K130998 · Siemens AG Medical Solutions · LLZ · Jan 3, 2014 · Radiology
Device Facts
Record ID
K130998
Device Name
SYNGO.VIA WEB VIEWER
Applicant
Siemens AG Medical Solutions
Product Code
LLZ · Radiology
Decision Date
Jan 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
syngo.via WebViewer is a software-only device indicated for reviewing medical images from syngo.via. It supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments (supported Image types: CT, MR, CR, DR, DX, PET). It is not intended for storage or distribution of medical images. syngo.via WebViewer is an option for the syngo.via system and cannot be run without it. It is client server architecture and the client is intended to run on web clients which are connected to the healthcare institution IT infrastructure where the customer will insure HIPAA compliance. The communication of syngo.via WebViewer with connected medical IT systems will be done via standard interfaces such as but not limited to DICOM. The system is not intended for the display of digital mammography images for diagnosis.
Device Story
syngo.via WebViewer is a software-only PACS extension for the syngo.via system; provides secure access to rendered medical images via web browsers and mobile devices (iPad). Inputs include DICOM-formatted medical images (CT, MR, CR, DR, DX, PET) from the syngo.via system. Operates via client-server architecture within hospital IT infrastructure; accessed by clinicians (Radiology, Nuclear Medicine, Cardiology) for image review and basic manipulation. Includes imaging algorithms (MPR, MIP, VRT, PET/CT, SUV). Output is displayed on standard web browsers or mobile clients; assists clinicians in image interpretation and evaluation. Does not store or distribute images. Benefits include fast, continuous access to radiological data for authorized personnel within the hospital network.
Clinical Evidence
Bench testing only. Software verification and validation (Unit, Integration, and System Test levels) performed according to standards including DICOM, ISO/IEC 15444-1, ISO/IEC 10918-1, HL7, IEC 62304, IEC 62366, ISO 14971, and IEC 60601-1-4.
Technological Characteristics
Software-only PACS; server runs on Windows Server 2008; clients run on standard web browsers (IE, Firefox, Safari) or mobile (iPad iOS 5.0/5.1). Supports DICOM images (CT, MR, CR, DR, DX, PET). Connectivity via standard IT interfaces (DICOM, HL7). Complies with IEC 62304 (software lifecycle), ISO 14971 (risk management), and IEC 62366 (usability).
Indications for Use
Indicated for reviewing medical images (CT, MR, CR, DR, DX, PET) within healthcare institutions for interpretation and evaluation. Not for digital mammography diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K 130998
JAN - 3 2014
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### I. GENERAL INFORMATION
# Establishment:
| Address: | Siemens AG, Medical Solutions<br>Henkestrasse 127<br>D-91052 Erlangen<br>Germany |
|----------------------|----------------------------------------------------------------------------------|
| Registration Number: | 3002808157 |
| Contact Person: | Joerg Teiche<br>Telephone: +49 (9131) 84-4687<br>Telefax: +49 (9131) 84-8691 |
Device Name and Classification:
| • Trade Name: | <i>syngo.via</i> WebViewer |
|-------------------------|---------------------------------------------|
| • Classification Name: | Picture Archiving and Communications System |
| • Classification Panel: | Radiology |
| • CFR Section: | 21 CFR §892.2050 |
| • Device Class: | Class II |
| • Product Code: | LLZ |
Date of submission:
.
.
April 2013
510(k) for syngo®.via WebViewer
December 31, 2013 - '
Page 1 of 5
.
{1}------------------------------------------------
#### II. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE SUBSTANTIAL EQUIVALENCE DETERMINATION
### Indications for Use:
syngo.via WebViewer is a software-only device indicated for reviewing medical images from syngo via. It supports interpretation and evaluations within healtheare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments (supported Image types: CT, MR, CR, DR, DX, PET). It is not intended for storage or distribution of medical images.
syngo.via WebViewer is an option for the syngo.via system and cannot be run without it. It is client server architecture and the client is intended to run on web clients which are connected to the healtheare institution IT infrastructure where the customer will insure HIPAA compliance. The communication of syngo.via WebViewer with connected medical IT systems will be done via standard interfaces such as but not limited to DICOM.
The system is not intended for the display of digital mammography images for diagnosis.
## Device Description:
The syngo.via WebViewer is a software-based Picture Archiving and Communications System (PACS) used with the syngo.via system. The syngo.via WebViewer provides secure access to rendered medical image data and basic image manipulation through web browsers and mobile devices within the reach of the hospital network.
It extends the syngo.via WebViewer software application previous cleared under K111079. New image types supported are PET and X-Ray images. It also provides functionality for displaying images via a mobile application on an iPad.
### syngo.via WebViewer Data Management:
...ensures all authorized personnel fast and continuous access to radiological data. It's main functionality ranges from availability of images with regard to data sceurity, open interfaces.
### Integration:
The Workflow Management enables by integration of any HL7- / DICOM-compatible RIS (IHE Year 5) to the syngo product family consistent workflow within the healthcare organization.
### Technological Characteristics:
syngo.via WebViewer server part is a "software only"-system, which will be delivered on CD-ROM / DVD to be installed on common 1T hardware. This hardware has to fulfil the defined requirements. The Software will be installed by Siemens service engineers only.
The backend communication and storage solution is based on Windows 2008 operating system. The client machines are based on standard computer (Mac / Windows PC / Linux PC). The web
510(k) for syngo®.via WebViewer
December 31, 2013
Page 2 of 5
{2}------------------------------------------------
and mobile client application runs in a standard web browser - such as but not limited to Internet Explorer, Firefox, Safari - please refer to the specification for the complete list of supported web browsers.
Any hardware platform, which complies to the specified minimum hardware and software requirements and with successful installation verification activities, can be supported.
The herewith described syngo.via WebViewer supports DICOM formatted images (CT, MR. CR, DR, DX, PET) and objects (SC, pdf).
The syngo.via WebViewer will be marketed as a software only solution for the end-user (with recommended hardware requirements). The server part will be installed by trained service engineers only. Any special needs such as integration in a specific environment and undates / upgrades will be covered by individual service contract and fulfilled by special trained service technicians.
#### Summary of Non-Clinical Tests
The software verification and validation (Unit Test Level. Integration Test Level and System Test Level) was performed for all newly developed components and the complete system according to the following standards:
- DICOM Standard [2011] -
- ISO/IEC 15444-1:2005+TC 1:2007 -
- ISO/IEC 10918-1:1994 + TC 1:2005 -
- HL7 [2006] -
- IEC 62304:2006 -
- -IEC 62366:2007
- -ISO 14971:2007
- -IEC 60601-1-4:2000
#### General Safety and Effectiveness Concerns:
The device labelling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize hazards, Siemens adheres to recognized and established industry practice and standards.
### Substantial Equivalence:
The syngo.via Web Viewer, addressed in this premarket notification, is substantially equivalent to the following commercially available device:
December 31, 2013
Page 3 of 5
{3}------------------------------------------------
| Manufacturer | Predicate Device Name | FDA Clearance<br>Number |
|--------------|-----------------------|-------------------------|
| Siemens | syngo.via WebViewer | K111079 |
The syngo.via Web Viewer described in this 510(k) has similar intended use and similar technical characteristics as the device listed below in regard to the specific functionalities.
| Functionality | syngo.via WebViewer VAI1 | syngo.via WebViewer VA10A |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | identical | identical |
| Intended use | syngo.via WebViewer is a software-<br>only device indicated for reviewing<br>medical images from syngo.via.<br>It supports interpretation and evalu-<br>ation of examinations within<br>healthcare institutions, for example,<br>in Radiology, Nuclear Medicine<br>and Cardiology environments (sup-<br>ported Image types: CT, MR, CR,<br>DR, DX, PET). It is not intended<br>for storage or distribution of medi-<br>cal images.<br>syngo.via WebViewer is an option<br>for the syngo.via system and cannot<br>be run without it. It is client server<br>architecture and the client is in-<br>tended to run on web clients which<br>are connected to the healthcare in-<br>stitution IT infrastructure where the<br>customer will insure HIPAA com-<br>pliance.<br>The communication of syngo.via<br>WebViewer with connected medi-<br>cal IT systems will be done via<br>standard interfaces such as but not<br>limited to DICOM.<br>The system is not intended for the<br>display of digital mammography<br>images for diagnosis. | syngo.via WebViewer is intended to<br>be a software-only solution for re-<br>viewing medical images from syn-<br>go.via.<br>The system cannot be used as stand-<br>alone device. It is intended to be an<br>option for syngo.via system only.<br>syngo.via WebViewer is not intended<br>for storage or distribution of medical<br>images from one medical device to<br>another.<br>syngo.via WebViewer is a client<br>server architecture and the client is<br>intended to run on web clients which<br>are connected to the healthcare insti-<br>tution IT infrastructure where the<br>customer has to ensure HIPAA com-<br>pliance.<br>syngo.via WebViewer supports inter-<br>pretation and evaluation of examina-<br>tions within healthcare institutions,<br>for example, in Radiology, Nuclear<br>Medicine and Cardiology environ-<br>ments.<br>The communication of syngo.via<br>WebViewer with connected medical<br>IT systems will be done via standard<br>interfaces such as but not limited to<br>DICOM.<br>The system is not intended for the |
Device Comparison Table between new device and predicate device:
510(k) for syngo®.via WebViewer
December 31, 2013
{4}------------------------------------------------
| Functionality | syngo.via WebViewer VA11 | syngo.via WebViewer VA10A |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| | | displaying of digital mammography<br>images for diagnosis in the U.S |
| Image<br>com-<br>munication | identical | identical |
| Image<br>Processing and<br>Evaluation | identical | identical |
| Supported Im-<br>age Types | MR, CT, SC, PDF, PET, CR, DX | MR, CT, SC, PDF |
| Image<br>data<br>compression | identical | identical |
| User admini-<br>stration | UI is syngo®-based web client and<br>mobile client | UI is syngo®-based web client |
| User<br>Interface | identical | identical |
| Hardware | Client: PC with Windows System<br>or MAC with Mac iOS Server:<br>Windows Server 2008 SP1<br>iPad (with iOS 5.0 or 5.1) | Client: PC with Windows System,<br>MAC with Mac OS Server: Windows<br>Server 2008 |
| Imaging Algo-<br>rithms | MPR, MIP, VRT, PET/CT, SUV | MPR, MIP, VRT |
| Quantitative<br>Algorithms | identical | identical |
In summary, Siemens is of the opinion that syngo.via WebViewer does not introduce any new significant potential safety risks and is substantially equivalent to and performs as well as the predicate device.
.
:
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract shapes or lines, possibly representing human figures or abstract forms.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
January 3, 2014
Siemens AG, Medical Solutions % Ms. Sabine Schroedel Regulatory Affairs Manager Henkestrasse 127 Erlangen, Bavaria 91052 GERMANY
Re: K130998
Trade/Device Name: syngo.via WebViewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 19, 2013 Received: December 20, 2013
Dear Ms. Schroedel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K130998
# Device Name
syngo®.via WebViewer
### Indications for Use (Describe)
syngo.via WebViewer is a software-only device indicated for reviewing medical innages from syngo.via. It supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments (supported Image types: CT, MR, CR, DR, DX, PET). It is not intended for storage or distribution of medical images.
syngo.via WebViewer is an option for the syngo.via system and cannot be run without it. It is client server architecture and the client is intended to run on web clients which are connected to the healtheare institution IT infrastructure where the customer will insure IIIPAA compliance.
The communication of syngo.via WebViewer with connected medical IT systems will be done via standard interfaces such as but not limited to DICOM.
The system is not intended for the display of digital manimography images for diagnosis.
Type of Use (Select one or both, as applicable)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Smh.7)
FORM FDA 3881 (9/13)
PSC Publishing Services (101) 461-4740 EG
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.