K131204 · Cook, Inc. · DQY · Jan 29, 2014 · Cardiovascular
Device Facts
Record ID
K131204
Device Name
ADVANCE CS CORONARY SINUS INFUSION CATHETER
Applicant
Cook, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Jan 29, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Advance® CS Coronary Sinus Infusion Catheter is intended for temporary occlusion of the coronary sinus for infusion of contrast media, drugs, or therapeutic agents; or possible introduction of devices into the coronary venous system.
Device Story
Advance® CS Coronary Sinus Infusion Catheter; over-the-wire balloon catheter; 5.0 Fr size; compatible with 0.035 inch wire guide. Used by clinicians for temporary occlusion of coronary sinus; enables infusion of contrast media, drugs, or therapeutic agents; facilitates introduction of other devices into coronary venous system. Sterile, single-use device. Performance verified via bench testing (burst strength, sheath compatibility, inflation/deflation times, fatigue, torque, kink resistance) and acute porcine in vivo evaluation. Output is mechanical occlusion of the vessel; allows physician to perform targeted venous delivery or access.
Clinical Evidence
Bench testing included balloon burst strength, sheath compatibility, profile measurement, compliance, flow rate characterization, leakage, fatigue, bond/material strength, inflation/deflation time, torque, kink/flexibility, and particulate testing. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility) performed per ISO 10993-1. Acute in vivo animal evaluation (porcine coronary sinus) characterized safety and biologic response 48 hours post-inflation.
Technological Characteristics
5.0 Fr over-the-wire balloon catheter; compatible with 0.035 inch wire guide; 5, 6, or 7 Fr sheath compatibility. Mechanical sensing/actuation via balloon inflation. Sterile, single-use. Biocompatible materials per ISO 10993-1.
Indications for Use
Indicated for temporary coronary sinus occlusion to facilitate infusion of contrast media, drugs, or therapeutic agents, or to assist in the introduction of devices into the coronary venous system.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on a black background. Below "COOK" is the word "MEDICAL" in smaller, white letters. The logo is simple and modern.
COOK INCORPORATED
750 DANIELS WAY, P.O, BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
612.339,2235 10 : 800.457.4500
WWW.COOKMEDICAL.COM
# JAN 2 9 2014
### 510(k) SUMMARY
#### Submitted By:
Jennifer Richardson, MA Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47402 Phone: (812) 335-3575 x102370 Fax: (812) 332-0281 Date Prepared: December 20, 2013
#### Device:
Advance® CS Coronary Sinus Infusion Catheter Trade Name: Common Name: Percutaneous Catheter Class II, Product Code: DQY, (21 CFR §870.1250) Classification:
# Indications for Use:
The Advance® CS Coronary Sinus Infusion Catheter is intended for temporary occlusion of the coronary sinus for infusion of contrast media, drugs, or therapeutic agents; or possible introduction of devices into the coronary venous system.
#### Device Description:
The Advance® CS Coronary Sinus Infusion Catheter is an over-the-wire catheter. The 5.0 Fr balloon catheter is compatible with a 0.035 inch diameter wire guide. The device will be supplied sterile and intended for one-time use.
### Comparison to Predicate Device:
The proposed device is substantially equivalent to the predicate in terms of intended use, duration of use, principles of operation, and technological characteristics.
# Discussion of Tests and Test Results:
The device was subjected to the following tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of:
- Balloon Minimum Burst Strength Testing shows the balloons will burst at or above the 1. minimum rated burst pressure, with all failure modes being linear tears. The predetermined acceptance criteria were met.
- Sheath Compatibility Qualitative and quantitative evaluations show that the balloons 2. are compatible with a 5, 6, or 7 Fr sheath. The predetermined acceptance criteria were met.
- 3. Balloon Profile Measurement of the diameter of catheter shaft, bonds, and folded balloon shows that the device profile is compatible with a 5, 6. or 7 Fr sheath (0.074. 0.087, or 0.100 inch profile). The predetermined acceptance criteria were met.
K131204 Page 1 of 2
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Cook Medical. The logo is black and white and features the word "COOK" in large, bold letters on the top line. Below that, the word "MEDICAL" is written in smaller letters. The logo is simple and professional.
COOK
INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
HONE: 812.339.2235 TOLL FREE: 800.457.4500
•WWW.COOKMEDICAL.COM
- 4. Balloon Compliance Testing shows that the balloon will inflate to the stated balloon size. The predetermined acceptance criteria were met.
- 5. Dynamic Burst of the Distal Lumen Testing characterized saline and contrast agent flow rate vs. pressure curves. Testing shows that all test articles withstood at least pressures in excess of 300 psig regardless of injection media type. The predetermined acceptance criteria were met.
- 6. Static Burst of the Distal Lumen Testing shows that similar test articles can be expected to fail at or above 100 psig. The predetermined acceptance criteria were met.
- 7. Freedom from Leakage Testing shows that the balloon catheter will withstand pressurization to 200 psi for a time period of 15 seconds without leaking or significantly increasing in outside diameter. The predetermined acceptance criteria were met.
- Balloon Fatigue Testing shows that balloons are free from leakage and damage on 8. inflation, withstanding 10 cycles of inflation. The predetermined acceptance criteria were met.
- 9. Bond and Materials Strength Testing shows the tensile force during normal use should not fracture or rupture the balloon catheter bonds or materials. The predetermined acceptance criteria were met.
- 10. Inflation / Deflation Time Testing shows that the balloon will inflate within 60 seconds and fully deflate within 60 seconds. The predetermined acceptance criteria were met.
- 11. Torque Strength Testing shows that the catheter will withstand two full rotations without failure.
- 12. Kink and Flexibility The testing shows that the catheter will not kink at a radius equal to or greater than 13mm.
- 13. Particulate The testing shows that the balloon will not create particulates that exceed that of the reference limits.
- 14. Dimensional Verification The testing shows that the devices will meet dimensional specifications.
- 15. Animal Evaluation In vivo testing characterized the safety and biologic response approximately 48 hours following the acute inflation of the balloon in the porcine coronary sinus.
- 16. Biocompatibility Testing (i.e., cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility) shows the device is biocompatible. In conformance with the applicable sections of ISO 10993-1, the predetermined acceptance criteria were met.
# Conclusions Drawn from the Tests:
The results of these tests provide reasonable assurance that the device is as safe and effective as the predicate device and support a determination of substantial equivalence.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare, featuring a staff with a snake winding around it.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 29, 2014
Cook Incorporated c/o Ms. Jennifer Richardson, MA 750 Daniels Way . P.O. Box 489 Bloomington, IN 47402
Re: K131204
Trade/Device Name: Advance® CS Coronary Sinus Infusion Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Advance® CS Coronary Sinus Infusion Catheter Regulatory Class: Class II Product Code: DQY Dated: December 30, 2013 Received: December 31, 2013
Dear Ms. Richardson:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Jennifer Richardson
.
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
.fda.gov/MedicalDevices/ResourcesforYou/Industry/ucm141.htm
Sincerely yours,
M.A. Hilleman
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Cook Incorporated Advance® CS Coronary Sinus Infusion Catheter Traditional 510(k) 26 April 2013
# Indications for Use Statement
510(k) Number (if known): _K131204
Device Name: Advance CS Coronary Sinus Infusion Catheter
# Indications for Use:
The Advance CS Coronary Sinus Infusion Catheter is intended for temporary occlusion of the coronary sinus for infusion of contrast media, drugs, or therapeutic agents; or possible introduction of devices into the coronary venous system.
Over-The-Counter Use X Prescription Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.A. Hillerman
1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.