The myPatch is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, presyncope, syncope, fatigue, chest pain and/or anxiety.
Device Story
myPatch is a small, lightweight, reusable ambulatory ECG recorder; records 2 channels of ECG data for up to 3 days. Device consists of a plastic casing with battery, microprocessor, memory, and LED; clips onto a single-use adhesive electrode placed on the patient's sternum. Patient wears device continuously; device turns on only when connected to electrode. Post-recording, device connects to PC via USB; data transferred as a file for analysis by qualified healthcare providers using third-party processing software. Provides continuous ECG monitoring to assist clinicians in diagnosing transient cardiac symptoms. Benefits include patient comfort and extended monitoring duration compared to traditional Holter monitors.
Clinical Evidence
Bench testing only. Device verified against IEC 60601-1 (general safety), IEC 60601-2-47 (ambulatory ECG systems), EN 60601-1-11 (home healthcare environment), and EN 60601-1-2 (EMC). No clinical trial data provided.
Technological Characteristics
Reusable ECG recorder with rechargeable lithium polymer battery; USB connectivity for data transfer. Single-use adhesive electrode with printed conductor paths and gel. Records 2-channel ECG. Complies with IEC 60601-2-47:2012. Firmware controls data collection; software facilitates transfer to external processing systems.
Indications for Use
Indicated for patients, asymptomatic or symptomatic (palpitations, dyspnea, dizziness, lightheadedness, presyncope, syncope, fatigue, chest pain, anxiety), requiring ambulatory ECG monitoring.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "DMS SERVICE" in a bold, sans-serif font. The letters are black, and there is a black line underlining the text. A black dot is between DMS and SERVICE. The text appears to be a logo or heading.
ーストート アイトライト アート アイト アイト アイト アイト アイト アイト
2800 Neilson Wy Sto 1 Santa Monica, CA 90405 ph: {775} 315-0660 fax: (B00] 441-3437
OCT 2 9 2013
1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1
tervice.com
K131190 pg 1 of 3
510(k) Summary
March 26, 2013
- - - - - - - - - - - - - - -
.
Contact Person: Lynda Cole, Manager, Icole@dms-service.com,
ph: (775) 315-0660 fax: (800) 441-3437
Submitter: dms-service IIc, 2800 Neilson Way, Ste. 1, Santa Monica, CA 90405 Trade
Name: myPatch
Model #: Amors1000
Common Name: Ambulatory ECG Recorder
Classification: 21 CFR 870.2800, Medical Magnetic Tape Recorder, Cardiovascular Panel
Device Class: II
Product Code: DSH
Substantially Equivalent Devices
| Manufacturer | Substantially<br>equivalent devices | 510(k) |
|----------------------------|-------------------------------------|---------|
| iRhythm | Zio Patch | K113862 |
| Cardiac Science Corp. | mySense Heart | K113176 |
| Scole Engineering Co. Inc. | Scole 300 | K923664 |
{1}------------------------------------------------
### Device Description:
The myPatch ambulatory electrocardiograph(ECG) recorder is a small, lightweight monitor that records ECG continuously. The unit records 2 channel of ECG data up to 3 days. The device is composed of a large ECG electrode and ECG recorder. The ECG recorder clips onto the electrode and records until the recorder is removed from the electrode. At the end of the recorder can be plugged into a PC via the USB connection. The ECG recording can be transferred in an available file that can be processed through recording processing software provided by recording processing software manufacturers and analyzed by qualified healthcare providers.
The myPatch records ECG in accordance to the standard IEC 60601-2-47:2012 for ambulatory ECG recorders.
During the recording the patient wears the myPatch connected to an electrode on the sternym. The
myPatch consists of a plastic casing containing a battery and electronics consisting of a microprocessor, measuring circult, memory, data storage, light-emitting dlode (transilluminates the plastic casing), and contact to the electrode. In addition to that, myPatch contains firmware to control the collection of the ECG-data and software to transfer to processing software.
The battery is a rechargeable lithium polymer battery with a built-in protection circuit. The charging of the myPatch is by use of a USB cable, which can only be connected to myPatch when it is not connected to the electrode. At the bottom of myPatch is an electrical and click-on connection for the electrode. myPatch must be connected to the electrode to turn on and to acquire data. myPatch has a serial number to secure the traceability of the equipment and the data during and after the investigation. myPatch is reusable.
The electrode ( accessory to myPatch, ) is the passive part. The electrode consists of one layer of onesided adhesive tape that runs the full length of the electrode. This adhesive tape allows the electrode to be placed on the sternum. The electrode gel is placed on the underneath side of the adhesive tape. Through this electrode gel, the ECG of the patient is measured. The top side of the electrode has an electrical and click-on connection to myPatch. The conductor paths are printed on the adhesive tape. These conductor paths make the connection between the electrode gel on the underside and the connection to myPatch on the top side of the electrode.
Release liners are placed on the underside and on the top side of the electrode. These release liners cover the adhesive tape and the electrode gel and are torn off when placing the myPatch on the sternum. The electrode has to be connected to myPatch to have any practical use. The electrode is a single- use device and is disposed of after one single use.
Indications for Use:
The myPatch is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, presyncope, syncope, fatigue, chest pain and/or anxiety,
Technological Characteristics/Substantial Equivalence:
myPatch has the same technological characteristics as the ZioPatch, mySense Heart and Scole 300 ambulatory monitors. myPatch, mySense and ZioPatch are all patch based monitors. The Scole 300 and myPatch can both continuously record 2 channels of ECG data. All monitors are powered by a battery, the frequency response will not degrade the system performance and all monitors can record a minimum of 24 hours. myPatch will record up to 3 days and Zio Patch will record up to 14 days of continuous ECG.
{2}------------------------------------------------
## K131190 pg 3 of 3
All of the above mentioned devices are intended for the same use, recording continuous ECG signals. Any differences in the hardware design of the monitors and myPatch does not affect the safety or the effectiveness of the myPatch monitor. myPatch has met the design verification testing standards as set forth in IEC 60601-2-47, which is equivalent to ANSI/AAMI EC38.
Testing:
The myPatch ECG Recorder has undergone extensive testing, including testing listed below. All testing demonstrated acceptable results:
IEC 60601 1:2005 + CORR. 1 (2006) + CORR. 2 (2007) IEC 60601-1/A11:2011 Medical electrical
equipment - Part 1: General requirements for basic safety and essential performance same as
IEC 60601-1:2012 ed 3.1 Consol. with am1
EN 60601-1-11: 2010 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential perfor-mance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 60601-2-47: 2012 Medical electrical equipment - Part 2-47: Particular requirements for
the basic safety and essential performance of ambulatory electrocardiographic systems
EN 60601-1-2:2007 Collateral standard: Electromagnetic compatibility - Requirements and tests
EN 62366:2008 Application of usability engineering to medical devices
EN 1041:2009 Information supplied by the manufacturer of medical devices
EN 980:2008 Symbols for use in the labeling of medical devices
EN 980 Bil.1:2008 CD-ROM Graphical symbols for use in the labeling of medical devices – Fact Sheet 1: Graphical symbols in electronic format
ISO 15223-1:2012 Medical devices - Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General requirements
The myPatch ECG Recorder has the CE Marking (CE 1008). The CE Mark is a statement that the myPatch ECG Recorder fulfills the guidelines established by the EU for medical devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and three wavy lines underneath. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
October 29, 2013
DMS-Service LLC C/O Lynda Cole 2800 Neilson Way, Ste. 1 Santa Monica, CA 90405 US
Re: K131190
> Trade/Device Name: MyPatch Amores 1000 Ambulatory ECG Regulation Number: 21 CFR 870.2800 Regulation Name: Ambulatory Electrocardiograph (ECG) Recorder Regulatory Class: Class II Product Code: DSH Dated: September 26, 2013 Received: September 27, 2013
Dear Ms. Cole:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Lynda Cole
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97), For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Owen注册_aris -S
for
Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use Statement
510(k)
Number:_K131190_
Device Name: myPatch
Indications For Use:
The myPatch is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, presyncope, syncope, fatigue, chest pain and/or anxiety.
Prescription Use _ _ X
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/5/Figure/13 description: The image shows a digital signature. The signature includes the text "Digitally signed by Owen P. Faris -S Date: 2013.10.29 13:00:42-04'00'". The signature also includes a stylized design to the left of the text.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.