K131165 · Alliance Global · DZE · Nov 14, 2013 · Dental
Device Facts
Record ID
K131165
Device Name
ANKER DENTAL IMPLANT SYSTEM
Applicant
Alliance Global
Product Code
DZE · Dental
Decision Date
Nov 14, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Anker Dental Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. It is intended for delayed loading.
Device Story
Anker Dental Implant System consists of a titanium (grade IV) fixture with an SLA (sand-blasted, large grit, acid-etched) surface, various abutments (healing, fixed, dual, angle, o-ring, temporary), and a cover screw. Fixtures range from 3.4-5.0 mm in diameter and 7.0-15.0 mm in length. Abutments are primarily titanium alloy, with temporary abutments made of SUS316 stainless steel. The system is surgically placed in the jaw bone by a dental professional to support prosthetic devices like artificial teeth. It is designed for delayed loading to restore chewing function. All components are provided sterile.
Clinical Evidence
No clinical data. Safety and effectiveness were established through non-clinical bench testing, including compressive force and fatigue testing (ISO 14801), corrosion testing (ASTM G3-89), and comprehensive biocompatibility testing (ISO 10993 series).
Technological Characteristics
Fixture: Grade IV pure titanium with SLA surface. Abutments: Titanium alloy (temporary: SUS316 stainless steel). Dimensions: 3.4-5.0 mm diameter, 7.0-15.0 mm length. Sterilization: Gamma irradiation. Standards: ISO 14801 (fatigue), ASTM G3-89 (corrosion), ISO 10993 (biocompatibility), ISO 11137-1 (sterilization), ASTM F88/F1140/F1929 (shelf life).
Indications for Use
Indicated for patients requiring dental implants in the upper or lower jaw arches to support prosthetic devices and restore chewing function. Intended for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Alliance Global Technology Co. Ltd. The logo features the word "Alliance" in large, bold letters, with the words "Global Technology Co. Ltd" in smaller letters underneath. Above the word "Alliance" is a graphic of a globe. The logo is black and white.
# 510(k) SUMMARY
## Anker Dental Implant system
Company Name: Alliance Global Technology Co., Ltd. 2F., No.92, Luke 5th Rd., Luzhu Dist., Kaohsiung City 821, Taiwan (R.O.C.) Telephone: +886-7-695-6688 Fax: +886-7-695-5329
Contact Person: Ms. Yayuan Chang Telephone: +886-7-695-6688 Fax: +886-7-695-5329 E-mail: ketty@anchorfast.com.tw
Date prepared: Sep 12nd, 2013
Trade Name: Anker Dental Implant System
Common Name: Dental Implant
Classification Name: Root-form endosseous dental implant
Device Classification: Class II
Regulation Number: 21 CFR 872.3640
Panel: Dental
Product Code: DZE
#### Predicate Device:
(1)Predicate Device Name: Dentium Co., Ltd Implantium Manufacturer: Dentium Company Limited 510(k) Number: K041368 (2)Predicate Device Name: Osstem GS-III system Manufacturer: OSSTEM Implant Co., Ltd.
510(k) Number: K091208
510(k) K131165-Anker Dental Implant System Section 5 - Page 5-1
**NOV** 14 2019
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for ALLIANCE GLOBAL TECHNOLOGY CO. LTD. The logo features the word "ALLIANCE" in large, bold, black letters. Above the word "ALLIANCE" is a graphic of a globe. Below the word "ALLIANCE" is the text "GLOBAL TECHNOLOGY CO. LTD." in smaller, black letters.
#### Device Description:
Anker Dental Implant System consists of fixture, abutment (healing abutment, fixed abutment, dual abutment, angle abutment, o-ring abutment, temporary abutment) and cover screw. Fixture is made of pure titanium (grade IV) and its surface was treated by SLA (Sand-blasted, Large grit, Acid-etched) process. Diameters of fixtures are including 3.4 to 5.0 mm and lengthes are including 7.0 to 15.0 mm. Most abutments are made of titanium alloy and their diameters are including 4.0 to 7.0 mm. Temporary abutment is made of SUS316 stainless steel instead of titanium alloy. All products are sterilized as finished products.
### Indications for Use:
Anker Dental Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. It is intended for delayed loading.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for ALLIANCE GLOBAL TECHNOLOGY CO., LTD. The word "ALLIANCE" is in large, bold, black letters. Below that is the text "GLOBAL TECHNOLOGY CO., LTD." in smaller, thinner, black letters. Above the word "ALLIANCE" is a black and white image of a globe.
## Substantial Equivalence:
| Item | | Anker Dental Implant<br>System | Osstem GS-III system | Dentium Co., Ltd<br>Implantium |
|------|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Classification | Class II | Class II | Class II |
| 2. | Code or<br>Federal<br>Regulations | 872.3640 | 872.3640 | 872.3640 |
| 3. | 510K No. | K131165 | K091208 | K041368 |
| 4. | Design | Image: Anker Dental Implant | Image: Osstem GS-III system | Image: Dentium Co., Ltd Implantium |
| 5. | Intended Use | Anker Dental Implant<br>System is intended to be<br>surgically placed in the bone<br>of the upper or lower jaw<br>arches to provide support for<br>prosthetic devices, such as<br>artificial teeth, and to restore<br>the patient's chewing<br>function. It is intended for<br>delayed loading. | The GS III System is<br>indicated for use in<br>partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal<br>or intermediate abutment<br>support for fixed<br>bridgework. The GS III<br>System is for single and<br>two stage surgical<br>procedures. It is not for<br>immediate load. | The Dentium Co., Ltd<br>Implantium is intended to<br>be surgically placed in the<br>bone of the upper or<br>lower jaw arches to<br>provide support for<br>prosthetic devices, such<br>as artificial teeth, and to<br>restore the patient's<br>chewing function. |
| 6. | Body<br>Diameter | 3.4mm, 3.5mm, 3.8mm,<br>4.0mm, 4.3mm, 4.5mm,<br>4.8mm, 5.0mm | 3.7 mm,4.2 mm,4.6 mm,<br>5.1mm | 3.4mm, 3.8mm, 4.3mm,<br>4.8mm |
| 7. | Length | 7mm, 8 mm, 8.5mm, 10mm,<br>11.5mm, 12mm, 13mm,<br>14mm,<br>15mm | 7 mm, 8.5 mm, 10mm,<br>11.5 mm, 13mm, 15mm | 8 mm, 10 mm, 12mm,<br>14mm |
| 8. | Surface | S.L.A. (Sand-blasted, Large<br>grit, Acid-etched surface) | RBM (Resorbable<br>Blasting Media) | S.L.A. (Sandblasting with<br>large grit and acid<br>etching) |
| 9. | Sterilization | y-ray (Radiation) | y-ray (Radiation) | y-ray (Radiation) |
| 10. | Material | Titanium | Titanium | Titanium |
| 11. | Abutment<br>angulations | 0. 17 deg | 0. 17 deg | 0. 15, 25 deg |
| 12. | Material of<br>Abutments | Titanium Vanadium Alloy | Titanium Vanadium Alloy | Titanium Vanadium Alloy |
510(k) K131165-Anker Dental Implant System Section 5 - Page 5-3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ALLIANCE GLOBAL TECHNOLOGY CO., LTD. The word "ALLIANCE" is written in large, bold, black letters. Above the word "ALLIANCE" is a black and white image of a globe. Below the word "ALLIANCE" is the text "GLOBAL TECHNOLOGY CO., LTD" in smaller, black letters.
Comparing to the predicate devices, Dentium Co., Ltd Implantium (510K No. K041368) and Osstem GS-III system (510K No. K091208), Anker Dental Implant System is equivalent in surface treatment, intended use, method of operation, material and design.
### Non-clinical Testing:
Non-clinical test was used to support the decision of safety and effectiveness.
Non-clinical testing consisted of performance of testing in accordance with the FDA guidance "Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." The results of the non-clinical testing demonstrate that the Anker Dental Implant System is substantially equivalent to the predicate devices.
| Testing Item | Reference |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Compressive forces and Fatigue tests | ISO14801 |
| Compatibility test of dental implant/abutment<br>interface | N/A |
| Corrosion test | ASTM G3-89 |
| Residual of Acidic Substances Test | ISO10993-12 |
| Biocompatibility test | ISO10993-3<br>ISO10993-5<br>ISO10993-6<br>ISO10993-10<br>ISO10993-11<br>Pharmacopeia US<br>OECD guideline #473 |
Overview for all non-clinical testing:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ALLIANCE GLOBAL TECHNOLOGY CO., LTD. The word "ALLIANCE" is written in large, bold, black letters. Above the word "ALLIANCE" is a black and white image of the earth. Below the word "ALLIANCE" is the text "GLOBAL TECHNOLOGY CO., LTD" in smaller, black letters.
| | OECD guideline #474 |
|-----------------------------------------------|---------------------|
| Sterilization Validation of Gamma Irradiation | ISO11137-1 |
| Shelf life Validation | ASTM F88/F88M-09 |
| | ASTM F1140-07 |
| | ASTM F1929-98 |
| | ISO11737-2 |
## Clinical Testing:
Non-clinical test was used to support the decision of safety and effectiveness.
## Conclusion:
The evaluation of the Anker Dental Implant system does not raise any additional . concerns regarding safety and effectiveness and Anker Dental Implant system may therefore be considered substantially equivalent to their predicate device.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
November 14, 2013
Alliance Global Company Technology Company, Limited Ms. Yayuan Chang Manager 2F., No.92, Luke 5th Road Luzhu District, Kaohsiung City 82151 TAIWAN (R.O.C.)
Re: K131165
Trade/Device Name: Anker Dental Implant System Regulation Number: 21 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: October 4, 2013 Received: October 7, 2013
Dear Ms. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
MaryFDAmer-S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ALLIANCE GLOBAL TECHNOLOGY CO., LTD. The logo is in black and white and features the word "ALLIANCE" in large, bold letters. Below the word "ALLIANCE" is the text "GLOBAL TECHNOLOGY CO., LTD." in smaller letters. Above the word "ALLIANCE" is an abstract design.
# Indications for Use
510(k) Number: K131165
Device Name: Anker dental implant system
Indications for Use:
Anker Dental Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. It is intended for delayed loading,
Prescription Use ____________ Over-The-Counter Use _ AND/OR (Part 21 CFR 801 Subpart D) (2) CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Page I of ____________________________________________________________________________________________________________________________________________________________________
510(k) K131165-Anker Dental Implant System Section 4 - Page 4-1
Andrew I. Steen -S - . - . 2013.11.13 11:25:52 -05'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.