← Product Code [LYZ](/productcode/LYZ) · K131033

# POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED) (K131033)

_Decent Plastic Co., Ltd. · LYZ · Feb 12, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K131033

## Device Facts

- **Applicant:** Decent Plastic Co., Ltd.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Feb 12, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Device is a non-sterile, disposable vinyl patient examination glove. Functions as a physical barrier against body fluids and bloodborne pathogens. Material is PVC rubber, which is water-tight and flexible to conform to the hand during medical procedures. Used in clinical settings by healthcare providers to prevent contamination between patient and examiner. No electronic or software components.

## Clinical Evidence

No clinical data provided; substantial equivalence is supported by bench testing including physical property testing (tensile strength, elongation), water leak testing (pinholes), and biocompatibility testing (skin irritation and sensitization) per ASTM and ISO standards.

## Technological Characteristics

Material: PVC with surface coating agent (PU). Standards: ASTM D5250-06 (Reapproved 2011), ASTM D5151-06 (Reapproved 2011), ASTM D6124-06 (Reapproved 2011), ISO 10993-10:2002/Amd.1:2006. Physical properties: Tensile strength ≥14MPa, Elongation ≥300%. Dimensions: 230mm length. Non-sterile. No energy source or software.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder free Vinyl Patient Examination Gloves, Clear (Non-colored), Shijiazhuang Fuguan Plastic Products Co., Ltd. ([K032908](/device/K032908.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Section C

## 510(k) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is:___ K131033__" (applicant leave blank)

Premarket Notification [510(k)] Summary

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared:

| Submitter's name :             | Decent Plastic Co., Ltd.                       |
|--------------------------------|------------------------------------------------|
| Submitter's address :          | No.44, Dajing St., Shijiazhuang, 050000, China |
| Phone number :                 | (86) 311-67699616                              |
| Fax number :                   | (86) 311-67699906                              |
| Name of contact person:        | Ms. Qi Fang                                    |
| Date the summary was prepared: | 2013-12-31                                     |

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

| Device Name:           | Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)                         |  |  |  |
|------------------------|-------------------------------------------------------------------------------------------|--|--|--|
|                        | Proprietary/Trade name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) |  |  |  |
| Common Name:           | Patient examination glove                                                                 |  |  |  |
| Classification Name:   | Patient examination glove                                                                 |  |  |  |
| Device Classification: |                                                                                           |  |  |  |
| Regulation Number:     | 21 CFR 880.6250                                                                           |  |  |  |
| Panel:                 | General Hospital (80)                                                                     |  |  |  |

Product Code: LYZ [(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) that meets all of the requirements of ASTM standard D 5250-06 (Reaffirmation 2011).

Predicate device: Powder free Vinyl Patient Examination Gloves, Clear (Non-colored), Shijiazhuang Fuguan Plastic Products Co., Ltd. K032908.

{1}------------------------------------------------

#### [(a)(4)| A description of the device

Device Description : Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011).

-- How the device functions:

PVC films form a barrier to body fluids and bloodborne Pathogens

-- Scientific concepts that form the basis for the device The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

-- Physical and performance characteristics such as design, materials and physical properties:

PVC gloves are known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

#### [(a)(5)| The summary describes the intended use of the device

Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or linger to prevent contamination between patient and examiner.

#### [{a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

| Features &<br>Description                                                                           | Predicate Device                                                                                                                                                                                                                                | Subject Device                                                                                                                                                                                                                                   | Result of<br>Comparison     |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Company                                                                                             | Shijiazhuang Fuguan Plastic<br>Products Co., Ltd.                                                                                                                                                                                               | Decent Plastic Co., Ltd.                                                                                                                                                                                                                         | --                          |
| 510(K) Number                                                                                       | K032908                                                                                                                                                                                                                                         | K131033                                                                                                                                                                                                                                          | --                          |
| Product name                                                                                        | Powder Free Vinyl Patient<br>Examination Gloves, Clear<br>(Non-colored)                                                                                                                                                                         | Powder Free Vinyl Patient<br>Examination Gloves, Clear<br>(Non-colored)                                                                                                                                                                          | Same                        |
| Product Code                                                                                        | LYZ                                                                                                                                                                                                                                             | LYZ                                                                                                                                                                                                                                              | Same                        |
| Size                                                                                                | Small/ Medium/<br>Large/X large                                                                                                                                                                                                                 | Small/ Medium/<br>Large/X large                                                                                                                                                                                                                  | Substantially<br>equivalent |
| Intend for use                                                                                      | Powder free Vinyl Patient<br>Examination Gloves, Clear(Non-<br>colored)is a disposable device<br>intended for medical purposes<br>that is worn on the examiner's<br>hand or finger to<br>prevent contamination between<br>patient and examiner. | Powder free Vinyl Patient<br>Examination Gloves, Clear<br>(Non-colored) is a disposable<br>device intended for medical<br>purposes that is worn on the<br>examiner's hand or finger to<br>prevent contamination between<br>patient and examiner. | Substantially<br>equivalent |
| Device Description<br>and Specifications                                                            | Meets ASTM D5250-06<br>(Reapproved 2011)                                                                                                                                                                                                        | Meets ASTM D5250 -06<br>(Reapproved 2011)                                                                                                                                                                                                        | Substantially<br>equivalent |
| Dimensions<br>-- Length                                                                             | Meets ASTM D5250-06<br>(Reapproved 2011)<br>≥230mm min.                                                                                                                                                                                         | 230mm min for all sizes                                                                                                                                                                                                                          | Substantially<br>equivalent |
| Dimensions<br>-- Width                                                                              | Meets ASTM D5250-06<br>(Reapproved 2011)                                                                                                                                                                                                        |                                                                                                                                                                                                                                                  | Substantially<br>equivalent |
|                                                                                                     | Small 80-90 mm<br>Medium 90-100mm<br>Large 100-110mm<br>X large 110-120 mm                                                                                                                                                                      | Small 80-85 mm<br>Medium 95-97 mm<br>Large 102-108mm<br>X large 114-118 mm                                                                                                                                                                       |                             |
| -- Thickness                                                                                        | (Reapproved 2011)                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                  |                             |
|                                                                                                     | Finger 0.05mm min.<br>Palm 0.08mm min.                                                                                                                                                                                                          | Finger 0.05mm min.<br>Palm 0.08mm min.                                                                                                                                                                                                           |                             |
| Physical Properties                                                                                 | Meets ASTM D5250-06<br>(Reapproved 2011)                                                                                                                                                                                                        | Before aging/after aging                                                                                                                                                                                                                         | Substantially<br>equivalent |
|                                                                                                     | Before aging/after aging<br>Elongation ≥300%<br>Tensile Strength≥14MPa                                                                                                                                                                          | Elongation ≥300%<br>Tensile Strength≥14MPa                                                                                                                                                                                                       |                             |
| Freedom<br>from                                                                                     | Meets                                                                                                                                                                                                                                           | Meets ASTM<br>D5151-06<br>(Reapproved 2011)                                                                                                                                                                                                      | Substantially<br>equivalent |
| Pinholes                                                                                            | • 21 CFR 800.20<br>• ASTM D5250-06<br>(Reapproved 2011)<br>• ASTM D 5151-06<br>(Reapproved 2011)                                                                                                                                                | Holes<br>Inspection Level 1<br>AQL2.5                                                                                                                                                                                                            |                             |
| Residual Powder                                                                                     | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)                                                                                                                                                                                                    | D 6124-06<br>(Reapproved 2011)                                                                                                                                                                                                                   | Substantially<br>equivalent |
|                                                                                                     |                                                                                                                                                                                                                                                 | Results generated values below<br>2mg of residual powder                                                                                                                                                                                         |                             |
| Compare all<br>materials used to<br>fabricate the devices                                           | PVC                                                                                                                                                                                                                                             | PVC                                                                                                                                                                                                                                              | Substantially<br>equivalent |
| Dusting or Donning<br>Powder:                                                                       | PU                                                                                                                                                                                                                                              | PU                                                                                                                                                                                                                                               | Substantially<br>equivalent |
| Dusting or Donning<br>Powder: name                                                                  | PU                                                                                                                                                                                                                                              | Surface Coating Agent                                                                                                                                                                                                                            | Substantially<br>equivalent |
| Compare<br>performance data<br>supporting<br>substantial<br>equivalence                             | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reapproved 2011)                                                                                                                         | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reapproved 2011)                                                                                                                          | Substantially<br>equivalent |
| Single Patient Use                                                                                  | Single Patient Use                                                                                                                                                                                                                              | Single Patient Use                                                                                                                                                                                                                               | Substantially<br>equivalent |
| Biocompatibility                                                                                    | SKIN IRRITATION DERMAL<br>and SENSITIZATION<br>STUDIES Meets ISO<br>10993-10 :2002/Amd.1:<br>2006                                                                                                                                               | The test article was a non-<br>irritant or non- sensitizer.<br>SKIN IRRITATION DERMAL<br>and SENSITIZATION<br>STUDIES Meets ISO<br>10993-10 :2002/Amd.1:2006                                                                                     | Substantially<br>equivalent |
| Labeling for the<br>legally marketed<br>device to which<br>substantial<br>equivalence is<br>claimed | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot                                                                                                                                                  | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot                                                                                                                                                   | Substantially<br>equivalent |

{2}------------------------------------------------

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence.

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)meet requirements per ASTM D5250-06(Reaffirmation 2011), per ASTM D6124-06(Reaffirmation 2011), per 21 CFR 800.20 and ISO 10993-10: 2002/Amd. 1:2006(E).

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence.

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

Section C(rev.02)

{3}------------------------------------------------

((b)(3)) The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims and the Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is as safe, as effective, and performs as well as the predicate device, Powder free Vinyl Patient Examination Gloves, Clear(Non-colored), Shijiazhuang Fuguan Plastic Products Co., Ltd. K032908 .

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing the feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 12, 2014

Decent Plastic Co., Ltd. C/O Mr. Chu Xiaoan Official Correspondent Room 1606 Bldg.1 Jianxiang Yuan No.209 Bei Si Huan Zhong Road Haidian District Beijing, 100083 CHINA

Re: K131033

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: December 31, 2013 Received: January 8, 2014

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

### Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/5 description: The image shows a logo with the letters "FDA" in a stylized font. Next to the logo is the text "Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID". Below the logo is the word "FOR".

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K 131033

#### Device Name

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)

#### Indications for Use (Describe)

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

[x] Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/6/Picture/13 description: The image shows the name "Elizabeth F. Claverie -S" in bold, black font. Below the name is the date "2014.02.11" followed by the time "16:48:21" and timezone "-05'00'" also in bold, black font. The text appears to be extracted from a document or record.

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K131033](https://fda-staging.innolitics.com/device/K131033)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).