The Transcend Heated Humidifier is indicated for the humidification of the air delivered from a compatible Transcend positive airway pressure therapy device. The Humidifier is intended for single patient re-use in the home environment and in a hospital/institutional environment. The Humidifier is for use only as recommended by a physician.
Device Story
Transcend Heated Humidifier is an accessory for Transcend positive airway pressure (PAP) devices; it humidifies air delivered to the airway during PAP therapy. The device connects to a compatible PAP unit; it uses a heater plate to warm water in a reservoir, increasing air moisture. It is intended for single-patient use in home or clinical settings, operated by the patient or healthcare provider. The device features adjustable humidity settings (1-5) and includes safety features like out-of-water shutdown. The healthcare provider or patient adjusts settings based on comfort; the device improves patient compliance with PAP therapy by reducing airway dryness.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-1), and performance testing per ISO 8185:2007 (including flow resistance, spillage, flow sensing, sound power level, humidification output, and out-of-water shutdown).
Technological Characteristics
Respiratory humidifier; 325 ml water capacity; 1-5 humidity settings; 4-20 cm H2O operating pressure range. Compliant with IEC 60601-1 (Type BF), IEC 60601-1-2, and ISO 8185:2007. Powered by 100-240 VAC, 50/60 Hz. Biocompatible materials used. Not sterilized.
Indications for Use
Indicated for adult patients requiring humidification of air delivered by a compatible Transcend positive airway pressure (PAP) therapy device. Intended for single-patient reuse in home and hospital/institutional settings. Use only as recommended by a physician.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
# 31388
SEP 2 6 2013
Somnetics International Inc.
Traditional 510(k) Premarket Submission
Transcend Heated Humidifier
#### 5 510(k) Summary
[As required by 21 CFR 807.92]
#### 510(k) Number: K131388
Date Prepared: June 3, 2013
| Submitter/Manufacturer | Somnetics International, Inc.<br>33 5th Ave NW, Suite 500<br>New Brighton, MN 55112 |
|------------------------------|------------------------------------------------------------------------------------------------------|
| Establishment Registration # | 3008770104 |
| Contact Person | Melinda Swanson<br>Regulatory Consultant<br>Telephone: 651-621-1800<br>Email: mswanson@somnetics.com |
| Trade Name | Transcend Heated Humidifier |
| Common/Usual Name | Respiratory Humidifier |
| Classification | 21 CFR 868.5450, product code BTT<br>Humidifier, Respiratory Gas, (Direct Patient Interface) |
| Product Code | BTT |
| Predicate Device | ResMed HumidAire 2iTM, K080797 |
#### Device Description
The Transcend Heated Humidifier is a humidifier that is designed to humidify the air delivered to the airway during positive airway pressure (PAP) therapy with a compatible Transcend PAP device.
#### Indications for Use
The Transcend Heated Humidifier is indicated for the humidification of the air delivered from a compatible Transcend positive airway pressure therapy device. The Humidifier is intended for single patient re-use in the home environment and in a hospital/institutional environment. The Humidifier is for use only as recommended by a physician.
{1}------------------------------------------------
:
## Substantial Equivalence and Summary of Studies
The Transcend Heated Humidifier was tested and shown to be compliant with the following standards.
| Document Number | Title |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 | Medical Electrical Equipment, Part 1: General Requirements for Safety,<br>Ed. 3 |
| IEC 60601-1-2 | Medical Electrical Equipment – Collateral Standard: Electromagnetic<br>compatibility - Requirements and tests |
| IEC 60601-1-6 | Medical Electrical Equipment - Part 1-6: General requirements for basic<br>safety and essential performance - Collateral Standard: Usability |
| IEC 60601-1-11 | Medical Electrical Equipment – Collateral Standard: Requirements for<br>home health care environment |
| ISO 8185:2007 | Respiratory tract humidifiers for medical use -- Particular requirements<br>for respiratory humidification systems |
| ISO 10993-1 | Biological evaluation of medical devices |
| ISO 14971 | Medical devices -- Application of risk management to medical devices |
The Transcend Heated Humidifier is substantially equivalent to the predicate device based on comparisons of indications for use and technological characteristics.
| Summary of Predicate Comparisons | | | |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Transcend Heated<br>Humidifier | ResMed Ltd<br>HumidAire 2iTM | Comparison |
| Indications for use: | The Transcend Heated<br>Humidifier is indicated for<br>the humidification of the air<br>delivered from a compatible<br>Transcend positive airway<br>pressure therapy device.<br>The Humidifier is intended<br>for single patient re-use in<br>the home environment and<br>in a hospital/institutional<br>environment. The<br>Humidifier is for use only as<br>recommended by a<br>physician. | The HumidAire 2i is<br>indicated for the<br>humidification of the air<br>delivered from a ResMed<br>compatible CPAP therapy<br>device.<br>The HumidAire 2i is<br>intended for single patient<br>re-use in the home<br>environment and multi-<br>patient re-use in a hospital /<br>institutional environment.<br>The HumidAire 2i is for use<br>only as recommended by a<br>physician. | Similar. The Transcend<br>Heated Humidifier is<br>intended for use only by a<br>single patient. |
| Intended Population of Use | Adult | Adult | Identical |
| Summary of Predicate Comparisons | | | |
| Characteristic | Transcend Heated<br>Humidifier | ResMed LtdTM<br>HumidAire 2iTM | Comparison |
| Compatible Devices | Transcend PAP device and<br>associated accessories | ResMed PAP device and<br>associated accessories | Similar. The humidifiers are<br>made to be used with<br>manufacturers' existing PAP<br>devices. |
| Dimensions | 9 in X 5.5 in X 4.7 in | 4.8 in X 7.2 in X 8.6 in | Similar. Overall dimensions<br>do not effect substantial<br>equivalence determination. |
| Weight | 2.2 lbs | 2 lb | Similar. Overall weight does<br>not affect substantial<br>equivalence determination. |
| Disinfection | None | Should be disinfected when<br>used between patients. | The Transcend Heated<br>Humidifier is for single<br>patient re-use only. |
| Cleaning | Distilled water using a mild<br>detergent. | Warm water using a mild<br>detergent. | Identical |
| Sterilization | Not sterilized | Not sterilized | Identical |
| Biocompatible | Yes | Yes | Identical |
| Pressure Regulation | Determined by compatible<br>PAP device. | Determined by compatible<br>PAP device. | Identical |
| Working pressure range | 4-20 cm H₂O | 4-20 cm H₂O | Identical |
| Humidifier Settings | 1-5 | 1-6 | Similar. Both devices deliver<br>a range of humidity based<br>on setting level. |
| Inspiratory / Expiratory<br>Pressure Drop | 0.375 cmH₂O at 50 L/min | Not reported | Subject device meets<br>requirements of ISO 8185. |
| Gas Leak at Max Operating<br>Pressure | 4.85 L/min | Not reported | Subject device meets<br>requirements of ISO 8185. |
| Sound power level at 10 cm<br>H₂O static pressure | 37 dB | Not reported | Subject device meets<br>requirements of ISO 8185. |
| Maximum Heater Plate<br>Temperature | 131 °F | 167 °F | Subject device meets<br>requirements of ISO 8185. |
| Maximum output humidity | >95% | 95% | Similar |
| Maximum gas temperature | 77 °F | Not reported | Subject device meets<br>requirements of ISO 8185. |
| Water Capacity | 325 ml | 400 ml | Similar. Devices are<br>intended to have enough<br>water to last a minimum of<br>8 hours (equivalent to a full<br>night of sleep) |
| Operating Temperature | +41°F to +95°F | +41°F to +104°F | Similar. Operating<br>temperature is based on<br>compatible PAP device. |
| Operating Humidity | 10 - 80% relative humidity,<br>non-condensing | 10-95% relative humidity,<br>non-condensing | Similar. Operating humidity<br>is based on compatible PAP<br>device. |
| Summary of Predicate Comparisons | | | |
| Characteristic | Transcend Heated<br>Humidifier | ResMed Ltd<br>HumidAire 2i™ | Comparison |
| Shipping/Storage<br>Temperature | -4° F to +140° F | -4°F to +140°F | Identical |
| Gas inlet temperature<br>range | +41°F to +95°F | +41°F to +104°F | Similar. Inlet temperature is<br>based on compatible PAP<br>device. |
| Shipping/Storage Humidity | 10 - 90% relative humidity,<br>non-condensing | 15 - 95% relative humidity,<br>non-condensing | Similar |
| Power Supply | 100-240 VAC, 50/60 Hz | 100-240 VAC, 50/60 Hz | Identical |
| IEC 60601-1 Classification | Type BF | Type CF | Subject device is compliant<br>with ISO 8185 Section 6.8.2-<br>14 which states the device<br>should be either BF or CF. |
| Electromagnetic<br>compatibility | Compliant with IEC 60601-<br>1-2 | Compliant with IEC 60601-<br>1-2 | Identical |
| ISO 8185 compliant | Yes | Unknown - not stated in<br>510k summary or Users<br>Manual | Consensus standards can be<br>used to determine<br>substantial equivalence to a<br>legally marketed device. |
{2}------------------------------------------------
Somnetics International Inc.
.
.
Traditional 510(k) Premarket Submission
Transcend Heated Humidifier
י
"
·
: .
:
{3}------------------------------------------------
Somnetics International Inc.
Transcend Heated Humidifier Traditional 510(k) Premarket Submission
The device design was qualified through the following tests and assessments:
- . Electrical Safety
- . Electromagnetic Compatibility
- Biocompatibility Assessment .
- . Cleaning Validation
- . Software Validation
- Packaging and Shipping .
- Performance: physical (cycling), resistance to flow, spillage, flow sensing, sound . power level, humidification output, out of water shutdown, temperature, runtime, warm-up duration, and reservoir gas leak
These tests and assessments did not raise new safety or efficacy questions.
#### Conclusion
The Transcend Heated Humidifier is substantially equivalent to the HumidAire 2i (K080797). The subject and predicate devices are used in conjunction with CPAP devices to humidify air. They are equivalent in terms of technology and intended use. Risk assessments, biocompatibility evaluation, software, electromagnetic compatibility and electrical safety, bench testing, and compliance with recognized standards demonstrate that any differences do not raise new
{4}------------------------------------------------
Somnetics International Inc.
Transcend Heated Humidifier
Traditional 510(k) Premarket Submission
.
.
.
:
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
:
. .
-
.
・・
:
.
.
.
.
:
:
.
.
questions of safety or effectiveness. The Transcend Heated Humidifier is, therefore, substantially equivalent to the predicate HumidAire 2i device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services (USA). The logo features a stylized eagle with three wavy lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular pattern around the eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### September 26, 2013
Somnetics International, Incorporated Ms. Melinda Swanson Regulatory Consultant 33 510 Avenue NW, Suite 500 NEW BRIGHTON MN 22112
Re: K131388
Trade/Device Name: Transcend Heated Humidifier Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: Il Product Code: BTT Dated: August 22, 2013 Received: August 26, 2013
#### Dear Ms. Swanson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note
the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use
510(k) Number (if Known): どころは38
Device Name: Transcend Heated Humidifier
Indications for Use:
The Transcend Heated Humidifier is indicated for the humidification of the air delivered from a compatible Transcend positive airway pressure therapy device. The Humidifier is intended for single patient re-use in the home environment and in a hospital/institutional environment. The Humidifier is for use only as recommended by a physician.
Prescription Use _ X _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anya C. Harry -S
Division Sign-Off) wision of Anesthesiology, General Hospital .ection Control, Dental Devices
.i0(k) Number:_ K1313888
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.