates. Results showed subject device had the least variance from true CO2 values within 95% confidence intervals. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, skin irritation) confirmed non-cytotoxic, non-sensitizing, and non-irritating properties.
Technological Characteristics
Phthalate-free PVC construction; not made with natural rubber latex. Dual-lumen tubing for gas sampling and O2 delivery. Manufacturing via injection molding, extrusion, and assembly. Passive device; no energy source. Limited patient contact (up to 24 hours).
Indications for Use
Indicated for patients requiring end-tidal CO2 sampling and/or supplemental oxygen therapy for up to 24 hours. Suitable for clinical use via prescription.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
# KI31410
JAN 1 3 2014
## 510(k) Summary
| Date: | January 8, 2014 |
|----------------------|--------------------------------------------------------|
| Trade Name: | CO2/O2 Nasal Cannula |
| Common Name: | Oxygen Delivery, Carbon Dioxide Sampling Nasal Cannula |
| Classification Name: | Carbon Dioxide Gas Analyzer |
| Regulation: | Class II per 21CFR 868.1400 |
| Product Code: | CCK |
| Sponsor: | Southmedic Inc. |
| | 50 Alliance Blvd. |
| | Barrie, ON |
| | L4M 5K3 |
| | Canada |
| | (705) 720-1902 |
| Contact: | Tish Anger |
| Phone: | (705)720-1902 |
| Fax: | (705)728-9537 |
## LEAGALLY MARKETED PREDICATE DEVICES
This premarket notification will demonstrate that Southmedic's CO2/02 Sampling Nasal Cannula is substantially equivalent to the devices listed below:
- . Salter Labs #4706 CO2/O2 Nasal Cannula (K892406)
- . Hudson RCI BiFlow 1850 (K961150)
- . UnoMedical/Hospitak 368E (K915228)
#### DEVICE DESCRIPTION
Southmedic's CO2/02 Nasal Cannula incorporates sampling nares that channel expired carbon dioxide to a capnograph. A port located between the two nares is used to deliver oxygen to the patient as needed from an oxygen source. Its unique design delivers 02 to both the nose and mouth area. This device when in use involves surface contact to the patient's intact skin. There
{1}------------------------------------------------
is also gas pathway contact between patient and tubing/nasal cannula. Contact is considered limited exposure (up to 24 hours).
## INTENDED USE
This device is intended to provide a means for sampling end tidal carbon dioxide with the option to deliver supplemental O2 therapy to patients for up to 24 hours.
## TECHNOLOGICAL CHARACTERISTICS
Southmedic's CO2/O2 Nasal Cannula and the predicate devices have similar technological characteristics. Specifically Southmedic's CO2/O2 Nasal Cannula and the predicate devices are all designed to draw expired end-tidal CO2 from the patient through the nasal cannula and tubing to a capnograph, and administer medical grade USP O2 to patients from an oxygen supply.
Southmedic's CO2/O2 Nasal Cannula's cannula has been designed with slight differences from the predicate devices. All predicate cannulae have two separate systems built into the part to allow for the delivery of oxygen and sampling of end-tidal carbon dioxide. Each predicate takes a different approach. Exhaled carbon dioxide is collected through the nasal prongs and a single port was created central to the nasal prongs for O2 delivery in Southmedic's model. Modification to the outer portion helps direct O2 down towards the mouth as well as the nose.
| Characteristic | Southmedic's CO2<br>Sampling Nasal<br>Cannula | Salter Labs<br>#4706<br>CO2/O2<br>Nasal<br>Cannula | Hudson RCI<br>BiFlow 1850 | UnoMedical/Hospitak<br>368E |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| 510(k) | K131410 | K892406 | K961150 | K915228 |
| Intended Use | This device is intended<br>to provide a means for<br>sampling ETCO2 with<br>the option to deliver<br>supplemental O2<br>therapy to patients. | Same | Same | Same |
| Prescription | Yes | Yes | Yes | Yes |
| Technological<br>Characteristics | Provides a means to<br>deliver exhaled CO2 to<br>a capnograph via | Same | Same | Same |
| Characteristic | Southmedic's CO2<br>Sampling Nasal<br>Cannula | Salter Labs<br>#4706<br>CO2/O2<br>Nasal<br>Cannula | Hudson RCI<br>BiFlow 1850 | UnoMedical/Hospitak<br>368E |
| Design | Collects CO2 from<br>nasal prongs and<br>delivers O2 though a<br>single port between<br>nasal prongs | Delivers<br>oxygen from<br>one nasal<br>prong and<br>samples CO2<br>from the other. | Delivers<br>oxygen and<br>samples CO2<br>from both<br>nasal prongs. | Same as Southmedic's,<br>however, O2 delivery<br>through two ports<br>rather than one. |
| Material | Phthalate free PVC<br>Not made with natural<br>rubber latex. | PVC<br>containing<br>Phthalates<br>Not made with<br>natural rubber<br>latex. | PVC<br>containing<br>Phthalates<br>Not made<br>with natural<br>rubber latex. | PVC containing<br>Phthalates<br>Not made with natural<br>rubber latex. |
| Energy Used<br>or Delivered | Not Applicable | Not Applicable | Not<br>Applicable | Not Applicable |
| Manufacturing<br>Process | Injection molding,<br>extrusion, assembly | Injection<br>molding,<br>extrusion,<br>assembly | Injection<br>molding,<br>extrusion,<br>assembly | Injection molding,<br>extrusion, assembly |
| Performance | Same | Same | Same | Same |
| Labelling | Same | Same | Same | Same |
## SUBSTANTIAL EQUIVALENCE SUMMARY
{2}------------------------------------------------
{3}------------------------------------------------
# TESTING
In order to support the specific safety of Southmedic's CO2/02 Nasal Cannula, as well as confirm the suitability of the materials used in the manufacture of this product, extensive biocompatibility studies were contracted to an independent laboratory. The following table summarizes the biocompatibility testing conducted by Nelson Labs:
| Biocompatibility Testing | |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test/Method: | Agar Overlay -- Test for Cytotoxicity |
| Objective: | To determine the cytotoxicity of diffusible components from<br>materials. |
| Acceptance Criteria | The United States Pharmacopeia & National Formulary (USP 87)<br>states that the test article meets criteria if the reactivity grade is not<br>greater than 2 or a mild reactivity. The ANSI/AAMI/ISO 10993-5<br>standard states the achievement of a numerical grade greater than 2 is<br>considered a cytotoxic effect. |
| Results: | Three samples were investigated obtaining an average score of 1.<br>Results were compared to both positive and negative controls<br>(average scores of 4 and 0 respectively). |
| Discussion: | Southmedic's CO2/O2 Nasal Cannula is composed of materials that<br>do not have a cytotoxic effect. This supports the safety of the device<br>as well as the suitability of the materials used in its production. |
| Test/Method: | ISO 10993 Part 10 Guinea Pig Buehler Sensitization Test |
| Acceptance/Evaluation<br>Criteria | Test results were based on incidence and severity of the sensitization<br>reaction. Incidence was defined as the percentage of animals<br>exhibiting a sensitization reaction at each observation time point (24<br>and 48 hours) post challenge. Severity was calculated by dividing the<br>sum of the scores of one (1) or greater in the test group generally<br>indicated sensitization, provided grades of less than one ( 1 ) were<br>observed in the control animals. If severity of one (1) or greater was<br>noted on the controls, then the reaction of the test animals exceeded<br>the most severe control reaction to be considered due to sensitization.<br>In the final analysis of data consideration was given to overall<br>patterns, intensity, duration and character of reactions of the test as<br>compared to the control animal. |
| Results: | No sensitization reaction was observed in either the control or test<br>group. |
| Discussion | Southmedic's CO2/O2 Nasal Cannula is composed of materials that<br>do not have a sensitization effect. This supports the safety of the<br>device as well as the suitability of the materials used in its production. |
{4}------------------------------------------------
| Test/Method: | ISO 10993 Part 10 - Primary Skin Irritation Test in Rabbits | | |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--------------------------------|
| Acceptance/Evaluation Criteria: | After the 72 hour scoring, all erythema plus edema scores generated during the 24±2, 48±2, and 72±2 hour observations were totaled separately for each test sample and control for each animal. The resulting totals for each animal were divided by 6 (two tests/observation site, three time points) to determine separate test article and control primary irritation scores for each animal. The primary irritation index is characterized by the number (score) and description (response category) given in below: | | |
| | | Response Category | Primary Irritation Index (PII) |
| | | Negligible | >0 to 0.4 |
| | | Slight | 0.5 to 1.9 |
| | | Moderate | 2 to 4.9 |
| | | Severe | 5 to 8 |
| Results: | The test subject received a Primary Irritation Index Score of 0 in all areas. | | |
| Discussion: | Since no skin irritation was observed in the test subjects, this test demonstrates that Southmedic's CO2/O2 Nasal Cannula is composed of materials that are a non-irritant to skin. This supports the safety of the device as well as the suitability of the materials used in its production. | | |
Risk evaluation performed by Southmedic determined that because of the well characterized safe and effective use of nasal cannulas for oxygen delivery and end tidal CO2 sampling, animal or clinical testing were not necessary to support this application. In order to confirm effectiveness of Southmedic's CO2/O2 Nasal Cannula for its intended use, a comparative performance bench test against predicate devices was contracted from an independent laboratory. The following table summarizes the performance testing conducted by Piper Laboratory:
| Study Attribute | Description |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test subjects | Southmedic's CO2/O2 Nasal Cannula was tested and results compared<br>against predicate devices (see above) |
| Objective | To measure the end tidal CO2 values of four oxygen delivering nasal<br>cannulae under simulated patient conditions |
| Acceptance Criteria | End tidal CO2 values for Southmedic Nasal Cannula shall not be<br>statistically different, or shall vary less from the true end tidal CO2<br>value than the predicate products |
| Apparatus | Harvard Respiratory pump and mannequin head to simulate patient<br>head. Nasal cannula connected to oxygen supply line, CO2 sensing<br>line and CO2 detector (capnograph) |
{5}------------------------------------------------
| Simulated<br>Respiratory Settings | 1. Adult, Respiratory Rate 8, Tidal Volume 800<br>2. Adult, Respiratory Rate 16, Tidal Volume 600<br>3. Pediatric, Respiratory Rate 30, Tidal Volume 300<br>All conditions had I:E ratio of 1:1 (inspiration:expiration) |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Deviations | No deviations were noted during the study. All equipment met<br>predetermined operation and calibration testing before and after testing |
| Oxygen delivery for<br>study | Source flows of 0, 1, 3, and 5 liters per minute for each setting was<br>used. The system was allowed at least 3 minutes with oxygen deliver<br>prior to sampling CO2 to allow the system to equilibrate. |
| Measurements | Each test was sampled three times at each combination of settings for a<br>total of 144 tests ( 4 samples x 4 tests per sample x 3 respiratory<br>settings x 4 oxygen flow rate settings = 144 tests total) |
| Results | The end-tidal CO2 values for the Southmedic CO2/O2 Nasal Cannula<br>was not statistically different, or varied less from the true end tidal<br>CO2 value than predicate products.<br>A comparison of variances of the end tidal CO2 measurements from<br>the actual value for all four devices indicates that the Southmedic nasal<br>cannula has the least variance from true value when compared to the<br>three predicate devices within 95% confidence intervals. |
| Discussion | The Southmedic CO2/O2 Nasal Cannula met the predetermined<br>acceptance criteria of the test.<br>Both Southmedic's CO2/O2 Nasal Cannula and predicate devices are<br>intended to provide a means for sampling ETCO2 with the option to<br>deliver supplemental O2 therapy to patients. This performance test<br>demonstrated that Southmedic's CO2/O2 Nasal Cannula performs<br>equally to, or better than, the predicate devices in measuring end tidal<br>CO2. This supports the substantial equivalence of Southmedic's<br>CO2/O2 Nasal Cannula to the predicate products. |
## CONCLUSIONS
In summary, Southmedic's CO2/O2 Nasal Cannula path through the 510(k) regulatory analysis is as follows:
- . The Southmedic CO2/02 Nasal Cannula and the predicate devices are all intended for use as a means for sampling ETCO2 with the option to deliver supplemental O2 therapy to patients.
- The Southmedic CO2/O2 Nasal Cannula has similar key technological characteristics as . the predicate devices. All three are manufactured by extrusion, injection molding, and assembly. Designs are based on the same principle of one tube providing oxygen from a source and the other delivering expired CO2 to a capnograph.
- In order to verify the safety of the materials used in the manufacture of Southmedic's . CO2/02 Nasal Cannula, extensive biocompatibility testing was conducted. The testing demonstrated that materials used have no cytotoxic effect. The testing also confirmed
б
{6}------------------------------------------------
that the materials used are a non-irritant and do not have a sensitization effect. These results help support the safety of Southmedic's CO2/O2 Nasal Cannula
- . The use of nasal cannulas for sampling CO2 and delivering oxygen is well established in the medical community. In order to address any difference between Southmedic's CO2/02 Nasal Cannula and the predicate devices in their design, additional studies were performed. In order to demonstrate the effectiveness of the new design, a comparative bench test was conducted on Southmedic's behalf. The results demonstrated that Southmedic's CO2/02 Nasal Cannula varied less or equal to the true values when compared to the predicate device. This supports the effectiveness of the device.
Therefore, Southmedic's CO2/02 Nasal Cannula meets the criteria for substantial equivalence to Salter Lab's CO2/O2 Nasal Cannula, Hudson RCI Biflow Cannula, and the UnoMedical/Hospitak Nasal Cannula.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2014
Southmedic Incorporated Ms. Tish Anger Vice President of Quality and Regulatory Assurance 50 Alliance Blvd. Barrie, Ontario Canada L4M 5K3
Re: K131410
Trade/Device Name: CO2/02 Nasal Cannula Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: December 12, 2013 Received: December 13, 2013
#### Dear Ms. Anger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Ms. Anger
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
FOR
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
#### Indications for Use
CO2/O2 Nasal Cannula
510(k) Number:
K131410
Device Name:
Intended Use:
This device is intended to provide a means for sampling end tidal carbon dioxide with the option to deliver supplemental O2 therapy to patients for up to 24 hours.
Prescription Use_ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
-05'00'
Page 1 of 1
Page 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.