# TX WHEELCHAIR CHASSIS (K131336)

_Jcm Seating Solutions, Ltd. · IOR · Oct 28, 2013 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K131336

## Device Facts

- **Applicant:** Jcm Seating Solutions, Ltd.
- **Product Code:** [IOR](/productcode/IOR.md)
- **Decision Date:** Oct 28, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The TX Wheelchair Chassis is designed and manufactured by JCM Seating Solutions for the mobilization of occupants ages 2 to adult, and intended primarily for general indoor and outdoor (smooth surfaces) use. The chassis and seating system combination can also be used for transporting the occupant in a suitable motor vehicle.

## Device Story

TX Wheelchair Chassis is a mechanical wheelchair frame designed to mobilize postural seating systems. Input: user/occupant weight (up to 110 kg). Operation: manual propulsion by attendant via push handle or by occupant via rear wheels; features 'tilt in space' (-30°/+5°) for postural stability. Output: mobility for occupant. Used in indoor/outdoor environments; operated by attendant or occupant. Includes tie-down points for motor vehicle transport. Benefits: provides mobility, postural support, and safe transport for individuals requiring specialized seating.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing per ANSI/RESNA WC-1:2009 and ISO 7176-8:1998, including static stability, brake effectiveness, dimensions, and fatigue strength.

## Technological Characteristics

Tubular steel frame; manual propulsion; hand-operated rear wheel parking brakes; drum brakes for attendant; anti-tip mechanism; tilt-in-space functionality (-30°/+5°); WTORS tie-down fittings. Tested per ANSI/RESNA WC-1:2009 and ISO 7176-8:1998.

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Multi Frame Wheelchair ([K100084](/device/K100084.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### K131336

# 510(k) Summary for TX Wheelchair Chassis

#### 1. Submission Sponsor

JCM Seating Solutions Ltd 15-18 Maxwell Road Woodston Industrial Estate Peterborough Cambridgeshire PE2 7HU United Kingdom Phone: 44 1733 405830 Fax: 44 1733 405838 Contact: Wayne Killick, Operations Director Website: www.jcmseating.co.uk

OCT 2 8 2013

## 2. Submission Correspondent

Emergo Group 816 Congress Avenue Suite 1400 Austin, TX 78701 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Stuart R. Goldman, Senior Consultant, RA/QA Email: project.management@emergogroup.com

## 3. Date Prepared

March 19, 2013

## 4. Device Identification

Trade/Proprietary Name: TX Wheelchair Chassis Common/Usual Name: Wheelchair Classification Name: Wheelchair, Mechanical Classification Regulation: 890.3850 Product Code: IOR Device Class: Class I Physical Medicine Classification Panel:

## 5. Predicate Device

Multi Frame Wheelchair (K100084)

## 6. Device Description

The TX Wheelchair Chassis is designed and manufactured by JCM Seating Solutions for the mobilization of both our own postural seating systems and those manufactured by other specialist seating providers, and is indicated for occupants

{1}------------------------------------------------

ages 2 to adult, and intended primarily for general indoor and outdoor (smooth surfaces) use.

The chassis consist of a fabricated steel framework with two rear wheels and two front castors where both front and rear wheel sizes can be selected at point of order from a pre-determined range. The TX Wheelchair Chassis also has a "tilt in space" feature that allows for the user's position in space to be changed between -30°/+5°, which may lead to improvements in postural stability and comfort, contributing to enhanced functional activity. A push handle is provided to enable an attendant to propel the chassis and the seated occupant and under certain circumstances the product could be setup to enable the occupant to self-propel through the rear wheels themselves. The chassis and seating system combination can also be used for transporting the occupant in a suitable motor vehicle and has tie-down points to facilitate this.

# 7. Indication for Use

The TX Wheelchair Chassis is designed and manufactured by JCM Seating Solutions for the mobilization of occupants ages 2 to adult, and intended primarily for general indoor and outdoor (smooth surfaces) use. The chassis and seating system combination can also be used for transporting the occupant in a suitable motor vehicle.

# 8. Substantial Equivalence Discussion

The following table compares the TX Wheelchair Chassis to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

| Device Manufacturer | Snug Seat                 | JCM Seating<br>Solutions | Substantial<br>Equivalence<br>Comparison                                                                                                                 |
|---------------------|---------------------------|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name   | Multi Frame<br>Wheelchair | TX Wheelchair<br>Chassis | -                                                                                                                                                        |
| FDA 510(k)          | K100084                   | Pending                  | -                                                                                                                                                        |
| FDA Product Code    | IOR                       | IOR                      | IOR                                                                                                                                                      |
| FDA Regulation      | § 890.3850                | § 890.3850               | § 890.3850                                                                                                                                               |
| Frame Construction  | Tubular aluminum          | Tubular steel            | While the subject<br>chassis is made<br>from tubular steel<br>and the predicate<br>chassis is made<br>from tubular<br>aluminum, these<br>differences are |

| Table 5-A – Comparison of Characteristics        |
|--------------------------------------------------|
| Multi Frame Wheelchair vs. TX Wheelchair Chassis |

{2}------------------------------------------------

|                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                              | insignificant as both<br>the subject and<br>predicate device<br>passed performance<br>testing in<br>accordance with<br>FDA recognized<br>consensus<br>standards for<br>mechanical<br>wheelchairs to<br>prove their safety<br>and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Frame Models/Sizes<br>(overall width)        | Size 1/Size 2/Size 3<br>(590/640/690 mm)                                                                                                                                                                                                                                                                                                                                                               | TX 30/TX 35<br>(550/620 mm)                                                                                                                                                                                                                                                                                                                                                                  | Both the subject<br>and predicate<br>chassis are offered<br>in several different<br>models/frame sizes<br>based on their<br>overall width. The<br>slight difference in<br>size does not affect<br>the intended use of<br>these devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Front Wheel Options<br>(diameter)            | 180 mm                                                                                                                                                                                                                                                                                                                                                                                                 | 150/200 mm                                                                                                                                                                                                                                                                                                                                                                                   | The subject chassis<br>is offered with front<br>wheels (castors) in<br>two different sizes,<br>while the predicate<br>chassis is offered<br>with front wheels in<br>only one size. The<br>slight difference in<br>sizes does not affect<br>the intended use of<br>these devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Rear Wheel Options<br>(diameter)             | 320/560 mm                                                                                                                                                                                                                                                                                                                                                                                             | 300/400/600 mm                                                                                                                                                                                                                                                                                                                                                                               | The subject chassis<br>is offered with rear<br>wheels in three<br>different sizes,<br>while the predicate<br>chassis is offered<br>with rear wheels in<br>two different sizes.<br>The slight difference<br>in sizes does not<br>affect the intended<br>use of these<br>devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Chassis Weight                               |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                              | Size 1/Size 2/Size 3<br>(320 mm)<br>= 13 kg<br>Size 1/Size 2/Size 3<br>(560 mm)<br>= 14 kg                                                                                                                                                                                                                                                                                                             | TX 30<br>(300/400/600 mm)<br>= 16-17 kg<br>TX 35<br>(300/400/600 mm)<br>= 17-18 kg                                                                                                                                                                                                                                                                                                           | While the subject<br>chassis is made<br>from tubular steel<br>and the predicate<br>chassis is made<br>from tubular<br>aluminum, these<br>differences are<br>considered to be<br>insignificant in their<br>overall operation by<br>either the user or<br>the user's assistant,<br>as both the subject<br>and predicate<br>devices passed<br>performance testing<br>in accordance with<br>FDA recognized<br>consensus<br>standards for<br>mechanical<br>wheelchairs to<br>prove their safety<br>and effectiveness.                                                                                                                                                                                                                                                   |
| Max. Chassis Load<br>(user/seat/accessories) | 100 kg<br>(220 lbs)                                                                                                                                                                                                                                                                                                                                                                                    | 110 kg<br>(243 lbs)                                                                                                                                                                                                                                                                                                                                                                          | While the subject<br>chassis is made<br>from tubular steel<br>and can handle a<br>slightly larger<br>maximum load<br>capacity when<br>compared to the<br>predicate chassis<br>that is made from<br>tubular aluminum,<br>these differences<br>are insignificant as<br>both the subject<br>and predicate<br>devices passed<br>performance testing<br>in accordance with<br>FDA recognized<br>consensus<br>standards for<br>mechanical<br>wheelchairs to                                                                                                                                                                                                                                                                                                              |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                              | prove their safety<br>and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Tilt/Prone Seat Angle                        | 0-30°/35°/40°                                                                                                                                                                                                                                                                                                                                                                                          | -30°/+5°                                                                                                                                                                                                                                                                                                                                                                                     | Both the subject<br>and predicate<br>chassis are offered<br>with the ability of<br>their postural<br>seating system (an<br>accessory offered<br>with the chassis by<br>their manufacturer)<br>to be tilted back. In<br>the case of the<br>subject device, the<br>amount of rearward<br>tilt is set at -30°,<br>while in the case of<br>the predicate device<br>the amount of<br>rearward tilt can be<br>positioned to -<br>30°/35°/40°<br>depending on how<br>the seating system<br>is mounted to the<br>frame. While the<br>predicate device<br>also has the ability<br>to tilt forward +5°.<br>The slight difference<br>in forward and rear<br>tilt options between<br>the subject and<br>predicate devices<br>does not affect their<br>overall intended<br>use. |
| Age of Occupant                              | Children                                                                                                                                                                                                                                                                                                                                                                                               | 2-adult                                                                                                                                                                                                                                                                                                                                                                                      | The subject chassis<br>is designed for<br>occupants ages 2 -<br>adult (18) up to 243<br>lbs in weight, while<br>the predicate<br>chassis is designed<br>for children up to<br>220 lbs in weight.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| User Environment                             | General indoor and<br>outdoor use                                                                                                                                                                                                                                                                                                                                                                      | General indoor and<br>outdoor use                                                                                                                                                                                                                                                                                                                                                            | Both the subject<br>and predicate<br>chassis are designed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                              | for general indoor and outdoor use under normal environmental conditions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Parking Brakes                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                          | Both the subject and predicate chassis are offered with a hand operated barking system that engages the rear wheels and can be adjusted to accommodate the different size rear wheel options that each model of device is offered in.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Hand (Drum) Brakes                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                          | Both the subject and predicate chassis are offered with drum brakes that can be operated by the occupant's assistant, and these brake levers are mounted (one on each side) of the device's push brace that is attached to the chassis frame.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Anti-Tip Mechanism                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                          | Both the subject and predicate chassis are offered with an anti-tip mechanism by way of a frame mounted anti-tip bar that is attached to the rear of the chassis that can be rotated to either engage or disengage into position.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Seat Plate                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                          | Both the subject and predicate chassis are offered with an interface                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                              | (seat) plate to which<br>the postural seating<br>system is attached.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Foot Support                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                          | Both the subject<br>and predicate<br>chassis are offered<br>with a frame<br>mounted foot<br>support (rest) for<br>the occupant to rest<br>their feet on when<br>using the device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Transport Tie-Down<br>Fittings               | Yes                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                          | Both the subject<br>and predicate<br>chassis are offered<br>with a WTORS<br>(Wheelchair Tie<br>Down and Restraint<br>Systems) fittings<br>that allow the<br>chassis/seating<br>system/occupant to<br>be safely secured<br>(tied down) during<br>transport.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Indications for Use                          | The Multi Frame's<br>intended function<br>and use is to<br>provide mobility to<br>children limited to a<br>sitting position. The<br>Wheelchair consists<br>primarily of an<br>aluminum frame,<br>large rear wheels<br>with hand rims for<br>propelling the<br>wheelchair or<br>smaller rear wheels<br>for attendant-only<br>use, and smaller<br>front pivoting<br>casters for steering<br>and turning. | The TX Wheelchair<br>Chassis is designed<br>and manufactured<br>by JCM Seating<br>Solutions for the<br>mobilization of<br>occupants ages 2 to<br>adult, and intended<br>primarily for general<br>indoor and outdoor<br>(smooth surfaces)<br>use. The chassis and<br>seating system<br>combination can<br>also be used for<br>transporting the<br>occupant in a<br>suitable motor<br>vehicle. | Some minor<br>differences<br>between the subject<br>and predicate<br>device are:<br>•The frame of the<br>subject chassis is<br>made of steel, while<br>the frame of the<br>predicate chassis is<br>made from<br>aluminum.<br>•Both the subject<br>and predicate<br>chassis are offered<br>in several frame<br>sizes.<br>•Both the subject<br>and predicate<br>chassis are offered<br>with different<br>options for their<br>rear wheel<br>diameter while the                                                                                                                                                                                                                                                                                                       |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                        | subject chassis is<br>offered with front<br>wheels in two size<br>diameters, and the<br>predicate device is<br>offered in only one<br>size.<br>•Both the subject<br>and predicate<br>chassis allow for the<br>occupant to be<br>tilted back to a<br>prone position.                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{3}------------------------------------------------

{4}------------------------------------------------

{5}------------------------------------------------

・

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{6}------------------------------------------------

.

{7}------------------------------------------------

# 9. Non-Clinical Performance Data

As part of demonstrating the safety and effectiveness of its TX Wheelchair Chassis and in showing substantial equivalence to the predicate device, JCM Seating Solutions submitted its device for extensive performance testing in accordance with the following FDA guidance document and recognized consensus standards shown below.

- . Guidance Document for the Preparation of Premarket Notification [510k]] Applications for Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles
- ANSI/RESNA WC-1:2009, American National Standard for Wheelchairs -. Volume 1: Requirements and Test Methods for Wheelchairs (including Scooters)

{8}------------------------------------------------

- o WC Section 1 Static Stability (Results: Passed)
- o WC Section 3 Determination of the Effectiveness of Brakes (Results: Passed)
- o WC Section 5 Overall Dimensions, Mass, and Turning Space (Results: Passed)
- o WC Section 7 Measurement of Seating and Wheel Dimensions (Results: Passed)
- O WC Section 8 Static, Impact, and Fatigue Strength Dimensions (Results: Passed)
- o WC Section 15 Requirements for Information, Disclosure, Documentation, and Labeling (Results: Passed)
- . ISO 7176-8:1998, Wheelchairs -- Part 8: Requirements and Test Methods for Static, Impact and Fatigue Strengths (Results: Passed)
- ISO 14971: 2007, Medical Devices -- Application of Risk Management to . Medical Devices

The TX Wheelchair Chassis passed all the testing in accordance with national and international standards stated above as shown by the acceptable results obtained.

# 10. Clinical Testing

There was no clinical testing required to support the TX Wheelchair as the indications for use is equivalent to the predicate device. Mechanical wheelchairs, including the predicate device, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

# 11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared device, or has the same intended use and different technological characteristics, but it can be demonstrated that the new device is substantially equivalent to the predicate device, and that the new device does not raise any questions regarding its safety and effectiveness when compared to the predicate device.

It has been shown in this 510(k) submission that the differences between the TX Wheelchair Chassis and the Multi Frame Wheelchair do not raise any questions regarding its safety and effectiveness when used as indicated. Furthermore, the TX Wheelchair Chassis has undergone extensive performance testing in accordance with the applicable sections of ANSI/RESNA WC-1 and ISO 7176-8 to demonstrate its safety. The TX Wheelchair Chassis, as designed and manufactured by JCM Seating

{9}------------------------------------------------

Solutions, is therefore determined to be substantially equivalent to the predicate device manufactured by Snug Seat, Inc., previously cleared under K100084.

{10}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 28, 2013

JCM Seating Solutions. Ltd c/o Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313

Re: K131336

Trade/Device Name: TX Wheelchair Chassis Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: September 30, 2013 Received: October 1, 2013

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{11}------------------------------------------------

Page 2 - Mr. Mark Job

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

v.ru.207/MedicalDevices/ResourcesforIndustry/ucm400142.htm

Sincerely yours,

Joyce M. Whang -S

- for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{12}------------------------------------------------

# Indications for Use

510(k) Number (if known): K131336

Device Name:_TX Wheelchair Chassis

Indications For Use:

The TX Wheelchair Chassis is designed and manufactured by JCM Seating Solutions for the mobilization of occupants ages 2 to adult, and intended primarily for general indoor and outdoor (smooth surfaces) use. The chassis and seating system combination can also be used for transporting the occupant in a suitable motor vehicle.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

# Joyce M. Whang -S

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K131336](https://fda-staging.innolitics.com/device/K131336)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).