K131082 · Next Orthosurgical · MAX · Aug 26, 2013 · Orthopedic
Device Facts
Record ID
K131082
Device Name
INTERFORM INTERBODY CAGE SYSTEM
Applicant
Next Orthosurgical
Product Code
MAX · Orthopedic
Decision Date
Aug 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as an intervertebral body fusion device, the InterForm Lumbar Interbody Cage System is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft to facilitate fusion. Additionally, the InterForm Lumbar Interbody Cage System is intended for use with supplemental spinal fixation systems cleared for use in the lumbar spine. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.
Device Story
InterForm Interbody Cage System consists of PEEK implants in various footprints, lengths, and heights with lordotic angles (0°, 5°, 7°, 12°). Designed for lumbar/lumbosacral interbody fusion via PLIF, TLIF, OTLIF, LLIF, or ALIF approaches. Hollow geometry allows packing with autogenous bone graft to facilitate fusion. Implants provide structural support and maintain disc space until fusion occurs. Used in clinical settings by surgeons; requires supplemental spinal fixation. Packaged non-sterile for hospital sterilization. Benefits include spinal stabilization and correction in patients with degenerative disc disease.
Clinical Evidence
No clinical studies were performed. Evidence is based on bench testing only, including static and dynamic compression (ASTM F2077), subsidence (ASTM F2267), and expulsion testing.
Technological Characteristics
Implants manufactured from PEEK Optima LT1 (ASTM F2026) and unalloyed tantalum (ASTM F560). Various footprints, lengths, and heights available with 0°, 5°, 7°, or 12° lordosis. Hollow geometry for bone graft. Non-sterile, hospital-sterilized. Mechanical performance validated via ASTM standards.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD defined as discogenic back pain with disc degeneration confirmed by history/radiography; may include Grade 1 spondylolisthesis or retrolisthesis. Requires six months of failed non-operative treatment. Used with autogenous bone graft and supplemental spinal fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
# 510(k) Summary for the InterForm Interbody Cage System
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the InterForm Interbody Cage System
#### 1. GENERAL INFORMATION
| Date Prepared: April 16, 2013 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: InterForm Interbody Cage System |
| Common Name: intervertebral body fusion device |
| Classification |
| Name: Intervertebral body fusion device - lumbar |
| Class: II |
| Product Code: MAX |
| CFR section: 21 CFR section 888.3080 |
| Device panel: Orthopedic |
| Legally Marketed: Spinal Elements, Lucent Straight Interbody Cage (K071724/K081968) |
| Predicate Device: DePuy, Brantigan I/F CAGE (P960025)<br>SEASPINE, Pacifica Cage (K082310)<br>DiFusion Technologies, XIPHOS Interbody Fusion System (K100042)<br>Synthes T-PAL (K100089)<br>Nuvasive CoRoent XL (K071795/ K081611)<br>Medtronic Capstone (K073291)<br>Synthes OPAL (K072791)<br>GS Medical ANYPLUS (K100516)<br>Medtronic Boomerang (K023570)<br>Synthes Oracle (K062933)<br>Spinal USA ShurFit (K092193)<br>Amedica Valeo (K091278) |
| Submitter: Next OrthoSurgical<br>3270 Corporate View, Suite A<br>Vista CA. 92081<br>760-295-3600 Tele<br>760-295-3610 Fax |
| Contact: J.D. Webb<br>1001 Oakwood Blvd<br>Round Rock, TX 78681<br>512-388-0199 Tele<br>512-692-3699 Fax<br>e-mail: idwebb@orthomedix.net |
111-2
{1}------------------------------------------------
#### 2. DEVICE DESCRIPTION
The InterForm PLIF Interbody Fusion Implants consists of cage footprint widths of 9, 11, 13mm , lengths of 20, 22, 24, 26, 30mm, ranging in height from 7mm to 14mm with lordosis of 5°, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm PLIF Interbody Fusion devices are intended to be implant via an open or minimally invasive, posterior approach and used with supplemental fixation and autogenous bone graft.
The InterForm TLIF Interbody Fusion Implants consists of cage footprints widths of 9, 11, 13mm , lengths of 22, 24, 26, 28, 30, 32mm, ranging in height from 7mm to 14mm with lordosis of 5°, which can be inserted between two lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm TLIF Interbody Fusion devices are intended to be implant via an open or minimally invasive, transforaminal approach and used with supplemental fixation and autogenous bone graft.
The InterForm OTLIF Interbody Fusion Implants consists of cage footprint widths of 9, 11, 13mm , lengths of 20, 22, 24, 26, 30, 32, 34mm, ranging in height from 7mm to 14mm with lordosis of 0°, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm OTLIF Interbody Fusion devices are intended to be implant via an open or minimally invasive, posterior or transforaminal approach and used with supplemental fixation and autogenous bone graft.
The InterForm LLIF Interbody Fusion Implants consists of cage footprint widths of 18, 20, 22, 24mm, lengths of 40, 45, 50, 55, 60mm, ranging in height from 8mm to 18mm with lordosis of 0° or 7°, which can be inserted between two lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm LLIF Interbody Fusion devices are intended to be implant via a lateral approach and used with supplemental fixation and autogenous bone graft.
The InterForm ALIF Interbody Fusion Implants consists of cage footprints of 30x24mm, and 39x30mm (DxW), ranging in height from 10mm to 20mm with lordosis of 0°, 7° or 12° which can be inserted between two lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm ALIF Interbody Fusion devices are intended to be implant via an open or minimally invasive, anterior approach and used with supplemental fixation and autogenous bone graft.
All implants are packaged non-sterile to be sterilized at the hospital.
#### Materials:
PEEK Optima LT1 conforming to ASTM F2026. Unalloyed tantalum conforming to ASTM F560.
{2}------------------------------------------------
# 3 of 3
# Function:
Maintain adequate disc space until fusion occurs.
#### 3. SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES
The InterForm Interbody Cage System is substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performances.
#### 4. INTENDED USE
When used as an intervertebral body fusion device, the InterForm Lumbar Interbody Cage System is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft to facilitate fusion. Additionally, the InterForm Lumbar Interbody Cage System is intended for use with supplemental spinal fixation systems cleared for use in the lumbar spine. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.
#### NON-CLINICAL TEST SUMMARY 5.
The following tests were conducted:
- . Static and dynamic compression per ASTM F2077
- Subsidence per ASTM F2267 .
- . Expulsion
The results of this testing indicate that the InterForm Interbody Cage System is equivalent to predicate devices.
## 6. CLINICAL TEST SUMMARY
No clinical studies were performed
## 7. CONCLUSIONS NONCLINICAL AND CLINICAL
Next OrthoSurgical considers the InterForm Interbody Cage System to be equivalent to the predicate device listed above. This conclusion is based upon the devices' similarities in principles of operation, technology, materials and indications for use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 26, 2013
Next OrthoSurgical % Mr. J.D. Webb The OrthoMedix Group. Incorporated 1001 Oakwood Boulevard Round Rock, Texas 78681
Re: K131082
Trade/Device Name: InterForm Lumbar Interbody Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 22, 2013 Received: July 31, 2013
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
# Page 2 - Mr. J.D. Webb
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Erin I. Keith
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known): _ K131082
Device Name: __InterForm Lumbar Interbody Cage System
Indications for Use:
When used as an intervertebral body fusion device, the InterForm Lumbar Interbody Cage System is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These palients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft to facilitate fusion. Additionally, the InterForm Lumbar Interbody Cage System is intended for use with supplemental spinal fixation systems cleared for use in the lumbar spine. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.