STAXX IBL SYSTEM

K131071 · Spine Wave, Inc. · MAX · Aug 20, 2013 · Orthopedic

Device Facts

Record IDK131071
Device NameSTAXX IBL SYSTEM
ApplicantSpine Wave, Inc.
Product CodeMAX · Orthopedic
Decision DateAug 20, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The StaXx® IBL System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-L5. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The StaXx® IBL System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of nonoperative treatment prior to treatment with an intervertebral body fusion device.

Device Story

StaXx® IBL System is an expandable intervertebral body fusion device; composed of stackable wafers; used for lumbar spinal fusion. Implant manufactured from PEEK-OPTIMA with 6% Barium Sulfate and tantalum markers. System includes delivery device for implantation and expansion. Used in surgical setting by physicians. Requires autogenous bone graft and supplemental fixation. Provides structural support to intervertebral space to facilitate fusion; benefits patients with discogenic back pain.

Clinical Evidence

Bench testing only. Mechanical performance demonstrated via static and dynamic axial compression (ASTM F2077), static and dynamic compression shear (ASTM F2077), and subsidence testing (ASTM F2267).

Technological Characteristics

Intervertebral body fusion device; stackable wafer design; expandable. Materials: PEEK-OPTIMA with 6% Barium Sulfate and tantalum markers. Mechanical testing per ASTM F2077 and ASTM F2267. Requires supplemental fixation and autogenous bone graft.

Indications for Use

Indicated for skeletally mature patients with lumbar degenerative disc disease (DDD) at one or two contiguous levels (L2-L5), including those with up to Grade I spondylolisthesis or retrolisthesis, who have failed at least 6 months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K131071 Page 1 of 2 AUG 2 0 2013 Image /page/0/Picture/1 description: The image shows the word "SPINEWAVE" in bold, sans-serif font. A curved line is present under the second half of the word, specifically under "WAVE". The text is black against a white background. # 510(k) Summary StaXx® IBL System ## 1. Submitter Information | Submitter: | Spine Wave, Inc. | |----------------|----------------------------------------------------------| | Address: | Three Enterprise Drive<br>Suite 210<br>Shelton, CT 06484 | | Telephone: | 203-712-1839 | | Telefax: | 203-944-9493 | | Contact: | Roaida Rizkallah | | Date Prepared: | July 12, 2013 | ## 2. Device Information | Trade Name: | StaXx® IBL System | |----------------------|------------------------------------------------------| | Common Name: | Intervertebral Body Fusion Device | | Classification: | Class II (special controls) per 21 CFR 888.3080 | | Classification Name: | Intervertebral Fusion Device with Bone Graft, Lumbar | | Product Code: | MAX | ## 3. Purpose of Submission The purpose of this submission is to gain clearance for additional sizes of implants utilizing the StaXx® technology for intervertebral body fusion. ### 4. Predicate Device Information The StaXx® IBL System described in this submission is substantially equivalent to the following predicates: | Predicate Device | Manufacturer | 510(k) No. | |------------------|----------------------|------------| | StaXx® IB System | Spine Wave, Inc. | K123461 | | Patriot® Spacer | Globus Medical, Inc. | K072970 | {1}------------------------------------------------ K131071 Page 2 of 2 ### 5. Device Description The StaXx® IBL System is an intervertebral body fusion device composed of wafers that are stacked into an expandable implant to adjust the implant. The implant. The implants are to be used with autogenous bone graft material. The implant components are manufactured from PEEK-OPTIMA with 6% Barium Sulfate and tantalum markers. The system also includes a delivery device to both implant and expand the implant. ## 6. Intended Use The StaXx® IBL System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-L5. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The StaXx® IBL System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of nonoperative treatment prior to treatment with an intervertebral body fusion device. ## 7. Comparison of Technological Characteristics The substantial equivalence of the StaXx® IBL System to the predicates is shown by similarity in intended use. indications for use, materials and performance. #### 8. Performance Data The following mechanical tests were performed to demonstrate the substantial equivalence of the StaXx® IBL System to its predicate: - t Static and dynamic axial compression (per ASTM F2077) - . Static and dynamic compression shear (per ASTM F2077) - Subsidence (per ASTM F2267) . ### 9. Conclusion Based on the indications for use, technological characteristics, performance testing and comparison to the predicates, the StaXx® IBL System has been shown to be substantially equivalent to the predicate devices identified in this submission, and does not present any new issues of safety or effectiveness. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 20, 2013 Roaida Rizkallah Regulatory Affairs Manager Spine Wave, Incorporated Three Enterprise Drive, Suite 210 Shelton, Connecticut 06484 Re: K131071 Trade/Device Name: StaXx® IBL system Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 12, 2013 Received: July 15, 2013 Dear Roaida Rizkallah: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ ## Page 2 - Roaida Rizkallah forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/deliault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use K131071 510(k) Number (if known): StaXx® IBL System Device Name: _____________________ Indications for Use: The StaXx® IBL System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-L5. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The StaXx" IBL System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device. Prescription Use X (Part 21 CFR 801 Subpart D) And / Or Over-The-Counter-Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anton E. Dmitriev, PhD Division of Orthopedic Devices Page _1_ of _1_
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...