← Product Code [MEB](/productcode/MEB) · K131247

# MULTIBOLUS II DISPOSABLE PAIN MANAGEMENT SYSTEM (K131247)

_Mfs - Medical Flow Systems, Ltd. · MEB · Aug 14, 2013 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K131247

## Device Facts

- **Applicant:** Mfs - Medical Flow Systems, Ltd.
- **Product Code:** [MEB](/productcode/MEB.md)
- **Decision Date:** Aug 14, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The MultiBolus II™ is intended to provide intermittent delivery of medication on patient demand, using patient control module allowing bolus doses, at relatively rapid velocity in parallel to a continuous basal flow or without continuous basal flow, to/or around surgical wound site and/or close proximity to nerves for preoperative, perioperative and postoperative regional anesthesia and pain management. Routes of administration may be intraoperative, perineural, percutaneous, subcutaneous, intramuscular and epidural.

## Device Story

MultiBolus II is a 100% mechanical, semi-automatic disposable device; integrates with MFS elastomeric pumps (SmartBlock/SmartInfuser) to create a PCA system. Input: medication from elastomeric pump reservoir. Operation: hydraulic mechanism responds to pressure differential between bolus reservoir and pump reservoir; enables automatic filling when empty; seals flow when inactive. Output: bolus dose of medication on patient demand. Used in hospital, clinic, or home; operated by patient. Healthcare providers set bolus volume and lock-out parameters. Benefits: provides on-demand pain relief; prevents unintentional delivery via sealing mechanism.

## Clinical Evidence

Bench testing only. Evaluated functional performance (operating cycles, priming efficiency, bolus volumes, delivery velocity, lock-out time), leak resistance, tube bonding strength, hydraulic check valve function, external force resistance, shelf life (1-year real-time), and packaging integrity. Usability validated with physicians and patients.

## Technological Characteristics

100% mechanical, semi-automatic infusion accessory. Materials: biocompatible polymers/elastomers (ISO 10993). Energy: hydraulic pressure differential. Connectivity: none (mechanical). Sterilization: EtO (SAL 10^-6). Form factor: disposable module for elastomeric pumps.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Baxter Healthcare Corporation Multirate Infusor (LV and SV) with PCM ([K051253](/device/K051253.md))
- I-Flow Corporation ON-Q Pump ([K063530](/device/K063530.md))

## Reference Devices

- MFS FlowSplitter ([K072053](/device/K072053.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters "mfs" in a stylized font, with a water droplet shape incorporated into the letter "f". The text "Medical Flow Systems Ltd." is written in a smaller font below the letters.

Image /page/0/Picture/3 description: The image is a circular seal with the words "Quality Systems Certified" around the edge. In the center of the seal, the text "ISO 9001:2008" and "ISO 13485:2003" are stacked on top of each other. The seal appears to be used to indicate that a company or organization has met the requirements of the ISO 9001 and ISO 13485 standards.

# 510(K) SUMMARY FOR MFS - MEDICAL FLOW SYSTEMS' MULTIBOLUS IITM

### DATE PREPARED: MARCH 31, 2013

- 1. 510(K) OWNER NAME
AUG 1 4 2013

MFS - Medical Flow Systems Ltd. ELT Building, Dora Industrial Zone, Shlomi 22832, Israel Tel: +972 4 9808280, Fax: +972-4-9875565, E mail: info@mfs-medical.com.

Contact person name: Mr. Ofer Shay, Managing Director. Telephone: +972-4-9114876, Fax: +972-4-9875565, E mail: offers@mfs-medical.com.

- DEVICE NAME 2.
Common/Usual Name: Disposable Pain Management System Proprietary/Trade name: MultiBolus II™ ('Parallel' and 'In-line' modules) Classification: MFS's MultiBolus IITM device has been classified as Class II device under the following classification names:

| Classification Name         | Product Code | 21 CFR Ref. | Panel            |
|-----------------------------|--------------|-------------|------------------|
| Pump, Infusion, Elastomeric | 80-MEB       | 880.5725    | General Hospital |
| Pump, Infusion, Pca         | 80-MEA       |             |                  |

#### PREDICATE DEVICES 3.

MFS's MultiBolus II™ device is substantially equivalent to the following Predicate Devices:

- 3.1 Baxter Healthcare Corporation Multirate Infusor (LV and SV) with PCM, cleared under 510(k) number K051253 and;
- 3.2 I-Flow Corporation ON-Q Pump, cleared under 510(k) number K063530.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters "mfs" in a stylized font, with a teardrop shape replacing the vertical line of the "f". The text "Medical Flow Systems Ltd." is written below the letters in a smaller font.

Image /page/1/Picture/2 description: The image is a circular seal with the words "Quality Systems Certified" around the outer edge. Inside the circle are the text "ISO 9001:2008" and "ISO 13485:2003" stacked on top of each other. The seal appears to be a certification mark related to quality management systems.

#### DEVICE DESCRIPTION র্বা

MFS's MultiBolus II™ is a semiautomatic disposable 100% mechanical device used as a complementary to MFS's pain management system to be controlled by the patient as prescribed and set by an authorized medical team member using either one or two catheters.

The device is to be integrated with the existing cleared MFS's elastomeric pumps and is designed to provide a bolus delivery of pain relief medication with or without parallel continuous basal flow.

The MultiBolus IITM, together with the elastomeric pump creates a PCA (Patient Control Analgesia) infusion pump.

The MultiBolus II™ with Parallel set configuration is used when the delivery of a bolus medication is required to be executed parallel to a continuous basal flow.

The MultiBolus II™ with In-Line set configuration is used when the delivery of a bolus medication is required only.

Both configurations may be used during a dual catheter administration when a flow splitter such as; MFS's FlowSplitter™ (cleared within MFS's 510(k) number: K072053) device is assembled at the set exit port.

The device was developed by MFS to assist patients with pain management following surgical procedures or other pain managements. The device may be used in the hospital, at the clinic and at the home environment for out-patients subject to the physician decision.

MFS's MultiBolus II™ has the same indication for use and is substantially equivalent to the predicate devices selected by MFS in order to determine substantial equivalence as following detailed in this submission. The combination of these predicate-devices indication for use, technologies and performances supports our MultiBolus II™ substantial equivalency.

Body contact materials were evaluated for biocompatibility with accordance to FDA's Memorandum - #G95 1, May 1, 1995 and ISO 10993-1:2009.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters 'mfs' in a stylized font, with a water droplet incorporated into the 'f'. The words 'Medical Flow Systems Ltd.' are written below the letters in a smaller font. The logo is black and white.

Image /page/2/Picture/2 description: The image is a circular seal with the words "Quality Systems Certified" around the edge. In the center of the seal, the text "ISO 9001:2008" and "ISO 13485:2003" are stacked on top of each other. The seal is black and white, and the text is in a simple sans-serif font.

#### ડ. INTENDED USE

The MultiBolus II™ is intended to provide intermittent delivery of medication on patient demand, using patient control module allowing bolus doses, at relatively rapid velocity in parallel to a continuous basal flow or without continuous basal flow, to/or around surgical wound site and/or close proximity to nerves for preoperative, perioperative and postoperative regional anesthesia and pain management. Routes of administration may be intraoperative, perineural, percutaneous, subcutaneous, intramuscular and epidural.

### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.

The MultiBolus II™ is a semi-automatic 100% mechanical device to be activated by the patient after integrating with one of MFS's cleared elastomeric pumps (SmartBlock™ or SmartInfuser™).

MultiBolus II™ controls the medication flow through the system utilizing a built-in hydraulic mechanism that responds to the pressure differential between the pressure at the bolus device internal reservoir and the pressure at the pumps' reservoir. This mechanism enables an automatic medication filling into the bolus' reservoir when becomes empty, and to eliminate any flow out of the bolus device when not being activated and/or once the bolus' reservoir is empty ("stand-by" position).

### The comparison table with the predicate devices is following presented:

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo features the letters "mfs" in a stylized font, with a water droplet replacing the dot on the "i". The text "Medical Flow Systems Ltd." is written in a smaller font below the letters.

Image /page/3/Picture/2 description: The image is a circular seal with the words "Quality Systems Certified" around the outer edge. In the center of the seal are the text "ISO 9001:2008" and "ISO 13485:2003", stacked on top of each other. The text is in a simple, sans-serif font and is surrounded by a thick black border.

| Feature                                                 | MFS's<br>MultiBolus IITM                                                                                                                                                                                                                                                                                                                                                                                                                                    | Baxter Helthcare<br>Multirate Infusor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | I-Flow Corporation<br>ON-Q Pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                         | -New Device -                                                                                                                                                                                                                                                                                                                                                                                                                                               | 510(k) no. K051253                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 510(k) no. K063530                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Intend of use                                           | The MultiBolus IITM is<br>intended to provide<br>intermittent delivery of<br>medication (such as local<br>anesthetic and/or narcotics) on<br>patient demand at relatively<br>rapid velocity, parallel to a<br>continuous basal flow or with<br>no continuous basal flow, to/or<br>around surgical wound site<br>and/or close proximity to<br>nerves for preoperative,<br>periopertative and<br>postoperative regional<br>anesthesia and pain<br>management. | The intended use of the<br>Multirate Infusor with<br>Patient Control Module<br>includes slow, continuous,<br>subcutaneous or epidural<br>administration of pain<br>medications. It may also<br>include the slow,<br>continuous infusion of<br>pain medications directly<br>into an intra-operative<br>site, subcutaneously for<br>postoperative pain<br>management or the<br>continuous infusion of a<br>local anesthetic near a<br>nerve for regional<br>anesthesia. The Patient<br>Control Module allows<br>for intermittent bolus<br>doses of pain medication<br>on patient demand.<br>Prescription | 1. The ON-Q Pump is<br>intended for continuous<br>and/or intermittent<br>delivery of medication<br>(such as local<br>anesthetics or narcotics)<br>to or around surgical<br>wound sites and/or close<br>proximity to nerves for<br>preoperative,<br>perioperative and<br>postoperative regional<br>anesthesia and pain<br>management. Routes of<br>administration may be<br>intraoperative,<br>perineural, percutaneous<br>and epidural.<br>[2. The ON-Q Pump<br>is...] - Note: Section 2 is<br>irrelevant to MultiBolus<br>IITM since it does not<br>contain pump. |
| Operation<br>Principle                                  | Mechanical, semi-automatic                                                                                                                                                                                                                                                                                                                                                                                                                                  | Mechanical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Mechanical, semi-<br>automatic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Safety                                                  | • No alarm<br>• Secure/lock of selected<br>volume<br>• Sealing by Hydraulic<br>independent valve to avoid<br>continuous flow when not<br>activated.                                                                                                                                                                                                                                                                                                         | • No alarm<br>• Per-define volume<br>• Mechanical<br>• Mechanical closure<br>(not necessarily<br>sealed) thus<br>continuous flow when<br>not activated may be<br>possible.                                                                                                                                                                                                                                                                                                                                                                                                                             | • No alarm<br>• Per-define volume<br>• Sealing by<br>mechanical valve<br>(                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Feature                                                 | MFS's<br>MultiBolus IITM                                                                                                                                                                                                                                                                                                                                                                                                                                    | Baxter Helthcare<br>Multirate Infusor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | I-Flow Corporation<br>ON-Q Pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                         | -New Device -<br>0ml, 3ml, 4ml, 5ml & 6ml.<br>• Lock Out flow rate:<br>6 ml/hr<br>• Velocity (through 21G<br>Catheter): 6ml/minute.<br>• Accuracy -<br>Volume (manual): +5%/-<br>10%;<br>Lock-Out: ±20%                                                                                                                                                                                                                                                     | 510(k) no. K051253<br>device: 0.5ml or 2.0 ml.<br>• Lock-out flow rate<br>0.5ml/hr, 2ml/hr<br>• Velocity (through 21G<br>Catheter): no data<br>available (estimation:<br>1 ml/minute).<br>• Accuracy -<br>No data                                                                                                                                                                                                                                                                                                                                                                                      | 510(k) no. K063530<br>5.0ml<br>• Lock-out flow rate<br>per device:<br>5ml/hr or 10ml/hr<br>• Velocity (through<br>21G Catheter):<br>1.5 ml/minute (tested<br>by us).<br>• Accuracy -<br>Volume:<br>+10% / -20%<br>Lock-Out: ±20%                                                                                                                                                                                                                                                                                                                                   |
| Bolus Injection<br>on patient<br>demand                 | Yes. User needs to press once.                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes. User needs to hold<br>until bolus is completed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Yes. User needs to press<br>once.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Parallel<br>Continuous flow<br>and 'on-demand'<br>bolus | Yes (the 'Parallel' model only)                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Materials                                               | Biocompatible Polymers,<br>Elastomers per: ISO 10993<br>parts 1, 4, 5, 6, 10 & 11.                                                                                                                                                                                                                                                                                                                                                                          | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| PCM - Patient<br>Control Module                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Use<br>Environment                                      | Hospital, clinic, home use                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Sterility                                               | Sterile. EtO                                                                                                                                                                                                                                                                                                                                                                                                                                                | Sterile, EtO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Sterile, EtO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

### Table 1: Comparison table with Predicate Devices

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo features the letters 'mfs' in a stylized font, with the 'f' being larger and having a teardrop shape incorporated into it. Below the letters, the full name 'Medical Flow Systems Ltd.' is written in a smaller, simpler font.

Image /page/4/Picture/2 description: The image is a black and white circular seal. The seal has the words "Quality Systems Certified" around the edge of the circle. In the center of the seal are the words "ISO 9001:2008" and "ISO 13485:2003".

Based on the above comparison, it can be shown that in all major features MFS's MultiBolus III" is substantial equivalent to both legally marketed devices selected as predicate devices while our device infuse the medication at a relative higher velocity and also provides a complete sealing preventing any unintentional drug delivery through the bolus device when not activated or when being empty.

..............................................................................................................................................................................

..............................................................................................................................................................................

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters 'mfs' stacked on top of each other, with the 'f' being the largest and having a teardrop shape in the middle. Below the letters is the text 'Medical Flow Systems Ltd.' in a smaller font. The logo is black and white.

Image /page/5/Picture/2 description: The image shows a circular stamp with the words "Quality Systems Certified" around the edge. In the center of the stamp are the text "ISO 9001:2008" and "ISO 13485:2003". The stamp is black and white and appears to be a certification mark.

### 7. PERFORMANCE DATA

MFS's MultiBolus II'M has been successfully tested through bench, usability and safety tests to support the determination of substantial equivalence with predicate devices.

## The summary of performance data is following presented in order to support substantial equivalency and safety/efficacy of our device:

| No. | Test Name           | Purpose                                                                                                                                                     | Method                                                                                        | Acceptance Criteria                                                                                                                                                                                                                                                                                                             | Results                                                                                                                                                                                                                                                                                                                                               | Conclusion                                                                                                                                                |
|-----|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
|     | Functional Tests    |                                                                                                                                                             |                                                                                               |                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                           |
| 1   | Operating<br>Cycles | Assuring the<br>device<br>capability to<br>function in<br>multi iteration<br>(repeatedly)                                                                   | 30 activation iteration of<br>each device.                                                    | Activate semi-<br>automatically<br>according to spec in<br>terms of functionality<br>and in all volumes 3,<br>4, 5, & 6ml):<br>• Lockout time<br>according to spec<br>(not completely<br>filled after the<br>specified time).<br>• No parallel flow is<br>allowed when not<br>activated (scale<br>monitor should show:<br>0.0). | • Device<br>performed as<br>intended.<br>• Under the<br>test conditions<br>the maximum<br>volume<br>tolerance was:<br>-0.75ml.<br>No result<br>exceeding<br>+0.0ml was<br>detected.<br>• When<br>nominal set to<br>6ml. no<br>complete<br>filling after<br>the specified<br>time occurred.<br>• Before<br>activation scale<br>monitor<br>showed: 0.0. | Under the<br>Test's<br>conditions all<br>acceptance<br>criteria were<br>met.                                                                              |
| 2   | Priming             | Verifying<br>priming<br>efficiency and<br>0.2µ filter<br>effectiveness<br>in trapping air<br>bubbles in<br>order to make<br>priming for<br>the parallel set | Conducting priming<br>according to IFU in the<br>basal path flow only<br>(parallel set model) | • No air bubbles after<br>the 0.2µ filter are<br>allowed.                                                                                                                                                                                                                                                                       | • No air<br>bubbles after<br>the 0.2µ filter<br>were detected.                                                                                                                                                                                                                                                                                        | Under the test<br>condition, all<br>results met<br>acceptance<br>criteria. All air<br>bubbles were<br>trapped by the<br>0.2µ filter in<br>priming for the |

MFS 510(k) MultiBolus IITM – Revised Summary

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters 'mfs' in a stylized font, with a teardrop shape replacing the dot on the 'i'. The text 'Medical Flow Systems Ltd.' is written below the letters in a smaller font.

Image /page/6/Picture/2 description: The image is a circular seal with the words "Quality Systems Certified" around the edge. In the center of the seal, it says "ISO 9001:2008" and "ISO 13485:2003". The text is in black and the background is white. The seal is used to indicate that a company's quality management system has been certified to meet the requirements of the ISO 9001 and ISO 13485 standards.

| No.              | Test Name                    | Purpose                                                                            | Method                                                                                                                                           | Acceptance Criteria                                                                                                                                                     | Results                                                                                                                     | Conclusion                                                     |
|------------------|------------------------------|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Functional Tests |                              |                                                                                    |                                                                                                                                                  |                                                                                                                                                                         |                                                                                                                             |                                                                |
|                  |                              | model (option to exclude priming of the MultiBolus IITM device).                   |                                                                                                                                                  |                                                                                                                                                                         |                                                                                                                             | parallel set model.                                            |
|                  | Verifying priming efficiency |                                                                                    | Conducting priming according to IFU (in-line set model with the MultiBolus IITM)                                                                 | • No air bubbles in the tubing between the 0.2µ filter and the exit port after 1st activation are allowed.<br>• Volume of 2nd & 3rd activations should be as specified. | • No air bubbles after the 0.2µ filter were detected.<br>• Results were according to spec.                                  | Under the test condition, all results met acceptance criteria. |
| 3                | Volumes                      | Verifying that actual volume is within the spec:                                   | Measuring the actual infused bolus volumes with and without catheter after semi-automatic, manual activation (from set volumes of 3, 4, 5, 6 ml. | Actual infused bolus volume range should be according to spec.                                                                                                          | Under the test's conditions the actual delivered volume in both semi-automatic and manual activation was according to spec. | Under the test condition, all results met acceptance criteria. |
| 4                | Velocity                     | Verifying actual velocity is within the spec                                       | Measuring the actual velocity at 6ml volume with 21G catheter                                                                                    | Velocity should be at-least 5 ml/minute                                                                                                                                 | Under the test's conditions all results were greater than 5ml/minute.                                                       | Under the test condition, all results met acceptance criteria. |
| 5                | Lock-Out Time                | Verifying that actual re-filling time at nominal volume of 6ml is within the spec. | Verifying the actual filling time of 6 ml volume is between 48-72minutes (state chosen as worst case representing all volumes)                   | Filling time was between 48-72 minutes of nominal volume.                                                                                                               | After the specified time reservoirs were not completely filled. After 72 min. all reservoirs were completely                | Under the test condition, all results met acceptance criteria. |

MFS 510(k) MultiBolus II™ – Revised Summary

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters 'mfs' in a stylized font, with a water droplet shape incorporated into the 'f'. The text 'Medical Flow Systems Ltd.' is written below the letters. The logo is black and white.

Image /page/7/Picture/2 description: The image is a circular seal with the words "Quality Systems Certified" around the perimeter. In the center of the seal are the certifications "ISO 9001:2008" and "ISO 13485:2003". The text is in a bold, sans-serif font and is arranged to fit within the circular shape of the seal. The seal appears to be used to indicate that a product or organization has met certain quality standards.

| No. | Test Name                       | Purpose                                                                                               | Method                                                                                                                                                                                       | Acceptance Criteria                                                                                                                                                                               | Results                                                                                                                                                    | Conclusion                                                                                |  |
|-----|---------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|--|
|     | Functional Tests                |                                                                                                       |                                                                                                                                                                                              |                                                                                                                                                                                                   |                                                                                                                                                            |                                                                                           |  |
| 6   | Leaks                           | Verify the<br>device design<br>withstand<br>operation<br>pressure                                     | Reservoir and tubing<br>connection withstand air<br>pressure of 3 bar at 10<br>seconds                                                                                                       | No leaks are allowed<br>within at-least 10<br>seconds under 3 bar.                                                                                                                                | filled.<br>Volumes in<br>semi-automatic<br>mode were as<br>specified.<br>No leaks<br>(stable<br>pressure level)<br>occurred<br>during test's<br>procedure. | Under the test<br>condition,<br>acceptance<br>criteria was<br>met.                        |  |
| 7   | Tube<br>bonding                 | Verify the<br>device's<br>tubing<br>connection<br>withstand the<br>forces during<br>use.              | Apply puling force of<br>1.5Kg on each tube for at-<br>least 10 seconds.                                                                                                                     | · Tube remains<br>connected to the<br>device.<br>· No leaks after<br>applying 1.5Kg.                                                                                                              | During test, no<br>tubing<br>disconnections<br>were detected<br>and no leaks<br>afterwards<br>(according to<br>same method<br>as in test #6).              | Under the test<br>conditions,<br>acceptance<br>criteria was<br>met.                       |  |
| 8   | Hydraulic<br>check valve        | Verify no<br>flow through<br>the device<br>when not<br>activated                                      | Connect the device to<br>filled-out reservoir for 14<br>days in a static position<br>and evaluate against<br>acceptance criteria.                                                            | During test period, the<br>following are not<br>allowed:<br>· Observed leaks;<br>· Gross weight of the<br>device after 14 days is<br>equal or up to 10 gram<br>less than day 0<br>(baseline).     | During the<br>test's period,<br>no leaks<br>occurred.<br>Maximum<br>gross weight<br>reduction was<br>within defined<br>acceptance<br>criteria.             | Under the test<br>conditions, all<br>tested samples<br>met the<br>acceptance<br>criteria. |  |
| 9   | External<br>force<br>during use | Verify the<br>device<br>withstand<br>reasonable<br>external<br>forces may be<br>applied during<br>use | Tested samples went<br>through two free falls from<br>one meter height. Then,<br>they were filled with<br>saline, activated according<br>to IFU and observed<br>against acceptance criteria. | During test, the<br>following are not<br>allowed:<br>· Cracks in the device<br>cover (shells), or other<br>plastic parts,<br>· Leaks,<br>· Device shall be<br>properly activated and<br>function. | Following the<br>test conditions,<br>no cracks, leaks<br>or<br>malfunctioning<br>were detected.                                                            | Under the test<br>conditions, all<br>acceptance<br>criteria were<br>met.                  |  |

..............................................................................................................................................................................

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters 'mfs' in a stylized font, with a teardrop shape replacing the dot on the 'i'. The text 'Medical Flow Systems Ltd.' is written in a smaller font below the letters.

Image /page/8/Picture/2 description: The image is a circular seal with the words "Quality Systems Certified" around the edge. In the center of the seal, the text "ISO 9001:2008" and "ISO 13485:2003" are stacked on top of each other. The seal appears to be used to indicate that a company's quality management system has been certified to meet the requirements of the ISO 9001:2008 and ISO 13485:2003 standards.

| No.              | Test Name                                                                     | Purpose                                                                          | Method                                                                                                                                                                                                                                                                                                      | Acceptance Criteria                                                                                                                                                                                                                                                                                                       | Results                                                                                                                                                                  | Conclusion                                                                                                            |
|------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Functional Tests |                                                                               |                                                                                  |                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                          |                                                                                                                       |
| 10               | Shelf life<br>equal to 1<br>year real-<br>time (post<br>accelerated<br>aging) | Verify that the<br>device will<br>function well<br>throughout its<br>shelf life. | After accelerate aging for<br>45 days under 52°C to<br>simulate 1 year shelf life,<br>products were tested using<br>19-21G catheters by filling<br>6ml, priming, delivered<br>volume weight, testing<br>volume post<br>45&72minutes and<br>activate bolus delivery<br>(simulating the real use 5<br>times). | Under the test method<br>the product should be<br>operated and<br>functioned according<br>to its spec.<br>• Volume filled (per<br>test 3),<br>• Lockout time (per<br>test 5),<br>• No parallel flow,<br>when not activated<br>(per test 8),<br>• No visible cracks<br>(per test 9),<br>• No visual leaks (per<br>test 9). | Under the test<br>method all<br>filled volumes;<br>lockout times;<br>sealing state<br>(no-parallel<br>flow); no-<br>cracks and<br>leaks tests<br>passed<br>successfully. | All tested<br>samples met<br>the acceptance<br>criteria and the<br>product was<br>qualified for 1<br>year shelf life. |

| No.           | Test Name                          | Purpose                                                                                                                                                                       | Method                                                                    | Acceptance Criteria                                                                                                                                                                                                                                                                                                                                                                                       | Results                                                                                                                                                                        | Conclusion                                                                                                                      |
|---------------|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Sterilization |                                    |                                                                                                                                                                               |                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                |                                                                                                                                 |
| 11            | EtO<br>Sterilization<br>validation | Validate that<br>the device may<br>be sterilized<br>for SAL 10-6<br>through EtO<br>established<br>cycle using<br>overkill<br>approach (also<br>called: Half<br>Cycle Method). | Per written protocol with<br>accordance to ISO 11135-<br>1:2007, Annex B. | With accordance to the<br>following standards:<br>Bio Burden:<br>ISO 11737-1:2006<br>ISO 11135-1:2007<br>ETO/ECH Residues:<br>ISO 10993-7:2008<br>LAL Endotoxin:<br>USP 32:2009 <161><br>Transfusion and<br>Infusion Assemblies<br>and Similar Medical<br>Devices and;<br>FDA's Guidance for<br>Pyrojen & Endotoxins<br>testing (Q&A),<br>June 2012<br>Sterility:<br>ISO 11135-1:2007<br>ISO 11737-2:2010 | All tests' results<br>were<br>determined to<br>be within the<br>standards'<br>acceptance<br>criteria.<br>More details are<br>available in<br>section 14 of<br>this submission. | Following the<br>sterilization<br>validation, SAL<br>10-6 was<br>established for<br>MFS's<br><i>MultiBolus II</i> TM<br>device. |

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters 'mfs' stacked on top of each other, with the 'f' being larger than the other letters. A water droplet is placed in the middle of the 'f'. The text 'Medical Flow Systems Ltd.' is written in a smaller font below the letters.

Image /page/9/Picture/2 description: The image is a circular seal with the words "Quality Systems Certified" around the perimeter. In the center of the seal, the text "ISO 9001:2008" and "ISO 13485:2003" are stacked on top of each other. The text is in bold and black, and the seal is also black and white.

| No.                 | Test Name              | Purpose                                                                                                                                      | Method                                                                                                                                                                                                                                                                                                                                         | Acceptance Criteria                                                                                                                                         | Results                                                                                                                                                                                                                                                              | Conclusion                                                                                                                                                                                                    |
|---------------------|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Packaging Integrity |                        |                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                             |                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                               |
| 12                  | Packaging<br>Integrity | • Verify that<br>the shipping<br>packaging the<br>device<br>provides<br>sufficient<br>protection to<br>the device<br>during<br>shipping      | Based on ASTM D5276-<br>98 (drop) & ISO 2248<br>(vertical impact by<br>dropping)                                                                                                                                                                                                                                                               | • All packages shall<br>remain whole,<br>unopened and<br>complete.<br>• No physical<br>damages to the sealed<br>sterile unit are<br>allowed.                | None of the<br>below defects<br>have been<br>occurred:<br>• No<br>significant<br>damage to<br>shipping cases<br>and internal<br>boxes.<br>• No<br>significant<br>(tears, open)<br>damage to<br>sterile<br>pouches.<br>• No<br>functional<br>damage to the<br>device. | Following<br>sterile unit's<br>package<br>integrity tests,<br>all acceptance<br>criteria were<br>met and it was<br>concluded that<br>sterility will be<br>maintained<br>along the<br>product's shelf<br>life. |
| No.                 | Test Name              | Purpose                                                                                                                                      | Method                                                                                                                                                                                                                                                                                                                                         | Acceptance Criteria                                                                                                                                         | Results                                                                                                                                                                                                                                                              | Conclusion                                                                                                                                                                                                    |
| Packaging Integrity |                        |                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                             |                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                               |
| 12.1                | Packaging<br>Integrity | • Verify<br>sterile single<br>unit package<br>integrity in<br>order to<br>maintain<br>product's<br>sterility along<br>3 years shelf<br>life. | Exposing the packaging<br>unit (single and shipping<br>box) to all treatments<br>undergo or might undergo<br>in real-life (sterilization<br>and drop). Products<br>undergo real-time aging<br>for 3 years under room<br>temp: 22 $\pm$ 1°C for three<br>years and then tested for:<br>visual defects, sterility,<br>peel, dye and burst tests. | • Sterile package<br>integrity tests per each<br>test acceptance criteria<br>(as further detailed in<br>section 18 and<br>appendix C of this<br>submission. | All tested<br>items met<br>acceptance<br>criteria for<br>visual test,<br>sterility test,<br>and package<br>integrity tests<br>(peel, dye,<br>burst) as<br>further<br>detailed in<br>section 18 and<br>appendix C of<br>this<br>submission.                           | Shelf life of 3<br>years real-time<br>was<br>established for<br>the sterile unit<br>package.                                                                                                                  |

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters 'mfs' in a stylized font, with a teardrop shape replacing the dot on the 'i'. The text 'Medical Flow Systems Ltd.' is written in a smaller font below the letters.

Image /page/10/Picture/2 description: The image shows a circular seal with the words "Quality Systems Certified" around the edge. Inside the circle, there are two lines of text: "ISO 9001:2008" and "ISO 13485:2003". The seal indicates that the organization has been certified to meet the requirements of these two ISO standards, which relate to quality management systems.

| No. | Test Name                 | Purpose                                                                                                                                                                                                                                   | Method                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Acceptance<br>Criteria                                                                                                                                              | Results                                                                                                                                        | Conclusion                                                                                                                                                                                                                                          |
|-----|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 13  | Usability<br>(Physicians) | Validate that<br>users of each<br>group<br>(physicians<br>and home<br>users) are<br>capable to<br>properly,<br>safely and<br>effectively<br>set and<br>operate the<br>device and<br>that the<br>device is<br>intuitive for<br>activating. | Provide dummy regulating set with<br>the device and filled reservoir<br>(saline) to a person who "simulates<br>a physician" (educated participant)<br>accompany by physician IFU.<br>Train this person on the device and<br>then ask him to perform the<br>following activating missions per<br>IFU:<br>1. Assemble the regulating set to<br>the reservoir (already filled with<br>saline).<br>2. Prime the system.<br>3. Set and secure of the active<br>volume.<br>4. Activate the device.<br>5. Re-set the device volume.<br>6. Permanently lock the active<br>volume. | All<br>'physicians'<br>participates<br>shall operate<br>the device<br>through the<br>"activation<br>missions"<br>successfully<br>with according<br>to the IFU.      | • All<br>participants<br>have fulfilled<br>successfully<br>all "activation<br>missions"<br>according to<br>the relevant<br>accompanied<br>IFU. | It was<br>concluded<br>from the<br>usability<br>results that the<br>device is<br>"usable" by its<br>intended users<br>(both<br>physicians and<br>patients) with<br>accordance to<br>IFU. The<br>device met all<br>test's<br>acceptance<br>criteria. |
|     | Usability<br>(Patients)   |                                                                                                                                                                                                                                           | Provide dummy devices connected<br>to an elastomeric pump and filled<br>with Saline.<br>Provide general information on the<br>device and review the IFU for<br>patients. Then, ask the patient to<br>activate the device.                                                                                                                                                                                                                                                                                                                                                 | All 'patients'<br>participants<br>shall<br>appropriately<br>activate the<br>device, i.e.,<br>squeeze the<br>activating<br>lever all the<br>way until it<br>'clicks' |                                                                                                                                                |                                                                                                                                                                                                                                                     |

Tests results are supporting all labeling claims in order to establish substantial equivalency.

#### 8. CONCLUSIONS

The evaluation of our device performances demonstrates that it is as safe and as effective as the predicate devices.

{11}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/11/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract representation of a human figure or a caduceus, with three curved lines that could be interpreted as arms or wings.

Public Health Service

Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-God Silver Spring, MD 20993-0002

### August 14, 2013

MFS-Medical Flow Systems Limited Mr. Ofer Shay Managing Director ELT Building, Dora Industrial Zone, Shlomi Shlomi Israei 22832

Re: K131247

Trade/Device Name: MultiBolus IITM Disposable Pain Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MEB, MEA Dated: July 9, 2013 Received: July 12, 2013

Dear Mr. Shay:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the cnclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{12}------------------------------------------------

Page 2 - Mr. Shay

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

**Mary S. Runner -S**

Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows a logo for Medical Flow Systems Ltd. The logo consists of the letters 'mfs' in a stylized font, with a water droplet replacing the dot on the 'i'. The text 'Medical…

---

**Source:** [https://fda-staging.innolitics.com/device/K131247](https://fda-staging.innolitics.com/device/K131247)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).