K131213 · Sonoscape Company Limited · IYN · Jun 21, 2013 · Radiology
Device Facts
Record ID
K131213
Device Name
DIGITAL COLOR DOPPLER ULTRASOUND SYSTEM
Applicant
Sonoscape Company Limited
Product Code
IYN · Radiology
Decision Date
Jun 21, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SonoScape S40 device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.
Device Story
The SonoScape S40 is a general-purpose, software-controlled digital color Doppler ultrasound system. It acquires ultrasound data via various transducers (curved, micro-curved, linear, and phased arrays) and displays images in B-Mode, M-Mode, TDI, Color-Flow Doppler, Pulsed Doppler, Power Doppler, and 3D/4D modes. The system utilizes a 128-channel digital beamformer with progressive dynamic receive focusing to enhance image quality. Operated by qualified physicians in clinical settings, the device features a touch panel interface, cine playback, and image archiving. The system processes inputs using autocorrelation for color and FFT for Doppler analysis. Output images assist clinicians in diagnostic evaluations across multiple specialties, including OB/GYN, cardiology, and urology, potentially improving diagnostic confidence through high-resolution imaging and advanced measurement tools.
Clinical Evidence
No clinical testing was required. Substantial equivalence was established through bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), performance (IEC 60601-2-37), and biocompatibility (ISO 10993-5/10).
Technological Characteristics
Digital ultrasound system with 128-channel beamformer. Supports linear, curved, and phased array transducers (2-15 MHz). Features 18.5-inch LCD, touch panel, and 4 probe ports. Connectivity via USB for software updates. Biocompatible materials per ISO 10993. Operates on embedded Linux. Acoustic output compliant with NEMA UD 2/UD 3.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal/adult cephalic, trans-rectal, trans-vaginal, peripheral vascular, conventional/superficial musculo-skeletal, and neonatal/adult cardiac applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SonoScape Diagnostic Ultrasound System, Model S6 (K112602)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Submission
# 510(k) Summary
## JUN 2 1 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number: K131213
- 1. Date of submission: April 26, 2013
- 2. Submitter
SonoScape Company Limited Address: Yizhe Building, Yuquan Road,Nanshan, Shenzhen 518051, P.R.China Tel: (86) 755-26722890 Fax: (86) 755-26722850 Contact Person: Toki Wu E-mail:Faith@sonoscape.net/wusq@sonoscape.net
#### 3. Proposed Device Identification
Trade/Proprietary Name: S40 Digital Color Doppler Ultrasound System Common Name: Diagnostic Ultrasound System and Transducers Classification:
21 FR 892.1550 Ultrasonic Pulsed Doppler Imaging System (90-IYN)
21 FR 892.1560 Ultrasonic Pulsed Echo Imaging System (90-IYO)
21 CFR 892.1570 Diagnostic Ultrasound Transducer (90-ITX)
Classification Panel: Radiology
Device Class: Il
#### 4. Legally Marketed Predicate Device
SonoScape Company Limited, Diagnostic Ultrasound System, Model S6 has been cleared by FDA through 510(k) No.K112602 (Decision Date - November 07, 2011).
{1}------------------------------------------------
#### 5. Device Description
The SonoScape S40 Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.
The all digital architecture with progressive dynamic receive focusing allows the system to maximize the utility of all imaging transducers to enhance the diagnostic utility and confidence provided by the system. The exam dependent default setting allows the user to have minimum adjustment for imaging the patient, while the in-depth soft-menu control allows the advanced user to set the system for different situations. The architecture allows cost-effective system integration to a variety of upgrade-able options and features.
This SonoScape system is a general purpose, software controlled, diagnostic ultrasound system. Its basic function is to acquire ultrasound data and display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Doppler and Power Doppler, or a combination of these modes, 3D/4D.
#### 6. Intended Use Statement
The SonoScape S40 device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.
Transducer Information is shown in the table 1.
| No. | Probe | Type | Frequency<br>Range | Intended Use |
|-----|-------|-----------------------|--------------------|-------------------------------|
| 1 | C344 | curved Array | 2.0-5.0 MHz | Fetal / Abdominal/ Ob/GYN |
| 2 | C353 | curved Array | 2.0-6.0 MHz | Fetal / Abdominal/ Ob/GYN |
| 3 | C322 | curved Array | 2.0-6.0 MHz | Fetal / Abdominal/ Ob/GYN |
| 4 | VC6-2 | curved Array | 2.0-6.0 MHz | Fetal / Abdominal/ Ob/GYN |
| 5 | 6V1 | Micro-curved<br>Array | 4.0-8.0 MHz | Trans-rectal<br>Trans-vaginal |
| 6 | 6V3 | Micro-curved | 5.0-9.0 MHz | Trans-rectal |
Table 1 Transducer Information
510(k) Summary
{2}------------------------------------------------
510(k) Submission
| No. | Probe | Type | Frequency<br>Range | Intended Use |
|-----|-------|--------------|--------------------|---------------------------------------------------------------------------------------------------------------------------------|
| | | Array | | Trans-vaginal |
| 7 | L741 | Linear Array | 5.0-10.0 MHz | Small Organ (breast, thyroid, testes)<br>Musculo-skeletal (Conventional)<br>Peripheral vessel |
| 8 | L742 | Linear Array | 5.0-12.0 MHz | Small Organ (breast, thyroid, testes)<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Peripheral vessel |
| 9 | L752 | Linear Array | 5.0-12.0 MHz | Small Organ (breast, thyroid, testes)<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Peripheral vessel |
| 10 | 2P2 | Phase Array | 1.0-5.0 MHz | Abdominal<br>Cephalic(neonatal and adult)<br>Cardiac (neonatal and adult) |
| 11 | 3P1 | Phase Array | 1.0-5.0 MHz | Abdominal<br>Cephalic(neonatal and adult)<br>Cardiac (neonatal and adult) |
| 12 | 5P2 | Phase Array | 3.0-8.0 MHz | Pediatric<br>Neonatal Cephalic<br>Cardiac Pediatric |
| 13 | 8P1 | Phase Array | 4.0-12.0 MHz | Pediatric<br>Neonatal Cephalic<br>Cardiac Pediatric |
#### 7. Testing
Laboratory testing was conducted to verify that the S40 system with added transducer met all design specification and was substantially equivalent to the Predicate Device. The device has been found to conform to applicable medical device safety standards in regards to thermal, mechanical and electrical safety as well as biocompatibility. The acoustic output is measured and calculated per "NEMA UID 2: 2004 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment".
#### 510(k) Summary
{3}------------------------------------------------
| Standards<br>No. | Standards Title | Version | Date |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|------------------------|
| IEC 60601-1 | Medical Electrical Equipment - Part1.<br>General Requirements for Safety | 1988+A1:<br>1991+A2:<br>1995 | 10/31/2005 |
| IEC<br>60601-1-2 | Medical Electrical Equipment, Part 1-2:<br>General Requirements for Safety -<br>Collateral Standard: Electromagnetic<br>Compatibility - Requirements and Tests | 2007 | 03/01/2007 |
| IEC<br>60601-2-37 | Medical Electrical Equipment, Part 2-37:<br>Particular Requirements for the Safety of<br>Ultrasonic Medical Diagnostic and<br>Monitoring Equipment | 2007 | 08/01/2007 |
| NEMA UD 2 | Acoustic Output Measurement Standard for<br>Diagnostic Ultrasound Equipment Version 3 | 2004 | 01/01/2004<br>(R 2009) |
| NEMA UD3 | Standard for Real-Time Display of Thermal<br>and Mechanical Acoustic Output Indices on<br>Diagnostic Ultrasound Equipment | 2004 | 01/01/2004<br>(R 2009) |
| ISO 10993-5 | Biological evaluation of medical devices -<br>Part 5: Tests for In Vitro cytotoxicity | 1999 | 05/15/1999 |
| ISO<br>10993-10 | Biological evaluation of medical devices -<br>Part 10: Tests for irritation and delayed-type<br>hypersensitivity | 2002 | 09/01/2002 |
### Tab 2 Applicable Safety Standards
#### 8. Clinical Test:
No clinical testing was required.
#### 9. Comparison Table
The differences between the S40 and the predicate device S6 in almost every part are listed in the tables below.
{4}------------------------------------------------
| ID | Items | Proposed Device<br>SonoScape S40 | Predicate Device<br>SonoScape S6 | Remark |
|----|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 1 | Intended Use | The device is a<br>general-purpose ultrasonic<br>imaging instrument<br>intended for use by a<br>qualified physician for<br>evaluation of Fetal,<br>Abdominal, Pediatric, Small<br>Organ (breast, testes,<br>thyroid), Cephalic (neonatal<br>and adult), Trans-rectal,<br>Trans-vaginal, Peripheral<br>Vascular, Musculo-skeletal<br>(Conventional and<br>Superficial), Cardiac<br>(neonatal and adult),<br>Urology and OB/Gyn. | The device is a<br>general-purpose ultrasonic<br>imaging instrument<br>intended for use by a<br>qualified physician for<br>evaluation of Fetal,<br>Abdominal, Pediatric, Small<br>Organ (breast, testes,<br>thyroid), Cephalic(neonatal<br>and adult), Trans-rectal,<br>Trans-vaginal, Peripheral<br>Vascular, Musculo-skeletal<br>(Conventional and<br>Superficial), Cardiac<br>(neonatal and adult),<br>OB/Gyn and Urology. | Same |
### Table 3 Intended Use Comparison
·
### Table 4 General Comparison
| ID | Comparison Items | Proposed Device<br>SonoScape S40 | Predicate Device<br>SonoScape S6 | Remark |
|----|----------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|---------------------|
| 2 | Classification Name | Ultrasonic Pulsed Doppler<br>Imaging System<br>Ultrasonic Pulsed Echo<br>Imaging System<br>Diagnostic Ultrasound<br>Transducer | Ultrasonic Pulsed Doppler<br>Imaging System<br>Ultrasonic Pulsed Echo<br>Imaging System<br>Diagnostic Ultrasound<br>Transducer | Same |
| 3 | Product Code | 90-IYN/90-IYO/90-ITX | 90-IYN/90-IYO/90-ITX | Same |
| 4 | Regulation Number | 892.1550/892.1560/892.1570 | 892.1550/892.1560/892.1570 | Same |
| 5 | Panel | Radiology | Radiology | Same |
| 6 | Class | II | II | Same |
| 7 | Probe Type &<br>Connectors | L741 Linear Array, 5.0-10.0 MHz<br>L742 Linear Array, 5.0-12.0 MHz<br>L752 Linear Array, 5.0-12.0 MHz | L741 Linear Array, 5.0-10.0 MHz<br>L742 Linear Array, 5.0-12.0 MHz<br>L743 Linear Array, 5.0-10.0 MHz | SE<br>Analysis<br>1 |
510(k) Summary
8-5
{5}------------------------------------------------
510(k) Submission
| ID | Comparison<br>Items | Proposed Device<br>SonoScape S40 | Predicate Device<br>SonoScape S6 | Remark |
|----|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| | | C344 Curved Array, 2.0-5.0<br>MHz<br>C322 Curved Array, 2.0-6.0<br>MHz<br>C353 Curved Array, 2.0-6.0<br>MHz | C344 Curved Array, 2.0-5.0<br>MHz<br>C362 Curved Array, 2.0-6.0<br>MHz | |
| | | | C611 Micro-curved Array,<br>4.0-8.0 MHz | |
| | | VC6-2 Curved Array, 2.0-6.0<br>MHz<br>6V1 Micro-curved Array,<br>4.0-8.0 MHz<br>6V3 Micro-curved Array,<br>5.0-9.0 MHz | VC6-2 Curved Array, 2.0-6.0<br>MHz<br>6V1 Micro-curved Array,<br>4.0-8.0 MHz<br>6V3 Micro-curved Array,<br>5.0-9.0 MHz<br>EC9-5 Micro-curved Array,<br>5.0-9.0 MHz | |
| | | 2P2 Phased Array, 1.0-5.0<br>MHz<br>3P1 Phased Array, 1.0-5.0<br>MHz<br>5P2 Phased Array, 3.0-8.0<br>MHz<br>8P1 Phased Array, 4.0-12.0<br>MHz<br>Multi-port connector connects | 2P1 Phased Array, 2.0-4.0<br>MHz<br>5P1 Phased Array, 4.0-7.0<br>MHz<br>Multi-port connector connects | |
| | | 4 transducers | 2 transducers | |
| 8 | Acoustic<br>Track | TRACK 3 | TRACK 3 | Same |
### Table 5 Functions Comparison
| ID | Comparison<br>Items | Proposed Device<br>SonoScape S40 | Predicate Device<br>SonoScape S6 | Remark |
|----|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| 9 | | Based on an embedded Linux<br>operating system. | Based on an embedded Linux<br>operating system. | Same |
| | Design | Based on a 128 channel full<br>digital beam former. | Based on a 64 channel full<br>digital beam former. | SE<br>Analysis<br>2 |
| | | Autocorrelation for color<br>processing and FFT for pulse | Autocorrelation for color<br>processing and FFT for pulse | Same |
| ID | Comparison<br>Items | Proposed Device<br>SonoScape S40<br>and CW Doppler processing. | Predicate Device<br>SonoScape S6<br>and CW Doppler processing. | Remark |
| | | Supporting Linear, Curve linear<br>and Phase array probes from 2<br>to 15 MHz. | Supporting Linear, Curve linear<br>and Phase array probes from 2<br>to 15 MHz. | Same |
| | | Cine play back capability | Cine play back capability | Same |
| | | Image file archive | Image file archive | Same |
| | | Software upgrade with USB<br>flash drive. | Software upgrade with USB<br>flash drive. | Same |
| | | Digital Scan Converter<br>800×600 | Digital Scan Converter<br>800×600 | Same |
| | | With touch panel | With full keyboard | SE<br>Analysis<br>3 |
| | | TGC 8 slider | TGC 8 slider | Same |
| | | Depth Range: 3 to 32 cm | Depth Range: 3 to 32 cm | Same |
| | | Image sector size: 32 lines to<br>full B (512 lines) | Image sector size: 32 lines to<br>full B (256 lines) | SE<br>Analysis<br>4 |
| | | Image Sector position: Steering<br>within full maximum | Image Sector position: Steering<br>within full maximum | Same |
| | | B orientation flip :L/R key with<br>marking on the screen | B orientation flip :L/R key with<br>marking on the screen | Same |
| | | B Dynamic range control:<br>preset 14 curves over 140 dB | B Dynamic range control:<br>preset 14 curves over 140 dB | Same |
| | | Gray Scale Control: 7 Settings | Gray Scale Control: 7 Settings | Same |
| | | Focal Number: 12 focal zone<br>setting | Focal Number: 12 focal zone<br>setting | Same |
| 10 | Operation<br>Controls | B persistence: 0-95% | B persistence: 0-95% | Same |
| | | Image Processing: Smoothing,<br>edge enhancement | Image Processing: Smoothing,<br>edge enhancement | Same |
| | | PW sweeping speed 2,4,6,8<br>sec over display | PW sweeping speed 2,4,6,8<br>sec over display | Same |
| | | PW Wall filter setting:16<br>settings,25 to 750 HZ | PW Wall filter setting:16<br>settings,25 to 750 HZ | Same |
| | | PW sample volume:0.5 to<br>20mm | PW sample volume:0.5 to<br>20mm | Same |
| | | PW/B update: with UPDATE<br>key | PW/B update: with UPDATE<br>key | Same |
| | | PW cursor steering: Steer soft<br>key | PW cursor steering: Steer soft<br>key | Same |
| | | PW angle correction:0 to 80<br>degree user control | PW angle correction:0 to 80<br>degree user control | Same |
| ID | Comparison Items | Proposed Device SonoScape S40 | Predicate Device SonoScape S6 | Remark |
| | | PW spectrum dynamic range:10 preset curve over 15-48 dB | PW spectrum dynamic range:10 preset curve over 15-48 dB | Same |
| | | Spectrum baseline shift and invert | Spectrum baseline shift and invert | Same |
| | | Color ROI setting: trackball and set key to control size and position | Color ROI setting: trackball and set key to control size and position | Same |
| | | Color steering on flat probe: $\pm$ 20, $\pm$ 16,0 | Color steering on flat probe: $\pm$ 20, $\pm$ 16,0 | Same |
| | | Color Wall Filter: Color wall filter with 16 selection, 25-750 of PRF | Color Wall Filter: Color wall filter with 16 selection, 25-750 of PRF | Same |
| | | Color priority-B priority soft menu | Color priority-B priority soft menu | Same |
| | | Color Packet size: preset per Exam, horizontal, vertical, off | Color Packet size: preset per Exam, horizontal, vertical, off | Same |
| | | Zoom adjustable | Zoom adjustable | Same |
| | | Freeze control: Toggling freeze key | Freeze control: Toggling freeze key | Same |
| | | Cine control: step, play backward, play continuously | Cine control: step, play backward, play continuously | Same |
| 11 | Operation Mode | B, M, PW, CW, CFM, DPI, TDI, Tissue Harmonic Image 3D/4D Mode Color M Mode | B, M, PW, CW, CFM, DPI, TDI, Tissue Harmonic Image 3D/4D Mode Color M Mode | Same |
| 12 | Display Modes | Dual B, Quad Display, B and M, B and Doppler B + Color, Dual B(Flow) Triplex mode: B,CFM, and PW/CW; B,DPI, and PW/CW;B,THI and Color M, steer M Dual B and Color in real time Compound Imaging, Panoramic Imaging, Trapezoid Imaging. | Dual B, Quad Display, B and M, B and Doppler B + Color, Dual B(Flow) Triplex mode: B,CFM, and PW/CW; B,DPI, and PW/CW;B,THI and Color M, steer M Dual B and Color in real time Compound Imaging, Panoramic Imaging, Trapezoid Imaging. | Same |
| 13 | Measurement Items | Distance; area; circumference; calipers; volume, velocity, HR, PI, RI. Cardiac. OB/GYN, Urology, Vascular and small | Distance; area; circumference; calipers; volume, velocity, HR, PI, RI. Cardiac. OB/GYN, Urology, Vascular and small | Same |
| ID | Comparison<br>Items | Proposed Device<br>SonoScape S40 | Predicate Device<br>SonoScape S6 | Remark |
| | | part package | part package | |
| 14 | Cine Loop | Automatic review/ manual<br>review | Automatic review/ manual<br>review | Same |
| | | Review speed can be adjusted | Review speed can be adjusted | Same |
510(k) Summary
8-6
{6}------------------------------------------------
:
:
·
·
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510(k) Submission
510(k) Summary
.
8-7
.
{7}------------------------------------------------
510(k) Submission .
510(k) Summary
:
.
8-8
.
{8}------------------------------------------------
#### 510(k) Submission
### Table 6 Specifications Comparison
.
| ID | Comparison<br>Items | Proposed Device<br>SonoScape S40 | Predicate Device<br>SonoScape S6 | Remark | | | | |
|----|-------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|---------------------|------------------|--------------------------------|----------------|------|
| 15 | Power<br>Supply | Voltage: 110-127/220-240 VAC<br>Frequency: 50/60 Hz<br>Power Consumption: 450VA | Voltage: 110-240 VAC<br>Frequency: 50/60 Hz<br>Power Consumption: 110-240V AC, 2.7-1.2A | SE<br>Analysis<br>5 | | | | |
| 16 | Operating<br>Condition | Temperature: 10~40°C<br>Relative humidity: 30~75%<br>Air pressure: 700hPa ~1060hPa | Temperature: 10~40°C<br>Relative humidity: 30~75%<br>Air pressure: 700hPa ~1060hPa | Same | | | | |
| 17 | Storage<br>Condition | Temperature: -20~55°C<br>Relative humidity: 20~90%<br>Air pressure: 700hPa ~1060hPa | Temperature: -20~55°C<br>Relative humidity: 20~90%<br>Air pressure: 700hPa ~1060hPa | Same | | | | |
| 18 | Screen Size | 18.5 inch Widescreen LCD<br>monitor | 15 inch Widescreen LCD monitor | SE<br>Analysis<br>6 | | | | |
| 19 | Measurement<br>Accuracy | Parameter | Value range | Error<br>range | Parameter | Value range | Error<br>range | |
| | | Display<br>depth | Max 32.9 cm; (Probe<br>depend) | ±3% | Display<br>depth | Max 32.9 cm; (Probe<br>depend) | ±3% | Same |
| | | Distance | 0~31.0 cm | ±3% | Distance | 0~31.0 cm | ±3% | Same |
| | | Area | Max. ≥855 cm² | ±7% | Area | Max. ≥855 cm² | ±7% | Same |
| | | Angle | 10~193° | ±3% | Angle | 10~193° | ±3% | Same |
| | | Circumference | 200 cm | ±3% | Circumference | 200 cm | ±3% | Same |
| | | Volume | Max. 25000 cm³ | ±10% | Volume | Max. 25000 cm³ | ±10% | Same |
| | | M-Mode<br>time | 2,4,6,8 S | ±1% | M-Mode<br>time | 2,4,6,8 S | ±1% | Same |
| | | Heart<br>Rate | 8 ~ 1000<br>beats/sec | ±3% | Heart<br>Rate | 8 ~ 1000<br>beats/sec | ±3% | Same |
| | Slope | 1300 cm/s | ±3% | Slope | 1300 cm/s | ±3% | Same | |
510(k) Summary
{9}------------------------------------------------
#### 510(k) Submission
| ID | Comparison Items | Proposed Device SonoScape S40 | | Predicate Device SonoScape S6 | | Remark |
|----|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|-------------------------------|-----------|---------------|
| | | Velocity | Angle | Velocity | Angle | |
| | Velocity (PW) | 0.04-2940 cm/s | ≤60°, ≤5% | 0.04-2940 cm/s | ≤60°, ≤5% | Same |
| | Velocity (CW) | 0.12-3795 cm/s | ≤60°, ≤5% | 0.13-3529 cm/s | ≤60°, ≤5% | SE Analysis 7 |
| | Velocity (Color) | 1-298 cm/s | ≤60°, ≤5% | 2-226cm/s | ≤60°, ≤5% | |
| 20 | Acoustic Output | Track 3:MI, TIS, TIC, TIB<br>Derated ispta: 720Mw/cm2 maximum.<br>TIS/TIB/TIC: 6.0 Maximum,<br>Mechanical Index: 1.9 Maximum,<br>or Derated Isppa: 190W/cm2 max | | | | |
| | | Same | | | | |
#### Table 7 Safety Comparison
| ID | Comparison Items | Proposed Device<br>SonoScape S40 | Predicate Device<br>SonoScape S6 | Remark |
|----|---------------------------------|-------------------------------------|-------------------------------------|--------|
| 21 | Electrical Safety | -IEC 60601-1 | -IEC 60601-1 | Same |
| 22 | EMC | -IEC 60601-1-2 | -IEC 60601-1-2 | Same |
| 23 | Performance | -IEC 60601-2-37 | -IEC 60601-2-37 | Same |
| 24 | Biocompatibility | -ISO 10993-5,<br>-ISO 10993-10 | -ISO 10993-5,<br>-ISO 10993-10 | Same |
| 25 | Level of Concern<br>Of Software | Moderate level of<br>concern system | Moderate level of<br>concern system | Same |
#### SE Analysis 1:
Probe Type & Connectors, Compare to the predicate device, the proposed device is with different probe type or frequency, such as L752, C353 etc. But no new intended use is added and all of them comply with IEC 60601-2-37; And the proposed device has 4 probe connection ports and the predicate device has 2 probe connection ports, but both of them comply with IEC 60601-1 and IEC 60601-1-2, therefore they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.
{10}------------------------------------------------
#### SE Analysis 2: ·
The proposed device is based on a 128 channel full digital beam former and the predicate device is 64, but the 128 channel is better than 64 channel in terms of the image quality. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.
#### SE Analysis 3:
The predicate device is with full keyboard and the proposed device is not, but both of them comply with IEC 60601-1 and IEC 60601-1-2. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.
#### SE Analysis 4:
The proposed device is with 512 lines image sector size and the predicate device is 256, but the 512 lines is better than 256 lines in terms of the image quality. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.
#### SE Analysis 5
The Power Supply of the proposed device and the predicate device are 110-127/220-240 VAC, 450VA and 110-240VAC, 2.7-1.2A respectively, but both of them comply with IEC60601-1 and IEC 60601-1-2. Therefore, power supply can be considered Substantially Equivalent in safety and effectiveness.
#### SE Analysis 6
The screen size of the proposed is larger than that of the S6. This difference is considered to have no effect on effectiveness and safety.
#### SE Analysis 7:
The proposed device and the predicate device are with different measurement accuracy in Velocity (CW/ Color), but the proposed device is better. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.
#### Discussion of Substantially Equivalent
The subject device has same intended use, similar product design, and same performance effectiveness, performance safety as the predicate device. The differences above between the subject device and predicate device do not affect the basic design
8-11
{11}------------------------------------------------
K131213
Page 12 of 12
510(k) Submission
principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.
#### 10. Substantially Equivalent Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, SonoScape Company Limited concludes that S40 Digital Color Doppler Ultrasound System is substantially equivalent to predicate devices with regard to safety and effectiveness.
510(k) Summary
{12}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three wing-like shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 21, 2013
SonoScape Company Limited % Ms. Toki Wu Yizhe Building, Yuquan Road, NanShan Shenzhen, Guangdong 518051 P.R. CHINA
Re: K131213
Trade/Device Name: S40 Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYN, IYO, and ITX Dated: April 26, 2013 Received: May 2, 2013
Dear Ms. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the S40 Digital Color Doppler Ultrasound System, as described in your premarket notification:
#### Transducer Model Number
2P2 Phase Array 8P1 Phase Array C344 Curved Array VC6-2 Curved Array 3PI Phase Array 6V1 Micro-curved Array C353 Curved Array L741 Linear Array L752 Linear Array
5P2 Phase Array 6V3 Micro-curved Array C322 Curved Array L742 Linear Array
{13}------------------------------------------------
### Page 2 - Ms. Toki Wu
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050;
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/About/DA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH) s) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shing Chun Benny Lam, Ph.D. at (301) 796-9328.
Sincerely yours,

for
Janine M. Morris Director, Division of Radiological Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosures
{14}------------------------------------------------
# Indications for Use
| 510(k) Number: | K131213 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | S40 Digital Color Doppler Ultrasound System |
| Indications for Use: | The SonoScape S40 device is a general-purpose ultrasonic<br>imaging instrument intended for use by a qualified<br>physician for evaluation of Fetal, Abdominal, Pediatric,<br>Small Organ (breast, testes, thyroid), Cephalic(neonatal<br>and adult), Trans-rectal, Trans-vaginal, Peripheral<br>Vascular, Musculo-skeletal (Conventional and Superficial),<br>Cardiac (neonatal and adult), OB/Gyn and Urology. |
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
\$m(x-h)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
K131213 510(k) _
Indications for Use
Page 1 of 15
{15}------------------------------------------------
System: SonoScape S40
Diagnostic Ultrasound Pulsed Echo System
Diagnostic Ultrasound Pulsed Doppler Imaging System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | | |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal | N | N | N | | N | N | Note 1 | Notes 2,4,5 | |
| | Abdominal | N | N | N | | N | N | Note 1 | Notes 2,4,5 | |
| | Intra-operative Specify | | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| | Small Organ (specify) | N | N | N | | N | N | Note 1 | Notes 2,4,6 | |
| | Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| Fetal | Adult Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| Imaging& | Trans-rectal | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| Other | Trans-vaginal | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph.(non-Card) | | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| | Musculo-skeletal<br>(Superficial) | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| | Intravascular | | | | | | | | | |
| | Other (Ob/GYN) | N | N | N | | N | N | Note 1 | Notes 2,4,5 | |
| | Cardiac Adult | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| | Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| Cardiac | Intravascular(Cardiac) | | | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
| Peripheral | Peripheral vessel | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| Vessel | Other (specify) | | | | | | | | | |
E = added under this appendix P = previously cleared by FDA; N = new indication; Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) _
Indications for Use
Page 2 of 15
{16}------------------------------------------------
Transducer: 2P2 Phase Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|------------------------------|------------------------------------|-------------------|---|-----|---------|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CW<br>D | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal<br>Imaging&<br>Other | Fetal | | | | | | | | |
| | Abdominal | N | N | N | | N | N | Note 1 | Notes 2,4 |
| | Intra-operative Specify | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Adult Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph.(non-Card) | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (Ob/GYN) | | | | | | | | |
| Cardiac | Cardiac Adult | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Intravascular(Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (specify) | | | | | | | | |
P = previously cleared by FDA; E = added under this appendix N = new indication; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) _
Indications for Use
Page 3 of 15
{17}------------------------------------------------
Transducer: 3P1 Phase Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | | |
|------------------------------|------------------------------------|---|-------------------|-----|-----|------------------|----------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude<br>) Doppler | Other*<br>Combined | Other*<br>Specify | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| Fetal<br>Imaging&<br>Other | Fetal | | | | | | | | | |
| | Abdominal | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| | Intra-operative Specify | | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | | | | | | | | | |
| | Small Organ (specify) | | | | | | | | | |
| | Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| | Adult Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph(non-Card) | | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| | Intravascular | | | | | | | | | |
| | Other (Ob/GYN) | | | | | | | | | |
| Cardiac | Cardiac Adult | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| | Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| | Intravascular(Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) __
Indications for Use
Page 4 of 15
{18}------------------------------------------------
#### Transducer: 5P2 Phase Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------------|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.