The Respire Pink Series - Herbst is indicated to treat mild to moderate OSA.
Device Story
Respire Pink Series-Herbst is a custom-fitted, two-piece mandibular advancement splint (MAS) made of acrylic; used to treat snoring and mild to moderate obstructive sleep apnea. Device holds the mandible in a forward position to maintain an open airway during sleep. Features telescopic Herbst hardware on the buccal aspect allowing lateral and protrusive jaw movement while preventing retrusion; adjustable via titration keys and screws for incremental advancement. Available in Hard/Soft (dual laminate) or Hard (all-acrylic) configurations. Used by patients at home; prescribed by clinicians. Benefits include increased pharyngeal opening, improved airflow, and patient comfort due to tongue space design. Upper and lower trays unhook for easy removal.
Clinical Evidence
Clinical evaluation and observation results demonstrated success in reducing snoring and apneic events as measured by polysomnograms. Bench testing confirmed all test method acceptance criteria were met, demonstrating results equivalent to predicate devices.
Technological Characteristics
Customized two-piece mandibular advancement splint. Materials: Acrylic (polyethyleneterephthalate) trays, stainless steel hardware. Biocompatibility per ISO 7405:2008. Features telescopic Herbst hardware, ball clasps, and titration keys for up to 8 mm of mandibular advancement. Non-powered, mechanical device.
Indications for Use
Indicated for adult patients with mild to moderate obstructive sleep apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
ס״ד
### STERLING MEDICAL REGISTRATION
### SECTION 5 ......... 510(k) Summary (21 CER 807.92)
### 510(k) Number K
SEP 09 2013
| 1 | Submission Owner | Respire Medical® LLC | | |
|---|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|----------------------|
| | | 18 Bridge St., Ste. 4J | | |
| | | Brooklyn, 11201 NY | | |
| | | Phone 718-643-7326 | | |
| | | Fax 718-643-7322 | | |
| 2 | Official Correspondent<br>Contact Person | Sterling Medical Registration<br>Daniela Levy - Regulatory Consultant<br>22817 Ventura blvd. #161<br>Woodland Hills, CA 91364, USA<br>Phone 213-787-3026<br>Fax 213-447-5297<br>Web www.sterlingmedicalregistration.com | | |
| 3 | Submission Date | April 2013 | | |
| 4 | Device Trade Name | Respire Pink Series- Herbst | | |
| 5 | Regulation Description | Intraoral devices for snoring and intraoral devices<br>for snoring and obstructive sleep apnea (OSA) | | |
| 6 | Classification | Device Name | : | Device, Anti-Snoring |
| | | Product Code | : | LRK |
| | | Regulation No | : | 872.5570 |
| | | Class | : | II |
| | | Panel | : | Dental |
| 7 | Reason for the Premarket Notification Submission | | : | New Device |
- 8 Identification of Legally Marketed Predicate Devices :
- · Respire Pink Series Herbst is substantially equivalent to Respire Blue Series . K111207 ; SUAD K023836 in terms of intended use, indication for use, technological characteristics, performance and user interface. The predicate devices are Class II medical devices.
{1}------------------------------------------------
## FERLING MEDICAL REGISTRATION
#### ರಿ Device Description
Respire Pink Series - Herbst is a customized device for each patient which consists of two dental plates, upper and lower, made of Acrylic.
The Respire Pink Series – Herbst is a mandibular advancement splint that holds the jaw in a forward position to help keep the tongue and supporting tissues in a position to help maintain an open airway, which in turn helps in the treatment of snoring and mild to moderate obstructive sleep apnea.
The Herbst hardware on the side of the devices allow the patient to move forward and left and right, but not backwards. These movements give the patient some freedom to move which is important for their comfort and overall success of the device.
The upper and lower components are connected by an adjustable hinge, thus patient can open and close while wearing the appliances.
Respire Pink Series - Herbst are offered in two options: {1) Hard/Soft which has a dual laminate layer that provides a soft layer on the tooth surface (2) Hard devices which are all acrylic and retained with ball clasps, this allows the device to be tightened if it it becomes loose.
10 Intended use
- The Respire Pink Series Herbst is indicated to treat mild to moderate OSA. .
- 11 Performance Standards or Special Controls
- Recognized Consensus Standard: ISO 7405:2008 Dentistry Evaluation of . biocompatibility of medical devices used in dentistry
. •
| Substantial<br>Equivalent<br>Table | Respire Blue Series<br>(Hard/Soft Surface) | Respire Blue Series<br>(Hard Surface) | Respire Pink Series -<br>Herbst | SUAD |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K | K111207 | K111207 | | K023836 |
| Company<br>Name | Respire Medical LLC | Respire Medical LLC | Respire Medical LLC | Strong Dental Inc. |
| Regulation<br>Description | Intraoral devices for<br>snoring and intraoral<br>devices for snoring<br>and obstructive sleep<br>apnea (OSA) | Intraoral devices for<br>snoring and intraoral<br>devices for snoring and<br>obstructive sleep apnea<br>(OSA) | Intraoral devices for<br>snoring and intraoral<br>devices for snoring and<br>obstructive sleep apnea<br>(OSA) | Intraoral devices for<br>snoring and intraoral<br>devices for snoring and<br>obstructive sleep apnea<br>(OSA) |
| Device Name | Device, Anti-Snoring | Device, Anti-Snoring | Device, Anti-Snoring | Device, Anti-Snoring |
| Product Code | LRK | LRK | LRK | LRK |
| Classification | Class II | Class II | Class II | Class II |
| Intended Use | The Respire Blue<br>Series is indicated to<br>treat mild to moderate<br>OSA. | The Respire Blue<br>Series is indicated to<br>treat mild to moderate<br>OSA. | The Respire Pink<br>Series - Herbst is<br>indicated to treat mild<br>to moderate OSA. | A custom fitted<br>mandibular<br>repositioning device<br>intended to reduce or<br>alleviated nighttime<br>snoring and obstructive<br>sleep apnea |
| Single or<br>Multiple use | Multiple use | Multiple use | Multiple use | Multiple use |
| Target<br>population | Adults Patients | Adults Patients | Adults Patients | Adults Patients |
| Prescription /<br>OTC Use | Prescription only | Prescription only | Prescription only | Prescription only |
| Device<br>Components | Orthodontic Acrylic<br>trays, Expansion<br>Screws, Wire with<br>Ball Clasp | Orthodontic Acrylic<br>trays, Expansion<br>Screws, Wire with Ball<br>Clasp | Orthodontic Acrylic<br>trays, Telescopic<br>Herbst Hardware and<br>Ball Clasp | Orthodontic Acrylic<br>trays, Telescopic<br>Herbst Hardware and<br>Ball Clasp |
| Appliance<br>Design | Customized device<br>Rigid tray two pieces<br>Upper/Lower acrylic | Customized device<br>Rigid tray two pieces<br>Upper/Lower acrylic | Customized device<br>Rigid tray two pieces<br>Upper/Lower acrylic | Customized device<br>Rigid tray two pieces<br>Upper/Lower acrylic |
| Device<br>Functionality | Allows to increase<br>pharyngeal opening,<br>and to improve the<br>ability to exchange air<br>during sleep.<br>Upper and lower tray<br>unhook for easy<br>removal from mouth.<br>Works by mandibular<br>advancement.<br>Adjustable using<br>titration keys. | Allows to increase<br>pharyngeal opening,<br>and to improve the<br>ability to exchange air<br>during sleep.<br>Upper and lower tray<br>unhook for easy<br>removal from mouth.<br>Works by mandibular<br>advancement.<br>Adjustable using<br>titration keys. | Allows to increase<br>pharyngeal opening,<br>and to improve the<br>ability to exchange air<br>during sleep.<br>Upper and lower tray<br>unhook for easy<br>removal from mouth.<br>Works by mandibular<br>advancement.<br>Adjustable using<br>titration keys. | Allows to increase<br>pharyngeal opening,<br>and to improve the<br>ability to exchange air<br>during sleep.<br>Upper and lower tray<br>unhook for easy<br>removal from mouth.<br>Works by mandibular<br>advancement.<br>Adjustable using<br>titration keys. |
| Mandibular<br>Advancement<br>range | Up to 6 mm | Up to 6 mm | Up to 8 mm | Up to 8 mm |
| Raw Material:<br>Upper and<br>Lower Trays | Acrylic<br>(polyethyleneterephth<br>alate) | Acrylic<br>(polyethyleneterephtha<br>late) | Acrylic<br>(polyethyleneterephtha<br>late) | Acrylic<br>(polyethyleneterephtha<br>late) |
| Raw Material:<br>Metal<br>Components | Stainless Steel | Stainless Steel | Stainless Steel | Stainless Steel |
| Colorants | Red/Blue | Red/Blue | Pink | No colorant |
### 12 Substantial Equivalence
{2}------------------------------------------------
## MEDICAL REGISTRATION
,
Summary of Equivalence: The Respire Pink Series - Herbst is considered to be substantially equivalent to Respire Blue Series K111207 ; SUAD K023836 in terms of intended use, indication for use, technological characteristics, performance and user interface.
{3}------------------------------------------------
# EDICAL REGISTRATION
As similar to its predicate device Respire Pink Series is a customized device, consists of two parts, upper and lower trays, made of acrylic. The difference between the Respire Blue Series and the Respire Pink Series is related to the hardware/components that are used on the buccal aspect to hold the lower jaw in a forward position.
The differences between Respire Pink Series – Herbst to SUAD appliance is that Respire Pink acrylic contour is designed to increase tonque space which is more comfortable for patient use. Respire Pink appliance is adjustable using titration keys and a screw which enables the jaw to be brought forward in small advancements. The design differences emphasis the advantages of Respire Pink Series technology and thus, raise no new safety and/or effectiveness issues. Respire Pink Series the same technological characteristics as its predicate devices and raise no new issues of safety or effectiveness, thus, the Respire Pink Series is substantially equivalent to its predicate devices.
Risk Assessment performance has demonstrated no new safety and/or effectiveness issues.
Bench testing results have demonstrated that all test method acceptance criteria were met and demonstrated equivalent results to the predicated devices. Thus, Respire Pink Series shares similarity with its predicate devices and raise no new safety and/or effectiveness issues.
Clinical testing - Clinical evaluation and observation results have demonstrated the success rate of reduction of snoring and the success rate of reduction of apneic events measured by polysomnograms. Thus, Respire Pink Series shares similarity in the indication of use and raise no new safety and/or effectiveness issues.
### Conclusion:
As verified by clinical and non clinical data, bench testing and substantial equivalence table, Respire Pink Series shares similarity with its predicated device by term of intended use, raw material and technical design. The fundamental scientific technology of the device is identical or very similar to the referenced predicate devices, thus Respire Pink Series is as safe and effective for its intended use and performs as well the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, uppercase letters. The text is centered and appears to be the heading of a document or publication. The font is a simple, sans-serif typeface.
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### September 9, 2013
Respire Medical Limited Liability Company C/O Ms. Daniela Levy Regulatory Consultant Sterling Medical Registration 22817 Ventura Boulevard #161 WOODLAND HILLS CA 91364
Re: K131138
Trade/Device Name: Respire Pink Series-Herbst Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II Product Code: LRK Dated: August 4, 2013 Received: August 9, 2013
Dear Ms. Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be four of the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Levy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## TERLING MEDICAL REGISTRATION
SECTION 4 - Indication for Use Statement
Indications for Use
Indications for Use
510(k) Number (if known): _|113 |/ 3 8
Device Name:
Respire Pink Series - Herbst
Indications for Use:
The Respire Pink Series - Herbst is indicated to treat mild to moderate OSA.
Prescription Use _ كا (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Andrew I. Steen -5 2013.09.06 115:11 1:55 -04'00'
'Division Sign-Off) vision of Anesthestology, General Hospital qtion Control, Dental Devices
Page 1 of __
0(k) Number: _ Ki 31138
PDF processed with CutePDF evaluation edition www.CutePDF.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.