K131319 · Sirona Dental Systems GmbH · EFB · May 30, 2014 · Dental
Device Facts
Record ID
K131319
Device Name
T1/T2/T3 TURBINE FAMILY
Applicant
Sirona Dental Systems GmbH
Product Code
EFB · Dental
Decision Date
May 30, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Indications for Use
The turbines of the T1 / T2 / T3 Turbine family are intended for the - Preparation of cavities and crowns - Removal of carious material - Removal of fillings - Processing of tooth and restoration surfaces - Reducing hard tooth structure
Device Story
Sirona T1/T2/T3 Turbine family consists of reusable, air-powered, high-speed dental handpieces. Devices feature ergonomic designs; T1 and T2 models include fiber optic lighting. Water delivery system utilizes 4 pairs of water/air spray nozzles with one-way retraction valves to prevent ingress of external contaminants. Handpieces are compatible with various quick-couplings (Sirona, KaVo, W&H, NSK). Operated by dentists in clinical settings to perform cavity/crown preparation, caries/filling removal, and tooth structure reduction. Sterilizable via autoclave. Output is mechanical rotation of dental burs for tooth/restoration processing; light output assists visualization. Benefits include efficient removal of tooth structure and restoration materials during dental procedures.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ISO 14457 (handpieces/motors), ISO 1797 (rotary instruments), ISO 9168 (hose connections), and ISO 21531 (graphical symbols) was demonstrated. Additional testing validated sterilization processes, device effectiveness, and biocompatibility.
Technological Characteristics
Air-powered turbine handpieces; materials include metal components; push-button FG chuck; fiber optic light (25,000 lux on T1/T2); speed range 250,000-400,000 rpm; power 20-23W. Compatible with Sirona, KaVo, W&H, and NSK quick-couplings. Sterilization via autoclave. Standards: ISO 14457, ISO 1797, ISO 9168, ISO 21531.
Indications for Use
Indicated for dental procedures including cavity and crown preparation, removal of carious material and fillings, processing of tooth and restoration surfaces, and reduction of hard tooth structure in patients requiring dental restorative work.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
## 510(k) Summary
## MAY 3 0 2014
### for
## Sirona Dental Systems T1 / T2 / T3 Turbine family
### with serial no. > 600 000
#### Sponsor 1
Sirona Dental Systems GmbH Fabrikstrasse 31 D-64625 Bensheim Germany
Contact Person: Fritz Kolle Telephone: 49 6251 16 32 94
Date Prepared: May 1, 2013
#### ମଧ Device Name
Proprietary Name: T1 / T2 / T3 Turbine family with serial no. > 600 000
Common/Usual Name: High speed air turbine
Classification Name: Handpiece, air-powered, dental
#### 3 Predicate Devices
Nakanishi Ti-Max X Turbine (K113655) and the Morita Twin Power Turbine (K 043498)
Sirona Dental Systems T1 / T2 / T3 Turbine family with serial no. > 600 000 Traditional 510(k) Summary May 1, 2013
APPENDIX G . Page 1 of 3
{1}------------------------------------------------
#### 4 Intended Use
The turbines of the T1 / T2 / T3 Turbine family are intended for the
- Preparation of cavities and crowns
- Removal of carious material
- Removal of fillings
- Processing of tooth and restoration surfaces
- Reducing hard tooth structure
#### Device Description 4.1
The Sirona Dental Systems T1 / T2 / T3 turbine family consists of air powered, high speed handpieces. The handpieces are reusable, ergonomically shaped and include (T1 and T2 Turbines) a fiber optic light system. The water delivery geometry includes 4 pair of water / air spray nozzles just below the head with one way retraction valves to prevent the ingress of external water or air, when the spray system is shut down. The devices are sterilizable in an autoclave. The turbines are provided in three variants: control, boost and mini.
In addition, the turbines are compatible with the following quick-coupling:
- Sirona R coupling
- Sirona F coupling
- Sirona B coupling
- KaVo coupling
- W&H coupling
- NSK MachLite coupling
- NSK QD-J coupling
#### 5 Summary of the technological characteristics
The Sirona Dental Systems T1 / T2 / T3 turbine familar in the operating principle, technical data and performance to other high speed dental handpieces currently in the US commercial distribution. Examples of substantially equivalent devices include Nakanishi Ti-Max X Turbine (K113655) and the Morita Twin Power Turbine (K 043498). Below is a comparison of the proposed device to the predicate devices.
Sirona Dental Systems T1 / T2 / T3 Turbine family with serial no. > 600 000 Traditional 510(k) Summary May 1, 2013
APPENDIX G . Page 2 of 3
{2}------------------------------------------------
| | Sirona T1 / T2 / T3<br>turbine family | Nakanishi Ti-Max X<br>(K113655) | Morita Twin Power<br>(K043498) |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Principle of operation | Turbine<br>Air driven ¹ | Turbine<br>Air driven | Turbine<br>Air driven |
| Couplings to treatment<br>chair | Sirona, Kavo, NSK,<br>W&H | Sirona, Kavo, NSK,<br>W&H, Bienair | Sirona, Kavo, NSK,<br>W&H, Morita |
| Speed range | 250000-400000 | 300000-450000 | 320000-400000 |
| Power [W] | 22/ 23/ 20 | 16-22 | 18 -22 |
| Chuck | Push button FG | Push button FG | Push button FG |
| Light intensity [lux] | 25000 | Data not available | 25000 |
| Intended use | - Preparation of<br>cavities and crowns<br>- Removal of carious<br>material<br>- Removal of fillings<br>- Processing of tooth<br>and restoration<br>surfaces<br>- Reducing hard tooth<br>structure | ... intended for<br>removing carious<br>material, reducing<br>hard tooth structure,<br>cavity preparation,<br>finishing tooth<br>preparations and<br>restorations and<br>polishing teeth. | TWIN POWER<br>TUBINE is for use by<br>authorized persons in<br>the practice of the<br>dentistry |
| | | | | | Table 1: Comparison of the proposed device to the declared predicate devices. |
|--|--|--|--|--|-------------------------------------------------------------------------------|
| | | | | | |
1 As described in Sec. 11.1 ion page 9 and 10 in the original submission
#### 6 Performance Testing
The Sirona T1 / T2 /T3 turbine family complies with ISO 14457 Dentistry -Handpieces and motors, ISO 1797 Dental Rotary Instruments - Shanks, ISO 9168 Dental Handpieces - Hose Connections, ISO 21531 Dentistry - Graphical symbols for dental instruments, additional performance testing was conducted to validate the sterilization process, device effectiveness and biocompatibility. Reverse engineering of the quick coupling was also conducted.
#### 7 Clinical Testing
Clinical testing has not been performed.
#### 8 Conclusion
Based on a comparison of intended use, indications, principal of operations, features and technical data, the Sirona Dental T1 / T2 / T3 high speed turbine family is substantially equivalent to its Predicate Devices.
Sirona Dental Systems T1 / T2 / T3 Turbine family with serial no. > 600 000 Traditional 510(k) Summary May 1, 2013
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. Inside the circle is an abstract symbol that resembles a person with outstretched arms. The symbol is made up of three curved lines.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug, Administration 19903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### May 30, 2014
Sirona Dental Systems GmbH Mr. Fritz Kolle Regulatory Manager Fabrikstrasse 31 Bensheim. Germany D-64625
Re: K131319
Trade/Device Name: TI/T2/T3 Turbine family with serial n. > 600,000 Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB Dated: April 25, 2014 Received: April 28, 2014
Dear Mr. Kolle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmelic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Kolle
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mary S. Runner -S
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT 4
Indications for use
510(k) Number (if known): K131319
Device Name: T1 / T2 / T3 Turbine family with scrial n. > 600 000, Models:
TI CONTROL S ;TI CONTROL K; TI CONTROL W; TI CONTROL N; TI Boost S; TI Boost K; T1 Boost W; T1 Boost N; T1 mini K; T1 mini K; T1 mini N; T1 mini N; T2 CONTROL S; 12 CONTROL K; T2 CONTROL W; T2 CONTROL N; T2 Boost S; T2 Boost K; T2 Boost W; T2 Boost N; T2 mini S; T2 mini W; T2 mini N; T2 Boost S; T3 Boost K; T3 Boost W: T3 Boost NO: T3 mini S; T3 mini K; T3 mini W; T3 mini NQ;
Indications for Use:
The turbines of the T1 / T2 / T3 Turbine family are intended for the
- Preparation of cavities and crowns
- Removal of carious material
- Removal of fillings
- Processing of tooth and restoration surfaces
- Reducing hard tooth structure
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
Andrew I. Steen -S 2014.05.30 08:48:08 -04'00'
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1
Sirona Dental Systems traditional 510(k) May 1, 2013 T1 / T2 / T3 Turbine family
CONFIDENTIAL Page 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.