K131272 · Medusa Medical Technologies, Inc. · DXJ · Jan 3, 2014 · Cardiovascular
Device Facts
Record ID
K131272
Device Name
SIREN EPCR SUITE
Applicant
Medusa Medical Technologies, Inc.
Product Code
DXJ · Cardiovascular
Decision Date
Jan 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2450
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Siren ePCR Suite™ is intended for the collection, storage and printing of patient data that is entered by a user (paramedics), or captured from specified medical devices, and integrated into a patient care report (patient electronic medical record). Siren ePCR Suite™ is intended for use by qualified medical personnel providing direct patient care in the pre-hospital environment to document the care provided. Siren ePCR Suite™ is indicated for use by health care providers whenever there is a need for generation of a patient record.
Device Story
Siren ePCR Suite is a software-only medical data collection system; operates on commercial off-the-shelf hardware (desktop/tablet PCs). Inputs include manual user entry and data captured from medical devices (e.g., monitor/defibrillators). System stores, prints, and integrates data into electronic patient care reports (ePCR). Used by paramedics/EMTs in pre-hospital settings. Features include data encryption, audit trails for database changes, and interfaces with CAD/dispatch, billing, and hospital systems via HL7. Facilitates clinical documentation and record generation; benefits include standardized patient data management and improved information exchange between pre-hospital providers and hospitals.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical software verification and validation testing, including hazard analysis and risk mitigation, confirming the device performs as intended.
Technological Characteristics
Software-only system; runs on Windows XP/7 or Windows Server 2003/2008. Connectivity via internet for data sync; supports HL7 interoperability. Features include data encryption, electronic signature support, and integration with external drug monograph databases (e.g., MicroMedix). Non-alarming software.
Indications for Use
Indicated for use by qualified medical personnel (paramedics/EMTs) providing direct patient care in pre-hospital environments to document care and generate patient records.
Regulatory Classification
Identification
A medical cathode-ray tube display is a device designed primarily to display selected biological signals. This device often incorporates special display features unique to a specific biological signal.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Medusa Medical Technologies Inc. c/o Dr. Diane Sudduth Senior Consultant QA Emergo Group, Inc. 816 Congress Ave., Suite 1400 Austin. Texas 78701
September 1, 2023
Re: K131272
Trade/Device Name: Siren ePCR Suite™ Regulation Number: 21 CFR 870.2450 Regulation Name: Medical Cathode-ray Tube Display Regulatory Class: Class II Product Code: DXJ
Dear Dr. Diane Sudduth:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 3, 2014. Specifically, FDA is updating this SE Letter to remove the secondary product code NSX as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter, please contact Aneesh Deoras, OHT2: Office of Cardiovascular Devices, 240-402-4363, Aneesh.Deoras@fda.hhs.gov.
Sincerely,
# Aneesh S. Deoras -S
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's logo in the center. The logo consists of a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 3, 2014
Medusa Medical Technologies, Incorporated C/O Dr. Diane Sudduth Senior Consultant, QA Emergo Group, Incorporated 816 Congress Avenue, Suite 1400 Austin, TX 78701
Re: K131272
Trade/Device Name: Medical Technologies, Inc. Siren ePCR Suite Regulation Number: 21 CFR 870.2450 Regulation Name: Medical Cathode-Ray Tube Display Regulatory Class: II Product Code: DXJ, NSX Dated: November 18, 2013 Received: November 19, 2013
Dear Dr. Sudduth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
... .. ..
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Mary S. Runner -S
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use Statement
#### 510(k) Number (if known): K131272
Device Name: Siren ePCR Suite™
#### Indications for Use:
Siren ePCR Suite™ is intended for the collection, storage and printing of patient data that is entered by a user (paramedic), or captured from specified medical devices, and integrated into a patient care report (patient electronic medical record). Siren ePCR Suite™ is intended for use by qualified medical personnel providing direct patient care in the pre-hospital environment to document the care provided. Siren ePCR Suite™ is indicated for use by health care providers whenever there is a need for generation of a patient record.
Prescription Use_ ਮ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by Richard C. Chapman Date: 2014.01.02 1 1:33:26 -05'00'
{4}------------------------------------------------
Medusa Medical Technologies Inc. Traditional 510(k) Premarket Submission Siren ePCR Suite™
# EMERGO GROUP
## 510(k) Summary
#### 1. Submission Sponsor
Medusa Medical Technologies Inc. Suite 140-36 Solutions Dr Halifax, Nova Scotia B3S 1N2 Canada Phone: (902) 429-1200 Fax: (902) 484-5583 Contact: Craig Fraser, VP of Sales and Product Management
#### 2. Submission Correspondent
Emergo Group, Inc. 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (561) 305-5075 Office Phone: (512) 327-9997 Fax: (512) 327-9998 Contact: Diane Sudduth, Senior Consultant, QA Email: project_management@emergogroup.com
#### 3. Date Prepared
May 2, 2013
#### 4. Device Identification
Medusa Medical Technologies, Inc. Siren ePCR Suite™ Trade/Proprietary Name: Common/Usual Name: Siren ePCR Display, Cathode Ray Tube, Medical Classification Name: Classification Regulation: 870.2450 DX1: NSX Product Code: Device Class: Class II Classification Panel: Cardiovascular; General Hospital
#### 5. Predicate Devices
K103473 Zoll Medical Corporation - RescueNet ePCR
#### 6. Device Description
Siren ePCR Suite™ is a software-only product. Siren ePCR Suite™ is a medical data collection system used to collect, store and print patient data that is entered by a user (caregiver), or captured from specified medical devices, and is integrated into a patient care report (patient
{5}------------------------------------------------
electronic medical record). Siren ePCR Suite™ is non-alarming software that runs on a variety of commercial off-the-shelf hardware.
#### 7. Intended Use
Siren ePCR Suite™ is intended for the collection, storage and printing of patient data that is entered by a user (paramedics), or captured from specified medical devices, and integrated into a patient care report (patient electronic medical record). Siren ePCR Suite™ is intended for use by qualified medical personnel providing direct patient care in the pre-hospital environment to document the care provided. Siren ePCR Suite™ is indicated for use by health care providers whenever there is a need for generation of a patient record.
#### 8. Comparison of Technological Characteristics
The following table compares the Siren ePCR Suite™ to the predicate device with respect to intended use, overall technological and functional characteristics, providing more detailed information regarding the basis for the determination of substantial equivalence.
The Siren ePCR Suite™ is similar in design and function to the predicate device for the modes of operation and use.
| Manufacturer | Zoll Medical Corporation | Medusa Medical Technologies Inc. |
|-------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Predicate<br>RescueNet ePCR | New Device<br>Siren ePCR Suite™ |
| 510(k) Number | K103473 | Not assigned |
| Product Code | DJX<br>NSX | DJX<br>NSX |
| Regulation Number | 870.2450<br>Null | 870.2450<br>Null |
| Regulation Name | Display, Cathode Ray Tube,<br>Medical | Display, Cathode Ray<br>Tube, Medical |
| Indications for Use | Software, Transmission and<br>Storage, Patient Data<br>RescueNet ePCR is intended for<br>the collection, storage and<br>printing of patient data that is<br>entered by a user (caregiver), or<br>captured from specified medical<br>devices, and integrated into a<br>patient care report (patient<br>electronic medical record). | Software, Transmission and<br>Storage, Patient Data<br>Siren ePCR Suite™ is intended<br>for the collection, storage and<br>printing of patient data that is<br>entered by a user (paramedic),<br>or captured from specified<br>medical devices, and integrated<br>into a patient care report<br>(patient electronic medical record). |
| Manufacturer | Zoll Medical Corporation | Medusa Medical Technologies Inc. |
| Trade Name | Predicate<br>RescueNet ePCR | New Device<br>Siren ePCR Suite™ |
| | RescueNet ePCR is intended for use<br>by qualified medical personnel providing direct<br>patient care in the pre-hospital environment to document the<br>care provided. RescueNet ePCR<br>is indicated for use by health<br>care providers whenever there<br>is a need for generation of a<br>patient record. | record). Siren ePCR Suite™ is<br>intended for use by qualified<br>medical personnel providing<br>direct patient care in the pre-<br>hospital environment to<br>document the care provided.<br>Siren ePCR Suite™ is indicated<br>for use by health care providers<br>whenever there is a need for<br>generation of a patient record. |
| Material | Software | Software |
| System Requirements | | |
| Compatible operating<br>system (software only) | Windows XP Professional | Field User: Windows XP or XP<br>Tablet Edition SP3 or Windows<br>7 SP1; SQL Express 2005 SP3<br><br>Server: Windows Server 2003<br>R2 SP3; SQL Server 2005 SP3 /<br>SQL Server 2008 SP3<br><br>.NET Framework |
| Web-Based Application<br>(Locally installed vs.<br>vendor server based) | WebPCR module provides a<br>web based access when<br>connected to web browser or<br>WebPCR server. | Web based Administration and<br>Workflow. Store and forward<br>communication from tablet to<br>server. |
| Hardware | | |
| Compatibility and system<br>requirements | | |
| Desktop PC | Yes | Yes |
| Tablet PC | Yes | Yes |
| Pocket PC/Palm device<br>compatible | Yes | No |
| Internet connection<br>required at all time or<br>only during data sync | Only during data sync from<br>mobile devices | Only during data sync from<br>mobile devices |
| Printer compatibility | Yes, can print locally or across a<br>network | Yes, can print locally or across a<br>network |
| Wireless access supported<br>(ie Verizon/sprint WWAN) | Yes | Yes |
| Manufacturer | Zoll Medical Corporation | Medusa Medical Technologies Inc. |
| Trade Name | Predicate<br>RescueNet ePCR | New Device<br>Siren ePCR Suite™ |
| Security | | |
| Data encryption | Yes | Yes |
| Ability to lock PCR once completed | Yes | Yes |
| Tracks changes to module databases, including date, time, computer and user who made the change | Yes | Yes |
| Data Exchange | | |
| Interface to server (Data synchronization) | Yes, wireless or hard wired | Over internet connection |
| Interface with CAD/Dispatch | Yes | Yes |
| Interface with Hospitals (Fax, email/etc., direct sync, etc.) | Yes | Yes |
| Interface with Hospital Pre-Alert | Yes | Yes |
| Interface with Billing | Yes | Yes |
| Interface with Medical Equipment | Physio-Control<br>LifePak 11 monitor/defibrillator,<br>LifePak 12/15<br>monitor/defibrillator,<br>LifePak 500<br>monitor/defibrillator;<br>Philips<br>HeartStart MRx;<br>Zoll<br>1600,<br>AED Plus/AED Pro,<br>M Series/E Series | Physio-Control<br>LifePak 12/15<br>monitor/defibrillator<br>Philips<br>HeartStart MRx<br>Zoll<br>M Series/E Series |
| EKG Integration | Yes | Yes |
| Mobile-to-mobile data transfer | Yes | Yes |
| Interfaces to electronic record (EHR) / HL7 | Yes | Yes |
| Manufacturer | Zoll Medical Corporation | Medusa Medical Technologies<br>Inc. |
| Trade Name | Predicate | New Device |
| | RescueNet ePCR | Siren ePCR Suite™ |
| Additional Features | | |
| Electronic Signature<br>Support | Yes | Yes |
| Ability to link reference<br>documents (protocols) | Yes | Yes |
| Ability to populate patient<br>info from previous patient<br>contact | Yes | Yes |
| Drug Monograph<br>database | No | Yes via integration with existing<br>commercial drug monograph<br>providers (MicroMedix) |
| Use environment | EMT, paramedic | EMT, paramedic |
| Intended users<br>(target population) | Professional Users | Professional Users |
Table 5A - Comparison of Characteristics
{6}------------------------------------------------
### Medusa Medical Technologies Inc. Traditional 510(k) Premarket Submission Siren ePCR Suite™
# EMERGO G GROUP
·
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
•
:
{7}------------------------------------------------
•
{8}------------------------------------------------
#### 9. Non-Clinical Testing
The device's software development, verification and validation have been carried out in accordance with the FDA's guidance documents. The software was tested against the established Software Design Specifications for each of the test plans to assure the device performs as intended.
The device Hazard analysis was completed and risk control implemented to mitigate hazards. The testing results supports that all specifications have met the acceptance criteria of each module and interaction of processes. Siren ePCR Suite™ device passed all testing and supports the claims of substantial equivalence and safe operation.
#### 10. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
#### 11. Statement of Substantial Equivalence
By definition, a new device is substantially equivalent to a predicate device when the device has the same intended use as the previously cleared predicate device and either (i) the same technological characteristics as the predicate, or (ii) if the new device has different
{9}------------------------------------------------
technological characteristics, then those differences raise no new issues regarding the safety or effectiveness of the new device.
It has been shown in this 510(k) submission that Siren ePCR Suite™ has the same intended use as the predicate device and that any technological differences between the Siren ePCR Suite™ software and the predicate device do not raise any questions regarding Siren ePCR Suite™'s safety and effectiveness. Performance testing and compliance with voluntary standards demonstrate that Siren ePCR Suite™ software is substantially equivalent to the relevant aspects of the predicate device in terms of design, principals of operation, performance characteristics, and intended use. The Siren ePCR Suite™ software, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.