This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMG) signals and electrical stimulus of nerves. This device is indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2.
Device Story
Sentio MMG is a multichannel intraoperative monitor used during surgeries where motor nerves are at risk. System inputs include mechanomyographic (MMG) signals from muscles innervated by the target nerve, triggered by either operator-applied electrical stimulus or direct mechanical stimulus during surgery. The system consists of a touchscreen PC control unit, patient module, sensor connector module, and disposable probes/sensors. The device processes these signals to assist surgeons in identifying and mapping nerves, alerting them when a nerve is activated to minimize trauma. Used in clinical/surgical settings by surgeons; output is displayed on the touchscreen to guide real-time surgical decision-making and nerve preservation.
Clinical Evidence
No clinical data. Substantial equivalence was established through non-clinical comparative performance evaluation assessing statistical agreement between the subject and predicate devices.
Technological Characteristics
Multichannel intraoperative monitor; 10-channel capacity. Uses constant electrical current (1-15 mA) for nerve stimulation; monophasic square pulse (100μs, 2Hz). Employs skin-surface mounted electromechanical MMG sensors. System includes a touchscreen PC control unit and patient module. Connectivity via sensor connector module. Sterilization method not specified.
Indications for Use
Indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2 during surgical procedures.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "sentio" in a sans-serif font. To the left of the word is a four-pointed star-like symbol. The word is in lowercase letters and appears to be a logo or brand name. The overall image has a simple and clean design.
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
- · 1. Submitter's Identification:
Aashish Shah Vice President of Quality & Operations Sentio, LLC 50461 W Pontiac Trail Wixom, MI 48393 Tel: (248) 595-0420 Fax: (248) 956-8123
Date Summary Prepared: April 29th, 2014
#### 2. Name of the Device:
Trade or Proprietary Name: Classification Name: Device Class: Classification: Product Code:
Sentio MMG® Neurosurgical Nerve Locator Class II 21 CFR §874.1820 PDQ
#### Common or Usual Name: 3.
Avoidance Intraoperative Mechanomyographic (MMG) Mapping and System, Nerve Monitor/Stimulator, Nerve Locator/Stimulator
#### Predicate Device Information: .4.
The subject device Sentio MMG® is substantially equivalent to the following predicate device:
K100992 NeuralMAS™, Innovative Surgical Solutions, LLC d/b/a Sentio, LLC, Wixom, Michigan
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "sentio" in a sans-serif font, with a four-pointed star-like symbol to the left of the word. The word is in lowercase letters, and the symbol is slightly larger than the letters. The image is in black and white.
#### Device Description: 5.
The Sentio MMG® system is a multichannel intraoperative monitor for use during surgeries in which a motor nerve is at risk. The Sentio MMG® system records mechanomyographic (MMG) signals from muscles innervated by the affected nerve, which may originate from operator applied electrical stimulus or from direct mechanical stimulus occurring during the course of surgery. The monitor will assist early nerve identification by providing the surgeon with a tool to help locate and identify the particular nerve at risk to minimize trauma by alerting the surgeon when a particular nerve has been activated.
The Sentio MMG® system consists of a reusable Control Unit comprised of a touch-screen PC and Patient Module, a reusable Sensor Connector Module and an assortment of disposable conductive probes, stimulators, sensors, electrodes and electrode leads.
#### Intended Use: ર.
This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMG) signals and electrical stimulus of nerves. This device is indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2.
#### 7. Comparison to Predicate Devices:
The subject device has indications for use which are the same as the predicate device, K100992, is composed of the same materials as the predicate commercially marketed device, has the same or equivalent design features as the predicate device, is substantially equivalent in its method of . use, incorporates the same fundamental scientific technology and has functional characteristics which are the same or equivalent to those of the predicate device. Due to the equivalency of indications for use, materials of composition, design features, method of use, scientific technology and functional characteristics, the device raises no new safety or effectiveness issues.
A comparison of the technological characteristics to the predicate device is provided below:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "sentio" in lowercase letters next to a four-pointed star-like symbol. The word "sentio" is written in a simple, sans-serif font. The star symbol is positioned to the left of the word, slightly above the baseline of the text.
r
# Table 1: Overview & Indications
| Subject Device: | Predicate Device: |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sentio MMG® #K131304 | ISS NeuralMAS #K100992 |
| System Overview<br>Uses low level electrical current to stimulate<br>nerves sufficient to produce a detectable<br>mechanical signal in contracting muscle using<br>skin-surface mounted mechanomyographic<br>(MMG) sensors. | System Overview<br>Uses low level electrical current to stimulate nerves sufficient to produce a<br>detectable mechanical signal in contracting<br>muscle using skin-surface mounted<br>mechanomyographic (MMG) sensors. |
# Table 2: Physical Attributes
| Attribute | Sentio MMG®<br>#K131304 | ISS NeuralMAS<br>#K100992 |
|-----------------------------------------|--------------------------|---------------------------|
| Control Module: PC/Display | | |
| Type: | Touchscreen | Touchscreen |
| Number of Channels: | 10 | 10 |
| Channel<br>Enable/Disable<br>Controls: | Touchscreen controlled | Touchscreen controlled |
| Event Threshold<br>Control and Display: | Touchscreen controlled | Touchscreen controlled |
| Stimulator | | |
| Type Constant: | Electrical Current (mA) | Electrical Current (mA) |
| Range: Control: | 1-15 mA | 1-15 mA |
| Waveform: | Digitally controlled | Digitally controlled |
| Pulse Duration: | Monophasic, square pulse | Monophasic, square pulse |
| | 100μς | 100μς |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "sentio" in a bold, sans-serif font. To the left of the word is a four-pointed star-like symbol. The word is in lowercase letters and appears to be a logo or brand name.
| Rate: | 2Hz | 2Hz |
|-------------------|-----------------------------|-----------------------------|
| Stimulus Probe: | Monopolar | Monopolar |
| Activation: | Touch screen or hand switch | Touch screen or hand switch |
| Monitoring Sensor | | |
| Function: | Mechanomyographic (MMG) | Mechanomyographic (MMG) |
| Type: Attachment | Electromechanical | Electromechanical |
| Site: Attachment | Skin surface | Skin surface |
| Mode: | Adhesive | Adhesive |
## 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
A comparative performance evaluation was conducted to assess statistical agreement between the subject device and the predicate. Results for positive, and overall percent agreement have established that 1) the principle functions of the Sentio MMG® system are effective, such that spinal nerve status and location may be ascertained and monitored, and 2) the data processed by the subject device following surgical use correlate well with those acquired by the predicate device.
#### 9. Discussion of Clinical Tests Performed:
Not Applicable
#### 10. Conclusions:
The conclusions drawn from our non-clinical testing (to include our market data evaluation) of the subject device demonstrates that the subject device is as safe, as effective and performs better than the legally marketed predicate device. Any noted differences between the devices are minor and do not raise new issues of safety and effectiveness. The subject device is substantially equivalent to the legally marketed predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing and tail. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 29, 2014
Innovative Surgical Solutions, LLC Mr. Aashish Shah Vice President Quality & Operations 21520 Bridge Street Southfield, MI 48033
Re: K131304
Trade/Device Name: Sentio MMG Regulation Number: 21 CFR 874.1820 Regulation Name: Neurosurgical Nerve Locator Regulatory Class: Class II Product Code: PDQ Dated: April 29, 2014 Received: April 30, 2014
Dear Mr. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mr. Aashish Shah
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel - S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510{k} Number (if known) K131304
## Device Name
Sentio MMG
## Indications for Use (Describe)
This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMO) signals and electrical stimulus of nerves. This device is indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2.
Type of Use (Select one or both, as applicable)
∑ Prescription Use (Part 21 CFR 801 Subpart D)
■ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# ം Date: 2014.05.29
22:30:24 -04'00'
EF PSC Publishing Scriptocs (30)) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.