21.3 INCH MONOCHROME DIGITAL MAMMOGRAPHY LCD MONITOR MS55I2 (ML21055, MD211G5)
Applicant
Jvckenwood Corporation
Product Code
LLZ · Radiology
Decision Date
Jul 18, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
21.3 inch Monochrome Digital Mammography 5M pixel LCD Monitor with sub-pixel driving techniques enabling 15M subpixels to be driven independently, MS55i2 (ML21055, MD211G5) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is to be used in digital mammography PACS and modalities including FFDM.
Device Story
MS55i2 is a 21.3-inch monochrome LCD monitor for digital mammography; displays medical images from PACS and FFDM modalities. Device utilizes sub-pixel driving techniques to independently drive 15M subpixels; supports DVI and DisplayPort inputs. Operated by trained medical practitioners in clinical settings. Device transforms digital input signals into high-resolution grayscale images; provides 10.3-bit grayscale (1276 gradations) via 11-bit LUT. Includes 'Mediviser' software for DICOM GSDF calibration and QC/QA. Output allows clinicians to view diagnostic-quality mammographic images; aids in clinical decision-making by providing high-contrast, high-resolution visualization of medical data.
Clinical Evidence
Bench testing only. Performance verified against AAPM-TG18 guidelines for luminance response, uniformity, geometric distortion, and chromaticity. No clinical data presented.
Indicated for trained medical practitioners to display and view medical images for diagnosis in digital mammography PACS and modalities including FFDM.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
21.3 inch (54 cm) Monochrome Digital Mammography LCD Monitor MS53i2 (ML21053) (K111496)
Submission Summary (Full Text)
{0}------------------------------------------------
131137 2
## JVCKENWOOD
## JUL 1 8 2013
t
# 510(k) SUMMARY
| Submitted Information: | JVC KENWOOD Corporation<br>3-12, Moriya-cho, Kanagawa-ku, Yokohama-shi,<br>Kanagawa, 221-0022 Japan |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tsukasa Tashiro, General Manager<br>Email: tashiro.tsukasa@jvckenwood.com<br>Tel: +81.258.24.6611<br>Fax: +81.258.24.6617 |
| Date Prepared: | July 17, 2013 |
| Device Name: | 21.3 inch Monochrome Digital Mammography LCD Monitor MS55i2<br>(ML21055, MD211G5) |
| Common Name: | MS55i2, ML21055, MD211G5 |
| Classification Name: | Class II<br>(Part 892 Radiology Devices<br>Sec. 892.2050 Picture Archiving and Communication System) |
| Predicate Device: | 21.3 inch (54 cm) Monochrome Digital Mammography LCD Monitor<br>MS53i2 (ML21053) (K111496). |
| Device Description: | MS55i2 (ML21055, MD211G5) is a 21.3-inch monochrome LCD<br>monitor whose display resolution is 2048 x 2560 (landscape), 2560 x<br>2048 (portrait) supporting DVI (digital visual interface)<br>and<br>DisplayPort. |
| Intended Use: | 21.3 inch Monochrome Digital Mammography 5M pixel LCD Monitor<br>with sub-pixel driving techniques enabling 15M subpixels to be driven<br>independently, MS55i2 (ML21055, MD211G5) is intended to be used in<br>displaying and viewing medical images for diagnosis by trained medical<br>practitioners. It is to be used in digital mammography PACS and<br>modalities including FFDM. |
| Substantial Equivalence: | MS55i2 (ML21055, MD211G5) shares the same characteristics with our<br>predicate device MS53i2 (ML21053) (K111496) except for the main<br>board, LCD panel and power supply. |
{1}------------------------------------------------
| - |
|-----------------|
| |
| |
| |
| |
| |
| |
| |
| |
| |
| |
| |
| |
| |
| |
| |
| 1 |
| And Status Load |
| |
| |
| 1 |
| 1 |
| |
| 510(k) Number | 5M Monochrome Digital Mammography<br>LCD Monitor MS53i2 (ML21053) (K111496)<br>Conventional LCD Panel (IAQS80P) | 5M Monochrome Digital Mammography<br>LCD Monitor MS55i2 (ML21055, MD211G5)<br>New LCD Panel (TX54D84MM0BAA) |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display Area | K061447<br>Horizontal: 337.9mm, Vertical: 422.4mm | Not Known<br>Horizontal: 337.92mm, Vertical: 422.4mm |
| Input Signal | DVI, DisplayPort | DVI, DisplayPort |
| Maximum Display | 2048 x 2560 dots | 2048 x 2560 dots |
| Scanning Frequency | Horizontal: 129.6K, Vertical: 50Hz - Portrait<br>Horizontal: 103.93KHz, Vertical: 50Hz - Landscape | Horizontal: 129.6K, Vertical: 50Hz - Portrait<br>Horizontal: 103.5KHz, Vertical: 50Hz - Landscape |
| Maximum Image Clock | 148 MHz | 285 MHz |
| Maximum Luminance<br>(Calibrated Luminance) | Min.850 cd/m² Typ.1100cd/m²<br>(500 cd/m² - calibrated luminance) | Min.850 cd/m² Typ.1200cd/m² (as LCD component)<br>(500 cd/m² - calibrated luminance) |
| Contrast Ratio | Min.600 Typ.850 | Min.900 Typ. 1200 |
| Viewing Angle | CR>20 Typ.85deg | CR>50 Typ.176 horizontal/vertical |
| Luminance Calibration<br>(Optional) | Software (Standard): Medivisor<br>Calibration Sensor (Optional): Chroma5 (X-Rite) | Software (Standard): Medivisor<br>Calibration Sensor (Optional): Chroma5 (X-Rite) |
| Serial Communication | USB: Downstream port x 2, Upstream port x 1 | USB: Downstream port x 2, Upstream port x 1 |
| Grayscale | 10.3 bit (1276 gradation) | 10.3 bit (1276 gradation) |
| Safety Standard | UL60601-1, CSA C22.2 No. 601-1, FCC-B, VCCI-B, MDD/CE, | UL60601-1, CSA C22.2 No. 601-1, FCC-B, VCCI-B, MDD/CE, |
| Weight & Dimension | Net: 12.1kg<br>474.5(W) x 480(H) x 220(D) mm (Landscape)<br>389(w) x 522(H) x 220(D) mm (Portrait)<br>Packed: 17.0 Kg<br>470(W) x 685 (H) x 345(D) | Net: approx. 12.8kg<br>474.5(W) x 480(H) x 220(D) mm (Landscape)<br>390(w) x 522(H) x 220(D) mm (Portrait)<br>Packed: approx.16.0 Kg<br>470(W) x 685 (H) x 345(D) |
| Power Supply | AC100-240V, 1.5-0.6A, 50/60Hz | AC100-240V, 1.5-0.6A, 50/60Hz |
Similarities :
MS55i2 (ML21055) employs the same diver board, tilt stand, power supply, etc. except of those of LCD Panel (included inverter board
Differences: Differences:
ellinentees.
MS55i2 (ML21055) employs a different LCD backlight) and the Maximum Luminance, contrast ratio and viewing angle have been
improved. MD211G5 is the same as MS55
{2}------------------------------------------------
| Display System Description | | |
|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 1. Active-Matrix Liquid-Crystal Displays (AMLCD) panel manufacturer, technology Screen size and<br>pixel pitch: | | |
| | Screen size: 422.4mm(H) x 337.92mm(V) at landscape display. 337.92mm(H) x 422.4mm(V) at<br>portrait display. Pixel pitch: 0.055mm(H) (sub-pixel pitch) x 0.165mm(V) at landscape display.<br>0.165mm(H) x 0.055mm(V) (sub-pixel pitch) at portrait display | |
| | • Communicating ports:<br>Serial communication: USB (upstream x1, downstream x2) | |
| 2. Graphics card and software | | |
| | • Digital-to-Analog converters: speed and precision:<br>Only Digital operation, Digital to Analog converters are not provided.<br>• Software included: calibration, QC/QA (procedures with required frequencies and action<br>limits): | |
| | Totoku's calibration software called "Mediviser" optimizes the display's performance using<br>DICOM GSDF gamma and luminance. Calibration is recommended quarterly.<br>• On-screen GUI: On-screen GUI is not provided.<br>• Panel user controls: power switch | |
| Technical Specification | | |
| 1. Measuring Equipment<br>Refer to List | | |
| 3. Luminance response | 2. Declared and actual (measured with test pattern in the screen) array size:<br>Declared array size (spec): 422.4mm(H) x 337.92mm(V)<br>Actual array size (measured): 422.35mm(H) x 337.95mm(V) | |
| • max and min achievable luminance: | [SPEC] Luminance Max (L.max)=1200cd/m2. Luminance Min (Lmin)= 0.5cd/m2 | |
| • max and min recommended (operational) luminance: | [SPEC] Luminance Max recommend: Lmax(r)=500cd/m2. Lmin(r)=0.8cd/m2 | |
| • intrinsic bit-depth of the panel: [SPEC] 8bit per each subpixel | | |
| | • true output bit-depth by performing visual test with gradient test pattern:<br>[SPEC] 10.3bit gray scale (1276 steps) achieved by 11bit LUT (Look-up Table) in the Display<br>hardware. | |
| | • intrinsic luminance response at 256 digital values:<br>[SPEC] dJND per dP<3.000, dJNDs/dP max error<2.000. | |
| | JNDs/dP RMS error<1.000 | |
| | Above spec is based on AAPM-TG18 Advanced Luminance Response. 4.3.5<br>Refer to actual luminance response data | |
| | • conformance to a grayscale function (i.e., DICOM GSDF) at 256 digital values and angular<br>dependence of such conformance: | |
| | [SPEC] Kσ within +- 10% deviation to DICOM GSDF based on AAPM-TG18<br>Luminance Response 4.3.4 | |
| | Refer to actual luminance response data including angular dependence (Digital Driving Level vs<br>Delta L, L.) | |
| | • angular dependency of luminance:<br>[SPEC] C*1 R*=175, K*=30% | |
| 4. Luminance uniformity | [SPEC] Less than 30% based on AAPM-TG18 4.4. Refer to actual luminance<br>uniformity data | |
·
.
:
.
.
..
{3}------------------------------------------------
K131137
Page 4 of 4
### [SPEC] Less than 2.0% based on A.APM-TG18, Refer to actual Geometrical Disturtion data 6. Display reflectance · Bi-directional reflection distribution function, or · specular and diffuse coefficients (ISO13406-2 by TUV) . [SPEC] Refer to Max allowable ambient luminance in Tables 4 and 5 on AAPM-TG18 4.2.4 Refer to actual Reflectance Data 7. Noise · Pixel fill factor (fefxos ISO 13406-2) ISPEC | 30% Min. Refer to Data of Pixel fill factor. · RMS (image variance) and noise power spectrum (weiner spectrum) ISPI C'10.1 Max at 0.6Hz. 3Hz. 30Hz Jitter. Refer to actual data 8. Veiling glare ISPI.C [Glare Ratio (GR) == 400 GR (I b=I m) (1 -1 m based on A \PM-1618 4,7,4 Reter to actual veiling glare ratio data 9. Chromaticity [SPEC] Delta (u', v') ≤ 0.01 measured at 809% Lmax based on AAPM-7018 4.8.4 Refer to Chromaticity actual data 10. Artifacts · pixel dropouts including spatial distribution (ISO13406-2 by TUV): [SPEC] Class (pixel) II. Refer to table 3 on 3.4.13 ISO13406-2 · phase/clock issues flicker · miscellaneous including ringing, ghosting, image sticking [SPEC] By visible check, no ringing, ghosting image sticking 11. Spatial resolution, spatial MTF DATA | Refer to actual MH data 12. Temporal response · Temporal MTF, or · rise and fall time constants for 5-95% and 40-60% transitions by { 'M(). |SPIC | Rise Time Tr. Full Time | f. Tr. 1 Fessons. Refer to actual data 13. Stability (possibly determined via temperature or time stress tests) by TUV · of luminance response, of temporal response, of described artifacts 18PEC] Ko within +2 10% deviation to DICOM (iSDF, based on AAPM-T(18 I uminance Response 4.3.4 via temperature stress 0 deg('. 25deg('. 40degC Refer to actual measured data
[fefxos = for example, following ... or similar]
5. Geometrical distortion
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a circular logo with an emblem in the center. The emblem appears to be a stylized representation of an eagle or bird-like figure with three curved lines forming its body and wings. Encircling the emblem is text that reads 'DEPARTMENT OF HEALTH & HUMAN SERVICES USA' in a circular arrangement around the emblem.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-G605 Silver Spring, MD 20993-0002
July 18, 2013
JVC KENWOOD Corporation % Mr. Tsukasa Tashiro General Manager 3-12, Moriya-cho, Kanagawa-ku, Yokohama-shi Kanagawa 221-0022 JAPAN
Rc: K131137
Trade/Device Name: 21.3 inch Monochrome Digital Mammography LCD Monitor MS5512 (ML21055, MD211G5)
Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 19, 2013 Received: April 23. 2013
Dear Mr. Tashiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (2) CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Tashiro
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Smh.7)
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
for
Enclosure
{6}------------------------------------------------
## Indications for Use
K131137 510(k) Number (if known):
Device Name: 21.3 inch Monochrome Digital Mammography LCD Monitor MS55i2 (ML21055, MD211G5)
Indications for Use:
21.3 inch Monochrome Digital Mammography 5M pixel LCD Monitor with sub-pixel driving techniques enabling 15M subpixels to be driven independently, MS55i2 (ML21055, MD211G5) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is to be used in digital mammography PACS and modalities including FFDM.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
\$\sinh p)\$
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> 210(k) K131137
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.