IC-1545-DL MULTI FLO DVT COMBO INTERMITTENT PNEUMATIC COMPRESSION DEVICE
K131306 · Bio Compression Systems, Inc. · JOW · Jul 22, 2013 · Cardiovascular
Device Facts
Record ID
K131306
Device Name
IC-1545-DL MULTI FLO DVT COMBO INTERMITTENT PNEUMATIC COMPRESSION DEVICE
Applicant
Bio Compression Systems, Inc.
Product Code
JOW · Cardiovascular
Decision Date
Jul 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used with GID-3045-T Calf/Thigh or GID-3045-K Calf garments: - Intended for prophylaxis of deep vein thrombosis When used with GI-3045-F Foot Garments - Intended for prophylaxis of deep vein thrombosis - Enhancement of venous and arterial circulation - Prevention of venous stasis ulcers - Reduction of acute or chronic edema - Reduction of lower limb pain due to surgery or trauma - Reduction of compartment pressures
Device Story
IC-1545-DL Multi-Flo DVT Combo is an intermittent pneumatic compression device for lower extremities. System comprises a pump, inflatable garments, and tubing. Pump delivers intermittent, rapid impulse pressurization to segmented chambers in garments to compress veins, expel blood, and overcome stasis. Device is used in clinical or home settings for bed-ridden patients; operated by clinicians or patients. Digital control system replaces analog regulator/gauge of predicate; features soft keypad for pressure adjustment (1 mmHg increments) and LED display. Pump maintains pressure via digital sensor; pre-set cycles (15s inflation/45s deflation). Alarm alerts user to low/no pressure (<30 mmHg). Benefits include DVT prophylaxis and improved circulation. Healthcare providers use device to manage venous/arterial conditions; output is mechanical compression of limbs.
Clinical Evidence
Bench testing only. Functional testing included electrical leakage, pressure adjustment, segment inflation pressure, display accuracy, and cycle times. Results demonstrated comparable inflation cycle profiles (rise times, inflation pressures, deflation times, cycle times) to the predicate device.
Technological Characteristics
Intermittent pneumatic compression pump; digital pressure sensor; LED display; soft keypad interface. Max pressure 150 mmHg. Dimensions: 6"H x 8"W x 8"D. Weight: 4.25 lbs. Power: 120VAC, 60Hz. Garments: non-sterile, single-patient use, segmented inflatable chambers. Connectivity: standalone. Sterilization: N/A (non-sterile).
Indications for Use
Indicated for prophylaxis of deep vein thrombosis (DVT) in patients using calf, thigh, or foot garments. Foot garments also indicated for enhancement of venous/arterial circulation, prevention of venous stasis ulcers, reduction of acute/chronic edema, reduction of lower limb pain due to surgery/trauma, and reduction of compartment pressures.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
K131306
# II. 510(k) Summary
APPLICANT'SINFORMATION: Jonathan Ross President Bio Compression Systems, Inc 120 West Commercial Avenue Moonachie, NJ, 07074 PH: 201 939-0716 FX: 201 939-4503 E-mail: jross@biocompression.com http://www.biocompression.com Internet: Medical Establishment Registration No .: 2424387
## SUBMITTER'SINFORMATION
Maureen Garner President New World Regulatory Solutions, Inc. P.O. Box 5374 Toms River, NJ 08754 732-779-7422 PH: 732-270-4829 Fax: E-mail: NWRSinc@gmail.com Internet: www.newworldreg.com
DATE: May 23, 2013
## DEVICEINFORMATION
## DEVICE NAME:
IC-1545-DL Multi-Flo DVT Combo Intermittent Pneumatic Compression Device Classification Panel: Cardiological and Respiratory Devices Classification Number: 870.5800 Product Nomenclature: Compressible Limb Sleeve Product Code(s): JOW Trade/Proprietary Name: IC-1545-DL Multi-Flo DVT Combo Common Name: IC-1545-DL Multi-Flo DVT Combo Intermittent Circulator
# DEVICECLASSIFICATION
Compressible Limb Sleeve Devices are classified as Class II devices, and reviewed by the Division of Cardiovascular Devices.
PREDICATEDEVICE Model Multi-Flo IC-1545-KT/F Intermittent Circulator
# JUL 2 2 2013
{1}------------------------------------------------
## DEVICEDESCRIPTION
The IC-1545-DL Multi-Flo DVT Combo (applicant device) provides intermittent pressure to the calf and thigh or foot through the use of inflatable garments. The applicant device is a digitally-
controlled version of the predicate device, with identical specifications. There is no change in
fundamental technology and no change in intended use from the predicate device, K610857. The digital modification provides identical call/thigh (KT) and foot (F) timing and pressure options as the predicate and it has the same intended use as the predicate, including prophylaxis of deep vein thrombosis (DVT) and enhancing venous and arterial circulation. Tubing connector design variations prevent interchangeability of garments between the models.
The user interface consists of a soft keypad where the predicate has a rocker on/off switch and a pressure adjustment knob. The analogue pressure gauge and regulator have been removed, making the applicant unit 27% smaller and 47% lighter than the predicate, and both are designed to hang on bedrails for easy access to bed-ridden patients.
The device consists of a pump, inflatable garments, and interconnection tubing. The pump compressor is capable of no more than 150 mmHg maximum pressure and has pre-set inflate/deflate cycle times. Default pressure and timing is pre-set at the factory to 50 mmHg, with a cycle of 15 seconds on and 45 seconds off (KT option). This is consistent with the majority of indications prescribed.
All controls and measurement functions are contained in the PCB assembly. When turned on, the timer sends a signal to start the pump sends air to the output ports, filling the garments with air. A digital pressure sensor maintains the pressure by turning the compression pump on and off. After 45 seconds, the timer stops the pump and opens a valve to deflate the garnents and 15 seconds later the cycle repeats.
In order to achieve its intended use, the device is attached via interconnecting tubing to sleeves (garments) applied externally and bilaterally over the lower extremities. The garments contain discrete, interconnected and segmented inflatable chambers. The pump provides intermittent, rapid impulse pressurization to the chambers. The pressure can be adjusted up or down in increments of I mmHg with up and down soft key arrows, and an alarm is provided for low/no pressure. When the garnents are inflated, they compress the veins in the calf, expelling blood from the leg, overcoming blood stasis and promoting circulation.
The device runs continuously until turned off. An LED display shows the pressure setting during the inflation cycle and "0" during the deflation cycle. Pressure setting can be changed by the user, while timing cycles can be changed with special instructions provided to the distributor. An alarm will alert the user if inflation pressures do not exceed 30 mmHg.
All garments are supplied non-sterile and for single patient use. DVT prophylaxis garments are available in a total of three anatomical configurations as indicated to the calves or the calves and the thighs (KT) and the feet (F).
#### INDICATIONSFORUSE
When used with GID-3045-T Calf/Thigh or GID-3045-K Calf garments:
- [ Intended for prophylaxis of deep vein thrombosis
When used with GI-3045-F Foot Garments
- I Intended for prophylaxis of deep vein thrombosis
- □ Enhancement of venous and arterial circulation
{2}------------------------------------------------
- O Prevention of venous stasis ulcers
- 7 Reduction of acute or chronic edema
- [ Reduction of lower limb pain due to surgery or trauma
- O Reduction of compartment pressures
# TECHNOLOGICALCHARACTERISTICS
The manufacturer believes that the technological characteristics of the modified IC-1545-DL Multi-Flo Combo device are substantially similar to those of the predicate IC-1545-KT and -F Multi-Flo devices. The user interface has been modified from a regulator and analogue dial gauge in the predicate to a digital pressure sensor and LED display in the applicant device. In place of a manual dial on a regulator in the predicate device, the pressure can be increased or decreased in increments of I mmHg via a soft keypad on the applicant device.
# PERFORMANCEDATA
Before being released every device is tested and must meet all performance specifications. In addition to aesthetic acceptance criteria, functional testing includes electrical leakage, pressure adjustment, inflation pressure in each segment, air pressure display accuracy, and inflation/deflation cycle times. The results demonstrate comparable inflation cvcle profiles (rise times. inflation pressures, deflation times and cycle times) between the applicant and predicate devices.
# STATEMENTOFSUBSTANTIALEQUIVALENCE
#### Similarities
Both the applicant and the predicate devices provide continuous intermittent pressure bilaterally to the foot, calf or calf and thighs using inflatable garments. The applicant and predicate devices have the same intended use and indications for use, both operate within the same clinicallyestablished parameters and both have the same performance specifications. The applicant and predicate devices use the same prescribed inflation pressures, inflation and deflation times and cycle times.
#### Differences
The applicant device utilizes a digital pressure sensor and LED pressure display while the predicate device utilizes an analogue gauge and needle display. The applicant has soft key up and down arrows to adjust pressure up or down, accessible by pressing an access sequence on the soft keys, while the predicate device has a locking pressure control knob attached to a regulator.
Both the predicate and the applicant devices operate within the same clinically-established parameters. The differences between the predicate and the applicant devices do not impact safety or effectiveness. A table illustrating the similarities and differences is provided below.
{3}------------------------------------------------
| Parameter | · Predicate K061857<br>Models IC-1545-KT and -F | Digital<br>IC-156-DL-Multi Flo<br>DVT Combo |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Intended Use | KT and F: Prophylaxis of deep vein thrombosis.<br>F: enhancement of venous and arterial<br>circulation, prevention of venous stasis ulcers, assist<br>in healing of cutaneous ulcers, reduction<br>of acute or chronic edema, reduction of lower<br>limb pain due to surgery or trauma, reduction of<br>compartmental pressures. | Same |
| Principal of Operation | Intermittent Pneumatic Compression | Same |
| Weight | 8 pounds | 4.25 pounds |
| Dimensions, inches | 5.5 H X 8 L X 12 W | 6 H x 8 W x 8 D |
| # of Segments in<br>garment(s) | 1 - 4 | Same. |
| Inflation Time, each<br>segment | 15 seconds bilateral (KT)<br>6 seconds bilateral (F) | Same |
| Deflation Time | 45 seconds(KT)<br>54 seconds (F) | Same |
| Recommended<br>Inflation Pressure | 45-60 mmHg (KT)<br>90-120 mmHg (F) | Same |
| Pressure Adjustment | Locking adjustable knob on regulator | Digital, soft keypad,<br>1mmHg increments |
| Pressure Gauge | 0-125 mmHg, analog | 0 - 120 mmHg, digital |
| Displayed Pressure<br>Accuracy | ± 5 mmHg of sleeve pressure,<br>real time pressure displayed on gauge face | ± 5 mmHg of sleeve<br>pressure, pressure set<br>point displayed on LED |
| Pause time between<br>inflation cycles | None (deflation time is pause time) | Same |
| Total Cycle Time | 60 seconds, 60 cycles per hour. | Same |
| Garments Available | Calf/thigh (Model KT)<br>Foot (Model F) | Same |
| Fail-safe hose<br>connectors | Yes | Yes |
| Bilateral Treatment<br>Option | Yes | Yes |
| Power Requirements | 120VAC, 60Hz, 0.5A | Same |
# Table of Similarities and Differences with the Predicate Device
# CONCLUSION
Based upon safety and performance testing, compliance with voluntary standards, and comparison to the predicate devices, the manufacturer believes that the IC-1545-DL-Multi Flo DVT Combo Intermittent Pneumatic Compression Device is substantially equivalent to the predicate device, and does not raise any new questions of safety or effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and has a simple, clean design. The eagle is facing to the right.
Public Health Service
June 11, 2014
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Bio Compression Systems, Inc. c/o Ms. Maureen Garner 1983 Hazelwood Road Toms River, NJ 08753 US
Re: K131306
Trade/Device Name: IC 1545-DL Multi-Flow DVT Combo Intermittent Pneumatic Compression Device Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: May 23, 2013 Received: May 24, 2013
Dear Ms. Garner:
This letter corrects our substantially equivalent letter of July 22, 2013.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Ms. Maureen Garner
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Fernando Aguel -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Statement
510(k) Number: K131306
Device Names: IC-1545-DL Multi-Flo DVT Combo Intermittent Pneumatic Compression Device
Indications for Use:
When used with GID-3045-T Calf/Thigh or GID-3045-K Calf garments:
- Intended for prophylaxis of deep vein thrombosis .
When used with GI-3045-F Foot Garments
- Intended for prophylaxis of deep vein thrombosis ●
- . Enhancement of venous and arterial circulation
- . Prevention of venous stasis ulcers
- Reduction of acute or chronic edema
- . Reduction of lower limb pain due to surgery or trauma
- . Reduction of compartment pressures
Prescription Use Over-The-Counter Use - AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. FDA
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.