The WarmTouch Surgical Access Blanket is indicated for use during abdominal, laparotomy, and spinal procedures. The Surgical Access Blanket is to be used only with Warm Touch convective warming units with model numbers 5015300A, 5015900 and 5016000. The warming system is for use only in a hospital or clinical environment, and only by trained and licensed caregivers. The WarmTouch Torso Blanket is indicated for use during procedures performed on the lower part of the body or lower extremities. The Torso Blanket is to be used only with WarmTouch convective warming units with model numbers 5015300A, 5015900 and 5016000. The warming system is for use only in a hospital or clinical environment, and only by trained and licensed caregivers.
Device Story
WarmTouch Surgical Access & Torso Blankets are convective warming accessories used with WarmTouch warming units. Surgical Access Blanket: sterile, full-body, rectangular opening for abdominal/spinal access, medial tape fixation. Torso Blanket: non-sterile, covers shoulders to hips, side flaps for tucking, distal tape fixation. Used in hospital/clinical environments by trained caregivers to prevent/treat hypothermia and maintain normothermia. Blankets distribute warm air from the unit to the patient. Benefits include thermal regulation during surgical procedures.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing: contact surface temperature, useful life, usability, anthropomorphic design, and biocompatibility.
Technological Characteristics
Convective warming blankets. Materials: medical-grade materials with integrated medical tape. Features: orientation labels, face shields, access windows, specific tabs. Connectivity: requires WarmTouch convective warming units (models 5015300A, 5015900, 5016000).
Indications for Use
Indicated for prevention and treatment of hypothermia and management of normothermia in hospital or clinical settings. Surgical Access Blanket for abdominal, laparotomy, and spinal procedures. Torso Blanket for lower body or lower extremity procedures. For use by trained, licensed caregivers only.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}------------------------------------------------
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Substantial Equivalence for the WarmTouch Surgical Access & Torso Blankets.
| Submitted By: | Covidien<br>6135 Gunbarrel Avenue<br>Boulder, CO 80301 |
|-----------------------------------|----------------------------------------------------------------------|
| Date: | July 26, 2013 |
| Contact Person: | Kelsey Lee<br>Senior Regulatory Affairs Specialist<br>(303) 305-2760 |
| Proprietary Name: | WarmTouch Surgical Access & Torso Blankets |
| Common Name: | Thermal Regulation System |
| Device Classification Regulation: | 21 CFR 870.5900 - Class II |
| Device Product Code & Panel: | DWJ |
| Predicate Device: | WarmTouch CareDrape Cardiac Blanket<br>(K014121) |
JUL 2 9 2013#### Device Description
The WarmTouch Surgical Access Blanket is a sterile full body blanket with a rectangular opening that allows access to the abdominal or lower back. This blanket stays in place via medial tape.
The WarmTouch Torso Blanket is a non-sterile blanket that is intended to cover the patient from the shoulders to the waist/hips. This blanket has flaps on either side that are tucked underneath the patient and medical tape at the distal end.
These blankets are intended to be used with the WarmTouch convective warming units.
{1}------------------------------------------------
### Indications for Use/Intended Use
The WarmTouch Surgical Access Blanket is indicated for use during abdominal, laparotomy, and spinal procedures. The Surgical Access Blanket is to be used only with Warm Touch convective warming units with model numbers 5015300A, 5015900 and 5016000. The warming system is for use only in a hospital or clinical environment, and only by trained and licensed caregivers.
The WarmTouch Torso Blanket is indicated for use during procedures performed on the lower part of the body or lower extremities. The Torso Blanket is to be used only with WarmTouch convective warming units with model numbers 5015300A, 5015900 and 5016000. The warming system is for use only in a hospital or clinical environment, and only by trained and licensed caregivers.
### Technological Characteristics Comparison
The subject WarmTouch blankets have similar materials, manufacturing processes, incorporation of a medical grade tape, and the Surgical Access Blanket has an equivalent sterilization method and shelf life as the predicate CareDrape Cardiac Blanket.
The subject WarmTouch blankets have the following new features when compared to the predicate CareDrape Cardiac Blanket:
Surgical Access: label identifying orientation, a face shield, a rectangular access window. different tabs and physical dimensions
Torso: label identifying orientation, a face shield, different tabs and physical dimensions These new technological and labelling features are not expected to raise new questions of safety or efficacy or be not substantially equivalent since the labeling change adds clarity to insure more effective use. The different physical dimensions face shield, tabs and access window have been shown to be substantially equivalent through performance and usability testing and a risk analysis has been performed for these products.
{2}------------------------------------------------
## Substantial Equivalence - Non-Clinical Evidence
キ
ﻤﻨﺘﻬﺎ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ
Substantial equivalence was shown through the following verification and validation testing:
| | Modification from<br>Predicate | Verification/Validation Test | Results |
|------------------------------------|------------------------------------------------------|--------------------------------|-----------|
| Surgical Access<br>Blanket | New blanket design | Contact Surface<br>Temperature | Pass |
| | New blanket design and<br>construction | Useful Life | Pass |
| | New blanket design | Usability | Pass |
| | New blanket design | Anthropomorphic Design | Pass |
| | New material and design | Cleanliness & Sterilization | No Impact |
| Torso Blanket | New blanket design | Contact Surface<br>Temperature | Pass |
| | New blanket design and<br>construction | Useful Life | Pass |
| | New blanket design | Usability | Pass |
| | New blanket design | Anthropomorphic Design | Pass |
| Surgical Access &<br>Torso Blanket | New blanket design,<br>construction and<br>materials | System Usability | Pass |
| | New materials | Biocompatibility | No Impact |
| | New blanket design | Design Validation | Pass |
The results of the tests show that the subject WarmTouch Surgical Access & Torso Blankets can be considered substantially equivalent to the legally marketed predicate.
## Substantial Equivalence - Clinical Evidence
N/A - Clinical evidence was not necessary to show substantial equivalence
## Substantial Equivalence - Conclusions
Substantial equivalence to the predicate is shown through usability, useful life testing, anthropomorphic verification, and contact surface temperature verification. From the evidence presented in the Premarket Notification, the subject devices can be considered substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles a bird or a stylized human figure.
Public Health Service
Food and Drug Administration 0903 New Hamoshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 29, 2013
Covidien Ms. Kelsey Lee Sr. Regulatory Affairs Specialist 6135 Gunbarrel Ave. Boulder, CO 80301
Re: K131361
> Trade/Device Name: WarmTouch Surgical Access & Torso Blankets Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ Dated: June 3, 2013 Received: June 4, 2013
Dear Ms. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Lee
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.Z. Hillman
for
for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number: K131361
Device Name: WarmTouch™ Surgical Access Blanket
Indications for Use:
The WarmTouch™ Surgical Access Blanket is intended for prevention and treatment of hypothermia, and for the management of appropriate normothermia. The warming system is for use only in a hospital or clinical environment, and only by trained and licensed caregivers.
> X Prescription Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
> > (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
M. Z. Hillerman
{6}------------------------------------------------
## Indications for Use
510(k) Number: K131361
Device Name: WarmTouch™ Torso Blanket
Indications for Use:
The WarmTouch™ Torso Blanket is intended for prevention and treatment of hypothermia, and for the management of appropriate normothermia. The warming system is for use only in a hospital or clinical environment, and only by trained and licensed caregivers.
> Prescription Use X _______________ AND/OR Over-The-Counter Use_ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
> > (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
## CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.