K131060 · Pursuit Vascular, Inc. · PEH · Dec 16, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K131060
Device Name
CLEARGUARD HD END CAP
Applicant
Pursuit Vascular, Inc.
Product Code
PEH · Gastroenterology, Urology
Decision Date
Dec 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5540
Device Class
Class 2
Indications for Use
ClearGuard HD is indicated for use as an end cap for use with hemodialysis catheter hubs.
Device Story
ClearGuard HD is a male luer lock end cap for hemodialysis catheter hubs. Device consists of a copolyester polymer plug with a 1.8 cm rod and a nylon lock ring, both coated with chlorhexidine acetate (CHA). When inserted into a liquid-filled catheter, CHA elutes into the lock solution to reduce microbial colonization within the hub. Catheter pinch clamps mechanically confine the CHA solution, preventing diffusion toward the patient's bloodstream. Used in clinical settings by healthcare providers to maintain catheter hub hygiene. Benefits include reduced microbial colonization in the hub. Not for treating existing infections.
Clinical Evidence
No clinical data. Evidence based on in vitro testing demonstrating antimicrobial effectiveness against specific microorganisms (e.g., VRE, MRSA, MRSE, P. aeruginosa, C. albicans). Correlation between in vitro activity and clinical effectiveness has not been tested.
Technological Characteristics
Copolyester polymer plug with 1.8 cm rod; nylon lock ring. Both components coated with chlorhexidine acetate (CHA). Sterile, non-pyrogenic, single-use. Mechanical luer lock interface. No software or electronic components.
Indications for Use
Indicated for use as an end cap for hemodialysis catheter hubs. Not intended for treatment of existing infections. Antimicrobial effect limited to catheter hub; no migration to distal catheter portions.
Regulatory Classification
Identification
A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
Special Controls
*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Boston Scientific Corporation Vaxcel™ Plus Chronic Dialysis Catheter
Reference Devices
Arrow Antimicrobial Pressure Injectable PICC
Submission Summary (Full Text)
{0}------------------------------------------------
Pursuit Vascular, Inc. Traditional 510(k): K131060
i
:
## 510(k) Summary
### Summary of Safety and Effectiveness for the ClearGuard HD
## General Information
| Submitter<br>Pursuit Vascular, Inc.<br>6901 E. Fish Lake Road, Suite 166<br>Maple Grove, MN 55369 | Contact Person<br>Laurie Lynch, PhD<br>Director of Quality/Regulatory/Clinical<br>Phone: (612) 424-9006<br>Fax : (763) 592-8202<br>Email : llynch@pursuitvascular.com |
|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Provisions | |
| Trade Name | ClearGuard HD |
| Common/Usual Name | Hemodialysis Catheter Luer End Cap |
| Classification Name | 21 CFR 880.5540 |
| Predicate Devices | Boston Scientific Corporation Vaxcel™ Plus<br>Chronic Dialysis Catheter |
#### Device Description
The ClearGuard HD end cap is male lucr lock end cap that incorporates an antimicrobial treatment on its surfaces. It is designed to reduce microbial colonization within a hemodialysis catheter hub.
The ClearGuard HD end cap consists of 1) a copolyester polymer plug, which has a rod extending from the luer region that is coated with the antimicrobial agent chlorhexidine actate (CHA) and 2) a nylon lock ring with threads that are also coated with CHA. When a ClearGuard HD end cap is inserted into a liquid-filled catheter, CHA clutes into the catheter lock solution. This CHA solution is designed to reduce the number of microorganisms in the hemodialysis catheter hub.
The catheter extension line pinch clamps are used to maintain the lock solution within the catheter lumens and minimize the risk of air embolism. These clamps, which are closed when the catheter is not in use, mechanically confine the CHA and prevent diffusion of CHA toward the catheter tip and the patient's bloodstream.
#### Intended Use / Indications for Use
ClearGuard HD is indicated for use as an end cap for use with hemodialysis catheter hubs.
Pursuit Vascular Confidential Information
DEC 16 2013
{1}------------------------------------------------
| Pursuit Vascular, Inc. | Page 2 of 2 |
|-----------------------------|-------------------|
| Traditional 510(k): K131060 | December 10, 2013 |
K131060 Page 2 of 2
The antimicrobial treatment on the ClearGuard HD has been shown to be effective at reducing microbial colonization in hemodialysis catheter hubs against the following microorganisms: Enterococcus faecium (VRE), Enterococcus faecalis (VRE), Acinetobacter baumannii, Escherichia coli, Staphylococcus aureus (MRSA), Staphylococcus aureus, Staphylococcus epidermidis (MRSE), Pseudomonas aeruginosa, Candida albicans and Candida parapsilosis and has not been shown to be effective against Candida paratropicalis and Klebsiella pneumoniae.
The antimicrobial effectiveness was evaluated using in vitro methods and correlation between in vitro antibacterial activity and any clinical effectiveness has not been tested. It is not intended to be used for the treatment of existing infections. The antimicrobial is only effective within the hub of the catheter, and does not migrate to distal portions of the catheter.
#### Substantial Equivalence Comparison
The ClearGuard HD end cap is substantially equivalent to a legally marketed 510(k) cleared device in intended use, indications for use, principles of operation, hub compatibility, and the fact that they are both provided as sterile, non-pyrogenic, single use devices. The predicate device is a chronic hemodialysis catheter that contains end caps. The technological differences between ClearGuard HD end caps and the end caps of the predicate device include 1) the use of an antimicrobial agent on the end cap and 2) a rod extending 1.8 cm (0.7 in) from the luer region. The antimicrobial agent on the ClearGuard HD end cap is the same agent used on the Arrow Antimicrobial Pressure Injectable PICC reference device. This reference device has two versions of chlorhexidine applied to the catheter surfaces, chlorhexidine base and chlorhexidine acctate. The ClearGuard HD end cap has chlorhexidine acetate applied to the cap surfaces.
Results of design verification and validation testing demonstrate that the ClearGuard HD end cap is safe for use with hemodialysis catheters and reliably seals catheter hubs. It also demonstrates that the chlorhexidine antimicrobial agent effectively reduces the number of microorganisms in hemodialysis catheter hubs as does the reference Arrow PICC and introduces no new or increased safety concerns. The risk assessment results, together with the results of design verification and validation testing presented in this submission, confirm that the ClearGuard HD end cap raises no new questions of safety or effectiveness compared to the predicate and reference devices. The ClearGuard HD end cap has, therefore, been shown to be substantially equivalent to a legally marketed device for the purpose of 510(k) clearance.
#### Summary of Non-Clinical Testing
The ClearGuard HD end caps have been tested and compared to the predicate device. All data gathered demonstrate this device is substantially equivalent. No new or increased safety risks or efficacy issues have been raised.
#### Summary of Clinical Performance Data
Not applicable - no clinical data was needed to demonstrate substantial equivalence to its predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo of the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 16, 2013
Pursuit Vascular, Inc. Laurie E. Lynch, Ph.D. Director Quality/Regulatory/Clinical 6901 E. Fish Lake Road, Suite 166 Maple Grove, MN 55369
Re: K131060
> Trade/Device Name: ClearGuard HD End Cap Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: PEH Dated: December 5, 2013 Received: December 6, 2013
Dear Laurie E. Lynch,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Laurie E. Lynch, Ph.D.
... ... ..
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K131060
Device Name ClearGuard HD End Cap
Indications for Use (Describe)
ClearGuard HD is indicated for use as an end cap for use with hemodialysis catheter hubs.
The antimicrobial treatment on the ClearGuard HD has been shown to be effective at reducing microbial colonization in hemodialysis catheter hubs against the following microorganisms: Enterococcus foecolis (VRE), Acinetobocter baumannii, Escherichia coli, Staphylococus aureus, Staphylococus aureus, Staphylococcus epidermidis (MRSE), Pseudomonas aeruginosa, Candida albicans and Candidosis and has not been shown to be effective against Candida paratropicalis and Klebsiella pneumoniae.
The antimicrobial effectiveness was evaluated using in vitro methods, and correlation between in vitro antibacterial activity and any clinical effectiveness has not been tested. The subject device is not intended to be used for the treatment of existing infections. The antimicrobial is only effective within the hub of the catheter, and does not migrate to distal portions of the catheter.
Herbert P. Lerner
2013.12.13 16:19:25 -05:00'
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
PROPERT STORE FREE STATE AND A FOR FDATUSE ONLY CONLY CONTRACT AT S Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.