The PadScan HD series Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder volume noninvasively. The PadScan HD series Bladder Scanner is intended to be used only by qualified medical professionals.
Device Story
Portable, battery-operated B-mode ultrasound device; projects ultrasound energy through lower abdomen to acquire multiple 2D plane images; calculates bladder volume noninvasively. Operated by qualified medical professionals in clinical settings. Features expert/easy operating modes; provides real-time imaging, volume measurement, and thermal printouts. Data transfer via USB flash disk. Benefits include rapid, non-invasive assessment of bladder volume, aiding clinical decision-making regarding urinary retention or post-void residual assessment.
Clinical Evidence
Bench testing only. Performance evaluated via acoustic output measurement per NEMA UD 2:2004, electrical/mechanical safety per IEC 60601-1:2005, EMC per IEC 60601-1-2:2007, and biocompatibility per ISO 10993-5:2009 and ISO 10993-10:2010. Bladder volume measurement accuracy is ±15%, ±15ml.
Technological Characteristics
Mechanical sector probe; 2.5 MHz resonant frequency; 120-degree sector angle; 12 scan planes. Portable form factor with 7" or 8" TFT-LCD display. Power: Li-ion battery or AC/DC adapter. Class II, Type B equipment. Biocompatible plastic/PE patient-contact materials. Connectivity: USB flash disk for data upload. Software performs real-time pre-scan imaging and volume calculation.
Indications for Use
Indicated for noninvasive bladder volume calculation in patients via B-mode pulsed-echo ultrasound. Contraindicated for fetal use, pregnant patients, patients with ascites, open/damaged skin, or suprapubic wounds. Intended for use by qualified medical professionals.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
page 1 of 6
# 510(K) SUMMARY
JUL 1 2 2013
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 8807.92.
The assigned 510(K) number is: K131227
#### 1. Submitter:
Caresono Technology. Co., Ltd. 4th Floor, No.11 Building Initiating Zone, Instruments and Meters Industry Base, Near Port Industry Zone, Dandong, Liaoning 118009, P.R. China
#### 2. Contact Person:
Mr. Yang Long Chief Operating Officer ﺴﻴﺮ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍ Shenzhen Hlongmed Biotech Co., Ltd. R1508, East Building, Yihai Plaza, Chuangye Road, Nanshan District, Shenzhen, P.R. China Tel: 0086-755-86664986 Fax: 0086-755-86664933 E-mail: yanglong@hlongmed.com
#### 3. Date Prepared: June 2, 2013
#### 4. Device Information
Device Name: PadScan HD series Bladder Scanner Models: PadScan HD 5, PadScan HD 3 Common Name: Diagnostic Ultrasound System with Accessories Regulatory Class: II Classification Name and Product Code: 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (90-IYO) 21 CFR 892.1570 Diagnostic Ultrasonic Transducer (90-ITX)
#### 5. Predicate Device
Manufacturer: Mcube Technology Co., Ltd. Device: CUBEscan BioCon-700 510(k) Number: K111021 (Decision Date-April 27, 2011)
{1}------------------------------------------------
### 6. Intended Use
The PadScan HD series Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder volume noninvasively. The PadScan HD series Bladder Scanner is intended to be used only by qualified medical professionals.
### 7. Device Description:
The PadScan HD series Bladder Scanner manufactured by Caresono Technology Co., Ltd. provides real-time ultrasound imaging and measuring, and also provides non-invasive volume measurement of the bladder. During image scanning, multiple 2D plane ultrasonic images are acquired in several seconds.
It features:
- Expert operating mode and Easy operating mode. .
- Correct, reliable, fast, and simple operation. ●
- Printouts of ultrasound images with many parameters. ●
- . Portable
- . Combined power supply with AC adapter and a built-in battery.
#### 8. Indications for Use
The PadScan HD series Bladder Scanner is B-mode pulsed-echo ultrasound device. It is intended as a portable battery-operated device. The PadScan HD series Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder volume noninvasively. The PadScan HD series Bladder Scanner is intended to be used only by qualified medical professionals.
#### 9. Contraindications
Do not use the PadScan HD series Bladder Scanner on following cases:
a) Fetal use or pregnant patients
- b) Patients with ascites
- c) Patients with open or damaged skin
- d) Wounds in the suprapubic region
#### 10. Compatible with predicate device:
The PadScan HD series Bladder Scanner is substantially equivalent in intended use and operation to following predicate devices:
CUBEscan Diagnostic Ultrasound System and Accessories K111021
The ultrasound power transmitted from the device is not user adjustable,
{2}------------------------------------------------
and PadScan HD series Bladder Scanner is Track 1 System and meets the FDA's pre-amendment acoustic output limits. So as the predicate devices are. Although there are some differences such as resonant frequency, power source and display screen, there is no significant differences in technological characteristics that affecting the safety and efficiently. These are evaluated by safety test and acoustic output test.
| Item | Proposed Device | Predicate Device |
|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | PadScan HD series Bladder<br>Scanner | CUBEscan BioCon-700 |
| Model | PadScan HD 5, PadScan<br>HD 3 | BioCon-700 |
| 510(K) Submitter | Caresono Technology Co.,<br>Ltd. | Mcube Technology Co.,<br>Ltd. |
| 510(K) Number | -- | K111021 |
| Classifications<br>Name & Citations | 21 CFR 892.1560 Ultrasonic<br>Pulsed Echo Imaging<br>System (90-IYO)<br>21 CFR 892.1570<br>Diagnostic Ultrasonic<br>Transducer (90-ITX) | 21 CFR 892.1560<br>Ultrasonic Pulsed Echo<br>Imaging System (90-IYO)<br>21 CFR 892.1570<br>Diagnostic Ultrasonic<br>Transducer (90-ITX) |
| Indications for<br>Use | The PadScan HD series<br>Bladder Scanner is B-mode<br>pulsed-echo ultrasound<br>device. It intended as a<br>portable battery-operated<br>device. The PadScan HD<br>series Bladder Scanner<br>projects ultrasound energy<br>through the lower abdomen<br>of the patient to obtain<br>images of the bladder which<br>is used to calculate bladder<br>volume noninvasively. The<br>PadScan HD series Bladder<br>Scanner is intended to be<br>used only by qualified<br>medical professionals. | The BioCon-700 is a<br>B-mode pulsed-echo<br>ultrasound device. The<br>BioCon-700 is intended as<br>a portable battery-operated<br>device. The BioCon-700<br>projects ultrasonic energy<br>through the abdomen of the<br>patient obtaining images of<br>the bladder in order to<br>calculate the urine volume<br>non-invasively BioCon-700<br>is intended to be used by a<br>qualified medical<br>professional.<br>Contraindications for the<br>BioCon-700 are fetal use<br>and use on pregnant<br>patients. |
| Item | Proposed Device | Predicate Device |
| Contraindications | Do not use the PadScan HD series Bladder Scanner on following cases:<br>a) Fetal use or pregnant patients.<br>b) Patients with ascites.<br>c) Patients with open or damaged skin.<br>d) Wounds in the suprapubic region | Do not use the BioCon-700 on following cases:<br>a) Fetal use or pregnant patients.<br>b) Patients with ascites.<br>c) Patients with open or damaged skin.<br>d) Wounds in the suprapubic region. |
| Modes of operation | B mode | B mode |
| System Characteristics | - Portable<br>- LCD Display<br>- Thermal Printer<br>- Power source: Battery or AD-DC adapter | - Portable<br>- LCD Display<br>- Thermal Printer<br>- Power source: Battery or AD-DC adapter |
| Display | PadScan HD5: 8" TFT-LCD<br>PadScan HD3: 7" TFT-LCD | 7" TFT LCD |
| Controls for Change of acoustic output during scan | No | No |
| Transducer Type | Mechanical Sector Probe | Mechanical Sector Probe |
| Measurement localization | Abdomen | Abdomen |
| Transducer Resonant Frequency | 2.5 MHz | 2.6 MHz |
| Number of elements | 1 | 1 |
| Sector Angle | 120 degrees | 120 degrees |
| No. of Scan Planes | 12 | 12 |
| FDA Limits | Track 1 | Track 1 |
| Product Safety Certification | IEC 60601-1:2005 +<br>CORR.1(2006) +<br>CORR.2(2007) | UL 60601 -1, 1st Edition<br>CAN/CSA-C22.2 No. 601.1<br>-M90, 2005 |
| Item | Proposed Device | Predicate Device |
| | IEC 60601-2-37:2007 | EN 60601-2-37 |
| EMC Compliance | IEC 60601-1-2 | EN 60601-1-2 |
| Patient<br>Contacting<br>Material | Plastic, PE (Skin Contact) | Plastic, PC (Skin Contact) |
| Range | - Bladder volume range:<br>0-999ml<br>- Accuracy: ±15%, ±15ml | - Bladder volume range:<br>0-999ml<br>- Accuracy: ±15%, ±15ml<br>(0-999ml) |
| Classification of<br>protection against<br>electric shock | - Class II equipment<br>- Type B equipment | - Class II equipment<br>- Type BF equipment |
| Real-time<br>scanning | Yes (Pre-scan) | Yes (Pre-scan) |
| PC S/W function | Data Review<br>Data Printing | Data Review<br>Data Printing |
| PC Data Upload | Using USB flash disk | Using SD card |
| Power | AC/DC Adapter:<br>Input: AC100-240V,<br>50/60Hz,<br>Output: DC14V±0.5V<br>Battery: Li-ion rechargeable | AC/DC Adapter:<br>Input: AC 100-240V,<br>50-60Hz<br>Output: DC 9Vdc, 3A<br>Battery: Li-ion<br>rechargeable |
{3}------------------------------------------------
510(k) Summary
:
4
{4}------------------------------------------------
Caresono Technology Co., Ltd. believes that the PadScan HD series Bladder Scanner is substantially equivalent to the CUBEscan BioCon-700 of Mcube Technology Co., Ltd.
### 11.Safety, EMC and Performance Data:
#### Safety:
Electrical, mechanical, environmental safety and performance data demonstrates that the device is in compliance with IEC 60601-1:2005 and IEC 60601-2-37:2007.
### EMC:
Electromagnetic Compatibility data demonstrates that the device is in compliance with IEC 60601-1-2:2007. Performance:
{5}------------------------------------------------
K131227
Page 6 of 6
The PadScan HD series Bladder Scanner have been tested as Track 1 device per the FDA Guidance document "Information for Manufactures Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" issued in September 2008. The acoustic output is measured and calculated per NEMA UD 2:2004 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment.
#### Biocompatibility:
The Biocompatibility testing conducted in according with standard Biocompatibility ISO 10993-5:2009 and ISO 10993-10:2010.
#### 12. Conclusion:
The PadScan HD series Bladder Scanner was evaluated with safety (IEC 60601-1:2005 and IEC 60601-2-37:2007), EMC (IEC 60601-1-2:2007), Biocompatibility (ISO 10993-1:2010, ISO 10993-5:2009, ાડવ 10993-10:2010) and Acoustic Output (NEMA UD2:2004).
The conclusions drawn from testing of the PadScan HD series Bladder Scanner demonstrate that the device is as safe and effective as the legally marketed predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 12, 2013
Caresono Technology Co., Ltd. % Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories, Inc. 333 Pfingsten Road NORTHBROOK IL 60062
Re: K131227
Trade/Device Name: PadScan HD series Bladder Scanner, Models: PadScan HD S and PadScan HD 3 Regulation Number: 21 CFR 892.1560 . Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: June 17, 2013 Received: June 21, 2013
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{7}------------------------------------------------
Page 2 - Mr. Devine
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDK regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safets/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hun.
Sincerely yours,
Sm-h-p)
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
for
Enclosure
{8}------------------------------------------------
### Indications for Use
510(k) Number (if known): K131227
Device Name: PadScan HD series Bladder Scanner
Indications for Use:
The PadScan HD series Bladder Scanner is B-mode pulsed-echo ultrasound device. It is intended as a portable battery-operated device. The PadScan HD series Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder volume noninvasively. The Pad Scan HD series Bladder Scanner is intended to be used only by qualified medical professionals.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k) K131227
Page 1 of _ 3
{9}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
System: PadScan HD 3 Bladder Scanner
- Transducer: N2
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|--------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General | Specific | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| (Track 1 Only) | (Track 1 & 3) | | | | | | | |
| Ophthalmic | Ophthaimic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non- | | | | | | | |
| | Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other(Bladder) | N | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N=new indication; P= previously cleared by FDA; E=added under this appendix
Page 2 of ____________________________________________________________________________________________________________________________________________________________________
.
.
Indications for Use
·
{10}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
System: PadScan HD 5 Bladder Scanner
Transducer: N2
Intended Use: Diagnostic ultrasound imaging or fluid flow anaiysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthaimic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify). | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.(non-<br>Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other(Bladder) | N | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| Cardiac | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N=new indication; P= previously cleared by FDA; E=added under this appendix

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k) K131227
Page 3 of 3
Indications for Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.