WIRED MOUTH PROTECTOR (WMP)

K131396 · Elba Laboratories, Inc. · PFL · Sep 30, 2013 · Dental

Device Facts

Record IDK131396
Device NameWIRED MOUTH PROTECTOR (WMP)
ApplicantElba Laboratories, Inc.
Product CodePFL · Dental
Decision DateSep 30, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6890
Device ClassClass 1
AttributesTherapeutic

Indications for Use

Provides a soft barrier between wired mouth hardware and oral mucosa to help relieve discomfort for convalescing oral and maxillofacial surgery patients.

Device Story

Wired Mouth Protector (WMP) is a non-sterile, non-latex foam strip with a central slit; designed for insertion between teeth/lip area and wired mouth hardware. Acts as a soft physical barrier to protect oral mucosa from irritation/damage caused by orthodontic or surgical wires. Used by patients; changed three times daily or as needed. Provides relief for convalescing oral and maxillofacial surgery patients.

Clinical Evidence

No clinical data. Evidence consists of bench-level biocompatibility testing: agar diffusion and MEM elution (non-cytotoxic), primary skin irritation (rabbit, index=0.0), ocular irritation (rabbit, non-irritant), and acute oral toxicity (mouse, LD50 > 40 mL/kg).

Technological Characteristics

Non-latex foam strip with central slit. Biocompatibility verified via agar diffusion, MEM elution, dermal/ocular irritation, and acute oral toxicity testing. Class I device.

Indications for Use

Indicated for convalescing oral and maxillofacial surgery patients experiencing discomfort from wired mouth hardware.

Regulatory Classification

Identification

Intraoral dental wax is a device made of wax intended to construct patterns from which custom made metal dental prostheses, such as crowns and bridges, are cast. In orthodontic dentistry, the device is intended to make a pattern of a patient's bite to make study models of the teeth.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 12,31,396 P 3 0 2013 510(k) Premarket Notification: Wired Mouth Protector Elba Laboratories, Inc. # 510(k) SUMMARY #### I. ADMINISTRATIVE Submitter: Elba Laboratories, Inc. 1925 West Maple Road Troy, MI 48084 (248) 288-6098 Contact Person: Michael Froehlich Date of Preparation: September 25, 2013 II. DEVICE NAME Proprietary Name: Wired Mouth Protector (WMP) Common Name: Wired Mouth Protector Classification Name: Wax, Dental, Intraoral; 21 CFR §872.6890 ### III.PREDICATE DEVICE Brace Eze; K020009 ; Udent, Inc. ### IV. DEVICE DESCRIPTION The Wired Mouth Protector consists of a non-sterile strip of non-latex foam with a slit in the middle that is inserted into the front of the mouth between the teeth and lip area with the slit in the center to provide a soft barrier to help protect the oral mucosa from contact with wired mouth hardware. It is changed three times daily or more often if needed. Biocompatibility testing of the device has been conducted. The device was shown to be noncytotoxic in the agar diffusion test as well as the MEM elution test using L929 mamalian cells. Primary skin irritation testing in rabbits showed that the device is not considered to be a {1}------------------------------------------------ primary skin irritant (Primary Dermal Irritation Index=0.0). Ocular irritation testing in rabbit eyes showed that the device is not an ocular irritant. An acute oral toxicity study in mice showed that extracts were not considered toxic at an oral dose of 40 mL/kg body weight; the LD50 was > 40 mL/kg b.w. ## V. INTENDED USE Provides a soft barrier between wired mouth hardware and oral mucosa to help relieve discomfort for convalescing oral and maxillofacial surgery patients. ## VI.COMPARISON TO PREDICATE DEVICE The Wired Mouth Protector provides a similar function as the predicate device; i.e., protecting oral mucosal tissue from discomfort and damage. The composition of the Wired Mouth Protector differs from that of the predicate (foam vs. gel); however, the safety of the foam material for oral use has been well established in standard biocompatibility testing. # VII. CONCLUSION Based on design considerations and the results of biocompatibility testing, we conclude that, in terms of protective function, the Wired Mouth Protector performs at least as well as the predicate device. The Wired Mouth Protector is therefore considered to be substantially equivalent to the above-mentioned predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure. Public Health Service Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 30, 2013 Elba Laboratories, Incorporated C/O Mr. Richard A. Hamer Vice President/Regulatory Affairs Ferndale Pharma Group, Incorporated 780 West 8 Mile Road FERNDALE MI 48220 Re: K131396 Trade/Device Name: Wired Mouth Protector Regulation Number: 21 CFR 872.6890 Regulation Name: Intraoral Dental Wax Regulatory Class: I Product Code: PFL Dated: July 9, 2013 Received: July 12, 2013 #### Dear Mr. Hamer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2) CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Image /page/3/Picture/8 description: The image shows the name "Mary S. Bunner -S" in a bold, sans-serif font. The name is written in black ink on a white background. The letters are clear and legible, and the name is easy to read. The "S." is smaller and more stylized than the rest of the name. Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Elba Laboratories, Inc. 510(k) Number (if known): _K131396 Device Name: Wired Mouth Protector Indications for Use: Provides a soft barrier between wired mouth hardware and oral mucosa to help relieve discomfort for convalescing oral and maxillofacial surgery patients. OR Prescription Use (Per 21 CFR §801.109) Over-the Counter Use __ X (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Joel M. Anderson 2013.09.30 13:03:13 -04'00'
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...