K131041 · Terumo Cardiovascular Systems Corp. · DTQ · Feb 20, 2014 · Cardiovascular
Device Facts
Record ID
K131041
Device Name
ADVANCED PERFUSION SYSTEM 1
Applicant
Terumo Cardiovascular Systems Corp.
Product Code
DTQ · Cardiovascular
Decision Date
Feb 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4220
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Terumo® Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The centrifugal pump is indicated for use in cardiopulmonary bypass procedures only.
Device Story
Advanced Perfusion System 1 is a configurable heart-lung machine used in clinical settings (e.g., OR) by perfusionists or qualified medical professionals. It features a distributed network architecture with a central control monitor and base chassis. Users customize the system by connecting various modules—including roller pumps (4" or 6"), centrifugal control units, air bubble detectors, level sensors, pressure/temperature/flow monitors, and gas blenders—to the base. The system provides power, data display, and control for these components. It monitors extracorporeal circuit parameters, generating alarms for conditions like backflow or air bubbles. The device supports cardiopulmonary bypass, arterial, and regional perfusion. By centralizing control and monitoring of perfusion components, it assists clinicians in managing blood circulation during surgery, aiming to maintain patient physiological stability.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including electromagnetic compatibility testing per IEC 60601-1 and IEC 60601-1-2, and software verification and validation testing for modified modules. Functionality testing was performed on new or modified components to ensure performance compatibility.
Technological Characteristics
Configurable heart-lung system with distributed network architecture. Components connect to a base chassis providing power (100/120V or 220/240V) and battery backup. Includes central touch screen monitor, roller pumps (4"/6"), centrifugal drive motors, and various monitoring modules (pressure, temperature, flow, air bubble, level). Connectivity via dedicated network points on the base. Complies with IEC 60601-1 and IEC 60601-1-2 standards.
Indications for Use
Indicated for patients requiring extracorporeal circulation of blood for arterial perfusion, regional perfusion, or cardiopulmonary bypass procedures for up to 6 hours. Centrifugal pump use restricted to cardiopulmonary bypass procedures. Must be operated by qualified medical professionals experienced with such equipment.
Regulatory Classification
Identification
A cardiopulmonary bypass heart-lung machine console is a device that consists of a control panel and the electrical power and control circuitry for a heart-lung machine. The console is designed to interface with the basic units used in a gas exchange system, including the pumps, oxygenator, and heat exchanger.
{0}------------------------------------------------
# Section 4: 510(k) Summary
This section includes a summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter Information | |
|---------------------------------------------------------------|-----------------------------------------------------------------------|
| Name | Terumo Cardiovascular Systems Corporation (TCVS) |
| Address | 6200 Jackson Road<br>Ann Arbor MI, 48103 |
| Name of Contact Person | Mark Bur |
| Phone number | Tel: (734) 663-4145 |
| Fax number | Fax: (734) 741-6069 |
| E-mail | Mark.Bur@terumomedical.com |
| Establishment Registration # | 1828100 |
| Date prepared | April 12, 2013 |
| Name of Device | |
| Trade or proprietary name | Advanced Perfusion System 1 |
| Common or usual name | Heart-Lung Machine |
| Classification name | Console, Heart-Lung Machine, Cardiopulmonary<br>Bypass |
| Classification panel | 74 Cardiovascular |
| Regulation | 21 CFR §870.4220 |
| Product Code(s) | DTQ |
| Legally marketed device(s) to<br>which equivalence is claimed | Advanced Perfusion System 1: K022947<br>APS1 6" Roller Pump: K112587 |
| Reason for 510(k) | Modifications to previously cleared system and<br>labeling changes |
{1}------------------------------------------------
Section 4: 510(k) Summary
# Device Information
Indication for Use: The Terumo® Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The centrifugal pump is indicated for use in cardiopulmonary bypass procedures only.
Device Description: The Advanced Perfusion System 1 is a configurable heart-lung system with a distributed network architecture that allows the user to customize the number and types of system components, which can then be configured, displayed, and controlled from a central monitor. The system is designed to enable users to choose from the TCVS supplied components to define and configure a heart-lung system to meet individual institution requirements.
The System 1 components are listed below.
- · System 1 Base:
- Chassis platform Provides operating power and back up battery power for all system components (100/120V or 220/240V)
- Central Control Monitor A touch screen display used for configuration o and control of system components
- Two roller pump hand cranks and hand crank bracket o
- · Pump(s) and pump mounting hardware -- Up to eight pumps can be used with System 1, including the following:
- 6" Roller Pump o
- 4" Roller Pump o
- Centrifugal Control Module with Centrifugal Drive Motor (up to 2) o
- · Modules
- Air Bubble Detection Module Used to detect air bubbles in the o extracorporeal circuit, in conjunction with the air sensor
- Level Detection Module Used to monitor liquid levels within a hard shell o reservoir.
- Pressure Module ~ Used to monitor the pressure in the extracorporeal O circuit
- Temperature Module Used to monitor the temperature in the o extracorporeal circuit and/or the patient
- Flowmeter Module Used to monitor flow volume and generate an o alarm if backflow is detected
Image /page/1/Picture/22 description: The image shows the logo for Terumo. The logo consists of a circle with a stylized "T" inside of it, followed by the word "TERUMO" in bold, sans-serif font. There is a small degree symbol to the right of the word.
{2}------------------------------------------------
# Section 4: 510(k) Summary
Image /page/2/Figure/3 description: This image shows a device illustration with several labeled components. The device includes a venous line occluder module, interface modules for data transfer, an electronic gas blender, flexible lamps, and mounting hardware. The illustration labels parts such as the centrifugal control unit, small and large roller pumps, a central control monitor, and an on/off switch.
Section 4 - Page 3 of 6
{3}------------------------------------------------
| Section 4: 510(k) Summary | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial Equivalence | | |
| The modified Advanced Perfusion System 1 is substantially equivalent to the currently cleared Advanced Perfusion System 1<br>because it has the same intended use, substantially equivalent indications for use, and the same or substantially equivalent<br>operating principles and technical specifications. The Advanced Perfusion System 1 has been modified to assure system<br>compliance with the IEC 60601-1-2 electromagnetic compatibility standard, and the Operators Manual has been revised to provide<br>additional warnings and revised instructions for use. | | |
| Item | Proposed Device<br>Modified Advanced Perfusion System 1 | Predicate Device<br>Advanced Perfusion System 1 (K022947) |
| Indication for Use | The Terumo® Advanced Perfusion System 1 is<br>indicated for use for up to 6 hours in the<br>extracorporeal circulation of blood for arterial<br>perfusion, regional perfusion, and cardiopulmonary<br>bypass procedures, when used by a qualified<br>medical professional who is experienced in the<br>operation of this or similar equipment. The<br>centrifugal pump is indicated for use in<br>cardiopulmonary bypass procedures only. | The Terumo® Advanced Perfusion System 1 is<br>indicated for use for up to 6 hours in the<br>extracorporeal circulation of blood for arterial<br>perfusion, regional perfusion, and cardiopulmonary<br>bypass procedures, when used by a qualified<br>medical professional who is experienced in the<br>operation of this or similar equipment. The<br>centrifugal pump is indicated for use in<br>cardiopulmonary bypass procedures only. |
| System<br>Components | • System 1 Base with Central Control Monitor and<br>various hardware accessories<br>• Large (6") and/or Small (4") Roller Pumps<br>• Centrifugal Control Unit and centrifugal drive<br>motor<br>• Modules and accessories, including:<br>o Air bubble detector<br>o Level sensor | • System 1 Base with Central Control Monitor and<br>various hardware accessories<br>• Large (6") and/or Small (4") Roller Pumps<br>• Centrifugal Control Unit and centrifugal drive<br>motor<br>• Modules and accessories, including:<br>o Air bubble detector<br>o Level sensor |
| tem | Proposed Device<br>Modified Advanced Perfusion System 1 | Predicate Device<br>Advanced Perfusion System 1 (K022947) |
| | Pressure monitor Temperature monitor Flow monitor Occluder Electronic Gas Blender Interface modules for external cardiac and data monitoring systems Lamp | Pressure monitor Temperature monitor Flow monitor Occluder Electronic Gas Blender Interface modules for external cardiac and data monitoring systems Lamp |
| Principles of Operation | Configurable heart-lung system with a distributed network architecture that allows the user to customize the number and types of system components, which can then be configured, displayed, and controlled from a central monitor.<br>Each component connects to the system network via points on the Base. There are six pump connections, two connections for the CCM, eighteen connections for modules, and two dedicated connections for a lamp. | Configurable heart-lung system with a distributed network architecture that allows the user to customize the number and types of system components, which can then be configured, displayed, and controlled from a central monitor.<br>Each component connects to the system network via points on the Base. There are six pump connections, two connections for the CCM, eighteen connections for modules, and two dedicated connections for a lamp. |
| Power Supply | The power system within the Base transforms AC power into the DC levels required by the system components. Integrated batteries provide backup power in the event of AC power loss during use and when power needed exceeds power available | The power system within the Base transforms AC power into the DC levels required by the system components. Integrated batteries provide backup power in the event of AC power loss during use and when power needed exceeds power available |
.
dvanced Perfusion System 1 510(k)
Terumo Cardiovascular Systems
{4}------------------------------------------------
rumo Cardiovascular System
# Section 4: 510(k) Summar
Section 4 - Page 5 of 6
TERUMO
February 18, 2014
{5}------------------------------------------------
Section 4: 510(k) Summary
## Performance Testing
Third party testing was conducted to confirm that, with the engineering modifications in place, the System 1 complies with IEC 60601-1 and IEC 60601-1-2.
Software verficiation and validation testing was conducted for those modules that have undergone software modification since initial clearance.
Moreover functionality testing was conducted on those new or modified components of the Advanced Perfusion System 1 to demonstrate performance compatibility.
All software and performance testing was successful.
### Conclusion
The modified Advanced Perfusion System 1 is substantially equivalent to the currently marketed Advanced Perfusion System 1 because it has the same intended use and substantially equivalent performance specifications.
Image /page/5/Picture/11 description: The image shows the Terumo logo. The logo consists of a circle with a stylized letter 'T' inside, followed by the word 'TERUMO' in bold, sans-serif font. A small circle is present at the upper right of the word.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized image of an eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 20, 2014
Terumo Cardiovascular Systems Corporation Mark Bur, Regulatory Affairs Manager 6200 Jackson Road Ann Arbor, MI 48103
Re: K131041
Trade Name: Advanced Perfusion System 1 Regulation Number: 21 CFR 870.4220 Regulation Name: CPB Heart Lung Machine Console Regulatory Class: Class II Product Code: DTQ Dated: January 27, 2014 Received: January 30, 2014
Dear Mr. Bur:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{7}------------------------------------------------
Page 2 - Mr. Mark Bur
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M. A. Freeman
for
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Terumo Cardiovascular Systems
Section 3: Indication for Use
510(k) Number: _ K131041
#### Device Name: Advanced Perfusion System 1
Indications for Use:
The Terumo® Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
The centrifugal pump is indicated for use in cardiopulmonary bypass procedures only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
. A. Hillman
April 12, 2013
**TERUMO**
Section 3 - Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.