← Product Code [JHI](/productcode/JHI) · K131166

# SOFIA(R) HCG FIA (K131166)

_Quidel Corp. · JHI · Aug 2, 2013 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K131166

## Device Facts

- **Applicant:** Quidel Corp.
- **Product Code:** [JHI](/productcode/JHI.md)
- **Decision Date:** Aug 2, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1155
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K131166 · Aug 2, 2013 | SOFIA(R) HCG FIA | Quidel Corp. | Multi-center point-of-care (POC) clinical study; Fresh patient urine specimens | The study evaluated the performance of the Sofia hCG FIA in a real-world clinical setting by comparing it to a commercially available qualitative test using 974 patient samples. | Point-of-care; Clinical performance; Pregnancy testing |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Multi-center POC field trial; Multi-center clinical study | Patients presenting for pregnancy testing; Sample Size: 974; Number of Sites: 5 | Commercially available qualitative test | Positive agreement, negative agreement, and overall agreement |

## Indications for Use

The Sofia hCG FIA is an immunofluorescence-based lateral flow assay intended for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine specimens for the early detection of pregnancy. The test is intended for prescription use only, including use at point-of-care sites.

## Device Story

Sofia hCG FIA is an immunofluorescence-based lateral flow immunoassay for qualitative hCG detection in urine. Device utilizes mouse monoclonal anti-hCG antibodies on a test cassette. User dispenses urine into cassette sample well; cassette is inserted into Sofia Analyzer. Analyzer performs automated development, scans test strip, processes fluorescent signals via method-specific algorithms, and displays results (Positive, Negative, Invalid). Used at point-of-care sites by healthcare professionals. Provides rapid pregnancy status determination to assist clinical decision-making. Benefits include objective, automated interpretation of lateral flow results, reducing potential for visual reading errors.

## Clinical Evidence

Multi-center POC field trial evaluated 974 fresh urine specimens. Compared to a commercially available qualitative test, the Sofia hCG FIA demonstrated >99% positive agreement (176/177; 95% CI: 97-100%) and >99% negative agreement (795/797; 95% CI: 99-100%). Overall agreement was >99% (971/974; 95% CI: 99-100%).

## Technological Characteristics

Immunofluorescence-based lateral flow assay. Uses monoclonal murine antibodies specific to the beta subunit of hCG. Form factor: Test cassette. Instrument: Sofia Analyzer. Connectivity: Integrated printer. Analytical sensitivity: 20 mIU/mL. Traceability: WHO International Standard 4th Edition 75/589.

## Regulatory Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

## Predicate Devices

- QuickVue+ One-Step hCG Combo test ([K973858](/device/K973858.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
K131166

B. Purpose for Submission:
New device

C. Measurand:
Human chorionic gonadotropin (hCG) in human urine

D. Type of Test:
Qualitative fluorescent immunoassay

E. Applicant:
Quidel Corporation

F. Proprietary and Established Names:
Sofia® hCG FIA

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  JHI | Class II | 21 CFR§ 862.1155, Human chorionic gonadotropin (HCG) test system | Chemistry (75)  |

{1}

H. Intended Use:

1. Intended use(s):

See below

2. Indication(s) for use:

The Sofia hCG FIA is an immunofluorescence-based lateral flow assay intended for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine specimens for the early detection of pregnancy.

The test is intended for prescription use only, including use at point-of-care sites.

3. Special conditions for use statement(s):

Prescription use only

4. Special instrument requirements:

For use with the Sofia Analyzer

I. Device Description:

The test kit consists of individually packaged test Cassettes—each containing mouse monoclonal anti-hCG antibodies for the capture and detection of hCG; disposable specimen transfer pipettes; and a package insert.

J. Substantial Equivalence Information:

1. Predicate Device Name:

CARDS® Q.S.® hCG Serum/Urine and Concise Performance Plus hCG-Combo

2. Predicate 510(k) Number:

K973858

3. Comparison with Predicate:

{2}

|  Item | Predicate Device | Candidate Device  |
| --- | --- | --- |
|  Features | CARDS® Q.S.® hCG Serum/Urine and Concise Performance Plus hCG-Combo (K973858) | Sofia hCG FIA  |
|  Intended Use | The QuickVue+ One-Step hCG Combo test is a one-step immunoassay intended for the qualitative detection of human Chorionic Gonadotropin (hCG) in serum or urine for the early detection of pregnancy. The test is intended for use by healthcare professionals | Same  |
|  Specimen Type | Urine/Serum | Urine  |
|  Format | Cassette | Same  |
|  Test Principle | Lateral flow immunoassay | Same  |
|  Instrument | No instrument; visually-read assay | Sofia Analyzer  |
|  Read Result Time | 3 minutes | Same for urine  |
|  Analytical Sensitivity | 20 mIU/mL for urine | Same for urine  |
|  Storage | 15 to 30°C | Same  |

K. Standard/Guidance Document Referenced (if applicable):

1. CLSI EP 5 A2: A Evaluation of Precision Performance of Quantitative MeasurementMethods; Approved Guideline—Second Edition 08/20/2004
2. CLSI EP 7A2: Interference Testing in Clinical Chemistry; Approved Guideline -Second Edition 11/23/2005
3. CLSI EP 12 A2: User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition 01/25/2008
4. CLSI EP 25A: Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline - 09/23/2009

{3}

5. CLSI LS101 A2: Specification for Low-Level Protocol to Transfer Messages Between Clinical Laboratory Instruments and Computer Systems; Approved Standard—Second Edition 04/28/2008
6. CLSI LS102 A2: Specification for Transferring Information Between Clinical Instruments and Computer Systems; Approved Standard—Second. Second Edition 10/20/2004
7. CENELEC. EN 61326-1 Electrical equipment for measurement, control and laboratory use -EMC requirements - Part 1: General requirements 12/2006
8. CENELEC EN 61326-2-6 Electrical equipment for measurement, control and laboratory use. EMC requirements. Particular requirements. Invitro diagnostic (IVD) medical equipment 12/2006
9. IEC 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use -Part 1: General requirements 02/2001
10. IEC 61010-2-081 Safety requirements for electrical equipment for measurement, control and laboratory use -Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes 08/2009
11. IEC 61010-2-101 Safety requirements for electrical equipment for measurement, control, and laboratory use -- Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment 12/2002
12. IEC 62304 Medical device software - Software life-cycle processes 8/20/2012
13. ISO 14971 Medical Devices - Application of risk management to medical devices 08/20/2012
14. ISO 15223-1 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied —Part 1: General requirements 07/01/2012

L. Test Principle:

The Sofia hCG FIA is an immunofluorescence-based lateral flow test for use with the Sofia Analyzer. The test uses a pair of monoclonal murine antibodies specific to the beta subunit of hCG to capture and detect hCG.

To perform the test, a urine specimen is collected and dispensed into the Sample Well on the Test Cassette. The Cassette is placed inside of the Sofia Analyzer for an automatically defined development time. The Sofia Analyzer then scans the test strip and analyzes the fluorescent signal, using method-specific algorithms. The Sofia Analyzer then displays the

{4}

test result (Positive, Negative, or Invalid) on the screen, and optionally prints the results on an integrated printer.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

The reproducibility of the Sofia hCG FIA was evaluated at three (3) different laboratories. Three (3) different operators at each site tested a series of coded, four contrived samples, prepared in negative urine, spiked with hCG traceable to WHO International 4th standard to generate samples ranging from 5 mIU/mL to 25 mIU/mL hCG.

A total of nine (9) Sofia Analyzers were used. For each level a total of 90 replicates were tested over five (5) different days at each site. The results are summarized as follows:

|  Site | Operator | (5 mIU/mL) | (9.5 mIU/mL) | (16 mIU/mL) | (25 mIU/mL)  |
| --- | --- | --- | --- | --- | --- |
|   |  | Negative/Total | Negative/Total | Positive/Total | Positive/Total  |
|  1 | 1 | 30/30 | 28/30 | 29/30 | 30/30  |
|   |  2 | 30/30 | 30/30 | 28/30 | 30/30  |
|   |  3 | 30/30 | 29/30 | 27/30 | 30/30  |
|   |  Total | 90/90 | 87/90 | 84/90 | 90/90  |
|  |   |   |   |   |   |
|  2 | 1 | 30/30 | 28/30 | 28/30 | 30/30  |
|   |  2 | 30/30 | 29/30 | 29/30 | 30/30  |
|   |  3 | 30/30 | 29/30 | 27/30 | 30/30  |
|   |  Total | 90/90 | 86/90 | 84/90 | 90/90  |
|  |   |   |   |   |   |
|  3 | 1 | 30/30 | 26/30 | 30/30 | 30/30  |
|   |  2 | 30/30 | 23/30 | 30/30 | 30/30  |
|   |  3 | 30/30 | 27/30 | 29/30 | 30/30  |
|   |  Total | 90/90 | 76/90 | 89/90 | 90/90  |
|  |   |   |   |   |   |
|  Total: |   | 270(-)/270 Total | 249(-)/270 Total | 257(+)/270 Total | 270(+)/270 Total  |

{5}

b. Linearity/assay reportable range:

Not applicable. This is a qualitative test.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The Sofia hCG FIA is standardized to the WHO Fourth International Standard 75/589.

All stability protocols were reviewed and found to be acceptable. The shelf-life stability study showed that the device is stable for 24 months when stored at room temperature (15 to 30°C).

d. Detection limit:

The sensitivity of the Sofia hCG FIA was tested by spiking pooled male urine with varying concentrations (0 to 100 mIU/mL) of hCG traceable to WHO International 4th Standard. Studies were conducted with 3 lots of devices. Representative results are listed in the table below. The positive/negative threshold at which 100% of the samples tested positive was confirmed at 20 mIU/mL hCG.

Detection Limit Testing

|  hCG Conc. (mIU/mL) | Number of Positives | Number of Negatives | Percent Positive  |
| --- | --- | --- | --- |
|  100 | 10/10 | 0/10 | 100%  |
|  50 | 10/10 | 0/10 | 100%  |
|  30 | 10/10 | 0/10 | 100%  |
|  25 | 60/60 | 0/60 | 100%  |
|  22.5 | 60/60 | 0/60 | 100%  |
|  20 | 60/60 | 0/60 | 100%  |
|  17.5 | 57/60 | 3/60 | 95%  |
|  15 | 48/60 | 12/60 | 80%  |
|  12.5 | 27/60 | 33/60 | 45%  |
|  10 | 5/60 | 55/60 | 8.3%  |
|  7.5 | 0/60 | 60/60 | 0%  |
|  5 | 0/60 | 60/60 | 0%  |
|  2.5 | 0/10 | 10/10 | 0%  |
|  0 | 0/10 | 10/10 | 0%  |

{6}

# e. Analytical specificity:

A hCG negative urine pool was initially spiked with each of the substances listed in the table below. Immediately prior to testing, five (5) levels of hCG (0, 5, 10, 20 and  $25\mathrm{mIU / mL}$ ) were then spiked into each solution containing the possible interfering or cross-reacting substance. Urinary pH and specific gravity was also evaluated. All samples were tested in replicates of five (5). The following compounds did not interfere with the test at the following tested concentrations.

|  Item | Substance/Microorganism | Concentration  |
| --- | --- | --- |
|  |   |   |
|  1 | Acetaminophen | 20 mg/dL  |
|  2 | Acetoacetic Acid | 1,600 mg/dL  |
|  3 | Acetylsalicylic Acid | 20 mg/dL  |
|  4 | Ampicillin | 2 mg/dL  |
|  5 | Ascorbic Acid | 20 mg/dL  |
|  6 | Atropine | 20 mg/dL  |
|  7 | β-Hydroxybutyrate | 2,000 mg/dL  |
|  8 | Benzoylecgonine | 8 mg/dL  |
|  9 | Bovine Serum | 10 mg/dL  |
|  10 | Caffeine | 20 mg/dL  |
|  11 | Cannabinol | 10 mg/dL  |
|  12 | Cellulose | 500 mg/dL  |
|  13 | Citric Acid | 500 mg/dL  |
|  14 | Clomiphene | 100 mg/dL  |
|  15 | Cow's Milk | 9 mg/dL  |
|  16 | DMSO | 0.90%  |
|  17 | EDTA | 80 mg/dL  |
|  18 | Ephedrine | 18 mg/dL  |
|  19 | Ethanol | 0.80%  |
|  20 | Gentisic Acid | 20 mg/dL  |
|  21 | Methanol | 0.90%  |
|  22 | Phenothiazine | 20 mg/dL  |
|  23 | Phenylpropanolamine | 20 mg/dL  |
|  24 | Salicylic Acid | 20 mg/dL  |
|  25 | Tetracycline | 20 mg/dL  |

{7}

|  Item | Substance/Microorganism | Concentration  |
| --- | --- | --- |
|  26 | Theophylline | 20 mg/mL  |
|  27 | Uric Acid | 18 mg/dL  |
|  |   |   |
|  28 | Albumin (serum) | 2,000 mg/dL  |
|  29 | Bilirubin | 1 mg/dL  |
|  30 | Glucose | 2,000 mg/dL  |
|  31 | Haptoglobin | 1 mg/dL  |
|  32 | Hemoglobin | 1 mg/dL  |
|  33 | Human Anti-Mouse Antibodies | 2.85 ng/mL  |
|  34 | Myoglobin | 1 mg/dL  |
|  35 | Rheumatoid Factor | 1.08 IU/mL  |
|  36 | Serum (negative human) | 1%  |
|  37 | Urine Peroxide | 10 mg/dL  |
|  38 | Urine pH | 5–9  |
|  39 | Urine Specific Gravity | 1.005-1.037  |
|  |   |   |
|  40 | hLH | 450 mIU/mL  |
|  41 | hFSH | 900 mIU/mL  |
|  42 | hTSH | 1,000 mIU/mL  |
|  43 | Estriol 17-beta | 28,000 μg/dL  |
|  44 | Pregnanediol glucuronide | 45,000 μg/dL  |
|  45 | Beta-core fragment, hCG | 5.1 x 10^{5} pmol/L  |
|  |   |   |

A high dose hook effect study was performed by spiking high levels of hCG concentrations 0 to 500,000 mIU/mL into negative urine and evaluating the test results. Positive results were observed up to 500,000 mIU/mL hCG. Therefore no hook effect was observed for urine samples with hCG concentrations up to 500,000 mIU/mL.

f. Assay cut-off:

Not applicable

{8}

9

2. Comparison studies:

a. Method comparison with predicate device:

A multi-center clinical study was conducted to establish the performance of the Sofia hCG FIA compared to results obtained from the predicate device. This study was conducted by health care personnel at five (5) different sites within the United States. In this multi-center, point-of-care (POC) field trial, 974 fresh urine specimens, collected from patients presenting for pregnancy testing, were evaluated. The study consisted of female subjects of childbearing age, who were being screened for pregnancy and who had given verbal consent to participation. The following subjects were excluded: post-menopausal subjects, subjects who within the past six weeks had experienced any of the following: delivered a newborn, had an abortion or a natural termination (miscarriage), or, received hCG supplements, subjects who had a hysterectomy, and subjects who were unable to understand and consent to participation. Results from the study are summarized below.

|  Predicate  |   |   |
| --- | --- | --- |
|   | Pos | Neg  |
|  Sofia Pos | 176 | 2  |
|  Sofia Neg | 1 | 795  |
|  Total | 177 | 797  |

Three discrepant results were from samples collected from patients that were too early in gestational age to obtain comparable results on the candidate and predicate devices.

b. Matrix comparison:

Not applicable

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

{9}

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Not applicable

N. Instrument Name:
Sofia analyzer.

O. System Descriptions:

1. Modes of Operation:
The Sofia Analyzer is a bench top instrument intended to be used with cassette based immunofluorescent *in vitro* diagnostic assays manufactured by Quidel Corporation. After the extracted patient sample has been added to the test cassette, the test is developed at room temperature for a pre-specified period of time. The cassette is then placed into the Analyzer where it is scanned, and the fluorescent signal of the test is processed.

2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐

3. Specimen Identification:
The user ID, patient ID, and order # can be entered via a handheld barcode scanner or by manually entering the information onto the keypad of the Sofia Analyzer.

4. Specimen Sampling and Handling:
Urine specimen is collected in a clean container. First morning specimens generally contain the highest concentrations of hCG and are recommended for early detection of pregnancy. However, any urine sample is suitable for testing.

Labeling states that if testing will not be performed immediately, the specimens may be kept at room temperature (15-30°C) or refrigerated (2-8°C) for up to 72 hours. For prolonged storage, specimens may be frozen once at -20°C or below.

5. Calibration:
The labeling provides details regarding the Calibration Check Procedure for the Sofia

10

{10}

analyzer. The Calibration Check Procedure should be performed every thirty (30) days. The Sofia Analyzer can be set to remind the user to complete the Calibration Check Procedure.

The Calibration Check is a required function that checks the Sofia Analyzer optics and calculation systems using a specific Calibration Cassette. This Calibration Cassette is shipped with the Sofia Installation Pack.

6. Quality Control:

Built-in Procedural Controls

The Sofia hCG FIA test strip contains a built-in procedural control feature. Each time a test is run, the Sofia Analyzer scans this part of the test strip, and the result is displayed on the Analyzer screen as "valid" or "invalid."

A valid result obtained with this procedural control demonstrates that the test flowed correctly and the functional integrity of the Cassette was maintained. The procedural control is interpreted by the Sofia Analyzer simultaneously with the end of the assay. If the test does not flow correctly, the Sofia Analyzer will indicate that the result is invalid. Should this occur, the labeling instructs the user to review the procedure and repeat the test with a new test Cassette.

External Quality Control

External controls are also recommended for use to demonstrate that the reagents and assay procedure perform properly. Labeling recommends that controls be tested once for each new lot, new shipment of kits, and every 30 days as a check on storage and if applicable, as deemed additionally necessary by laboratory internal quality control procedures, and in accordance with local, state and federal regulations or accreditation requirements.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Not applicable.

Q. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

11

---

**Source:** [https://fda-staging.innolitics.com/device/K131166](https://fda-staging.innolitics.com/device/K131166)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
