← Product Code [NRB](/productcode/NRB) · K131096

# AR1000 SERIES ULTRASONIC WOUND THERAPY SYSTEM (K131096)

_Arobella Medical, LLC · NRB · May 16, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K131096

## Device Facts

- **Applicant:** Arobella Medical, LLC
- **Product Code:** [NRB](/productcode/NRB.md)
- **Decision Date:** May 16, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4410
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing via: - Selective and non-selective dissection and fragmentation of soft and or hard tissue; - Surgical, excisional or sharp-edge wound debridement (acute and chronic wounds, burns) for the removal of nonviable tissue including but not limited to diseased tissue, necrotic tissue, slough and eschar, fibrin, tissue exudates, bacteria and other matter. Patient population is patients of any age with one or more wounds. Patient population may also exhibit diabetes mellitus (DM). The AR1000 Series and its variants is intended for Prescription use. The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing via: - Site cleansing irrigation and lavage of wound tissue (acute and chronic wounds, burns, diseased or necrotic tissue); - Contact and or non-contact maintenance debridement for the removal of debris, exudates, fragments, bacteria, slough, fibrin, excised or fragmented tissue, and other matter. IRRIGATION (LAVAGE) FLUID - Irrigation fluid may be sterile de-ionized water, sterile saline solution, other approved wound therapy or debridement solution. Patient population is patients of any age with one or more wounds. Patient population may also exhibit diabetes mellitus (DM). The AR1000 Series and its variants is intended for Prescription use. The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing by: - Preparing the wound bed for graft or other subsequent procedures using contact and or non-contact techniques to achieve wound debridement. Patient population is patients of any age with one or more wounds. Patient population may also exhibit diabetes mellitus (DM). The AR1000 Series and its variants is intended for Prescription use.

## Device Story

AR1000 Ultrasonic Wound Therapy System uses low-frequency (35kHz) ultrasonic energy to promote wound healing. System comprises a generator, handheld converter, and distal probe (Qurette). Generator converts wall power to ultrasonic signal; converter transforms signal into mechanical displacement at probe tip. Energy applied via direct contact or non-contact techniques, often coupled with liquid medium (e.g., sterile saline) to lavage wound bed. Used in hospital, clinic, home, or intra-operative settings by clinicians. Ultrasonic energy facilitates selective/non-selective tissue dissection, fragmentation, and debridement of necrotic tissue, slough, eschar, and bacteria. Cleansing irrigation removes debris and exudates. Clinicians use device to prepare wound beds for grafts or subsequent procedures. Benefits include efficient removal of nonviable tissue and wound bed preparation, potentially accelerating healing. Device is prescription-only.

## Clinical Evidence

Bench testing only. Compliance with electrical safety (IEC 60601-1, IEC 60601-1-2, FCC Part 18), ultrasonic output (IEC 61847, 21 CFR 1050.10), and biocompatibility (ISO 10993-1) standards. No new clinical data presented; expansion of indications supported by existing data from predicate devices.

## Technological Characteristics

Generator: 85-260VAC, 50/60Hz. Frequency: 35kHz (±3kHz). Amplitude: 0-75um. Materials: Titanium alloy (TI-6AL-4V) for converter horn and probe tips. Delivery: Contact/non-contact probe with liquid coupling medium. Sterilization: Steam (autoclave) for converter/tips. Connectivity: Standalone. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, FCC Part 18, 21 CFR 1050.10, IEC 61847, ISO 10993-1.

## Regulatory Identification

A low energy ultrasound wound cleaner is a device that uses ultrasound energy to vaporize a solution and generate a mist that is used for the cleaning and maintenance debridement of wounds. Low levels of ultrasound energy may be carried to the wound by the saline mist.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Low Energy Ultrasound Wound Cleaner.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- AR1000 Ultrasonic Wound Therapy System ([K062544](/device/K062544.md))
- SonicOne Ultrasonic Wound Debridement System ([K112782](/device/K112782.md))
- Celleration Mist Therapy System ([K122246](/device/K122246.md))
- SONOCA 180 / 185 Wound Care System ([K072904](/device/K072904.md))

## Reference Devices

- Ethicon ([K002906](/device/K002906.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## MAY 1 6 2014

#### 510(k) Summary (K131096)

In accordance with the requirements of 21 CFR 807.92, Arobella Medical is submitting the following 510(k) summary for the AR1000 Ultrasonic Wound Therapy System and its variants. (Reference previous 510(k) K062544)

#### 807.92(a)(1)

#### Submitter Information

Arobella Medical LLC 5929 Baker Road, Suite 470 Minnetonka, MN 55345 Main: (952) 345-6840 Fax: (952) 345-6841

Contact Person: Matthew Berscheid Director of Quality and Regulatory Affairs Office: (952) 288-2801 Fax: (952) 345-6841 Cell (612) 998-7416

Date: May 16, 2014

#### 807.92(a)(2)

Trade Names: AR1000/B Ultrasonic Wound Therapy System, AR1000/B Qoustic Wound Therapy System™ (QWTS), or Qoustic Choice Wound Therapy System Common Name: Ultrasonic Wound Therapy System Classification Name(s): Ultrasonic Surgical Instrument; Wound Cleaner, Ultrasound Product Classification: Class II or in the alternate Unclassified Product Code: LFL

#### 807.92(a)(3)

#### Predicate Device(s)

| Arobella Medical, LLC | AR1000 Ultrasonic Wound Therapy System       | K062544 |
|-----------------------|----------------------------------------------|---------|
| Misonix, Inc          | SonicOne Ultrasonic Wound Debridement System | K112782 |
| Celleration           | Celleration Mist Therapy System              | K122246 |
| Soring Gmbh           | SONOCA 180 / 185 Wound Care System           | K072904 |

{1}------------------------------------------------

#### 807.92(a)(4)

#### Device Description

The AR1000 Ultrasonic Wound Therapy System and its variants use ultrasonic energy from a generator delivered by a conductive cable to a hand held converter through the converter's distal end probe and liquid coupling medium: to perform wound therapy; to perform ultrasonic surgical, excisional, and or sharp-edge selective dissection procedures; with the intended end result to promote healing.

The AR1000 Ultrasonic Wound Therapy System generator utilizes commonly available wall outlet power (85 -260VAC, 50/60Hz) to produce the ultrasonic signal. A conductive cable connects the generator to the hand held converter. The hand held converter converts the signal into ultrasonic mechanical displacement in the distal end probe or Qurette. The ultrasonic mechanical displacement energy is applied to the treatment site by the distal end probe or Qurette through direct contact or noncontact techniques. The ultrasonic mechanical displacement energy is also transmitted as ultrasonic energy to the treatment site via the liquid coupling medium.

The AR1000 Ultrasonic Wound Therapy System distal end probe or Qurette is used with contact or non-contact modes to achieve intended wound therapy modalities to promote wound healing. This is achieved through the use of the low-frequency, controlled-intensity, ultrasonic energy to perform ultrasonic dissection and fragmentation of tissues, debridement of wound tissue (acute and chronic wounds, burns, diseased or necrotic tissue), and surgical, excisional or sharp-edge incisions.

The AR1000 Ultrasonic Wound Therapy System distal end probe or Qurette is used with contact or non-contact modes to achieve wound therapy and promotes wound healing through the ultrasonic dissection and fragmentation of tissues, debridement of wound tissue (acute and chronic wounds, burns, diseased or necrotic tissue) through the use of the low-frequency, controlled-intensity, non-thermal ultrasonic energy.

The AR1000 Ultrasonic Wound Therapy System ultrasonic energy, via the liquid coupling medium to the treatment site, promotes wound healing through the lavage of wound tissue (acute and chronic wounds, burns, diseased or necrotic tissue) and cleansing irrigation of the site for the removal of debris, exudates, fragments, bacteria, and other matter.

The AR1000 Ultrasonic Wound Therapy System uses continuous ultrasonic energy to propel the liquid medium (e.g. sterile saline, or other appropriate US FDA cleared medium) into a solution stream and couples the delivery of ultrasonic energy to the treatment site. This is accomplished by the ultrasonic energy application to the

Section 005 510(k) Summary

Page 005-2 of 005-13

{2}------------------------------------------------

treatment site where the ultrasonic energy is transmitted via the converter probe (Qurette) and also coupled via the liquid medium to the treatment site.

The converter has a transducer horn that is made from titanium alloy (TI-6AL-4V). The converter probe Qurette tips are made from titanium alloy (TI-6AL-4V),

No latex is used in this device, including any potential patient-contact areas, and testing has revealed no negative reactions to the materials used in this device.

Specifications:

| Generator Power Source                  | 85 -260V~ (VAC), 2.5A, 50/60Hz                                        |
|-----------------------------------------|-----------------------------------------------------------------------|
| Converter Power Output                  | <140 watts                                                            |
| Amplitude (adjustable)                  | 0um ~ 75um (0% to 100% power level)                                   |
| Frequency                               | 35kHz (±3kHz)                                                         |
| Weight                                  | Generator: ~12.5 lbs<br>Handheld Converter: ~1.0 lbs                  |
| Approximate Dimensions<br>inches Height | System 15 inches Deep x 12 inches Wide x 6                            |
| diameter                                | Handheld Converter: 9 inches x 1.3 inches                             |
| Storage Temperature<br>90%              | -20 to +50° C at a maximum relative humidity of<br>(non-condensing)   |
| Operating Temperature                   | 0 to +40° C at a maximum relative humidity of 90%<br>(non-condensing) |

{3}------------------------------------------------

#### 807 92(a)(5)

#### Intended Use(s)

The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing via:

- Selective and non-selective dissection and fragmentation of soft and or hard . tissue:
- Surgical, excisional or sharp-edge wound debridement (acute and chronic . wounds, burns) for the removal of nonviable tissue including but not limited to diseased tissue, necrotic tissue, slough and eschar, fibrin, tissue exudates, bacteria and other matter.

Patient population is patients of any age with one or more wounds. Patient population may also exhibit diabetes mellitus (DM).

The AR1000 Series and its variants is intended for Prescription use.

The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing via:

- Site cleansing irrigation and lavage of wound tissue (acute and chronic . wounds, burns, diseased or necrotic tissue);
- Contact and or non-contact maintenance debridement for the removal of ● debris, exudates, fragments, bacteria, slough, fibrin, excised or fragmented . tissue, and other matter.

#### IRRIGATION (LAVAGE) FLUID

- Irrigation fluid may be sterile de-ionized water, sterile saline solution, other . approved wound therapy or debridement solution.
Patient population is patients of any age with one or more wounds. Patient population may also exhibit diabetes mellitus (DM).

The AR1000 Series and its variants is intended for Prescription use.

{4}------------------------------------------------

The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing by:

- Preparing the wound bed for graft or other subsequent procedures using . contact and or non-contact techniques to achieve wound debridement.
Patient population is patients of any age with one or more wounds. Patient population may also exhibit diabetes mellitus (DM).

The AR1000 Series and its variants is intended for Prescription use.

#### Section 005 510(k) Summary

{5}------------------------------------------------

807.92(Δ)δ)
There have been no changes to the AR1000 Ultrasonic Wound Therapy System deared under 510K) K062644 a a
result title proposed dibeling changes.
The following Tabl

The following Tables illustrate similarities and differences between the submitted AR1000 Ultrasonic Wound Therapy
System and the predicate devices.

ection 005_510(k) Summar

{6}------------------------------------------------

....... ﺎﻟﻠﻐﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ L ifi F .
Innia t T ---C 141 T.

|                                                                                                                   | Device | 510(k) | Indications<br>for Use<br>per<br>indicated<br>US FDA<br>510(k) | Arobella AR1000 (Proposed<br>K131096)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Arobella AR1000                                                                                                                                                                                                                                                             | Misonix SonicOne                                                                                                                                                                                                                                                                                                                                                                                    | Celleration Mist                                                                                                                                                                                                                                                  |
|-------------------------------------------------------------------------------------------------------------------|--------|--------|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Table 1: Comparison of Technical Characteristics Table - Specified Indications For Use with Identified Predicates |        |        |                                                                | The AR1000 Series and its<br>variants produce and deliver low<br>frequency ultrasound used to<br>promote wound healing via:<br>• Selective and non-selective<br>dissection and fragmentation of<br>soft and or hard tissue;<br>• Surgical, excisional or sharp-<br>edge wound debridement<br>(acute and chronic wounds,<br>burns) for the removal of<br>nonviable tissue including but<br>not limited to diseased tissue,<br>necrotic tissue, slough and<br>eschar, fibrin, tissue exudates,<br>bacteria and other matter. | Selective dissection<br>and fragmentation of<br>tissue, wound<br>debridement (acute<br>and chronic wounds,<br>burns, diseased or<br>necrotic tissue), and<br>cleansing irrigation of<br>the site for the removal<br>of debris, exudates,<br>fragments, and other<br>matter. | The Misonix SonicOne is<br>indicated for use in the<br>debridement of wounds,<br>such as, but not limited<br>to: burn wounds, diabetic<br>ulcers, bedsores and<br>vaginal ulcers, soft tissue<br>debridement and<br>cleansing of the surgical<br>site in applications in<br>which, in the physician's<br>judgment would require<br>the use of an ultrasonic<br>aspirator with sharp<br>debridement. | The MIST Therapy<br>system produces a low<br>energy<br>ultrasound generated<br>mist used to promote<br>wound healing through<br>wound cleansing and<br>maintenance<br>debridement by the<br>removal of yellow<br>slough, fibrin, tissue<br>exudates and bacteria. |

Page 005-7 of 005-13

.

Section 005_510(k) Summar

{7}------------------------------------------------

| Device | 510(k) | Indications<br>for Use<br>per<br>indicated<br>US FDA<br>510(k) | Arobella AR1000 (Proposed<br>K131096)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Arobella<br>AR1000<br>K062544                                                                                                                                                                                                                                                                    | Misonix SonicOne                                                                                                                                                                                                                                                                                                                                                                                                      | K112782 | Celleration Mist                                                                                                                                                                                                                                                                       | K122246 | Soring Sonoca                                                                                                                                                                                                                                                                                                                                                                    | K072904 |
|--------|--------|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
|        |        |                                                                | The AR1000 Series and its<br>variants produce and deliver<br>low frequency ultrasound<br>used to promote wound<br>healing via:                                                                                                                                                                                                                                                                                                                                                                                                | Selective<br>dissection and<br>fragmentation<br>of tissue,<br>wound<br>debridement<br>(acute and<br>chronic<br>wounds, burns,<br>diseased or<br>necrotic tissue),<br>and cleansing<br>irrigation of the<br>site for the<br>removal of<br>debris,<br>exudates,<br>fragments, and<br>other matter. | The Misonix<br>SonicOne is<br>indicated for use in<br>the debridement of<br>wounds, such as,<br>but not limited to:<br>burn wounds,<br>diabetic ulcers,<br>bedsores and<br>vaginal ulcers, soft<br>tissue debridement<br>and cleansing of<br>the surgical site in<br>applications in<br>which, in the<br>physician's<br>judgment would<br>require the use of<br>an ultrasonic<br>aspirator with sharp<br>debridement. |         | The MIST<br>Therapy<br>system<br>produces a low<br>energy<br>ultrasound<br>generated mist<br>used to<br>promote wound<br>healing through<br>wound<br>cleansing and<br>maintenance<br>debridement by<br>the removal of<br>yellow slough,<br>fibrin, tissue<br>exudates and<br>bacteria. |         | The SONOCA 180<br>/ 185 Wound Care<br>system is an<br>instrument<br>intended for<br>selected<br>ultrasound<br>dissection and<br>fragmentation of<br>tissue, wound<br>debridement<br>(acute and chronic<br>wounds, burns,<br>diseased or<br>necrotic tissue)<br>and cleansing<br>irrigation of the site<br>for removal of<br>debris, exudates,<br>fragments, and<br>other matter. |         |
|        |        |                                                                | • Site cleansing irrigation<br>and lavage of wound tissue<br>(acute and chronic<br>wounds, burns, diseased<br>or necrotic tissue);<br><br>• Contact and non-contact<br>maintenance debridement<br>for the removal of debris,<br>exudates, fragments,<br>bacteria, slough, fibrin,<br>excised or fragmented<br>tissue, and other matter.<br><br>IRRIGATION (LAVAGE)<br>FLUID<br>• Irrigation fluid may be<br>sterile de-ionized water,<br>sterile saline solution, other<br>approved wound therapy<br>or debridement solution. |                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                       |         |                                                                                                                                                                                                                                                                                        |         |                                                                                                                                                                                                                                                                                                                                                                                  |         |

·

Section 005_510(k) Summar

{8}------------------------------------------------

Table 3: Comparison of Technical Characteristics Table -- Specified Indications For Use with Identified Predicates

| Device | 510(k) | Indications for Use per indicated US FDA 510(k)                                                                                                                                                                                                                       | Arobella AR1000 (Proposed K131096)                                                                                                                                                                                                         | Arobella AR1000<br>K062544                                                                                                                                                                                                                                                                                                                                | Misonix SonicOne<br>K112782                                                                                                                                                                                                      | Celleration Mist<br>K122246                                                                                                                                                                                                                                                                                          | Soring Sonoca<br>K072904 |
|--------|--------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
|        |        | The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing by:<br><br>• Preparing the wound bed for graft or other subsequent procedures using contact and or non-contact techniques to achieve wound debridement. | Selective dissection and fragmentation of tissue, wound debridement (acute and chronic wounds, burns, diseased or necrotic tissue), and cleansing irrigation of the site for the removal of debris, exudates, fragments, and other matter. | The Misonix SonicOne is indicated for use in the debridement of wounds, such as, but not limited to: burn wounds, diabetic ulcers, bedsores and vaginal ulcers, soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgment would require the use of an ultrasonic aspirator with sharp debridement. | The MIST Therapy system produces a low energy ultrasound generated mist used to promote wound healing through wound cleansing and maintenance debridement by the removal of yellow slough, fibrin, tissue exudates and bacteria. | The SONOCA 180 / 185 Wound Care system is an instrument intended for selected ultrasound dissection and fragmentation of tissue, wound debridement (acute and chronic wounds, burns, diseased or necrotic tissue) and cleansing irrigation of the site for removal of debris, exudates, fragments, and other matter. |                          |

Page 005-9 of 005-13

Section 005_510(k) Summar

{9}------------------------------------------------

Table 4: Comparison of Technical Characteristics Table continued

| Device             | Arobella AR1000                                                                                                                   | Arobella AR1000                                                                                                                   | Misonix SonicOne                                                                          | Celleration Mist                                                                            | Soring Sonoca                                                                                   |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| 510(k) number      | This submission<br>K131096                                                                                                        | K062544                                                                                                                           | K112782                                                                                   | K122246                                                                                     | K072904                                                                                         |
| Device Agent       | High Intensity<br>Ultrasound and<br>irrigation 35KHz                                                                              | High Intensity<br>Ultrasound and<br>irrigation 35KHz                                                                              | High Intensity<br>Ultrasound 22.5KHz                                                      | High Intensity<br>Ultrasound<br>40KHz                                                       | High Intensity<br>Ultrasound<br>40KHz                                                           |
| Agent Delivery     | Irrigation, atomized<br>irrigation, contact, and<br>non-contact probe                                                             | Irrigation, atomized<br>irrigation, contact, and<br>non-contact probe                                                             | Contact probe                                                                             | Atomized irrigant<br>and Non-contact<br>probe                                               | Contact probe                                                                                   |
| Agent Mechanism    | Ultrasound for<br>selective tissue<br>dissection,<br>fragmentation, and<br>irrigation of wound bed<br>over course of<br>treatment | Ultrasound for<br>selective tissue<br>dissection,<br>fragmentation, and<br>irrigation of wound bed<br>over course of<br>treatment | General cavitation<br>(thermal agitation<br>and fragmentation)<br>of wounds and<br>tissue | Ultrasound<br>wound cleansing<br>and maintenance<br>debridement                             | General<br>cavitation<br>(thermal agitation<br>and<br>fragmentation) of<br>wounds and<br>tissue |
| Treatment Variable | Applicator size,<br>positioning, contact<br>surface, frequency<br>and duration of<br>treatments, irrigation<br>flow rate          | Applicator size,<br>positioning, contact<br>surface, frequency<br>and duration of<br>treatments, irrigation<br>flow rate          | Applicator size,<br>positioning,<br>frequency and<br>duration of<br>treatments            | Applicator<br>positioning,<br>frequency and<br>duration of<br>treatments, Mist<br>flow rate | Applicator size,<br>positioning,<br>frequency and<br>duration of<br>treatments                  |
| Treatment Setting  | Hospital, clinic, in-<br>home, or intra-<br>operative settings                                                                    | Hospital, clinic, in-<br>home, or intra-<br>operative settings                                                                    | Hospital, clinic, in-<br>home, or intra-<br>operative settings                            | Hospital, clinic,<br>or in-home<br>settings                                                 | Intra-operative<br>settings                                                                     |

Page 005-10 of 005-13

Section 005_510(k) Summar

{10}------------------------------------------------

#### 807.92(b)

#### Safety and Effectiveness

There have been no changes to the AR1000 Ultrasonic Wound Therapy System cleared under 510(k) K062544 as a result of the proposed labeling changes.

This 510(k) submission supports an expansion of the device's Indications For Use. with respect to K062544, to incorporate expanded patient results and laboratory findings.

The expansion of the device's Indications For Use, with respect to K062544, do not raise any new issues of safety or effectiveness.

For safety and effectiveness purposes the AR1000 Ultrasound Wound Therapy System and its variants is substantially equivalent to the previously cleared AR1000 Ultrasound Wound Therapy System (K062544), the Misonix SonicOne (K112782) system, Celleration MIST (K122246) system, the Soring Sonoca (K072904) systems, and the Ethicon (K002906) system.

#### 1.0 Efficacy

This device, like the predicates, is directed for single (1) treatment, and total recommended treatments shall be no more than three (3) times per week or as directed on the order of the physician. A typical treatment regimen is approximately two (2) weeks or more, depending on the size and nature and or condition of the treated area as well as other subject dependent factors, but may be prescribed for more or less frequent use and or for shorter or longer treatment regimen durations on the order of a physician.

This device also compares with the predicates, as follows:

1.1. Controlled ultrasonic delivery with 0.9% sterile saline solution, or other approved medium, to lavage the wound tissue (e.g. acute and chronic wounds, burns, diseased or necrotic tissue), like the Arobella AR1000 (K062544) system and the Celleration (K122246) predicate ultrasonic wound system.

1.2. Maintenance debridement of the treatment site for the removal of loose fragments, debris, exudate, bacteria, and other matter, like Arobella AR1000 (K062544) system, the Misonix SonicOne (K112782) system, the Celleration (K122246) system, and the Soring Sonoca (K072904) predicate ultrasonic wound system.

Section 005 510(k) Summary

Page 005-11 of 005-13

{11}------------------------------------------------

1.3. Ultrasonic selective dissection and debridement of tissue through the use of the probe (aka Qurette), including parallel to the wound bed surface for shallow separation of tissue lavers, like the predicate Arobella AR1000 (K062544) system and the Misonix SonicOne (K112782) system.

#### 2.0 Safety

The AR1000 Ultrasonic Wound Therapy System and its variants consists of an ultrasonic power supply (hereto after referred to as a generator), an ultrasonic converter hand piece (or just converter), and distal end probe tip (Qurette), a liquid supply reservoir and delivery tubing. Substantial equivalence has been previously established between the AR1000 Ultrasonic Wound Therapy System (K062544) and the Celleration Mist, and the Misonix SonicOne predicate ultrasonic wound cleanser devices as outlined under K062544.

#### 2.1 Electrical Safety & Electromagnetic Compatibility

Similar to the relevant predicates, this device complies with the following standards:

· IEC 60601-1 Medical electrical equipment, Part 1: General requirements for basic safety and essential performance

· UL 60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety

• IEC 60601-1-2 Medical electrical equipment, Part 1-2: Electromagnetic compatibility

- Requirements and tests
· IEC 60601-1-4 Medical electrical equipment, Part 1-4: Programmable electrical medical svstems

· FCC Part 18 - EMC Requirements

#### 2.2 Ultrasonic Safety

Similar to the relevant predicates, this device complies with the following standards:

- · 21 CFR 1050.10 Ultrasonic therapy products
- · IEC 61847 Ultrasonics Surgical systems Measurement and declaration of the basic output characteristics

Section 005 510(k) Summary

{12}------------------------------------------------

#### 2.3 Biocompatibility

Similar to the relevant predicates, this device was tested for and complies with the standard ISO 10993-1:2003 - Biological evaluation of medical devices, Part 1: Evaluation and testing and its relevant annexes (based on nature and duration of patient contact). All testing was conducted in compliance to 21 CFR. Part 58 Good Laboratory Practice for Nonclinical Laboratory Studies.

The AR1000 system is routinely used with 0.9% sterile saline solution, which is widely available for purchase from medical suppliers in bottle or bag form and is to be supplied by the user, as well as the tubing and the valve (for saline flow control). In the alternate, the AR1000 can also be used with other appropriate sterile mediums and US FDA-approved irrigation liquids. This is substantially equivalent to all of the predicate devices and irrigation solutions.

#### 2.4 Transportation/Storage

Packaging systems for the non-sterile device and for the non-sterile accessories of the device (e.g. the converter, tips, shroud, et al) keep the product without deterioration and without a specific level of sterility stipulated, as do the predicate devices. Those accessories used specifically for treatment (e.g. converter, tip, shroud, etc) are to be sterilized prior to use to minimize the risk of microbial contamination, as established with the AR1000 K062544 device clearance.

As the device and the accessories are durable goods with little or no notable deterioration, the shelf life is the product lifecycle, barring storage or transportation damage, and/or use under extremes of environmental conditions (e.g. temperature, humidity, moisture, etc).

#### 2.5 Reprocessing

The AR1000 system, as well as the AR1000 ultrasonic converter hand piece, is the wound debridement device cleared under K062544 for which reprocessing instructions have been previously validated and cleared.

The variant AR1000 ultrasonic converter hand piece allows for steam sterilization (autoclave) in addition to the previously cleared and validated reprocessing methodology. This variant may therefore be reprocessed by steam sterilization (autoclave), as well as the previously cleared and validated reprocessing methodology.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. The logo is presented in black and white.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 16, 2014

Arobella Medical LLC Matthew Berscheid Director of Quality and Regulatory Affairs 5929 Baker Road, Suite 470 Minnetonka, MN 55345

Re: K131096

Trade/Device Name: AR1000 Ultrasonic Wound Therapy System Regulation Number: 21 CFR 878.4410 Regulatory Class: Class II Product Code: NRB Dated: April 03, 2014 Received: April 04, 2014

Dear Mr. Berscheid:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{14}------------------------------------------------

#### Page 2 - Matthew Berscheid

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

Binita Ashar, MD, MBA, FACS for Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

cc: DMC 510(k) Staff Division D.O.

{15}------------------------------------------------

# Indications for Use

510(k) Number (if known): _K131096_

Device Name: AR1000 Series Ultrasonic Wound Therapy System

Indications For Use:

The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing via:

- Selective and non-selective dissection and fragmentation of soft and or hard . tissue;
- Surgical, excisional or sharp-edge wound debridement (acute and chronic ● wounds, burns) for the removal of nonviable tissue including but not limited to diseased tissue, necrotic tissue, slough and eschar, fibrin, tissue exudates, bacteria and other matter.

Patient population is patients of any age with one or more wounds. Patient population may also exhibit diabetes mellitus (DM).

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

# Jiyoung Dang -S

Division Sign-Off Office of Device Evaluation

510(k) K131096

Page 1 of 3

{16}------------------------------------------------

# Indications for Use

510(k) Number (if known): K131096

Device Name: AR1000 Series Ultrasonic Wound Therapy System

Indications For Use:

The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing via:

- Site cleansing irrigation and lavage of wound tissue (acute and chronic wounds, . burns, diseased or necrotic tissue);
- Contact and or non-contact maintenance debridement for the removal of debris, . exudates, fragments, bacteria, slough, fibrin, excised or fragmented tissue, and other matter.

#### IRRIGATION (LAVAGE) FLUID

- Irrigation fluid may be sterile de-ionized water, sterile saline solution, other . approved wound therapy or debridement solution.
Patient population is patients of any age with one or more wounds. Patient population may also exhibit diabetes mellitus (DM).

× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

# Jiyoung Dang -S

Division Sign-Off Office of Device Evaluation

510(k)__K131096

Page 2 of 3

{17}------------------------------------------------

## Indications for Use

510(k) Number (if known): K131096

Device Name: AR1000 Series Ultrasonic Wound Therapy System

Indications For Use:

The AR1000 Series and its variants produce and deliver low frequency ultrasound used to promote wound healing by:

- Preparing the wound bed for graft or other subsequent procedures using contact . and or non-contact techniques to achieve wound debridement.
Patient population is patients of any age with one or more wounds. Patient population may also exhibit diabetes mellitus (DM).

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

# Jiyoung Dang -S

Division Sign-Off Office of Device Evaluation

510(k)_K131096__

Page 3 of 3

---

**Source:** [https://fda-staging.innolitics.com/device/K131096](https://fda-staging.innolitics.com/device/K131096)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).