K131392 · Etiam S.A.S.U. · LLZ · May 29, 2013 · Radiology
Device Facts
Record ID
K131392
Device Name
ETIAM PACS COMPONENTS
Applicant
Etiam S.A.S.U.
Product Code
LLZ · Radiology
Decision Date
May 29, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
ETIAM PACS Components™ is intended to be used by healthcare personnel to import, review, edit and send medical images. ETIAM PACS Components™ is not labeled for diagnostic use. ETIAM PACS Components™ is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants.
Device Story
ETIAM PACS Components™ is a software suite for capturing, storing, distributing, and networking medical images. It receives data from various imaging sources, including film digitizers and DICOM gateways. The system enables image manipulation, annotation, and measurement. It is used by trained healthcare professionals in distributed clinical locations. The device facilitates workflow by allowing users to import, review, edit, and send images/data across networks. It does not perform diagnostic interpretation; it serves as a tool for image management. Benefits include improved accessibility and communication of medical imaging data across clinical environments.
Clinical Evidence
No clinical data provided. Safety and effectiveness were established through bench testing, including software installation, configuration, patient-study information management, data acquisition/integration, DICOM conversion/connectivity, and data export/distribution. Clinical studies were deemed unnecessary as the device utilizes standard DICOM-compliant lossy compression techniques.
Technological Characteristics
Software-based medical image management system. Features include DICOM Query/Retrieve, DICOM Print, TWAIN interface, and support for JPEG lossy/lossless compression. Operates on standard computer networks. Does not contact patients. Classified as Moderate Level of Concern for potential hazards.
Indications for Use
Indicated for use by healthcare professionals (physicians, radiologists, nurses, technicians, assistants) to import, review, edit, and transmit medical images, including mammograms. Not for diagnostic use. Lossy compressed mammographic and digitized film images are contraindicated for primary interpretation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "etiam" in white letters with a black outline. The letters are set against a gray, textured background. There is a small, white shape above the letter "m".
K 131392 Secure Medical Networking
MAY 2 9 2013
# 510(k) Summary of Safety and Effectiveness
This 510(k) summary of Safety and Effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
#### Date Prepared:
March 22, 2013
### Submitter's Information: 21 CFR 807.92(a)(1)
Eric Le Bihan President ETIAM S.A.S .U. 2 rue Pierre-Joseph Colin 35000. Rennes - France Tel: +33 299 143388 Fax: +33 299 143380 Email: eric.lebihan@etiam.com
Name, title and phone number of contact: 21 CFR 807.92(a)(1)
Géraldine Tessens Quality Manager +33 299 143386 Tel: Fax: +33 299 143380 Email: qeraldine.tessens@etiam.com
#### Name of the device: 21 CFR 807.92(a)(2)
- ETIAM PACS Components™ Trade name: 网
- Common name: Picture, Archive and Communications System 网
- ם Classification name: system, image processing, radiological (21 CFR 892.2050, Product Code LLZ)
www.etiam.com
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "etiam" in white letters against a gray background. There is a small, white, abstract shape above the letter "m". The letters are bold and have a slight shadow effect, making them stand out from the background. The overall image has a textured appearance, possibly due to the printing or scanning process.
## Secure Medical Networking
### Predicate Devices: 21 CFR 807.92(a)(3)
| • | 510(k) number | K052358 |
|---|-----------------------------|----------------------------------------|
| | Regulation number | 892.2050 |
| | Device name | ETIAM STAR PACS COMPONENTS |
| | Applicant | ETIAM, S.A.S.U. |
| | Classification product code | LLZ |
| | Device classification name | SYSTEM, IMAGE PROCESSING, RADIOLOGICAL |
| | Class | II |
| | Panel | Radiology Device |
| • | 510(k) number | K062488 |
| | Regulation number | 892.2050 |
| | Device name | iQ-System PACS |
| | Applicant | IMAGE INFORMATION SYSTEMS, LTD. |
| | Classification product code | LLZ |
| | Device classification name | SYSTEM, IMAGE PROCESSING, RADIOLOGICAL |
| | Class | II |
| | Panel | Radiology Device |
| • | 510(k) number | K083618 |
| | Regulation number | 892.2050 |
| | Device name | PACS SCAN |
| | Applicant | PACSGEAR, INC |
| | Classification product code | LLZ |
| | Device classification name | SYSTEM, IMAGE PROCESSING, RADIOLOGICAL |
| | Class | II |
| | Panel | Radiology Device |
### Device Description: 21 CFR 807.92(a)(4)
ETIAM PACS Components™ are software applications that make possible the capturing, storage, distribution, and networking of medical images at distributed locations. In cases where DICOM images are not directly available to ETIAM PACS Components™, the system can acquire medical images using a DICOM gateway, which generates DICOM-type files. For example, film digitizers obtain images from old film and convert them to meet DICOM standards and stored in an archive. Stored files are transmitted using a network and can be viewed or manipulated from an imaging workstation.
### Intended Use: 21 CFR 807.92(a)(5)
ETIAM PACS Components™ is intended to be used by healthcare personnel to import, review, edit and send medical images.
ETIAM PACS Components™ is not labeled for diagnostic use.
{2}------------------------------------------------
# Secure Medical Networking
ETIAM PACS Components™ is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that meets other technical specifications reviewed and accepted by FDA.
Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants.
### Technological Characteristics: 21 CFR 807.92(a)(6)
ETIAM PACS Components 10 is a software product that handles digital medical images.
The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human interprets images and information being displayed and printed.
The design, fundamental technology, functionalities, intended use and performances of the ETIAM PACS Components™ are similar and therefore substantially equivalent to its three predicates like it is demonstrated in the table below.
| Spécification | ETIAM PACS<br>Components™ | ETIAM STAR<br>PACS<br>Components™ | iQ-System<br>PACS | PACS SCAN |
|---------------------------------------------------------------------------------------------|---------------------------|-----------------------------------|-------------------|-----------|
| Comparison of multiple studies | YES | YES | YES | YES |
| Measurements | YES | YES | YES | YES |
| Annotation of images | YES | YES | YES | YES |
| DICOM Query/Retrieve, DICOM Print<br>Client | YES | YES | YES | YES |
| JPEG lossy/lossless compression | YES | YES | YES | YES |
| TWAIN interface | YES | YES | YES | YES |
| Image file import | YES | YES | YES | YES |
| Patient CD/DVD import | YES | YES | YES | YES |
| Create Patient CD, DVD | YES | YES | YES | YES |
| Export to memory stick | YES | YES | YES | YES |
| Windows print | YES | YES | YES | YES |
| Image export to image file or AVI<br>video file | YES | YES | YES | - |
| Integration of direct interface to an<br>automatic publisher for patient<br>CD/DVD creation | YES | YES | YES | YES |
| Review report (SR formats) | YES | YES | YES | YES |
| Manipulating images | YES | YES | YES | YES |
| Interface to RIS worklist or DICOM<br>Archive | YES | YES | YES | YES |
| Video capture interface | YES | NO | YES | YES |
| Digitize film | YES | YES | NO | YES |
| Edit patient demographics | YES | YES | YES | YES |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "etiam" in white letters against a gray background. The letters are slightly stylized with a thick outline. There is a small white shape above the letter "m", possibly a design element or a blemish on the image. The background has a textured appearance, resembling a rough or grainy surface.
### Non-clinical performance
Every specification of the ETIAM PACS Components™ is validated by a test bench carried out during before release. The bench testing includes:
- o The package (documentation included)
- ם The installation of the software
- The configuration of the software
- The management of the patient-study information ם
- 그 The acquisition and integration of new data
- The manipulation of the data 0
- ם The DICOM conversion
- ם The DICOM connectivity
- 口 The export and distribution of data
### Cinical performance
Our device only utilizes standard lossy (irreversible) compression techniques defined and supported by DICOM Standard. Only the administrator can activate the compression function of the device.
So clinical studies were deemed unnecessary to evaluate the safety or effectiveness of ETIAM PACS Components TM.
### Conclusion: 21 CFR 807.92(b)(1)
The 510(k) Pre-Market Notification for ETIAM PACS Components™ contains adequate information and data to enable FDA-CDRH to determine substantial equivalence to the predicate device.
ETIAM PACS Components™ has been and will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the `\Level of Concern'' for potential hazard has been classified as "Moderate".
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 29, 2013
Etiam S.A.S.U. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street NW BUFFALO MN 55313
Re: K131392
Trade/Device Name: ETIAM PACS Components™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 10, 2013 Received: May 14, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. ' Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Mark Job
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
hup://www.lda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health · Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use
510(k) Number (if known): K131392
Device Name: ETIAM PACS Components ™
Indications for Use:
ETIAM PACS Components™ is intended to be used by healthcare personnel to import, review, edit and send medical images.
ETIAM PACS Components™ is not labeled for diagnostic use.
ETIAM PACS Components™ is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that meets other technical specifications reviewed and accepted by FDA.
Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Sm. 70.
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
K131392 510(k)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.