Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The system may be used wherever conventional screen-film systems are used. Agfa's Computed Radiography (CR) System With CR 12-X Digitizer is not indicated for use in mammography.
Device Story
Solid state x-ray imaging device; uses phosphor-coated imaging plates and cassettes for x-ray capture. CR 12-X digitizer uses laser scanning to convert latent images on plates into electronic data. NX workstation processes, previews, and routes images for storage or printing. Used in clinical settings; operated by radiology staff. Provides diagnostic quality images for general projection radiography; replaces conventional screen-film systems. Benefits include digital workflow, image processing (MUSICA/MUSICA2), and increased throughput compared to predecessors.
Clinical Evidence
Bench testing only. Verification and validation testing performed using phantom images to compare image quality against the CR 10-X predicate. No clinical images were required for evaluation.
Technological Characteristics
Solid state x-ray system; phosphor-coated imaging plates (BaSrFBrI:Eu); laser digitizer; 16-bit dynamic range (sq. root compressed); resolution options 100µm, 150µm, 200µm; throughput up to 90 plates/hr; NX workstation; DICOM connectivity; 100-240V power supply; mobile mounting option; software-based image processing (MUSICA/MUSICA2).
Indications for Use
Indicated for general projection radiographic applications in human anatomy. Not indicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K 131408
Page 1 of 5
Computed Radiography (CR) Systems with CR 12-X Digitizer
# 510(K) SUMMARY Agfa Computed Radiography (CR) Systems with CR 12-X Digitizer
Common Name: Computed Radiography System Classification Name: Solid State X-Ray System Regulatory Classification: 21 CFR 892.1680 Product Code: MOB Proprietary Name: Computed Radiography (CR) Systems with CR 12-X Digitizer Agfa HealthCare N.V. Septestraat 27 B-2640 Mortsel Belgium Contact: Koen Vervoort. Prepared: May 15, 2013 Telephone: +32-34444-7368
AUG 2 9 2013
#### LEGALLY MARKETED PREDICATE DEVICES A.
This is a 510(k) for Agfa's Computed Radiography (CR) System with CR 12-X Digitizer, a solid state x-ray imaging device. It is substantially equivalent to systems with Agfa's CR 10-X Digitizer (K121948) and 3Disc America's Fire CR+ Digitizer (K102619).
#### DEVICE DESCRIPTION B.
Agfa's Computed Radiography (CR) Systems with CR 12-X Digitizer is a solid state x-ray imaging device. Principles of operation and technological characteristics of the new and predicate devices are largely the same as other computed radiography systems:
Phosphor coated imaging plates and cassettes for image capture. Laser digitizer for generating the electronic image. NX workstation for image previewing, processing and routing.
#### C. INTENDED USE
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The system may be used wherever conventional screen-film systems are used.
Agfa's Computed Radiography (CR) System With CR 12-X Digitizer is not indicated for use in mammography.
The intended use has not changed as a result of any labeling modifications.
{1}------------------------------------------------
## D. SUBSTANTIAL EQUIVALENCE SUMMARY
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer has an Indications For Use statement identical to the statement for the predicate devices, K121948 and K102619. Intended uses are the same. The devices have the same technological characteristics. Descriptive characteristics and performance data are adequate to ensure equivalence.
Differences in devices do not alter the intended therapeutic/diagnostic effect.
| PRODUCT COMPARISON TABLE | | | |
|-------------------------------------|---------------------------------------------------|--------------------------------------------------------------|-------------------------------------------------|
| | Systems with CR 12-X<br>Digitizer<br>(New Device) | Systems with CR 10-X Digitizer<br>(PREDICATE-K121948) | 3Disc: FireCR+ Digitizer<br>(PREDICATE-K102619) |
| Communications | Same as both predicates | DICOM | DICOM |
| Cassette and Image Plates | | | |
| Cassettes and Image<br>Plate Sizes | Same as CR 10-X | 35x43 cm | 35x43 cm &<br>24x30 cm |
| Image Plate<br>Phosphors | Same as both predicates | BaSrFBrl:Eu | BaSrFBrl: Eu |
| Digitizer | | | |
| Scanning technology | Same as both predicates | Point at-a-time | Point at-a-time |
| Light collection | Same as CR 10-X | Solid, high efficiency | Unavailable |
| Scanning resolution | 100µm, 150µm, 200µm | 100µm | 100µm and 200µm |
| Active matrix | Same as CR 10-X | 3420 x 4218 | Unavailable |
| Dynamic range<br>(acquisition) | Same as both predicates | 16 bit, sq. root compressed | 16 bit, sq. root compressed |
| Throughput<br>(35x43 cm plates/hr.) | 60 @ 200µm<br>72 @ 150µm<br>90 @ 100µm | 34 @ 100µm | 20 @ 200µm<br>70 @ 100µm |
| Power Supply | Same as CR 10-X | 100-240V 50/60 Hz | 110-240V 50/60 Hz |
| Mobile installations | Same as CR 10-X | Mobile mounting option | Solid Transport case |
| Image stitching | Same as CR 10-X | Full Leg Full Spine | N/A |
| Workstation | | | |
| Image processing | Same as CR 10-X | MUSICA, MUSICA2 | QuantorMed+ Imaging<br>Software |
| Dynamic range<br>(display) | Same as CR 10-X | 12 bit | 32 bit & 64 bit |
| Operating system | Same as both predicates | Windows 7 | Windows 7 |
| Power Supply | Same as CR 10-X | 100 - 240V 50/60 Hz | 110-240V 50/60 Hz |
| Display System | Same as CR 10-X | Standard PC display or separately<br>cleared medical display | 1,280x900 pixel, DVI<br>interface |
{2}------------------------------------------------
### TECHNOLOGICAL CHARACTERISTICS E.
Agfa's Computed Radiography (CR) Systems with CR 12-X Digitizer is a solid state x-ray imaging device. It uses phosphor coated imaging plates to capture an x-ray exposure. Upon scanning of the image plates by the CR-12-X digitizer, the x-ray image is read and transmitted to the NX workstation where the image can be previewed, processed and transmitted to other devices for viewing, printing or storage.
The CR 12-X digitizer is identical to the CR 10-X with the exception of the following:
60 plates/hr with scanning resolution of 200um
72 plates/hr with added scanning resolution of 150um
Increase to 90 plates/hr with scanning resolution of 100um
Agfa CR 12-X and 3Disc America's FireCR+(K102619) both enable selection of resolution options. The remaining differences would not be expected to impact safety and effectiveness.
### F. TESTING
The device has completed verification and validation testing to confirm it meets specifications and operates as planned. Tests included image quality tests with internal and experts comparing the CR 12-X to the CR 10-X predicate. Clinical Images were unnecessary to evaluate modifications from the predicate device; therefore, phantom images were used to evaluate substantial equivalence with the CR 10-X predicate device (K121948).
The change in the design of the CR 10-X to include the image processing was verified and validated per the following:
Exhibit 3-2, Software Design Input Output Verification (DIOV) Report - Arcus Digitizer ..
Exhibit 5-1, Verification Plan - Consolidated Requirement & Testing (incl. SW), Arcus Solution.
Exhibit 5-2, Verification Report - Consolidated Requirement & Testing (incl. SW), Arcus Solution.
Exhibit 5-3, Validation Test Plan - Arcus Solution
Exhibit 5-4, Design Validation Report - Arcus Solution
Exhibit 5-5, Validation Test Report - Arcus Solution
Exhibit 5-6, Final Design Review for the CR 12-X Digitizer.
Exhibit 5-7, Design Complete Acceptance for the CR 12-X Digitizer
Exhibit 5-8, CRB (Change Review Board) Arcus Software Version ARC 1404 (This is the maintenance release for digitizer software version ARC 1404).
{3}------------------------------------------------
The product, manufacturing and development processes have been shown to conform to product safety, radiology, and imaging standards including:
# PRODUCT STANDARDS
IEC 60601-1: 2005 Medical Electrical Equipment - Part 1: General Requirements for Safety
IEC 60601-1-2: 2007 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
ACR/NEMA PS3.1-3.20: 2011 Digital Imaging and Communications in Medicine (DICOM)
## MANAGEMENT STANDARDS
ISO 14971:2007 Application of Risk Management to Medical Devices
ISO 13485:2003 Medical Devices - Quality Management Systems - Requirements For Regulatory purposes
### G. RISK ASSESSMENT AND MANAGEMENT SUMMARY
During the final risk analysis meeting, the risk management team concluded that the medical risk is not higher than previous released system to the field.
After review of these risks with the risk analysis team. it was decided that this risk could not be further mitigated and that for this risk, the benefits of the system outweigh the risks.
There are a total of 22 risks in the broadly acceptable region and 2 riskd in the ALARP region. Zero risks were identified in the not acceptable region.
The residual risk of the device is acceptable.
- · the risk control measures are effective
- · usability acceptance criteria for safety-related functions have been met (IEC 62366)
- · no residual risk is situated in the intolerable region
- · the residual risks are controlled as far as reasonably practicable
Of the 72 identified medical risks, 7 are in the ALARP region. The residual risks cannot be reduced any further with reasonable effort. Therefore is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk.
The term "Level of Concern" means the level of risk that the software device is determined to be if the software were to fail. The Level of Concern for the device has been determined to be moderate.
Refer to Section 16 Software, for more information regarding risk management.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a handwritten document with the text "K131408 Page 5 of 5". The text appears to be a document identifier and page number. The handwriting is legible, but the style is casual and not formal.
### H. CONCLUSIONS
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer has an Indications For Use statement is identical to the statement for the predicate devices. K121948 and K 102619. Intended uses are the same. The devices have the same technological characteristics in that they use the same imaging phosphor and the same laser diode for stimulation light operating with the same spot size as both predicates. Descriptive characteristics and performance data are adequate to ensure equivalence.
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is substantially equivalent to its CR 10-X System (K121948) in that it uses the same technology to capture and transmit images, use the NX user workstation, use the same MD1.0 image plate and cassette with manual cassette insertion, includes the full leg full spine function, provides a mobile mounting option, and uses 100-240V 50/60 Hz power supply.
Agfa CR 12-X and 3Disc America's FireCR+(K102619) both enable selection of resolution options from 100um to 200um.
The CR 12-X is equivalent to the predicate devices in regards to the plate speed.
This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes, with flowing lines suggesting movement or connection.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2013
AGFA HealthCare N.V. % Ms. ShaeAnn Cavanagh Regulatory Affairs Specialist NA AGFA HealthCare 10 South Academy Street GREENVILLE SC 29601
Re: K131408
Trade/Device Name: CR 12-X Digitizer Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: August 1, 2013 Received: August 7, 2013
Dear Ms. Cavanagh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Ms. Cavanagh
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/RcportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131408
Device Name: CR 12-X Digitizer
Indications For Use:
Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The system may be used wherever conventional screen-film systems are used.
Agfa's Computed Radiography (CR) System With CR 12-X Digitizer is not indicated for use in mammography.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Smh7)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.