The ASPIRON™ ACP System is intended for anterior inter-vertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: - Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration disc confirmed by patient history and radiographic studies); - Spondylolisthesis - Trauma (including fractures, dislocation) - Spinal stenosis - Tumors - Deformity (defined as scoliosis, kyphosis, or lordosis) - Pseudoarthrosis - Failed previous fusion WARNING: The device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.
Device Story
ASPIRON™ ACP System is a spinal intervertebral body fixation orthosis consisting of titanium alloy (Ti-6Al-4V ELI) bone plates, screws, and associated surgical instruments. Used by surgeons in an operating room setting via an anterior approach to the cervical spine. The system provides temporary stabilization of the cervical vertebrae (C2-T1) to facilitate fusion. The surgeon selects appropriate plate and screw sizes based on patient anatomy and secures the construct to the vertebral bodies. The device provides mechanical support to the spine, maintaining alignment during the healing process, which benefits patients by stabilizing the affected segments and promoting successful fusion. No software or electronic components are involved.
Clinical Evidence
Bench testing only. Testing included ASTM F1717 construct testing (static compression bending, static torsion, compression bending fatigue, torsion fatigue) and component screw back-out testing. Results met all acceptance criteria, demonstrating mechanical performance equivalent to predicate devices.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Components: Bone plates, screws, and surgical instruments. Principle: Mechanical fixation via anterior screw insertion into vertebral bodies. Connectivity: None (mechanical device). Sterilization: Not specified.
Indications for Use
Indicated for patients requiring temporary anterior cervical spine stabilization (C2-T1) during fusion. Conditions include degenerative disc disease, spondylolisthesis, trauma (fractures/dislocation), spinal stenosis, tumors, deformity (scoliosis, kyphosis, lordosis), pseudoarthrosis, and failed previous fusion. Contraindicated for posterior element (pedicle) fixation.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
K131200 Page 1 of 3
## 5. 510(k) Summary
| Manufacturer: | U & I Corporation<br>20, Sandan-ro, 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,<br>Korea, 480-859 | |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Sponsor: | U & I Corporation<br>20, Sandan-ro, 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,<br>Korea, 480-859 | |
| Sponsor Contact: | Gyeong-Je Kwon, Regulatory Affairs Specialist<br>+82 31 852 0102 (ext.610)<br>kkjane@youic.com | AUG 2 9 2013 |
| Date Prepared: | April 25, 2013 | |
| Device Name: | Trade Name: ASPIRONTM ACP System | |
| Classification Name: | Spinal Intervertebral Body Fixation Orthosis, per 21 CFR<br>888.3060 | |
| Common Name: | Anterior Cervical Plate | |
| Product Code: | KWQ | |
| Predicate Devices: | MaximaTM Anterior Cervical Plate System (K061002)<br>Blackstone TM III° Anterior Cervical Plating System<br>(K012184)<br>ZEPHIR TM Anterior Cervical Plate System (K994239) | |
## Description of Device:
ASPIRON™ ACP System consists of a variety of shapes and size of bone plates, screws and associated instruments. All implant components are made from a titanium alloy (Ti-6Al-4V ELI) in accordance with ASTM F136. ASPIRON ™ ACP System is intended to provide stabilization of the cervical vertebra for various indications. Fixation is provided by bone screws inserted into the vertebral body of the cervical spine using an anterior approach.
**U&i** CORPORATION
ASPIRON™ ACP System
{1}------------------------------------------------
#### Intended Use:
The ASPIRON™ ACP System is intended for anterior inter-vertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
- · Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration disc confirmed by patient history and radiographic studies);
- · Spondylolisthesis
- · Trauma (including fractures, dislocation)
- · Spinal stenosis
- · Tumors
- · Deformity (defined as scoliosis, kyphosis, or lordosis)
- · Pseudoarthrosis
- · Failed previous fusion
WARNING: The device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.
#### Substantial Equivalence:
The ASPIRON™ ACP System is substantially equivalent to Maxima™ Anterior Cervical Plate System (K061002), Blackstone™ III° Anterior Cervical Plating System (K012184) and ZEPHIR™ Anterior Cervical Plate System (K994239) in design, material, mechanical performance, function and intended use.
The mechanical performance of ASPIRON™ ACP System met the acceptance criteria which have been established from the predicate devices.
- 1. Comparison Technological Characteristics
The predicate and proposed devices have the similar intended use and basic fundamental scientific technology and share the following similarities;
- · The similar indications for use
- Similar design features .
- Incorporate the same or similar materials ●
- . The equivalent mechanical performance
- 2. Performance Testing
ASPIRON™ ACP System
**U&i** CORPORATION
Page 10 of 46
{2}------------------------------------------------
The ASPIRON™ ACP System was tested in a non clinical setting (bench testing) to assess that to no new safety and efficiency issues were raised with this device. The testing met all acceptance criteria and verifies that performance of the ASPIRON™ ACP System is substantially equivalent to the predicate devices.
The following tests were performed:
- 1) Construct Test (ASTM F1717)
- (1) Static compression bending test
- (2) Static torsion test
- (3) Compression bending fatigue test
- (4) Torsion fatigue test
- 2) Component Test
- (1) Screw back out test
#### 3. Conclusion
The data and information provided in this submission support the conclusion that the ASPIRON™ ACP System is substantially equivalent to its predicate devices with respect to indications for use and technological characteristics.
**U&i** CORPORATION
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The logo is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2013
Gyeong-Je Kwon Regulatory Affairs Specialist U & I Corporation 20, Sandan-ro, Uijeongbu-si, Gyeonggi-do, Korea 480-859
Re: K131200
Trade/Device Name: ASPIRIONTM ACP System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: July 1, 2013 Received: July 3, 2013
Dear Gyeong-Je Kwon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
## Page 2 - Gyeong-Je Kwon
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling reculation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MIDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal"roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDeviecs/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Erin I. Keith
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known):
Device Name: ASPIRON™ ACP System
#### Indications for Use:
The ASPIRON™ ACP System is intended for anterior inter-vertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
- Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration disc confirmed by patient history and radiographic studies);
- Spondylolisthesis
- Trauma (including fractures, dislocation)
- . Spinal stenosis
- Tumors
- . Deformity (defined as scoliosis, kyphosis, or lordosis)
- . Pseudoarthrosis
- Failed previous fusion
WARNING: The device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.
Over-The-Counter Use × Prescription Use ----------------------------------------------------------------------AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
ASPIRON™ ACP System
Image /page/5/Picture/19 description: The image shows the logo for U&I Corporation. The logo consists of the letters "U&I" in a stylized font, followed by the word "CORPORATION" in a blocky, sans-serif font. The text is black and the background is white.
Page 8 of 46
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.